The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
Device Story
System performs automated antimicrobial susceptibility testing (AST) and identification (ID) of Streptococcus species. Input: pure culture bacterial isolates suspended in AST-S broth (cation-adjusted, non-blood) at 0.5 McFarland density. Process: inoculated polystyrene trays with 136 micro-wells containing dried reagents/antibiotics; continuous incubation at 35°C; colorimetric oxidation-reduction redox indicator detects bacterial growth (blue to pink/colorless). Instrument reads wells every 20 minutes; software interprets biochemical reactions and MIC values. Output: ID result and MIC value/category interpretation (SIR). Used in clinical microbiology laboratories; operated by technicians. Healthcare providers use MIC data to guide antibiotic therapy decisions. Benefits: rapid, automated determination of susceptibility to levofloxacin (0.25-16 ug/mL) to inform treatment.
Clinical Evidence
Clinical performance evaluated by comparing Phoenix System results for clinical and challenge isolates against the CLSI reference broth microdilution method. Study included 1055 isolates. Metrics: Essential Agreement (EA) and Category Agreement (CA). Reproducibility assessed at three sites with streptococcal isolates, showing >90% intra-site and >95% inter-site reproducibility.
Technological Characteristics
Materials: molded polystyrene trays. Principle: broth microdilution with colorimetric oxidation-reduction (redox) indicator. Energy: automated incubation at 35°C. Connectivity: networked instrument. Software: rule-based 'EXPERT' system using CLSI guidelines. Inoculum: 0.5 McFarland standard. Detection: automated optical reading of color change every 20 minutes.
Indications for Use
Indicated for in vitro rapid identification and antimicrobial susceptibility testing (MIC) of Streptococcus species from pure culture. Prescription use only.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® System (PMA No. N50510)
BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321)
BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)
Related Devices
K073589 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-DAPTOMYCIN (0.0313-16 UG/ML) STEPTOCOCCUS ID/AST OR AST ONLY PHOENIX PANELS · Becton, Dickinson & CO · Mar 5, 2008
K052250 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - TOBRAMYCIN (GP) 0.5-16UG/ML · Becton, Dickinson & CO · Sep 30, 2005
K042932 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PIPERACILLIN-TAZOBACTAM 0.5/4-128/4 UG/ML · Becton, Dickinson & CO · Dec 2, 2004
K050881 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - TETRACYCLINE (STREP) 0.0625 - 16 UG/ML · Becton, Dickinson & CO · May 12, 2005
K051109 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - ERYTHROMYCIN (STREP) 0.0156 - 4UG/ML · Becton, Dickinson & CO · Jun 15, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
| 510(k) SUMMARY | APR 2 9 2005<br>K050747 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTED BY: | Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone: 410-316-4287<br>Fax: 410-316-4499 |
| CONTACT NAME: | Monica Evelyn Giguere,<br>RA Specialist |
| DATE PREPARED: | March 18, 2005 |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System -<br>Levofloxacin 0.25-16 µg/mL |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 |
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™<br>Automated Microbiology System with Gatifloxacin (K020321,<br>May 23, 2002), Ofloxacin (K020323, April 14, 2002), and<br>Levofloxacin (K020322, March 27, 2002). |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System is<br>intended for the rapid identification and in vitro antimicrobial<br>susceptibility testing of isolates from pure culture of most<br>aerobic and facultative anaerobic Gram-negative and Gram-<br>positive bacteria of human origin. |
### DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. For testing streptococcal species the system includes the following components:
- BD Phoenix instrument and software. ●
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents ● or AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
- BD Phoenix AST-S Broth used for performing AST tests only. .
- BD Phoenix AST-S Indicator solution added to the AST Broth to aid in bacterial growth � determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or nonsusceptible).
# DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Streptococcus ID/AST or AST only Phoenix panels with this antimicrobial agent.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003.
### Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of streptococcal isolates. Each site tested the isolates in triplicate on three different days using one lot of Streptococcus Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the streptococcal isolates tested.
BD Diagnostic Systems Becton, Dickinson and Company
{2}------------------------------------------------
# Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Streptococcus Phoenix Panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, resistant or nonsusceptible).
Table 1 summarizes the performance for the isolates tested in this study.
#### Performance of BD Phoenix System for Streptococcal Organisms by Drug Table 1:
| THE REAL PROPERTY OF CHARACT THE COLUMN OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLEAN<br>PRODUCT OF COLLECTION OF COLLECTION<br>Antimicrobial | LAND L BOLL CONTRACT THE FILM I 1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Concentration | LALE SECTION COLL OF C. OPEN CO. OF C.<br>B-8000 - Book Tournal Comments of Children Come of<br>EA (n)<br>PROPERTY LIGHT LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEA | Company Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come Come C<br>10/<br>10 A<br>11/2017<br>Canadian Call A C . A | (n<br>CA '<br>worker from annual or<br>The later an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an- | ())<br>C<br>70 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| 1000 000 000 000 000 000 000 000 000 000 0000 0000 0000 0000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000<br>and would and the mille as<br>Carreral<br>Levolloxacin | THE COLLECT COLLEGION COLL<br>STATIST I W . I<br>. 0 75 1 7<br>10 ug/mL | 1055<br>1 VW | C | 1055 | 00 -<br>سنڌ جي جي جي |
# Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
# SUMMARY INFORMATION FOR LEVOFLOXACIN Available Range 0.25-16 µg/mL
### Performance
### Accuracy
| Antimicrobial | Concentration | EA (n) | EA (%) | CA (n) | CA (%) |
|---------------|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|----------|--------|
| Levotloxacın | 0 75-17 | 1055 | 07 | 1055 | . QQ ~ |
| | 16 ug/mL | A 2011 11/2 11/2017 11/2017 11/11/2017 11/11/2017 11/11/2017 11/11/2017 11/11/2017 11/11/2017 11/11/2017 11/11/2017 11/11/2017 11/11/2017 11/11/2017 11/11/2017 11/11/2017 11/ | 11.6 | پ پ پ مم | 1100 |
### Reproducibility
Testing performed at multiple clinical sites demonstrated 95% reproducibility or greater within ± 1 dilution.
## Breakpoints - CLSI, FDA
| Organism | S | I | R |
|-----------------------------------------------------------|---------|---|---------|
| Streptococcus pneumoniae | $\leq2$ | 4 | $\geq8$ |
| Streptococcus spp. Other than<br>Streptococcus pneumoniae | $\leq2$ | 4 | $\geq8$ |
# Recommended Quality Control Organisms
| Quality Control Strain | MIC Range | Source |
|-------------------------------------|---------------------------------------------|--------|
| Streptococcus pneumoniae ATCC 49619 | 0.5-2 $ μg/mL $ | NCCLS |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three horizontal lines above it, representing the agency's mission to promote health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol. The logo is in black and white.
# APR 2 9 2005
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Monica Evelyn Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
k050747 Re:
R0507 . .
Trade/Device Name: BD PhoenixTM Automated Microbiology System Levofloxacin (0.25-16 µg/mL) - Streptococcus ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: March 18, 2005 Received: March 22, 2005
Dear Ms. Giguere:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bector of the device is substantially equivalent (for the indications felerenced above and have acterimes as a microse devices marketed in interstate for use stated in the encrosule) is regars and ment date of the Medical Device Amendments, or to comments proc to May 20, 1978, the occordance with the provisions of the Federal Food, Drug, devices that have occh recuire approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval application (PM allu Cosmetic Act (1101) that do not requesable to the general controls provisions of the Act. The r ou may, therefore, market and act include requirements for annual registration, listing of general controls provisions or ractice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (860 as 10) throls. Existing major regulations affecting your device It may be subject to Such additional come Regulations (CFR), Parts 800 to 895. In addition, FDA can be found in Thic 21, Cours ents concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that 1 Dri 3 issuated or woulder complies with other requirements of the Act that I DA has made a doted regulations administered by other Federal agencies. You must of ally if ederal statutes and regulance and limited to: registration and listing (21 Comply with the Free 5 require Parts 801 and 809); and good manufacturing practice CFR Part 807), labeling (21 CF quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ogin manismig of substantial equivalence of your device to a legally premarket notification: "The PDT Intamig cification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, of f If you desire specific information assuments of your device, please contact the Office of In of questions on the promotion and Safety at (240)276-0484. Also, please note the ville Diagnostic Device Device by reference to premarket notification" (21 CFR Part 807.97). Tegulation other general information on your responsibilities under the Act from the You may of amall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Sally a Form
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
510(k) Number: K050742
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Device Name: " BDTHochix" - Tureptococcus ID/AST or AST only Phoenix Panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of The BD Phoenix™ Automated witchology Uystem is mation (MIC) of most Gram-negative aerobice aetobic and antimicrobial susceptibility of inimal inincitory obiteriaceae and Non-Enterobacteriaceae
facultative anaerobic bacteria isolates from pure culture for Execuences of Starbul iacultative anaeroone bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.
Enterococcus and Streplococcus.
This premarket notification of the antimicrobial agent levolfoxacin at concern has been have to This premarket notification is for the addition of the untinue of the other his been shown to be 0.25-16 ug/mL to Streptococcus DIAST of AST only I notime parties "Every Lesseribed in the FDA-approved package insert for this antimicrobial agent.
# Active In Vitro and in Clinical Infections Against:
Streptococcus pneumoniae (including penicillin-resistant strains) Streptococcus pyogenes
#### Active In Vitro Against:
Streptococcus (Group C) Streptococcus (Group G) Streptococcus agalactiae Streptococcus viridans group (Viridans group streptococci, Streptococcus (Group F), and Streptococcus milleri)
Over-the-Counter Use
Prescription Use (Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K050747
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.