The Hipstar® hip stem is a single-use device intended for cementless fixation within the prepared femoral canal.
Device Story
The Hipstar Femoral Stem is a straight, titanium alloy (TMZF) implant designed for cementless fixation in the femoral canal. It features a proximal lateral flare for rotational stability and a narrow distal stem. The body is iron grit blasted to enhance bone-to-implant interface, excluding the trunnion, neck, and distal tip. Used by orthopedic surgeons in clinical settings for total or hemi-hip arthroplasty, the device replaces the femoral head and neck to restore joint function. It provides structural support and stability for patients suffering from degenerative joint disease, fractures, or failed previous implants. By facilitating bone ingrowth through its surface treatment, the stem aims to provide long-term fixation, reduce pain, and improve mobility for patients requiring hip reconstruction.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Material: TMZF titanium alloy. Surface: Iron grit blasted body (excluding trunnion, neck, distal tip). Design: Straight stem with proximal lateral flare and narrow distal stem. Fixation: Cementless. Sterilization: Not specified.
Indications for Use
Indicated for total and hemi hip arthroplasty in patients with noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, failed prior treatments/devices, and nonunion, femoral neck, or trochanteric fractures of the proximal femur. When used with a constrained acetabular liner, indicated for primary and revision total hip arthroplasty in patients at high risk of dislocation due to history of dislocations, bone loss, joint/soft tissue laxity, neuromuscular disease, or intraoperative instability.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
Alloclassic™ Zweymueller™ SL/SLL femoral hip stem (K030373)
Submission Summary (Full Text)
{0}------------------------------------------------
K05/223
**stryker**
Howmedica
OSTEONICS
1 10
## 510(k) Summary of Safety and Effectiveness for the HIPSTAR Femoral Stem
325 Corporate Drive Mahwah, NJ USA 07430
Proprietary Name:
Common Name:
Regulatory Class:
Device Product Code:
Classification Name and Reference
HIPSTAR Femoral Stem
Total Hip Joint Replacement Prosthesis
Hip joint, metal/ceramic/polymer semiconstrained cemented or nonporous uncemented prostheses, 21 CFR §888.3353
Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis 21 CFR §888.3360
Hip joint metal/polymer constrained cemented or uncemented prosthesis 21 CFR §888.3310
Class II
87 LZO - prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or nonporous, uncemented
87 LWJ - prosthesis, hip, semi-constrained, metal/polymer, uncemented
87 MEH - prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calcium-phosphate
87 KWL - prosthesis, hip, hemi-, femoral, metal
87 KWZ - prosthesis, hip, constrained, cemented or uncemented, metal/polymer
ЭЭДАЯ
02/120
{1}------------------------------------------------
For Information contact:
Tiffani Rogers Regulatory Affairs Specialist Stryker Orthopaedics 325 Corporate Drive Mahwah, New Jersey 07432 Phone: (201) 831-5612 Fax: (201) 831-6038 E-Mail: Tiffani.Rogers@stryker.com
Date Summary Prepared:
May 5, 2006
KOS 1223
### Device Description
The Hipstar® Femoral Stem is a straight hip stem manufactured from titanium alloy, TMZFTM. The Hipstar® hip features a proximal, lateral flare for rotational stability, with a narrow distal stem for implant stability. The body of the stem, with the exception of the trunnion, neck and distal tip, is iron grit blasted for increased bone to implant interface.
### Intended Use:
The Hipstar® hip stem is a single-use device intended for cementless fixation within the prepared femoral canal.
#### Indications
The Hipstar® Hip Stem is intended for use in total and hemi hip arthroplasty for the following indications:
- noninflammatory degenerative joint disease including osteoarthritis and avascular . necrosis;
- rheumatoid arthritis (excepting the Osteolock" HA Acetabular Cup and Peri-. Apatite coated prostheses);
- . correction of functional deformity;
- revision procedures where other treatments or devices have failed; and, ●
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal ▲ femur with head involvement that are unmanageable using other techniques.
When mated with a constrained acctabular liner the Hipstar® Hip Stern is indicated for use in primary and revision total hip arthroplasty for patients at high risk of hip dislocation due to a history of prior dislocations, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
#### Substantial Equivalence:
The determination of the substantial equivalence of the HIPSTAR® hip stem is based on its similarities in intended use, design and sterilization to Zimmer's Alloclassic™ Zweymueller™ SL/SLL femoral hip stem (K030373, cleared March 06, 2003).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name in a circular arrangement around a stylized caduceus symbol. The caduceus is a symbol of medicine and health, and it is composed of a staff with two snakes coiled around it.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 0 2006
Stryker® Orthopaedics % Ms. Tiffani Rogers Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07432
Re: K051223
Trade/Device Name: Hipstar® Hip Stem Regulation Number: 21 CFR 878.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis
Regulatory Class: II Product Code: LZO Dated: March 31, 2006 Received: April 3, 2006
Dear Ms. Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Tiffani Rogers
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Hubert Semeno
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# K051223
510(k) Number (if known): ___
Device Name: Hipstar® Hip Stem___
#### Indications
The Hipstar® Hip Stem is intended for use in total and hemi hip arthroplasty for the following indications:
- noninflammatory degenerative joint disease including osteoarthritis and avascular � necrosis:
- rheumatoid arthritis (excepting the Osteolock™ HA Acetabular Cup and Peri-. Apatite coated prostheses);
- correction of functional deformity; ●
- revision procedures where other treatments or devices have failed; and, ◆
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal . treatment of nonamon, remot that are unmanageable using other techniques.
When mated with a constrained acetabular liner the Hipstar® Hip Stem is indicated for use in primary and revision total hip arthroplasty for patients at high risk of hip dislocation due to a history of prior dislocations, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
The Hipstar® Hip Stem is intended for cementless use only.
X Prescription Use _____________________________________________________________________________________________________________________________________________________________
OR Over-the-Counter Use (Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Huker Lemer to
(Division Division of General. Restorative. and Neurological Devices
**510(k) Number** K051223
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.