SKIPPER AND SKIPPER RACE GUIDEWIRE

K050756 · Invatec Innovative Technologies, S.R.L. · DQX · May 5, 2005 · Cardiovascular

Device Facts

Record IDK050756
Device NameSKIPPER AND SKIPPER RACE GUIDEWIRE
ApplicantInvatec Innovative Technologies, S.R.L.
Product CodeDQX · Cardiovascular
Decision DateMay 5, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Intended Use

The SKIPPER and SKIPPE RACE guidewires are intended for use in the peripheral and coronary vasculature.

Device Story

SKIPPER and SKIPPER RACE are 0.014" diameter stainless steel guidewires; available in 175, 195, and 300cm lengths. Devices feature silicone or hydrophilic coatings. Introduced percutaneously into blood vasculature by clinicians; advanced to diseased areas to facilitate interventional procedures. Used in peripheral and coronary vasculature. Provides mechanical guidance for catheters. Benefits include facilitating navigation through complex vascular anatomy during interventional procedures.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993 Part 1 and FDA Memorandum #G95-1. Performance characteristics compared to predicate device to demonstrate equivalence.

Technological Characteristics

0.014" diameter stainless steel guidewires; 175, 195, and 300cm lengths. Surface features silicone or hydrophilic coatings. Mechanical device; no energy source or software.

Indications for Use

Indicated for use in the coronary and peripheral vasculature for patients requiring percutaneous vascular access.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### MAY - 5 2005 510(k) Summary SKIPPER™ and SKIPPER™ RACE guidewires ## 510(k) Number: KO50756 This 510(k) summary is being submitted in accordance with the requirements of 21 CFR క్తి807.92. | Submitter/Contact Person:<br>Applicant<br>(Manufacturer): | Invatec Innovative Technologies<br>Via Martiri della Libertà, 7<br>25030 Roncadelle (BS) Italy<br>Tel: +39 030 258 93 11<br>Fax: +39 030 258 93 12 | |-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter's Name: | ev3 Inc.<br>4600 Nathan Lane North<br>Plymouth, MN 55442<br>Tel: (763) 398-7000<br>Fax: (763) 398-7200 | | Official Contact: | Melissa Sommerfeld<br>International Regulatory Affairs Specialist<br>ev3 Inc.<br>4600 Nathan Lane North<br>Plymouth, MN 55442<br>Tel: (763) 398-7612<br>Fax: (763) 398-7200<br>msommerfeld@ev3.net | | Summary Preparation Date: | March 22, 2005 | | Device Name and Classification:<br>Trade Name:<br>Common Name/Usual Name:<br>Classification Name:<br>Class: | SKIPPERTM and SKIPPERTM RACE Guidewire<br>Guidewire<br>Catheter, Guidewire<br>Class II, 21 CFR 870.1330 | #### Predicate Device: ev3 Nitrex™ Nitinol Guidewire (K024021, K031864, K040345) This summary is provided in compliance with section 513(I)(3)(A) of the Act and summarizes the safety and effectiveness information contained in this premarket notification submission. #### Device Description: SKIPPER and SKIPPER RACE Guidewires {1}------------------------------------------------ The SKIPPER and SKIPPER RACE are 0.014" diameter stainless steel guidewires in usable lengths of 175, 195, and 300cm. The guidewires are designed for use in interventional surgical iongallo or 170, 1999, it is introduced percutaneously into the blood vasculature and advanced to a diseased area. The SKIPPER and SKIPPER RACE guidewires contain a silicone or a hydrophilic coating. #### Intended Use: The SKIPPER and SKIPPE RACE guidewires are intended for use in the peripheral and coronary vasculature. #### Summary of Testing: Biocompatibility: Biocompatibility testing in accordance with ISO 10993 Part 1, "Biological Evaluation of Medical Devices," 1997(E) and FDA Memorandum #G95-1, "Biological Evaluation of Medical Devices" was provided. Performance Data: Bench testing pertaining to performance characteristics was conducted on the SKIPPER and SKIPPER RACE guidewires and compared to the predicate device testing to demonstrate equivalency. #### Statement of Equivalence: The SKIPPER and SKIPPER RACE guidewires are substantially equivalent to the ev3 Nitrex™ Nitinol Guidewire (K024021, K031864, K040345) in intended use, materials, technological characteristics and performance. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white. MAY - 5 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ev3, Inc. c/o Ms. Melissa Sommerfeld International Regulatory Specialist 4600 Nathan Lane North Plymouth, MN 55442-2920 Re: K050756 SKIPPER and SKIPPER RACE Guidewires Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II (Two) Product Code: DQX Dated: March 22, 2005 Received: March 23, 2005 Dear Ms. Sommerfeld: We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 510(x) prematically equivalent (for the indications for the indications for referenced above and have decimined the ace a vices marketed in interstate comments, or to devices that use stated in the enclosure) to legally many in the Medice Amendments, or to devices that proof to May 28, 1976, the enactions date of the Federal Food, Drug, and Cosmetic have been reclassified in accordance with the provisions of the Federation (PMA) - You may have been reclassified in accordance with the provisions of the Act. The next Act (Act) that do not require approval of a premainconspisions of the Act. The general therefore, market the device, subject to the general controls provisions of devices p therefore, market the device, subject to the general sons and registration, list and advances, good controls provisions of the Act include requirements for annual registerior controls provisions of the Act morder requiritions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) into Entile regulations affecting your device can be may be subject to such additional controls. The 2004 - addition, FDA may woblish may be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may publish found in the Code of Peasa ening your device in the Federal Register. {3}------------------------------------------------ Page 2 -- Ms. Melissa Sommerfeld Please be advised that FDA's issuance of a substantial equivalence determination does not mean I teast be advised made a determination that your device complies with other requirements of the Act or that FDF has made a docemination administered by other Federal agencies. You must comply with ally rederal statutes and regulations annum timited to: registration and listing (21 CFR Part 807); an the Act 3 requirements) instructions practice requirements as set forth in the quality labeling (21 CFR Part 807) good 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and w you've ough finding of substantial equivalence of your device to a legally prematice notification - results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small general invermational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address (201) 115 0977 6da.gov/cdrh/dsma/dsmamain.html Sincerely yours, Dana R. Vines Image /page/3/Picture/6 description: The image shows a signature or symbol that appears to be handwritten. It consists of a combination of curved and angular lines, with a prominent loop-like structure on the left side. To the right of the signature, there are two words stacked vertically, with the top word being "Bra" and the bottom word being "Dir". The overall impression is that of a signed document or a marked piece of paper. A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications For Use 510(k) Number (if known): KO50756 Device Name: SKIPPER™ and SKIPPER™ RACE Guidewires Indications for Use: The SKIPPER and SKIPPER RACE guidewires are intended for use in the coronary and peripheral vasculature. ﻴ Over-The-Counter Use X Prescription Use _ AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Dwayne R. Vachner (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_k 050756 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...