← Product Code [NHA](/productcode/NHA) · K051011

# ENDOSSEOUS DENTAL IMPLANT ABUTMENT (K051011)

_Atlantis Components, Inc. · NHA · Jul 12, 2005 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K051011

## Device Facts

- **Applicant:** Atlantis Components, Inc.
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Jul 12, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

## Device Story

Atlantis Abutment for 3i Certain Implant serves as an interface between an endosseous dental implant and a prosthetic restoration. The device consists of an abutment and an abutment screw. It is placed over the implant shoulder and secured with the screw to provide support for cement-retained single or multiple tooth prostheses. Used in dental clinics by dental professionals. The device benefits patients by enabling the attachment of prosthetic teeth to dental implants, restoring function and aesthetics in edentulous areas.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Constructed from Titanium grade Ti-6A1-4V ELI (ASTM F-136). Compatible with 3i Certain Standard (4.1mm, 5.0mm, 6.0mm), OSSEOTITE XP Certain (5.0mm, 6.0mm), and OSSEOTITE NT Certain (4.0mm, 5.0mm, 6.0mm) implants. Mechanical fixation via screw.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- Atlantis Abutment and Abutment Screw ([K981858](/device/K981858.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a date, "JUL 12 2005". The month is July, the day is the 12th, and the year is 2005. The text is in a bold, sans-serif font.

Kostoll

ﻴﺮ ﻣ

Pre-market Notification Page - 0000009

- - - - - - - - - - - - - - -

## VII. SECTION 10 - 510(K) SUMMARY

This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

### Applicant's Name and Address 1.

Atlantis Components Inc. 25 First Street Cambridge, Massachusetts 02141 617-661-9799 Telephone Number: 617-661-9063 Fax Number: 1 Contact Person: Franklin Uyleman Manager of Quality and Regulatory Affairs

#### 2. Name of Device

| Trade Name:          | Atlantis™ Abutment for 3i Certain™ Implant |
|----------------------|--------------------------------------------|
| Common Name:         | Endosseous dental implant abutment         |
| Classification Name: | Endosseous dental implant abutment         |
|                      | 21 CFR 872.3630 Product code NHA           |

## Legally Marketed Device to which Equivalence is claimed (Predicate Device) 3.

| Manufacturer                | Device                                  | 510(k) Number |
|-----------------------------|-----------------------------------------|---------------|
| Atlantis Components<br>Inc. | Atlantis Abutment and Abutment<br>Screw | K981858       |

#### Description of the Device 4.

The devices covered in this submission are abutments which are placed into the dental implant to provide support for a prosthetic restoration. The subject abutments are indicated for cemented restorations.

{1}------------------------------------------------

Pre-market Page - 000010 - - - - - - - - - - - - - - - -

#### Description of the Device (continued) 4.

The Atlantis™ Abutments for 3i Certain™ Implant and abutment screws are made from Titanium grade Ti-6A1-4V ELI (Meets ASTM Standard F-136). The abutment is placed over the implant shoulder and is mounted into the implant with a screw. The abutments are compatible with 3i Certain Standard Implants with 4.1mm, 5.0mm, 6.0mm diameters; the OSSEOTITE® XP Certain Implants with 5.0mm and 6.0mm diameters; and the OSSEOTITE® NT® Certain™ Implants with 4.0mm, 5.0mm and 6.0mm diameters.

ﺮﺳﺘﮯ ﮨ

#### Intended Use of the Device న్న

The devices covered by this submission are abutments which are placed into a dental implant to provide support for a prosthetic reconstruction. The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

#### Basis for Substantial Equivalence 6.

The Atlantis™ Abutments for 3i Certain Implants are substantially equivalent in intended use, material, design and performance to the Atlantis Abutments cleared under K981858.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 2 2005

Atlantis Components, Incorporated C/O Betsy A. Brown B.A. Brown & Associates 8944 Tamaroa Terrace Skokie, Illinois 60076

Re: K051011

Trade/Device Name: Atlantis™ Abutment for 3i Certain™ Implant Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: April 20, 2005 Received: April 21, 2005

Dear Ms. Brown:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Brown

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and instills ( = = = = forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), nlease contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Aury Suleom for

Chiu Lin. Ph. B Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Pre-market Notification Page - 000205

# Indications for Use

510(k) Number (if Known)_KOS 1011

Device Name: Atlantis ™ Abutment for 3i Certain™ Implant

Indication for Use:

The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

Prescription Use 8 (Part 21 CFR 801 SubpartD) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kbtz DVS for Dr. S. Runner

(Division Sign-Off) 
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices

510(k) Number: K051011

---

**Source:** [https://fda.innolitics.com/device/K051011](https://fda.innolitics.com/device/K051011)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
