# Y2K2 ENTERNAL TUBE FLUID FILTER (K051206)

_Moss Medical Products · KNT · Nov 2, 2005 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K051206

## Device Facts

- **Applicant:** Moss Medical Products
- **Product Code:** [KNT](/productcode/KNT.md)
- **Decision Date:** Nov 2, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

The Y2K2 Enteral In-Line Filter is intended for use to filter liquid feeding solutions prior to delivery into a patient's pre-inserted feeding tube, to minimize the risk of tube occusion by macro particulates. This will be applicable to any size or length pediatric or adult feeding tube (e.g., nasogastric, nasoduodenal, nasojejunal, gastrostomy, jejunostomy, transgastric duodenal, or transgastic jejunal).

## Device Story

Y2K2 Enteral In-Line Filter is an in-line fluid filter for enteral feeding systems. Device connects to feeding tube lines to filter liquid nutrition before delivery to patient. Primary function is removal of macro particulates to prevent tube occlusion. Device utilizes inert biomaterials similar to predicate filters. Used in clinical or home settings where enteral feeding is administered. Healthcare providers or caregivers install the filter in-line; output is filtered nutritional fluid. Benefit is reduced risk of feeding tube blockage, ensuring consistent nutrient delivery.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

In-line fluid filter; inert biomaterials; larger pore size compared to predicate filters for coarse particulate removal; standalone device; non-powered.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Pall PharmAssure Capsule Filter ([K943127](/device/K943127.md))
- Cobe Cardiovascular Pre-bypass Filter ([K850139](/device/K850139.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## NOV - 1 2005

## 510(k) SUMMARY INFORMATION

| APPLICANT NAME:       | Moss Medical Products, Inc.                                                  |
|-----------------------|------------------------------------------------------------------------------|
| ADDRESS:              | 4049 NY 150<br>West Sand Lake, New York 12196                                |
| PHONE:                | 518-674-0904                                                                 |
| FAX:                  | 914-684-1464                                                                 |
| E-MAIL:               | mosstube@mossmed.com                                                         |
| ACTIVITY OF APPLICANT | Initial Distributor                                                          |
| CONTACT PERSON:       | GERALD MOSS, Ph.D., M.D.                                                     |
| ADDRESS:              | 1 Reynal Road<br>White Plains, New York 10605                                |
| PHONE:                | 914-997-0392                                                                 |
| FAX:                  | 914-684-1464                                                                 |
| E-MAIL:               | gerald_moss@mossmed.com                                                      |
| NAME OF DEVICE        |                                                                              |
| TRADE NAME:           | Y2K2 Enteral Tube Fluid Filter                                               |
| COMMON NAME:          | In-Line Filter                                                               |
| CLASSIFICATION NAME:  | {21 CFR 876.5980}<br>{Tubes, Gastrointestinal (and accessories)              |
| PRODUCT CODE:         | {KNT}                                                                        |
| MANUFACTURER:         | Filtertek, Inc.<br>P.O. Box 310<br>11411 Price Road<br>Hebron, IL 60034-0310 |
| PREDICATE DEVICES:    | Pall PharmAssure Capsule Filter (K943127                                     |

I will maintain records to support a claim (TL, 15 to 127)

Cobe Cardiovascular Pre-bypass Filter (K850139)

INDICATIONS FOR USE: The device will filter fluids to be delivered into a feeding tube to remove potentially obstructing particulates.

TECHNICAL CHARACTERISTICS: The proposed and predicate devices are of essentially the same size, construction, and function. They are made of comparable (Re: Pall) or identical (Re: Cobe Cardiovascular) inert biomaterials. The Y2K2 Enteral In-Line Filter differs in having a larger pore size, to remove coarse (but not fine) particulates from enterally delivered fluids.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows a logo with a stylized bird-like figure composed of three curved lines, suggesting movement or flight. The logo is encircled by text that reads 'Department of Health'. The text is arranged along the curve of the circle, with the bird figure positioned inside the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 1 2005

Moss Medical Products, Inc. c/o Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K051206

Trade/Device Name: Y2K2 Enteral In-Line Filter Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: 76 KNT Dated: October 15, 2005 Received: October 17, 2005

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. hogdon.

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known):

K051206

Device Name:_Y2K2 Enteral In-Line Filter

Indications For Use:

The Y2K2 Enteral In-Line Filter is intended for use to filter liquid feeding solutions prior to delivery into a patient's pre-inserted feeding tube, to minimize the risk of tube occusion by macro particulates.

This will be applicable to any size or length pediatric or adult feeding tube (e.g., nasogastric, nasoduodenal, nasojejunal, gastrostomy, jejunostomy, transgastric duodenal, or transgastic jejunal).

Prescription Use
X
(Part 21 CFR 801 Subpart D)

ANDIOR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

5

(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

---

**Source:** [https://fda.innolitics.com/device/K051206](https://fda.innolitics.com/device/K051206)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com