← Product Code [NWG](/productcode/NWG) · K051061

# KRONUS GLUTAMIC ACID DECARBOXYLASE ANTIBODY KIT (K051061)

_Kronus Market Development Associates, Inc. · NWG · Nov 4, 2005 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K051061

## Device Facts

- **Applicant:** Kronus Market Development Associates, Inc.
- **Product Code:** [NWG](/productcode/NWG.md)
- **Decision Date:** Nov 4, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5660
- **Device Class:** Class 2
- **Review Panel:** Immunology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K051061 · Nov 4, 2005 | KRONUS GLUTAMIC ACID DECARBOXYLASE ANTIBODY KIT | Kronus Market Development Associates, Inc. | Retrospective clinical serum samples from patients with Type I diabetes; Retrospective clinical serum samples from patients with non-diabetic autoimmune diseases (Graves', Hashimoto's, Addison's, Myasthenia Gravis, SLE, RA); Retrospective serum samples from healthy blood donors | The applicant used retrospective clinical cohorts to determine the clinical sensitivity and specificity of the GADAb RIA assay by comparing results against established clinical diagnoses. | Clinical sensitivity; Clinical specificity; Retrospective cohort; Autoimmune disease |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical Sensitivity and Specificity Evaluation; Retrospective clinical cohort study | 153 Type I diabetes patients, 98 patients with non-diabetic autoimmune diseases, and 150 healthy blood donors; Sample Size: 401 total subjects | Not applicable for this study | Clinical sensitivity and specificity for GAD autoantibodies |

## Indications for Use

The KRONUS GADAb RIA Assay Kit is for the semi-quantitative determination of glutamic acid decarboxylase antibody in human serum. The KRONUS GADAb RIA Assay is useful as an aid in the diagnosis of Type I diabetes mellitus (autoimmune mediated diabetes).

## Device Story

The KRONUS GADAb RIA Assay Kit is an in vitro diagnostic test for semi-quantitative detection of glutamic acid decarboxylase (GAD) autoantibodies in human serum. The device uses a radioimmunoassay (RIA) principle; patient serum is incubated with human recombinant I125-labeled GAD65 tracer. GAD antibodies present in the sample bind to the tracer. Protein A is added to precipitate the antigen-antibody complexes. After centrifugation and decanting, the radioactivity in the pellet is measured using a gamma counter. The amount of radioactivity is directly proportional to the concentration of GAD antibodies in the sample. Results are determined by interpolating from a calibration curve. The assay is performed in a clinical laboratory setting by trained personnel. The output provides clinicians with a semi-quantitative value to assist in the diagnosis of Type I diabetes mellitus.

## Clinical Evidence

Clinical performance evaluated in 401 subjects. Sensitivity for Type I diabetes was 69.3% (106/153; 95% CI: 62-76%). Specificity in non-diabetic autoimmune patients and healthy normals was 98.4% (244/248; 95% CI: 96-100%). Overall agreement was 87.3%. Supporting literature reported clinical sensitivity of 64-83% and specificity of 98-100%. Analytical precision (intra-assay CV 3.6-6.7%; inter-assay CV 4.7-10.8%) and interference testing (hemoglobin, bilirubin, lipids) were also provided.

## Technological Characteristics

Radioimmunoassay (RIA) using human recombinant I125-labeled GAD65 tracer. Components include lyophilized tracer, assay buffer, calibrators, and Protein A precipitating reagent. Requires external gamma counter for detection. Measuring range 0-300 U/mL; cut-off >1 U/mL. Manual liquid-phase binding assay.

## Regulatory Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k051061

B. Purpose for Submission:
New device

C. Measurand:
Glutamic acid decarboxylase (GAD) autoantibodies

D. Type of Test:
Semi-quantitative, radioimmunoassay (RIA)

E. Applicant:
KRONUS Market Development Associates, Inc.

F. Proprietary and Established Names:
KRONUS Glutamic Acid Decarboxylase Antibody (GADAb) RIA Assay Kit

G. Regulatory Information:
1. Regulation section:
21 CFR § 866.5660, Multiple Autoantibodies Immunological Test System
2. Classification:
Class II
3. Product code:
NWG, Autoantibodies, glutamic acid decarboxylase (GAD), used for the semi-quantitative determination of glutamic acid decarboxylase (GAD) antibodies in human serum as an aid in the diagnosis of Type I diabetes mellitus (autoimmune mediated diabetes).
4. Panel:
Immunology 82

H. Intended Use:
1. Intended use(s):
The KRONUS GADAb RIA Assay Kit is for the semi-quantitative determination of glutamic acid decarboxylase antibody in human serum. The KRONUS GADAb RIA Assay is useful as an aid in the diagnosis of Type I diabetes mellitus (autoimmune mediated diabetes).
2. Indication(s) for use:
Same as the intended use.
3. Special conditions for use statement(s):
The device is for prescription use only
4. Special instrument requirements:
Gamma counter for I¹²⁵

I. Device Description:
The KRONUS GADAb RIA Assay Kit consists of:
1. Lyophilized I¹²⁵ GAD₆₅ Tracer (human recombinant).
2. Assay buffer
3. GAD antibody calibrators – ready to use
4. Protein A precipitating reagent

{1}

5. Two control sera: A. close to the cut-off, and B. moderate to high

# J. Substantial Equivalence Information:

1. Predicate device name(s): KRONUS Acetylcholine Receptor Antibody
2. Predicate 510(k) number(s): k042248
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | KRONUS Glutamic Acid Decarboxylase Antibody (GADAb) RIA Assay Kit | KRONUS Acetylcholine Receptor Antibody (AChRAb) Assay Kit  |
|  Intended Use | Semi-quantitative detection of autoantibodies | Same  |
|  Matrix | Serum | Same  |
|  Test principle | Radioimmunoassay | Same  |
|  Test platform | Antibodies bind to labeled antigen in liquid phase (test tube), the antigen-antibody complexes are precipitated and radioactivity measured. | Same  |
|  Detection instrument | Gamma counter | Same  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Analyte | Anti-glutamic acid decarboxylase antibodies | Anti-acetylcholine receptor  |
|  Indications for Use | Aid in the diagnosis of Type I diabetes mellitus (autoimmune mediated diabetes) | Aid in the differential diagnosis of Myasthenia Gravis  |
|  Precipitating reagent | Protein A | Anti-human IgG  |

# K. Standard/Guidance Document Referenced (if applicable):

"Review Criteria for In Vitro Diagnostic Devices for the Assessment of Thyroid Autoantibodies Using Direct Immunofluorescence Assay (IFA), Indirect Hemagglutination (IHA), Radioimmunoassay (RIA), and Enzyme Linked Immunosorbent Assay (ELISA)"

# L. Test Principle:

Calibrators, controls and patient samples are incubated for two hours with human recombinant  $\mathrm{I}^{125}\mathrm{GAD}_{65}$ . During this incubation, antibody binds to the tracer. Protein A is added and the tubes are incubated for one hour during which time the antibodies present are bound by Protein A and removed from solution. Assay buffer is then

{2}

added and the tubes are centrifuged. After centrifugation, the supernatants are decanted or aspirated. The amount of radioactivity in the pellets is directly proportional to the amount of antibody contained in the samples. Calibrator concentrations are plotted on semi-log graph paper and the concentration of antibody in the unknowns is interpolated from the curve.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

**Intra-assay**

Two controls containing low and high levels of GAD antibodies were assayed in 25 replicates each. The mean, SD and coefficient of variation (%CV) were determined. For the low sample the mean was 6.4 U/mL and the %CV was 3.58%. For the high sample the mean was 42.66 U/mL and the CV % was 3.73%. In addition, data were submitted to demonstrate precision of the assay closer to the 1.0 U/mL cut-off by running 18 replicates of 2 samples with values of 2.6 and 1.2 U/mL.

|  Intra-Assay  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Sample | 1 | 2 | 3 | 4  |
|  N | 25 | 25 | 18 | 18  |
|  Mean | 6.4 | 42.6 | 2.6 | 1.2  |
|  SD | 0.23 | 1.59 | 0.1 | 0.1  |
|  %CV | 3.6 | 3.7 | 3.7 | 6.7  |

**Inter-assay**

Two controls containing low and high levels of GAD antibodies were assayed in 25 independent runs. The mean, SD and coefficient of variation (%CV) were determined. For the low sample (mean = 6.1 U/mL) the %CV was 4.9% and for the high sample (mean = 42.9 U/mL) the %CV was 6.99%. KRONUS submitted data for 5 samples closer to the cut-off with values ranging from 1.2 to 37.5 U/mL.

|  Inter-Assay  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample | A | B | C | D | E | F | G  |
|  N | 18 | 18 | 18 | 18 | 18 | 25 | 25  |
|  Mean | 2.3 | 37.5 | 11.4 | 2.8 | 1.2 | 2.3 | 4.5  |
|  SD | 0.2 | 3.5 | 1.1 | 0.2 | 0.1 | 0.17 | 0.21  |
|  %CV | 10.8 | 9.7 | 9.4 | 6.2 | 8.1 | 7.4 | 4.7  |

#### b. Linearity/assay reportable range:

**Linearity**

The measuring range for the assay is from 0 to 300 U/mL. Because each patient sample will have a different dilution curve due to the nature of autoantibody affinities and avidities, linearity is variable. Patient dilutions are not advised for samples with GAD antibody concentrations above 300 U/mL. The 300 U/mL calibrator represents the approximate maximum binding of the

{3}

tracer in the assay and allows for most samples to be read off the curve without need of dilution. Samples with results above the highest calibrator are reported as “&gt;300 U/mL”.

## Recovery

The calibrators were spiked with three serum samples of varying GAD antibody levels (5.2, 7.3 and 9.6 U/mL). Recoveries ranged from 83.6 to 117.6%.

c. Traceability, Stability, Expected values (controls, calibrators, or methods): There is no reference standard available and no claim was made for traceability.

d. Detection limit:

### Lower detection limit

The lower detection limit of the assay was determined by sequentially testing the zero calibrator 20 times. A calibration curve of %B/T (binding of tracer) versus concentration was constructed. The mean and standard deviation were calculated and the mean + 2 standard deviations (SD) and the mean + 3 SD were interpreted from the curve. The lower detection limit was computed to be 0.11 U/ml which is the mean of 20 determinations plus 2 SD. The cut-off for the assay is set at &gt;1 U/mL for positive.

### Functional sensitivity

The functional sensitivity (defined as the lowest level yielding an average inter-assay CV not greater than 20%) was determined to be 0.85 U/mL. KRONUS recommends that results below the functional sensitivity of the assay be reported as less than 0.85 U/mL.

e. Analytical specificity:

### Hemoglobin

Samples from patients positive for antibodies to GAD and normal healthy blood donors were spiked with hemoglobin levels up to 490 mg/dL and analyzed. Percent differences ranged from 0-12% for all positive samples.

### Bilirubin

Samples from GAD antibody positive patients and healthy blood donors were spiked with 2 to 20 mg/dL of bilirubin and analyzed. With the exception of one sample (difference 24%), the % difference for all positive samples ranged from 0-17%. Other samples with similar results to the outlier showed % differences ranging from 2-8%.

### Lipids

Samples positive for GAD antibodies as well as from healthy blood donors were spiked with lipid at approximately 3000 mg/dL and 1000 mg/dL. Percent differences ranged from 0-17%

f. Assay cut-off:

Fifty normal healthy controls with no family history of diabetes mellitus (DM) were assayed. There were 33 females and 17 males ranging in age from 20-61 years. Forty-eight samples (96%) contained less than 0.35 U/mL and all samples contained 0.5 U/mL of GAD antibodies or less. Given these results, values less than or equal to 1 U/mL are considered negative for GAD antibodies and values greater than 1 U/mL are considered positive.

4

{4}

In an additional study to validate the appropriateness of the assay cut-off, 100 human blood donor serum specimens were tested. Of the 100 tested, 91 (91%) yielded results indistinguishable from 0 U/mL, and 95% contained less than 0.35 U/mL GAD antibodies. Three sera gave values of 1.0 U/mL or greater and all three were reduced by incubation with unlabeled GAD suggesting they contained specific GAD antibodies.

## 2. Comparison studies:

a. Method comparison with predicate device:

Not applicable as there was no comparison to a predicate device. The new device claims were supported by clinical studies and published literature.

b. Matrix comparison:

Not applicable.

## 3. Clinical studies:

a. Clinical Sensitivity:

Sixty Type I diabetes serum samples (35 males aged 7 to 46 years and 25 females aged 10 to 35 years) were assayed. Using a cut-off of &gt;1 U/mL, 40 (67%) were found to contain antibodies to GAD. In another study, KRONUS tested samples from an additional 93 confirmed Type I diabetes patients. Of the sera tested, 66 (71%) were positive for GAD antibodies

b. Clinical specificity:

KRONUS submitted data for 98 patients with non-diabetic diseases: 27 Graves' disease, 12 Hashimoto's thyroiditis, 20 Addison's disease, 19 myasthenia gravis, 10 SLE, and 10 rheumatoid arthritis patients. Combining results from these groups with 153 serum results from Type I diabetes subjects and 150 normal subjects, the following table was submitted:

|   | Target = Diabetes present | Non-target = Diabetes absent = Other autoimmunity | Normals  |
| --- | --- | --- | --- |
|  POSITIVE:  |   |   |   |
|   | Study 1: 40/60 (67%)
Study 2: 66/93 (71%) | Graves': 1/27 (4%)
Hashimoto's: 0/12 (0%)
Addison's: 0/20 (0%)
Myasthenia Gravis: 0/19 (0%)
SLE: 1/10 (10%)
RA: 0/10 (0%) | Study A: 0/50 (0%)
Study B: 2/100 (2%)  |
|  Total positive | 106/153 (69%) | 2/98 (2%) | 2/150 (1%)  |
|  Combined | 106/153 (69%) |  | 4/248 (2%)  |
|  Negative:  |   |   |   |
|   | Study 1: 20/60 (33%)
Study 2: 27/93 (29%) | Graves': 26/27 (96%)
Hashimoto's: 12/12 (100%)
Addison's: 20/20 (100%)
Myasthenia Gravis: 19/19 (100%)
SLE: 9/10 (90%)
RA: 10/10 (100%) | Study A: 50/50 (100%)
Study B: 98/100 (98%)  |

{5}

|   | Target = Diabetes present | Non-target = Diabetes absent = Other autoimmunity | Normals  |
| --- | --- | --- | --- |
|  Total negative | 47/153 (31%) | 96/98 (98%) | 148/150 (99%)  |
|  Combined | 47/153 (31%) |  | 244/248 (98%)  |
|  GADAb assay result | Diabetes present/absent |   | Total  |
| --- | --- | --- | --- |
|   |  (+) | (-)  |   |
|  (+) | 106 | 4 | 110  |
|  (-) | 47 | 244 | 291  |
|  Total | 153 | 248 | 401  |

Clinical sensitivity:  $69.3\%$  (106/153)  $(95\% \mathrm{CI};62,76)$

Clinical specificity:  $98.4\%$  (244/248)  $(95\% \mathrm{CI}; 96, 100)$

Overall agreement:  $87.3\%$  (350/401)

Published literature included in the submission showed a clinical sensitivity for GAD antibodies ranging from 64 to  $83\%$  and a clinical specificity ranging from  $98 - 100\%$ .

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable.

4. Clinical cut-off:

See assay cut-off.

5. Expected values/Reference range:

The expected value for the normal population is  $\leq 1\mathrm{U / mL}$ . However,  $2 - 3\%$  of the general population are positive for GAD antibodies.

# N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

# O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K051061](https://fda.innolitics.com/device/K051061)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
