NANO-CHECK AMI 3 IN 1 CARDIAC DISEASE TEST FOR CARDIAC TROPONIN I, CREATINE KINASE MB AND MYOGLOBIN
Applicant
Nano-Ditech Co.
Product Code
MMI · Clinical Chemistry
Decision Date
Mar 2, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Indications for Use
The Nano-Check ™AMI 3 IN 1 Test is a rapid immunoassay for the qualitative determination of Cardiac Troponin I (cTnl), Creatine Kinase MB (CK-MB), and Myoglobin in human serum and plasma specimens at cutoff concentrations of 0.5 ng/ml, 5.0 ng/ml, and 80 ng/ml, respectively, as an aid in the diagnosis of Acute Myocardial Infarction (AMI). The Nano-Check™ AMI 3 IN 1 Test is a qualitative assay, which can not monitor the rise and fall of cTnl, CK-MB, and Myoglobin in single testing. Single testing is not recommended for AMI monitoring. Test results should be interpreted by the physician in conjunction with other laboratory test results and patient clinical findings.
Device Story
Nano-Check AMI 3 IN 1 is a single-use, lateral flow immunochromatography cassette for qualitative detection of cardiac markers (cTnl, CK-MB, Myoglobin) in serum or heparin plasma. Device contains membrane strip with immobilized capture antibodies (CK-MB, Myoglobin, streptavidin for cTnl, goat anti-mouse for control) and dye pad with gold-colloidal labeled antibodies. Sample application triggers migration of immune complexes; binding to test lines produces visible red bands if concentrations exceed cutoffs (0.5 ng/ml cTnl, 5 ng/ml CK-MB, 80 ng/ml Myoglobin). Used in clinical settings by professionals; results interpreted alongside clinical findings to aid AMI diagnosis. Provides rapid (15 min) visual qualitative assessment; does not monitor marker kinetics.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing only, including analytical performance (precision, reproducibility, detection limits, analytical specificity, assay cut-off) and method comparison studies against the predicate device.
Indicated for qualitative detection of cTnI, CK-MB, and Myoglobin in human serum/plasma as an aid in diagnosing Acute Myocardial Infarction (AMI). For prescription use only. Not for point-of-care settings.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
Spectral Cardiac Status CK-MB/Myoglobin/Troponin I 3-in-1 Test (k030057)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY
A. 510(k) Number:
k050975
B. Purpose for Submission:
New 510(k)
C. Measurand:
Troponin I, Creatinine Kinase MB (CK-MB) and Myoglobin
D. Type of Test:
Qualitative visually read single-use immunochromatographic assay
E. Applicant:
Nano-Ditech Corporation
F. Proprietary and Established Names:
In Vitro Nano-Check™ AMI 3 IN 1, Cardiac Marker Test cTnI, CK-MB, and Myoglobin
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1215, Creatine phosphokinase/creatine kinase or isoenzymes test system (troponin)
21 CFR 862.1215, Creatine phosphokinase/creatine kinase or isoenzymes test system (CK-MB)
21 CFR 866.5680, Myoglobin immunological test system
2. Classification:
All analytes are Class II
3. Product code:
MMI, JHT, and DDR, respectively.
4. Panel:
75 (Chemistry), 75, and 81 (Hematology), respectively.
H. Intended Use:
1. Intended use(s):
See indications for use.
{1}
2. Indication(s) for use:
The Nano-Check AMI 3 IN 1 Test is a rapid immunoassay for the qualitative determination of Cardiac Troponin I (cTnI), Creatine Kinase MB (CK-MB), and Myoglobin in human serum and plasma specimens at cutoff concentrations of 0.5 ng/mL, 5.0 ng/mL, and 80 ng/mL, respectively, as an aid in the diagnosis of Acute Myocardial Infarction (AMI).
The Nano-Check AMI 3 IN 1 Test is a qualitative assay, which can not monitor the rise and fall of cTnI, CK-MB, and Myoglobin in single testing. Single testing is not recommended for AMI monitoring. Test results should be interpreted by the physician in conjunction with other laboratory test results and patient clinical findings.
3. Special conditions for use statement(s):
The device is for in vitro diagnostic prescription use.
The sponsor has not evaluated this device in point-of-care settings, and they indicate the product will not be marketed to point-of-care settings.
The test system utilizes heparinized plasma.
4. Special instrument requirements:
Not applicable. The device is a visually read single-use device.
I. Device Description:
The Nano-Check TM AMI 3 IN 1 Test is a one-step single-use lateral flow immunochromatographic assay. It is a visually read in a plastic cassette housing. A membrane strip inside the plastic housing contains immobilized antibodies for each test and at the control line. A disposable sample dropper is provided.
No human source materials are present in the device.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Spectral Cardiac Status CK-MB/Myoglobin/Troponin I 3-in-1 Test
2. Predicate 510(k) number(s):
k030057
3. Comparison with predicate:
The Indications for Use, cut-off concentrations, and technology of the two devices are the same. Details of the differences appear below, i.e., the antibody types and acceptable matrices.
{2}
| Item | Device | Predicates |
| --- | --- | --- |
| | Nano-Check™ AMI 3 IN 1, Cardiac Marker Test, cTnI, CK-MB, and Myoglobin | Spectral Cardiac STATus™ CK-MB-Myoglobin/Troponin I 3-in-1 test |
| Differences | | |
| Capture molecule in test line | cTnI: Streptavidin Myo: Monoclonal mouse anti-Myo antibody CKMB: Monoclonal mouse anti-CKMB antibody | cTnI: Streptavidin Myo: Polyclonal rabbit anti-myo antibodies CKMB: Polyclonal goat anti-CKMM antibodies |
| Conjugate molecule | cTnI: Dye coupled with monoclonal mouse anti-cTnI antibody and biotinylated monoclonal mouse anti-cTnI antibody Myo: Dye coupled with monoclonal mouse anti-Myo antibody CKMB: Dye coupled with monoclonal mouse anti-CKMM | cTnI: Dye coupled with monoclonal anti-cTnI antibody and biotinylated polyclonal rabbit anti-cTnI antibodies Myo: Dye coupled with anti-Myo antibody CKMB: Dye coupled with monoclonal mouse anti-CKMB |
| Sample type | Serum or Plasma (heparin as anticoagulant) | Whole blood, serum or plasma (heparin as anticoagulant) |
# K. Standard/Guidance Document Referenced (if applicable):
The sponsor did not reference any standards or guidance documents.
# L. Test Principle:
The test is an immunochromatography assay for the qualitative determination of three cardiac markers simultaneously in serum and heparinized plasma. The membrane strip contains three test lines and one control line; immobilized monoclonal mouse antibody against CK-MB, monoclonal mouse antibody against Myoglobin, streptavidin for biotinylated cTnI antibody, and goat anti-mouse antibody for control line. A dye pad is placed at the end of the membrane containing biotinylated cTnI antibody and gold colloidal particles coupled with CK-MM, cTnI and Myoglobin antibodies. When a sample is applied into the sample well, the cardiac makers present in the sample bind to the specific antibodies coupled with gold particles. cTnI in a sample binds to both cTnI specific dye coupled antibody and biotinylated antibody. The immune complexes move along the nitrocellulose membrane through the test lines and bind to their corresponding capture antibodies immobilized on the test line. Unbound immune complexes pass through the test line and are captured by goat anti mouse antibody in the control line.
If the concentration of any of these three markers in the sample is above the cutoff level, red bands appear at the corresponding test lines and the control line. If the concentration of the markers in the sample is lower than the cutoff level, only the control line can be seen in the test window. The control line is present to validate that an adequate volume of sample has
{3}
been added.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Two Clinical sites and one in-house operator were provided with blindly labeled serum samples. All operators had experience with in vitro device testing.
Samples were prepared by diluting patient serum samples with normal serum to the desired concentrations. Normal human serum samples were also evaluated.
Each sample was tested using 5 devices at each testing site.
AMI 3 IN 1 Precision Results
| | | Sample 1 Concentration (ng/mL) | Sample 2 Concentration (ng/mL) | Sample 3 Concentration (ng/mL) |
| --- | --- | --- | --- | --- |
| | | cTnI 0
CK-MB 0
Myo 0 | cTnI 1.03
CK-MB 12.6
Myo 119.9 | cTnI 2.05
CK-MB 5.0
Myo 166.3 |
| | | Negative / Positive | Negative / Positive | Negative / Positive |
| cTnI | Site I | 5 / 0 | 0 / 5 | 0 / 5 |
| | Site 2 | 5 / 0 | 0 / 5 | 0 / 5 |
| | Site 3 | 5 / 0 | 0 / 5 | 0 / 5 |
| CK-MB | Site I | 5 / 0 | 0 / 5 | 0 / 5 |
| | Site 2 | 5 / 0 | 0 / 5 | 0 / 5 |
| | Site 3 | 5 / 0 | 0 / 5 | 0 / 5 |
| Myo | Site I | 5 / 0 | 0 / 5 | 0 / 5 |
| | Site 2 | 5 / 0 | 0 / 5 | 0 / 5 |
| | Site 3 | 5 / 0 | 0 / 5 | 0 / 5 |
Assay Cutoff and Precision Studies:
Similarly, patient plasma containing cTnI, CK-MB or Myoglobin were diluted in normal human serum to concentrations at or near the cutoff levels. Analyte concentrations in the diluted samples were confirmed by the Beckman Coulter Access quantitative test system. Fifteen devices were tested for each samples at the concentrations listed.
{4}
Assay Cutoff and Precision Studies
| | Concentration (ng/ml) | Test Result | |
| --- | --- | --- | --- |
| | | Negative | Positive |
| cTnI | 0.31 | 7 | 8 |
| | 0.57 | 0 | 15 |
| | 1.07 | 0 | 15 |
| CK-MB | 2.4 | 15 | 0 |
| | 4.9 | 0 | 15 |
| | 10.2 | 0 | 15 |
| Myoglobin | 56.5 | 14 | 1 |
| | 80.4 | 0 | 15 |
| | 141.3 | 0 | 15 |
b. Linearity/assay reportable range:
Not applicable. The assay is intended for qualitative use.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The test kit has an internal control to ensure that an adequate volume of sample is added. External Quality Controls are available from commercial sources. Specific controls are not identified in the labeling.
The sponsor does not indicate any degree of traceability for the device. However, the device is calibrated during the manufacturing process to working standards prepared to known concentrations of the analytes.
d. Detection limit:
Performance around the claimed cutoff concentration of the assay is characterized by the precision section, above.
e. Analytical specificity:
Potentially interfering endogenous substances were spiked into normal serums and into serums containing either cTnI, CK-MB or Myoglobin at concentrations approximately 1.5 times the cutoff concentration of the assay. The substances did not interfere with the expected results, i.e., spiked and unspiked sample results were the same.
AMI 3 IN 1 Endogenous Interference Studies
| | Substances | Concentration |
| --- | --- | --- |
| Endogenous substances | Bilirubin | 50 mg/dl |
| | Hemoglobin | 4000 mg/dl |
| | Human Serum Albumin | 10 g/dL |
| | Triglycerides | 1,250 mg/dl |
{5}
The device was tested for interference by potentially cross-reacting endogenous proteins. Proteins were added to normal human serum up to following concentrations, did not alter the expected negative result.
Potential Interference Study Results
| | Substances | Concentration |
| --- | --- | --- |
| Cross-reacting endogenous proteins | Cardiac myosin light chain | 1,000 ng/ml |
| | Cardiac Troponin T | 1,000 ng/ml |
| | Cardiac Troponin C | 1,000 ng/ml |
| | Skeletal Troponin I | 1,000 ng/ml |
| | CKMM | 5,000 ng/ml |
The sponsor did not evaluate the effects HAMA and other similar antibodies on their assay, although they describe potential interference described in the literature.
f. Assay cut-off:
See precision studies above for analytical characterization of the claimed cutoff concentrations.
2. Comparison studies:
a. Method comparison with predicate device:
Plasma samples were collected from 206 emergency room patients who were admitted because examination results suggested a cardiac event. Additionally, 50 samples were collected from outpatients who were not suspected of having a cardiac event. The 256 clinical samples were tested using the Nano-Ditech™ AMI 3 IN 1 Test and the Beckman Coulter Access test system. Results are summarized, below.
AMI 3 IN 1 TnI Results Compared to Quantitative Results (ng/mL)
| | Access Result 0.023-0.3 | Access Result 0.3-0.47 | Access Result 0.52-0.55 | Access Result 0.55- >100 |
| --- | --- | --- | --- | --- |
| Nano-Check Negative | 95 | 5 | 2 | 0 |
| Nano-Check Positive | 0 | 7 | 3 | 144 |
{6}
AMI 3 IN 1 CK-MB Results Compared to Quantitative Results (ng/mL)
| | Access Result 0.7-4.4 | Access Result 4.5-4.9 | Access Result 5-5.2 | Access Result 5.2- >300 |
| --- | --- | --- | --- | --- |
| Nano-Check Negative | 84 | 4 | 1 | 0 |
| Nano-Check Positive | 0 | 8 | 4 | 155 |
AMI 3 IN 1 Myoglobin Results Compared to Quantitative Results (ng/mL)
| | Access Result 10.2-61 | Access Result 61.5-79.1 | Access Result 80.4-85.3 | Access Result 0.55- >100 |
| --- | --- | --- | --- | --- |
| Nano-Check Negative | 56 | 15 | 2 | 0 |
| Nano-Check Positive | 0 | 7 | 5 | 171 |
b. Matrix comparison:
Paired serum and plasma samples were collected. Sample results are grouped below, according to the concentration of the targeted analyte.
Matrix Correlation Study Results
| | Concentration Range of Samples (ng/mL) | Serum Positive / Negative | Plasma Positive / Negative |
| --- | --- | --- | --- |
| TnI | 1.2-1.8 | 10 / 0 | 10 / 0 |
| | 0.5-1.0 | 10 / 0 | 10 / 0 |
| | 0.3-0.5 | 9 / 1 | 7 / 3 |
| | | | |
| CK-MB | 11-20 | 10 / 0 | 10 / 0 |
| | 5-10 | 10 / 0 | 10 / 0 |
| | 3.8-4.4 | 8 / 2 | 7 / 3 |
| | | | |
| Myo | 123-248 | 10 / 0 | 10 / 0 |
| | 85-118 | 10 / 0 | 10 / 0 |
| | 64-80 | 8 / 2 | 9 / 1 |
3. Clinical studies:
a. Clinical Sensitivity:
Clinical studies are not typically submitted for this device type and matrix. Clinical studies were not performed by the predicate device.
b. Clinical specificity:
Clinical studies are not typically submitted for this device type and matrix.
{7}
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Clinical performance of this device at the chosen cutoff concentration is characterized in the method comparison studies presented, above.
5. Expected values/Reference range:
Results of the Nano-Check Cardiac 3 IN 1 Test are interpreted according to the predetermined cutoff values of 0.5 ng/ml for cTnI, 5 ng/ml for CK-MB and 80 ng/ml for Myoglobin. These cutoff levels were determined by comparison to the quantitative Access assay system of Beckman Coulter, Access AccuTnI, Access CK-MB Assay and Access Myoglobin Assay.
The cutoff values of the Nano-Check device were determined by comparison to the Beckman Coulter quantitative assay, Access assays. Specimens containing cTnI, CK-MB and Myoglobin, at the concentration of equal or above established cutoff levels will give positive results using the Nano-Check test.
N. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None
O. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
P. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.