INSULIN PUMP

K051041 · Abbott Diabetes Care, Inc. · LZG · Dec 23, 2005 · General Hospital

Device Facts

Record IDK051041
Device NameINSULIN PUMP
ApplicantAbbott Diabetes Care, Inc.
Product CodeLZG · General Hospital
Decision DateDec 23, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Insulin Pumps are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.

Device Story

Abbott Diabetes Care Insulin Pump; external infusion pump for continuous insulin delivery; manages blood glucose levels in diabetes mellitus patients. Used in home and clinical settings. Operates by delivering insulin at set and variable rates. Healthcare providers and patients use the device to control glucose levels; output affects insulin dosing decisions; benefits patient through automated, precise insulin administration.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological comparison to predicate device.

Technological Characteristics

External insulin infusion pump; Class II device. Similar materials and basic design to predicate. Intended for continuous insulin delivery at set and variable rates.

Indications for Use

Indicated for continuous insulin delivery at set and variable rates for the management of diabetes mellitus in persons requiring insulin.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ : DEC 2 3 2005 #### K051041 510(k) Summary Common / Usual Name: Classification Name: Class: Panel: Product Code: External Insulin Infusion Pump Insulin Infusion Pump Class II; 21 CFR §880.5725 General Hospital 80 LZG | Sponsor: | TheraSense, Inc.<br>1360 South Loop Road<br>Alameda, CA 94502 | |-------------------------------------------------------|-----------------------------------------------------------------------------------------| | Contact Person:<br>Title:<br>Telephone:<br>Facsimile: | Andrea L. Ruth<br>Sr. Associate, Regulatory Affairs<br>(510) 749-6360<br>(510) 239-2799 | | Predicate Device | Medtronic MiniMed 515 | | Predicate Device Manufacturer | Medtronic MiniMed,<br>18000 Devonshire St.,<br>Northridge, CA 91325 | #### Establishment Information ### Performance Standards, Section 514 of the Act No performance standard(s) or special controls applicable to this device have been promulgated under Section 514 of the Act. ### Comparison of the Technological Features of the New Device and Predicate Device The new and predicate devices have similar materials and basic design and technology. #### Substantial Equivalence Comparison and Rationale The Abbott Diabetes Care Insulin Pump (Pump) is substantially equivalent to the Medtronic MiniMed Paridigm 515. The Pump has the same target users and operating environment as predicate devices. A comparison of product features and performance characteristics of the previously cleared Medtronic MiniMed Paridigm 515 [510(k) #K040676] and the TheraSense, Inc. Insulin Pump is provided within the substantial equivalence discussion below. PROPRIETARY AND CONFIDENTIAL PAGE 1 OF 2 57 {2}------------------------------------------------ 510(k) k051041 The insulin pumps are all intended for use in home and clinical settings for the control of blood glucose levels in diabetes. The indication for use statement for the proposed Insulin Pump has the same elements as the predicate devices. Page 2 of 2 {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's emblem, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The emblem is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" in a circular arrangement. DEC 2 3 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Mary Rose Supervisor, Regulatory Affairs Abbott Diabetes Care, Incorporated 1360 South Loop Road Alameda, California 94502 Re: K051041 Ros1011 Trade/Device Name: Abbott Diabetes Care Insulin Pumps Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: December 6, 2005 Received: December 6, 2005 Dear Ms. Rose: We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your becaller been is substantially equivalent (for the referenced above and nave determined to legally marketed predicate devices marketed in indications for use stated in the cholorary to engineers date of the Medical Device interstate comments, or to tria, 20, 1978, as ssified in accordance with the provisions of Amendinents, of to devices that have obci (Act) that do not require approval of a premarket the Federal Pood, Drug, and Connexe : reacefore, market the device, subject to the general approval application (1 Mr.). - 1 ou - general controls provisions of the Act include controls provisions of the Fet. "The girling of devices, good manufacturing practice, requirements for animons against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III If your device is classified (see as 10) in a sontrols. Existing major regulations affecting (PNA), it may of Subject to Sach adatof Federal Regulations, Title 21, Parts 800 to 898. In your device can be found in the Oods no receints concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Ms. Rose Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA s Issualice of a succearing with other requirements mean that FDA has made a determination that your device Fodard agencies mean that FDA nas made a decemmandin that your avinistered by other Federal agencies. of the Act or any Federal statutes and regulations administers and matter to: penstratio of the Act or any rederal statutes and regirements, including, but not limited to: registration You must comply with all the Act's requirements, and monufacturing proctice You must comply with an the Act 3 requirements and 801); good manufacturing practice and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing (21 CFR Par and listing (ZT CFK Part 807), labeling (21 CFR Par every go requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if requirements as set form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) the success of the same of any and can and capital and sections of section of o This letter will anow you to begin mailed.ing your as is equivalence of your device to a premarket notification. The PDA miding of basiants of your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific advice for your device of the Also, please note the regulation please contact the Office of Compliance at (240) 276-0115. please contact the Office of Comphation at (210) - 11 (21 CFR Part 807.97). You entitled, "Misbranding by reference to premarket notification" (21 carter Act from the entitled, " Misoranding by reference to premation on your responsibilities under the tall for may obtain offer general mionnation on Jour Forpol Popper Assistance at its toll-free DIVISION of Sman Manufacts of 1) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Syrita G. Michael Owens Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use ## 510(k) Number : K 05104 | Device Name: Abbott Diabetes Care Insulin Pumps Indications for Use: The Insulin Pumps are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Jd. Qull for ADW. 12/23/05 05/04/
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...