K051066 · Theratest Laboratories, Inc. · LKP · Oct 7, 2005 · Immunology
Device Facts
Record ID
K051066
Device Name
THERATEST EL-ANA PROFILES
Applicant
Theratest Laboratories, Inc.
Product Code
LKP · Immunology
Decision Date
Oct 7, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Indications for Use
The EL-ANA Profiles™ is an in vitro diagnostic test for the detection and measurement of autoantibodies directed against the following autoantigens: single-stranded DNA (ssDNA), double-stranded DNA (dsDNA), Smith, RNP/Sm, SSA (Ro) "R, SSB (La) "R (30217), Gould Carding Ribosomal Protein P, Centromere™, and Chromatin (Nucleosomes). This test system is intended as an aid in diagnosis of systemic lupus erythematosus, Sjogren's syndrome, progressive systemic sclerosis (scleroderma), drug-induced lupus and polymyositis.
Device Story
EL-ANA Profiles is an in vitro diagnostic test system; detects and measures specific autoantibodies in patient serum samples. Used in clinical laboratory settings by trained laboratory personnel. Provides qualitative or quantitative results to assist physicians in diagnosing systemic autoimmune diseases including systemic lupus erythematosus, Sjogren's syndrome, scleroderma, drug-induced lupus, and polymyositis. Results are interpreted by clinicians in conjunction with patient clinical presentation and other diagnostic findings to support clinical decision-making.
Clinical Evidence
No clinical data provided in the document; substantial equivalence is based on performance characteristics typical of immunological test systems for autoantibody detection.
Technological Characteristics
Semi-quantitative ELISA; 96-well microtiter plate format. Capture antigens: human recombinant (Sm, SS-B, Jo-1, centromere) and mixed calf thymus/recombinant (SS-A). Detection: goat anti-human IgG (Fcγ specific) conjugated with HRP. Chromogen: TMB/peroxide. Requires spectrophotometer (450 nm or 450/620-690 nm).
Indications for Use
Indicated for patients suspected of having systemic lupus erythematosus, Sjögren’s syndrome, progressive systemic sclerosis (scleroderma), drug-induced lupus, or polymyositis. For prescription use only.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k051066
B. Purpose for Submission:
Modifications to a cleared device. The applicant made modifications to the capture antigens for four of the test parameters by replacing the biological source (bovine) antigens with human recombinant antigens or by adding a human recombinant component to the biological source antigen. The modified antigens included Sm, SS-A, SS-B and Jo-1. They also modified three calibrators (anti-centromere, anti-ribosomal protein P and anti-Jo-1) by manufacturing them as ready to use. This memo will only address parameters of the device affected by the modifications.
C. Measurands:
Single-stranded DNA (ssDNA), double-stranded DNA (dsDNA), Sm (recombinant), RNP/Sm, SSA (Ro) (recombinant), SSB (La) (recombinant), histones, Scl-70, Jo-1 (recombinant), ribosomal protein P, centromere (human recombinant antigen), chromatin.
D. Type of Test:
Semi-quantitative enzyme-linked immunosorbent assay (ELISA)
E. Applicant:
TheraTest Laboratories, Inc.
F. Proprietary and Established Names:
TheraTest EL-ANA Profiles
G. Regulatory Information:
1. Regulation section:
21 CFR 866.5100 Antinuclear Antibody Immunological Test System
2. Classification:
Class II
3. Product code:
LKP Anti-Sm Antibodies, Antigen and Control
LJM Antinuclear antibody, (Enzyme Labeled), Antigen, Controls
MQA Anti-Ribosomal P Antibodies
LLL Extractable Antinuclear Antibody, Antigen and Control
4. Panel:
Immunology 82
H. Intended Use:
1. Intended use(s):
The EL-ANA Profiles is an in vitro diagnostic test for the detection and measurement of autoantibodies directed against the following autoantigens: single-stranded DNA (ssDNA), double-stranded DNA (dsDNA), $\mathrm{Sm}^{\mathrm{HR}}$, RNP/Sm, SSA (Ro)$^{\mathrm{HR}}$, SSB (La)$^{\mathrm{HR}}$, histones, Scl-70, Jo-1$^{\mathrm{HR}}$, ribosomal protein P, centromere and chromatin (nucleosomes). The test system is intended as an aid in diagnosis of systemic lupus erythematosus, Sjögren’s syndrome, progressive systemic sclerosis (scleroderma), drug induced lupus, and polymyositis.
{1}
2. Indication(s) for use: Same as Intended use.
3. Special conditions for use statement(s): For prescription use only
4. Special instrument requirements: Single (450 nm) or dual (450 nm test, 620-690 nm reference) wavelength spectrophotometer (ELISA reader) for 96 well microtiter plates.
I. Device Description:
The assay components consist of: several configurations of microtiter plates coated with antigens: SS-A 60kDa + 52kDa (calf thymus plus recombinant); and Sm (BB'), Jo-1 and SS-B (all human recombinant); IgG enzyme conjugate: goat anti-human IgG (Fcγ specific) conjugated with horseradish peroxidase (HRP); chromogen: containing both the peroxide substrate HRP and 3,3',5,5' tetramethylbenzidine (TMB) as chromogenic indicator; positive and negative controls: human serum; calibrators: antigen specific single calibrators for all antibodies, combined calibrators for ANA/6 and ANA/8; specimen diluent; wash buffer; and stop reagent.
J. Substantial Equivalence Information:
1. Predicate device name(s): TheraTest EL-ANA Profiles
2. Predicate 510(k) number(s): k040291
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay components: | | |
| Capture antigens: ssDNA, dsDNA, RNP/Sm, histones, Scl-70, ribosomal protein P, centromere and chromatin | Antigens are from thymus (calf or rabbit) or human recombinant (centromere) | Same |
| IgG enzyme conjugate | Anti-human IgG (Fγ specific conjugated with HRP | Same |
| Specimen diluent | Phosphate buffer with stabilizers | Same |
| Wash buffer | 10X phosphate buffer | Same |
| Chromogen | Peroxidase substrate of HRP and TMB | Same |
| Stop reagent | 2M phosphoric acid | Same |
| Positive control | Human sera containing antibodies | Same |
| Negative control | Normal human sera | Same |
{2}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Capture antigens: | | |
| Sm | Human recombinant BB’ | Calf thymus |
| SS-A (Ro) | Calf thymus (SS-A 60) plus human recombinant SS-A 52kDa | Calf thymus |
| SS-B (La) | Human recombinant | Calf thymus |
| Jo-1 | Human recombinant | Calf thymus |
| Calibrators: | | |
| Centromere | Ready to use | Dilute prior to use |
| Ribosomal protein P | Ready to use | Dilute prior to use |
| Jo-1 | Ready to use | Dilute prior to use |
K. Standard/Guidance Document Referenced (if applicable):
None referenced
L. Test Principle:
The device is an enzyme immunoassay with a 96 well plate format. Microplate wells, pre-coated with antigen, are incubated with calibrators, controls and patient samples. During the incubation, antibodies present in the test sample bind to the coated wells. The wells are washed to remove unbound antibodies and labeled goat anti-human IgG is incubated in the wells. The wells are washed again to remove unbound conjugate. As a final step, chromogen is added and autoantibodies are measured using a spectrophotometric plate reader.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Coefficients of variation (CV) for both the intra- and inter-assay reproducibility were all <10% for the modified antigen assays as they were for the predicate device assays.
b. Linearity/assay reportable range:
Semi-quantitative autoimmune antibody assays such as these are not expected to yield linear results in dilution studies as is evident in the results shown below.
Sm BB'
Serial dilutions of the Sm BB' calibrator were tested on the BB' antigen. Plotting absorbance versus the dilution factor yielded y = 1.139x + 0.034, r = 0.997.
SS-A 60 kDa + 52kDa
The in-house standard for SS-A was tested at two-fold dilutions. Plotting absorbance versus the dilution factor yielded y = 2.575x + 0.130, r = 0.993.
SS-B
The SS-B calibrator was tested at two-fold dilutions. Plotting absorbance
{3}
versus the dilution factor yielded y = 1.345x + 0.008, r = 0.998.
## Jo-1
The in-house standard for anti-Jo-1 was tested at two-fold dilutions. Plotting absorbance versus the dilution factor yielded y = 1.083x + 0.048, r = 0.997.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable
d. Detection limit:
Not determined or relevant for these assays.
e. Analytical specificity:
**Sm BB'**
Specificity was demonstrated by running a QC panel of 9 sera with various known specificities. Only the Sm antibody sample was positive. Also, a panel of 25 samples with high levels of Sm antibodies, along with an internal Sm standard and the Sm positive CDC serum specimen were tested. All samples showed a strong reaction with the modified Sm antigen.
**SS-A 60 kDa + 52 kDa**
Specificity was demonstrated by running a QC panel of 11 sera with various known specificities. Only the SS-A antibody sample was positive. Also, a panel of 10 samples with high levels of SS-A antibodies, along with the SS-A positive CDC serum specimen were tested. All samples showed a strong reaction with the modified mixed SS-A antigen.
**SS-B**
Specificity was demonstrated by running a QC panel of 8 sera with various known specificities. Only the SS-B antibody sample was positive. Also, a panel of 15 samples with high levels of SS-B antibodies, along with an internal SS-B standard and the SS-B positive CDC serum specimen were tested. All samples showed a strong reaction with the modified SS-B antigen.
## Jo-1
Specificity was demonstrated by running a QC panel of 12 sera with various known specificities. Only the Jo-1 antibody sample was positive. Also, a panel of 12 samples with high levels of Jo-1 antibodies, along with an internal Jo-1 standard and the Jo-1 positive CDC serum specimen were tested. All samples showed a strong reaction with the modified Jo-1 antigen.
f. Assay cut-off:
Assay cut-offs for the 4 assays with modified antigens did not change from that originally established with the predicate assays. The cut-offs were validated by running 100 normal sera.
2. Comparison studies:
a. Method comparison with predicate device:
**Sm**
A comparison between the new assay and the predicate assay for 222 positive and negative samples was presented. The samples included 27 internal QC
4
{4}
known positive samples, 100 normal blood bank donor sera and samples from patients with rheumatoid arthritis (n=20), scleroderma (n=20), and SLE (n=55).
| Sm BB' | | Native Sm | | |
| --- | --- | --- | --- | --- |
| | | + | - | Total |
| | + | 54 | 14 | 68 |
| | - | 0 | 154 | 154 |
| Total | | 54 | 168 | 222 |
Positive percent agreement 100% (54/54)
Negative percent agreement 92% (154/168)
Overall percent agreement 94% (208/222)
## SS-A 60 kDa + 52 kDa
A comparison between the new assay and the predicate assay for 102 positive and negative samples was presented. The samples included 50 blood bank donor sera and 52 samples from patients with SLE.
| SS-A 60 kDa + 52 kDa | | SS-A 60 kDa | | |
| --- | --- | --- | --- | --- |
| | | + | - | Total |
| | + | 39 | 3 | 42 |
| | - | 0 | 60 | 60 |
| Total | | 39 | 63 | 102 |
Positive percent agreement 100% (39/39)
Negative percent agreement 95% (60/63)
Overall percent agreement 97% (99/102)
## SS-B
A comparison between the new assay and the predicate assay for 192 positive and negative samples was presented. The samples included 17 internal QC known positive samples, 100 normal blood bank donor sera and samples from patients with rheumatoid arthritis (n=20), scleroderma (n=20), and 35 samples identified as sera from "abnormal patients".
| SS-B recombinant | | Native SS-B | | |
| --- | --- | --- | --- | --- |
| | | + | - | Total |
| | + | 38 | 2 | 40 |
| | - | 4 | 148 | 152 |
| Total | | 42 | 150 | 192 |
Positive percent agreement 90% (38/42)
Negative percent agreement 99% (148/150)
Overall percent agreement 97% (186/192)
{5}
Jo-1
A comparison between the new assay and the predicate assay for 192 positive and negative samples was presented. The samples included 17 internal QC known positive samples, 100 normal blood bank donor sera and samples from patients with rheumatoid arthritis (n=20), scleroderma (n=20), and 35 samples identified as sera from “abnormal patients”.
| Jo-1 recombinant | | Native Jo-1 | | |
| --- | --- | --- | --- | --- |
| | | + | - | Total |
| | + | 52 | 0 | 52 |
| | - | 0 | 140 | 140 |
| Total | | 52 | 140 | 192 |
Positive percent agreement 100% (52/52)
Negative percent agreement 100% (140/140)
Overall percent agreement 100% (192/192)
b. Matrix comparison:
The new and the predicate devices use serum as matrix.
3. Clinical studies:
a. Clinical Sensitivity:
Not done.
b. Clinical specificity:
Not done.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
See assay cut-off.
5. Expected values/Reference range:
A small percentage of the normal population will be positive for these antibodies. Based on testing 100 normal donors and using the established cut-offs, the percentage of negative results with the modified assays was: Sm BB’ 99%, SS-A 98%, SS-B, 99% and Jo-1 99%.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.