HUBER CLEAR SAFETY INFUSION SET

K051009 · Now Medical Distribution, Inc. · FPA · Jul 12, 2005 · General Hospital

Device Facts

Record IDK051009
Device NameHUBER CLEAR SAFETY INFUSION SET
ApplicantNow Medical Distribution, Inc.
Product CodeFPA · General Hospital
Decision DateJul 12, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The NOW Huber Clear™ Safety Infusion Set is a safety IV administration set with a noncoring, ninety-degree, right-angle Huber needle, used to access surgically implanted vascular ports. The Huber needle is used to administer fluids or to withdraw blood. The Huber Clear™ system facilitates safe removal of the needle by encapsulating the needle within the attachment wings to help prevent needlestick injuries when using the device for vascular port access.

Device Story

Non-coring, 90-degree right-angle Huber needle and administration set; includes passive needlestick prevention feature. Used to access surgically implanted vascular ports for fluid infusion or blood sampling. Operated by clinicians; needle inserted into port in standard manner. Upon removal, passive safety mechanism automatically encapsulates needle within attachment wings; provides tactile, visual, and audible confirmation of secure needle cover. Benefits include reduction of needlestick injuries during vascular port access.

Clinical Evidence

Bench testing only. Simulated use study performed to evaluate safety feature performance; results showed no incidences of sharps injuries and no failures of the safety mechanism activation.

Technological Characteristics

Biocompatible, medical grade materials; 90-degree right-angle Huber needle; passive mechanical safety feature for needle encapsulation. Single-use, sterile, nonpyrogenic. No software or electronic components.

Indications for Use

Indicated for patients requiring access to surgically implanted vascular ports for fluid administration or blood withdrawal.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL 12 2005 K051009 Page 1 of 2 ## NOW MEDICAL DISTRIBUTION HUBER CLEAR™ SAFETY INFUSION SET 510(k) PREMARKET NOTIFICATION #### 510(k) SUMMARY This 510(k) summary is being submitted in accordance with the requirements of SODA and 21CFR § 807.92 | Submitted by: | NOW Medical Distribution, Inc.<br>PO Box 936<br>Chadds Ford, PA 19317 | |-------------------|--------------------------------------------------------------------------------------------| | Contact Person: | Christine Emanuel<br>Regulatory Consultant<br>(805) 963-4312 (tel)<br>(805) 564-8642 (fax) | | Date Prepared: | April 15, 2005 | | Device Name: | | | Proprietary Name: | NOW Huber Clear™ Safety Infusion Set | | Common Name: | Huber Needle and Administration Set | |-----------------|-----------------------------------------------------------------| | Classification: | Class II: Product Code FPA<br>Regulation Number 21 CFR 880.5440 | Identification of Predicate Devices - NOW Huber Plus Safety Infusion Set, NOW Medical Distribution, Chadds Ford, PA ● K993848 ### Device Description: The NOW Huber Clear™ Safety Infusion Set is a standard non-coring right angle Huber needle and administration set with a needlestick prevention feature. The Huber Cleartist Safety Infusion Set is designed for use with a vascular access infusion system. The NOW Huber Clear™ Safety Infusion Set is fabricated from biocompatible, medical grade materials. The needle is inserted into the vascular access port in a standard manner for fluid infusion or for blood sampling. Removal of the needle is done as with any standard Huber needle, using a one-handed (dominant hand) technique to pull out the needle, and stabilizing the port with the nondominant hand. As the needle is removed, the passive needlestick 000026 {1}------------------------------------------------ K051009 ## NOW MEDICAL DISTRIBUTION HUBER CLEAR™ SAFETY INFUSION SET 510(k) PREMARKET NOTIFICATION prevention feature actuates automatically when the users hands are positioned correctly and can be easily seen, felt, and heard by the user to ensure that the needle cover is secure. The NOW Huber Clear™ Safety Infusion Set is supplied sterile and nonpyrogenic, for single use only. #### Indication for Use: The NOW Huber Clear™ Safety Infusion Set is a safety IV administration set with a noncoring, ninety-degree, right-angle Huber needle, used to access surgically implanted vascular ports. The Huber needle is used to administer fluids or to withdraw blood. The Huber Clear™ system facilitates safe removal of the needle by encapsulating the needle within the attachment wings to help prevent needlestick injuries when using the device for vascular port access. #### Technological Characteristics The Now Huber Clear™ Safety Infusion Set is a modification of the Huber Plus Safety Infusion Set, and encompasses no new technology. #### Performance Data Equivalency performance testing has been performed on the Now Huber Clear™ Safety Infusion Set components that have been modified from the predicate Huber Plus. In addition, a Simulated Use Study was performed showing that there were no incidences of sharps injuries or incidences where the safety feature failed to be activated during the course of the study. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows a circular logo. The logo features a stylized depiction of an eagle or bird in flight, with three curved lines representing its wings and body. The bird is facing to the right. Encircling the bird is text that reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA". JUL I 2 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Now Medical Distribution, Incorporated C/O Ms. Christine Emanuel Tecsa Technical Services 1205 De La Vina Street Santa Barbara, California 93101 Re: K051009 Trade/Device Name: NOW Huber Clear™ Safety Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: April 20, 2005 Received: April 26, 2005 Dear Ms. Emanuel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register. {3}------------------------------------------------ Page 2 - Ms. Emanuel Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CT R Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Qum Syllom be Chiu Lin, Ph. D Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ KD51009 # Indications For Use 510(k) Number (if known): ____K051009 Device Name:_NOW Huber Clear™ Safety Infusion Set Indications For Use: Indications For Use. The NOW Huber Clear™ Safety Infusion Set is a safety IV administration set with a ready Huber Clear™ Safety Infusion in and to seess surgically The NOW Huber Clear - Saicty innuole Huber needle, used to access surgically non-coring, hinety-degree, fight-alige in used to administer fluids or to withdraw implanted vascular ports. The Huber it's used to administer fluids or to withdraw Implanted vascular ports. The Huber notes safe removal of the needle by blood. The Huber Clear - System racklated cats vings to help prevent needlestick encapsulating the needle within the attachment wasses cheapsulating the news the device for vascular port access. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Cim Kem (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number . Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...