K050956 · Sudimplant SA · DZE · Feb 16, 2006 · Dental
Device Facts
Record ID
K050956
Device Name
TBR IDEA CONIC/CONIC
Applicant
Sudimplant SA
Product Code
DZE · Dental
Decision Date
Feb 16, 2006
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The T.B.R.® ide@ conic endosseous dental implant is a device intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function.
Device Story
T.B.R. ide@ conic is a self-tapping, threaded, screw-type endosseous dental implant system. Components include implants, abutments (tapered, castable, ball, ceramic/Ti), and surgical/laboratory tools. Implants are manufactured from Ti-6Al-4V alloy with a sandblasted/acid-etched surface to promote osseointegration. The device is surgically placed into the jawbone by a dental professional to anchor prosthetic teeth. The system provides mechanical support for dental restorations, enabling the restoration of chewing function. The device is provided sterile.
Clinical Evidence
Bench testing only. Mechanical static and dynamic strength testing performed in accordance with FDA guidance 'Class II Special Controls Guidance Document: Root-form endosseous dental implants and endosseous dental abutments' (May 12, 2004). Biocompatibility testing was also conducted to demonstrate safety.
Technological Characteristics
Material: Ti-6Al-4V. Design: Self-tapping, threaded, screw-type, tapered shape. Surface: Sandblasted/acid-etched. Connection: Internal octagon. Dimensions: 8-15.5mm length, 3.5-5mm diameter. Sterilization: Gamma radiation. System includes various abutment options.
Indications for Use
Indicated for patients requiring surgical placement of endosseous dental implants in the upper or lower jaw to support prosthetic devices (artificial teeth) and restore chewing function.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
# K050956
## FEB 1 6 2006
#### 510(k) Summary
- Sudimplant, SA Submitter: Parc de la Plaine 24, impasse René Couzinet 31500 Toulouse, FRANCE Tel: +335 62 16 71 00 Fax: +335 61 80 84 02
- Mr. Didier Sailhan Contact: Requlatory Affairs Manager E-mail: production@tbridea.com
June 22, 2005 Date:
Device Name: T.B.R.6 ide@ conic
Endosseous dental implant (21 CFR Classification Name: 872.3640) and Endosseous dental implant abutment (21 CFR 872.3630)
## Legally marketed device (predicate devices):
- 3i Osseotite Certain NT (K031475) u
- Bio-Lok Silhouette & Silhouette IC (K032454) L
- Zimmer Dental (formerly Sulzer Dental) Tapered Screw-Vent 트 (K013227, K011028)
#### Description of the device:
Description of the donic/Conic dental implant system consists of selftapping threaded screw-type implants (made from Ti-6AI-4V ) and tapping througou overs with many options such as Ti-6Al-4V abutments, tapered abutments, castable abutments, ball abutments and ceramic/Titapered abutments. The system also include surgical and laboratory ord av abutiments. Implants are double-packaged and provided sterile. Implants surface is roughened to promote osseointegration.
#### Intended use:
The T.B.R.® ide@ conic endosseous dental implant is a device intended to be surgically placed in the bone of the upper or lower jaw arches to to be ourglean; prosthetic devices, such as artificial teeth, and to restore the patient's chewing function.
#### Summary of testing:
Mechanical testing was done in accordance with the FDA guidance "Class Mechaniour controls guidance document:Root-form endosseous dental implants and endosseous dental abutments" issued on May 12, 2004. Implans and ondessendent laboratory showed T.B.R.R.® ide@ conic to have sufficient mechanical static and dynamic strength. Additionnal test
{1}------------------------------------------------
report include biocompatibility testing. Test conclusions demonstrate the report include blocompatistiky tooling."
safety and effectiveness of T.B.R.® ide@ conic implant system.
## Technological characteristics:
The overall design and characteristics of the T.B.R.® ide@ conic/Conic The overall design and charactoriotics of the and as safe, as effective implants are similar to the production as nowledgedly predicate devices.
| | New device | Predicate Devices | | |
|-----------------------------------|-----------------------------------|--------------------------------------------------|-------------------------------------------------|-----------------------------------|
| | | Bio-Lok<br>Silhouette | Zimmer<br>Tapered<br>Screw-Vent | 3i<br>Osseotite<br>Certain NT |
| | Conic | | | |
| 510(k) number | | K032454 | K013227<br>K011028 | K031475 |
| Type | Self-Tapping<br>Threaded<br>Screw | Self-Tapping<br>Threaded<br>Screw | Self-Tapping<br>Threaded<br>Screw | Self-Tapping<br>Threaded<br>Screw |
| Lengths (mm) | 8-15,5 | 8-15 | 8-16 | 8,5-15 |
| Diameters (mm) | 3,5-5 | 3,45-6,5 | 3,5-5,7 | 3,25-6 |
| Connection | Internal<br>Octagon | Internal<br>Connection or<br>External<br>hexagon | Internal<br>Hexagon | Internal<br>Hexagon |
| Shape | Tapered | Tapered | Tapered | Tapered |
| Materials | Ti-6Al-4V | Ti-6Al-4V | Ti-6Al-4V | Ti-6Al-4V |
| Surface | Sandblasted/<br>Acid-etched | Osseo-Lok<br>HA-coated<br>Laser-Lok | Machined /<br>blasted-<br>etched<br>Machined/HA | Acid-etched |
| Sterility | Gamma | Unknown | Unknown | Gamma |
| Abutment<br>containing<br>ceramic | Available | Available | No | No |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of human figures.
FEB 1 6 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Didier Sailhan Regulatory Affairs Manager Sudimplant SA 24 Impasse Rene Couzinet Parc De La Plaine Toulouse, France 31500
Re: K050956
Trade/Device Name: T.B.R.® ide@-conic Regulation Number: 872.3640 Regulation Name: Endosseous dental implant Regulatory Class: II Product Code: DZE Dated: January 20, 2006 Received: January 20, 2006
Dear Ms. Sailhan:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your becater mined the device is substantially equivalent (for the icitications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device microlie conments, or to devices that have been reclassified in accordance with the provisions of Amendinents, or to de rous that the Act (Act) that do not require approval of a premarket the Federal F ood, Drag, the Commay, therefore, market the device, subject to the general approvincianon (Printers) (The general controls provisions of the Act include controls provisions or annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 MA), it may of subject to back add of Federal Regulations, Title 21, Parts 800 to 898. In your device can be found in the over announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Ms. Didier Sailhan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Sengate Y. Michie Davis
Chíu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a handwritten alphanumeric string. The string is "K050956". The characters are written in black ink on a white background.
## Indications for Use
510(k) Number (if known):
Device Name: T.B.R.® ide@ conic
#### Indications For Use:
The T.B.R.® ide@ conic endosseous dental implant is a device intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function.
### (Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
Susan Quares
Case Cit
s
KCB50756
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.