BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-TRIMETHOPRIM-SULFAMETHOXAZOLE 0.0625/1.2-16/304 UG/ML

K050865 · Becton, Dickinson & CO · LON · May 20, 2005 · Microbiology

Device Facts

Record IDK050865
Device NameBD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-TRIMETHOPRIM-SULFAMETHOXAZOLE 0.0625/1.2-16/304 UG/ML
ApplicantBecton, Dickinson & CO
Product CodeLON · Microbiology
Decision DateMay 20, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative and facultative anaerobic bacteria isolates from pure culture for Emero Cram-Regarter accedes and and most Gram-positive bacteria isolates from pure culture belonging to the genera Stanylococcuses , Enterococcus, and Streptococcus. This premarket notification is for the addition of the antimicrobial agent trimethoprim-sulfamethoxazole at concentrations of 0.0625/1.2-16/304 ug/mL to Streptococcus ID/AST or AST only Phoenix panels. Trimethoprimsulfamethoxazole has been shown to be active in vitro against most strains of microrganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Device Story

System performs automated antimicrobial susceptibility testing (AST) for Streptococcus pneumoniae. Input: pure culture bacterial isolates suspended in AST-S broth (cation-adjusted, non-blood). Process: broth microdilution in 136-well polystyrene trays; redox indicator (colorimetric oxidation-reduction) detects growth; instrument incubates at 35°C with continuous monitoring; readings taken every 20 minutes. Output: MIC values and categorical interpretations (SIR). Used in clinical microbiology laboratories; operated by technicians. Software-driven 'EXPERT' system applies CLSI-derived rules to interpret reactions. Results assist clinicians in selecting appropriate antibiotic therapy. Benefits: rapid, standardized, automated susceptibility determination.

Clinical Evidence

Performance evaluated using clinical and challenge isolates across multiple US sites. Comparison against CLSI reference broth microdilution method. Results for trimethoprim-sulfamethoxazole: 906 isolates tested, 95.8% Essential Agreement (EA), 95.3% Category Agreement (CA). Reproducibility study at three sites showed >90% intra-site and ≥95% inter-site reproducibility within ±1 dilution.

Technological Characteristics

Automated colorimetric oxidation-reduction growth-based system. Uses 136-well polystyrene trays with dried reagents. AST-S broth contains Tween 80, pancreatic digest of casein, and peptones. Instrument incubates at 35°C; optical reader monitors color change (blue to colorless) every 20 minutes. Software-based 'EXPERT' system for rule-based interpretation. Standalone instrument.

Indications for Use

Indicated for testing Streptococcus pneumoniae using trimethoprim-sulfamethoxazole at concentrations of 0.0625/1.2 – 16/304 µg/mL. For prescription use only.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ CONFIDENTIAL AND PROPRIETARY MAY 20 2005 # 510(k) SUMMARY | SUBMITTED BY: | Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone: 410-316-4287<br>Fax: 410-316-4499 | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | CONTACT NAME: | Monica Evelyn Giguere,<br>RA Specialist | | DATE PREPARED: | April 1, 2005 | | DEVICE TRADE NAME: | BD PhoenixTM Automated Microbiology System -<br>Trimethoprim-sulfamethoxazole 0.0625/1.2-16/304 µg/mL | | DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation | | DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 | | PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD PhoenixTM<br>Automated Microbiology System with Gatifloxacin (K020323<br>May 23, 2002), Ofloxacin (K020323, April 14, 2002), and<br>Levofloxacin (K020322, March 27, 2002). | | INTENDED USE: | The BD PhoenixTM Automated Microbiology System is<br>intended for the rapid identification and <i>in vitro</i> antimicrobial<br>susceptibility testing of isolates from pure culture of most<br>aerobic and facultative anaerobic Gram-negative and Gram-<br>positive bacteria of human origin. | #### DEVICE DESCRIPTION: The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. For testing Streptococcus species the system includes the following components: - BD Phoenix instrument and software. ● - BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . or AST determinations. - of TLS Forethinations: . - BD Phoenix AST-S Broth used for performing AST tests only. . - BD Phoenix AST-S Indicator solution added to the AST Broth to aid in bacterial growth . determination. {1}------------------------------------------------ The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth. The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations. The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible). ## DEVICE COMPARISON: The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Streptococcus ID/AST or AST only Phoenix panels with this antimicrobial agent. ## SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING: The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. ## Site Reproducibility Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of streptococcal isolates. Each site tested the isolates in triplicate on three different days using one lot of Streptococcus Phoenix panels containing this antimicrobial agent and associated reagents. The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than or equal to 95% for the streptococcal isolates tested. {2}------------------------------------------------ ## Clinical Studies Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Streptococcus Phoenix Panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method. The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, resistant or nonsusceptible). Table 1 summarizes the performance for the isolates tested in this study. #### Performance of BD Phoenix System for Streptococcal Organisms by Drug Table 1: | Antimicrobial | Concentration | EA (n) | EA (%) CA (n) | | I CA (%) | |------------------|---------------------------|--------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|----------| | Trimethoprim- | ' 0.0625/1.2-16/304 μg/mL | 906 | તેરી જેવી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી | 906 | 95.3 | | sulfamethoxazole | | | | | | #### Conclusions Drawn from Substantial Equivalence Studies The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002). {3}------------------------------------------------ # SUMMARY INFORMATION FOR TRIMETHOPRIM-SULFAMETHOXAZOLE Available Range 0.0625/1.2-16/304 µg/mL #### Performance #### Accuracy | Antimicrobial | Concentration | EA (n) | EA (%) | CA (n) | CA (%) | |-----------------------------------|-------------------------|--------|--------|--------|--------| | Trimethoprim-<br>sulfamethoxazole | 0.0625/1.2-16/304 μg/mL | 906 | 95.8 | 906 | 95.3 | #### Reproducibility Testing performed at multiple clinical sites demonstrated 95% reproducibility or greater within ± 1 dilution. #### Breakpoints - CLSI, FDA | THE THE THE LABOUR LA S<br>1<br>And Address of Acres<br>Jrganism | | A<br> | A 10 10 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0<br>----- | |------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <br>1<br>Streptococcus pneumoniae | <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000<br>Vivi<br>Canadian Parcel Products of the Property of Children | 0190 | ≥4/7<br>Acres of the Art | #### Recommended Quality Control Organisms | Quality Control Strain | MIC Range | Source | |-------------------------------------|-----------------------|--------| | Streptococcus pneumoniae ATCC 49619 | 0.12/2.4 – 1/19 µg/mL | CLSI | | | | | {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized caduceus symbol. The caduceus is represented by a series of three curved lines that resemble a stylized staff with intertwined snakes, a symbol often associated with medicine and healthcare. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 MAY 2 0 2005 Ms. Monica Evelyn Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 k050865 Re: Trade/Device Name: BD Phoenix™ Automated Microbiology System Trimethoprim-sulfamethoxazole (0.0625/1.2-16/304 ug/mL) -Streptococcus ID/AST or AST only Phoenix Panels Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: April 1, 2005 Received: April 6, 2005 Dear Ms. Giguere: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass surfal in the enactment date of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may; aroney manage of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is classified dditional controls. Existing major regulations affecting your device it may be subject to Sach as Federal Regulations (CFR), Parts 800 to 895. In addition, FDA ound be louits in The Enouncements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean i trass be actived that I bermination that your device complies with other requirements of the Act that I Dr Has intatutes and regulations administered by other Federal agencies. You must or any I edital statutes and equirements, including, but not limited to: registration and listing (21 compry with an and the Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {5}------------------------------------------------ #### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to Iogally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Sale, a Hogg Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health #### Enclosure {6}------------------------------------------------ Page 1 of 1 510(k) Number: K050865 Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent trimethoprim-sulfamethoxazole (0.0625/1.2-16/304 ug/mL) - Streptococcus ID/AST or AST only Phoenix Panels. Indications for Use: The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative and facultative anaerobic bacteria isolates from pure culture for Emero Cram-Regarter accedes and and most Gram-positive bacteria isolates from pure culture belonging to the genera Stanylococcuses , Enterococcus, and Streptococcus. This premarket notification is for the addition of the antimicrobial agent trimethoprim-sulfamethoxazole at concentrations of 0.0625/1.2-16/304 ug/mL to Streptococcus ID/AST or AST only Phoenix panels. Trimethoprimsulfamethoxazole has been shown to be active in vitro against most strains of microrganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent. # Active In Vitro and in Clinical Infections Against: Streptococcus pneumoniae Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Ludden. Poole Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K050865 BD Diagnostic Systems Becton, Dickinson and Company
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...