K-ASSAY FERRITIN (2) & FERRITIN CALIBRATOR SET

K050944 · Kamiya Biomedical Co. · DBF · Dec 23, 2005 · Immunology

Device Facts

Record IDK050944
Device NameK-ASSAY FERRITIN (2) & FERRITIN CALIBRATOR SET
ApplicantKamiya Biomedical Co.
Product CodeDBF · Immunology
Decision DateDec 23, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5340
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K050944 · Dec 23, 2005K-ASSAY FERRITIN (2) & FERRITIN CALIBRATOR SETKamiya Biomedical Co.Clinical laboratory patient serum samplesRetrospective clinical patient samples were used to perform a method comparison against the predicate device and to establish reference ranges for the assay.Method comparison; Reference range; Clinical laboratory samples

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective method comparison64 patient serum samples (Asian patients); Sample Size: 64Roche Tina-Quant (predicate device)Correlation with predicate device (R=0.9969)
Retrospective reference range determinationNormal male and female serum samples; Sample Size: 431 (226 male, 205 female)Not applicable for this studyEstablishment of reference ranges (Male: 7-253 ng/mL; Female: 2-110 ng/mL)

Indications for Use

The K-ASSAY® Ferritin (2) assay is an in vitro diagnostic reagent for the quantitative determination of ferritin (an iron-storing protein) in human serum and plasma by immunoturbidimetric assay on the Roche/Hitachi 917 analyzer. Measurements of ferritin aid in the diagnosis of diseases affecting iron overload and iron deficiency anemia. For in vitro diagnostic use. The K-ASSAY® Ferritin Calibrator Set is an in vitro diagnostic reagent for calibration of the K-ASSAY® Ferritin (2) Assay. For in vitro diagnostic use.

Device Story

K-ASSAY® Ferritin (2) is an in vitro diagnostic reagent kit for quantitative ferritin measurement in human serum/plasma. Input: patient serum or plasma samples. Principle: latex-enhanced immunoturbidimetric assay; sample mixed with reagent containing latex particles sensitized with rabbit anti-human ferritin antibody; agglutination occurs proportional to ferritin concentration. Output: absorbance change measured via turbidimetry on Roche/Hitachi 917 analyzer; concentration determined by interpolation from a calibration curve. Used in clinical laboratories by trained personnel. Output aids clinicians in diagnosing iron overload and iron deficiency anemia, potentially benefiting patients through timely identification of iron-related disorders.

Clinical Evidence

No clinical prospective or retrospective studies provided. Performance supported by bench testing: precision (CV < 7%), linearity (R ≥ 0.99, range 2-1000 ng/mL), and analytical specificity (no interference from hemoglobin, bilirubin, lipids, or rheumatoid factor). Method comparison with predicate (n=64) showed correlation coefficient R=0.9969, slope 0.89. Matrix comparison (n=20) confirmed equivalence between serum, EDTA, and sodium heparin plasma.

Technological Characteristics

Latex-enhanced immunoturbidimetric assay. Reagents: reaction buffer and latex particles coated with rabbit anti-human ferritin antibodies. Instrument: Roche/Hitachi 917 analyzer. Calibration: 5-point curve using liquid calibrators. Traceability: 1st WHO International reference Preparation Human Ferritin (BIBSC 80/602).

Indications for Use

Indicated for quantitative determination of ferritin in human serum and plasma to aid in the diagnosis of iron overload and iron deficiency anemia. No specific age or gender contraindications provided.

Regulatory Classification

Identification

A ferritin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the ferritin (an iron-storing protein) in serum and other body fluids. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, such as hemochromatosis (iron overload) and iron deficiency amemia.

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k050944 B. Purpose for Submission: New Device C. Measurand: Ferritin D. Type of Test: Quntitative, latex-enhanced immunoturbidimetric assay E. Applicant: Kamiya Biomedical Company F. Proprietary and Established Names: K-ASSAY® Ferritin (2) and K-ASSAY® Ferritin Calibrator Set G. Regulatory Information: 1. Regulation section: 21 CFR §866.5340, Ferritin Immunological Test System 21 CFR 862.1150, Calibrator 2. Classification: Class II 3. Product code: DBF, Ferritin immunological test system JIT, Calibrator, Secondary 4. Panel: Immunology (82), assay Chemistry (75), Calibrator H. Intended Use: 1. Intended use(s): The K-ASSAY® Ferritin (2) assay is an *in vitro* diagnostic reagent for the quantitative determination of ferritin (an iron-storing protein) in human serum and plasma by immunoturbidimetric assay on the Roche/Hitachi 917 analyzer. Measurements of ferritin aid in the diagnosis of diseases affecting iron overload and iron deficiency anemia. For *in vitro* diagnostic use. The K-ASSAY® Ferritin Calibrator Set is an *in vitro* diagnostic reagent for calibration of the K-ASSAY® Ferritin (2) Assay. For *in vitro* diagnostic use. 2. Indication(s) for use: Same as Intended use. 3. Special conditions for use statement(s): The device is for prescription use only. 4. Special instrument requirements: Roche/Hitachi 917 analyzer I. Device Description: The device K-ASSAY® Ferritin (2) assay is a latex enhanced immunoturbidimetric {1} assay for the quantitative in vitro diagnostic determination of ferritin levels in serum and plasma (EDTA and sodium heparin) samples. The reagent kit contains 2 reagents, reagent 1 (1x18 mL) and reagent 2 (1x18 mL). Reagent 1 is a buffer and reagent 2 is the latex suspension with rabbit anti-human ferritin antibody. The K-ASSAY® Ferritin Calibrator contains four calibrators (A through D) each with a specific assigned value. Calibrator A: 100 ng/mL, Calibrator B: 200 ng/mL, Calibrator C: 500 ng/mL and Calibrator D: 1000 ng/mL J. Substantial Equivalence Information: 1. Predicate device name(s): Roche Tina-quant® (a) Ferritin 2. Predicate 510(k) number(s): k964282 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | K-ASSAY® Ferritin (2) assay | Roche Tina-quant® (a) Ferritin | | Intended Use | Quantitative in vitro diagnostic determination of ferritin | Same | | Methodology | latex Enhanced Immunoturbidimetry | Same | | Uses calibration curve to determine the ferritin levels | Yes | Yes | | Reagents | Reaction buffer and Latex particles coated with rabbit anti-human ferritin antibodies | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Sample type | Serum, and Plasma (Sodium EDTA or Sodium Heparin) | Serum, Plasma (Li-Heparin, citrated or EDTA) | | Instrument | Hitachi 917 | Roche/Hitachi 717/902, 904/911/912/917/MODULAR | | Range of the assay | 2 to 1000 ng/mL | 15 to 800 ng/mL | | Limit of detection | 2 ng/mL | 3 ng/mL (with automatic re-run) | | Calibration | K-Assay® Ferritin Calibrator | c.f.a.s proteins | | Calibrators | liquid | lyophilized | {2} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Calibration | Five point calibration curve | Six-point calibration curve | | Quality Control | Each laboratory is encouraged to run their own controls (normal and abnormal samples) | Control materials are recommended | # K. Standard/Guidance Document Referenced (if applicable): M29-A2 and H3-A2 (NCCLS documents) # L. Test Principle: The K-ASSAY® Ferritin (2) assay is a latex enhanced immunoturbidimetric assay. When a sample containing ferritin is mixed with the reagent 2 which is a suspension of latex particles sensitized with rabbit anti-human ferritin antibody, a clear agglutination occurs. This agglutination is measured by turbidimetry and is detected as an absorbance change, with the magnitude of the change being proportional to the quantity of ferritin in the sample. The actual concentration of ferritin is then determined by interpolation from a calibration curve prepared from calibrators of known ferritin concentration. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: i. Precision: Precision was performed on Hitachi 917 using three samples. The samples were low, medium and high controls (Immuno Q sera II from Denka Seiken, Japan). For calculating within-run precision, the samples were assayed twenty one times on the same day. For calculating between-day precision, samples were assayed in duplicate over twenty different days. The precision study met the acceptance criteria for all three samples which was $&lt; 7\%$ CV. | Within-Run | Mean (ng/mL) | Low (ng/mL) | High (ng/mL) | SD | CV (%) | | --- | --- | --- | --- | --- | --- | | Low control | 14.9 | 13.8 | 15.9 | 0.600 | 4.03 | | Medium control | 100.0 | 98.7 | 101.2 | 0.647 | 0.65 | | High Control | 431.1 | 427.6 | 435.5 | 2.203 | 0.51 | | Between-day | Mean (ng/mL) | Low (ng/mL) | High (ng/mL) | SD | CV (%) | Total Precision (%) | | --- | --- | --- | --- | --- | --- | --- | | Low control | 16.5 | 15.4 | 18.1 | 0.733 | 4.45 | 3.44 | | Medium | 105.2 | 102.7 | 107.6 | 1.522 | 1.45 | 1.54 | | High Control | 428.7 | 421.0 | 432.6 | 2.510 | 0.59 | 0.83 | {3} ii. Lot to lot reproducibility: Lot to lot reproducibility was performed using three different lots of reagents. The samples were assayed in duplicate over twenty different days. The acceptance criteria for 10-50 ng/mL and 51-1000 ng/mL are %CV ≤ 15% and ≤ 10% respectively. | | Mean (ng/mL) | Low (ng/mL) | High (ng/mL) | SD | CV (%) | | --- | --- | --- | --- | --- | --- | | Low control | 16.91 | 15.1 | 18.4 | 0.867 | 5.12 | | Medium control | 105.79 | 103.0 | 108.6 | 1.573 | 1.49 | | High Control | 431.05 | 425.2 | 435.0 | 2.538 | 0.59 | b. Linearity/assay reportable range: Linearity testing was performed using serially diluted in-house controls made from spiked serum pool (Denka Seiken, Japan) with ferritin in low, medium and high range. The acceptance criteria for linearity is ± 10% of theoretical value and R ≥ 0.99 To calculate the 'low range' linearity, a sample with 98 ng/mL ferritin was diluted down to zero. The theoretical measurements of the different levels of dilutions were 88.2, 78.4, 68.6, 58.8, 49, 39.2, 29.4, 19.6, 9.8 and 0.0 ng/mL. The regression equation for this measurement is y=0.9858x-0.0591 with an R=0.9997. To calculate the 'medium range' linearity, a sample with 630 ng/mL ferritin was diluted down to zero. The theoretical measurements of the different levels of dilutions were 567, 504, 441, 378, 315, 252, 189, 126, 63 and 0.0 ng/mL. For this measurement, y=0.9888x + 3.1818 with an R=0.9998. To calculate the 'high range' linearity, a sample with 1000 ng/mL ferritin was diluted down to zero. The theoretical measurements of the different levels of dilutions were 900, 800, 700, 600, 500, 400, 300, 200, 100 and 0.0 ng/mL. For this measurement, y=0.9513x + 18.377 with an R=0.9993. The linear assay range is from 2-1000 ng/mL. c. Traceability, Stability, Expected values (controls, calibrators, or methods): The reference standard for Ferritin calibrator is the 1st WHO International reference Preparation Human Ferritin, established in 1984 (BIBSC code 80/602). Calibrator: Secondary calibrators (In-house master standards) are produced from commercially available human ferritin prepared from human liver and value assigned to the WHO reference standard. The secondary calibrators are used to assign values to the product calibrators. The unopened calibrators are stable for one year from the date of manufacturing and Opened reagents are {4} found to be stable for 2 months if tightly capped and stored at 2-10°C. The stability of the calibration curve is found to be at least 4 weeks. d. Detection limit: The lower limit of detection was determined by assaying the absorbance of serial dilutions of diluted Ferritin Calibrator A (10 ng/mL Ferritin) with saline. The lower detection limit is claimed as 2 ng/mL. This was found to be the lowest level of ferritin that had an absorbance range (mean +2 standard deviation (SD)) above the absorbance range (mean +2 SD) of saline. The 2 ng/mL sample has a within run precision (n=10) of CV = 2.9%. Using a clinical acceptance criteria of CV ≤ 20% at the lower limit of detection, the 2 ng/mL lower limit of detection meets the clinical acceptance criteria. e. Analytical specificity: Interference testing was performed by adding the interfering substances at highest concentration to the base serum. These preparations were diluted serially with saline. Ferritin was measured in these diluted samples and the percent recovery calculated based on ferritin concentration of the serum diluent. The acceptance criterion for the percent recovery is within ± 10% of serum diluent. Based on the acceptance criterion, no significant interference was found with hemoglobin up to 1040 mg/dL, bilirubin F up to 62 mg/dL, bilirubin C up to 62 mg/dL, intra-fat (which contains 10% soybean oil) up to 3% and intra-lipid (which contains 10% soybean oil, different manufacturer) up to 5%. No interference with rheumatoid factor (RF) up to 520 IU/mL. f. Assay cut-off: No assay cut off was provided. 2. Comparison studies: a. Method comparison with predicate device: The Method Comparison study was performed with 64 patient serum samples on the new device and the predicate device, Roche Tina-Quant, both on Roche/Hitachi 917 analyzer. The samples were obtained from clinical laboratories by the manufacturer. The samples were from Asian patients. The samples covered a range from 11.2 to 766.8 ng/mL. The required specification was R value ≥ 0.95. Linear regression analysis showed a slope of 0.89 (95% CI: 0.872 to 0.907) and an intercept of -9.4 (95%CI: -15.846 to -2.96). The correlation coefficient was 0.9969. b. Matrix comparison: Matrix comparison was generated for serum, EDTA and sodium heparin, with 20 samples. The samples covered a range from 7.5 to 161 ng/mL. The required specification was R value ≥ 0.99. Linear regression analysis (serum vs. Na EDTA plasma) showed a slope of 0.989 and an intercept of 0.21 and the correlation coefficient was 0.997. Linear regression analysis (serum vs. Na heparin plasma) showed a slope of 1.0612 and an intercept of -0.5524 and the correlation coefficient was 0.998. 3. Clinical studies: a. Clinical Sensitivity: Not provided b. Clinical specificity: {5} Not provided c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not provided 5. Expected values/Reference range: The package insert provides the reference range which was determined at the manufacturer's site using real samples. Two hundred and twenty six normal male serum samples and two hundred and five normal female serum samples were assayed for ferritin. The reference range for male was found to be 7-253 ng/mL. The reference range for the female was found to be 2-110 ng/mL. Each laboratory is recommended to establish its own expected range. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...