TEGRESS IMPLANT NEEDLE

K050688 · C.R. Bard, Inc. · FBK · Apr 6, 2005 · Gastroenterology, Urology

Device Facts

Record IDK050688
Device NameTEGRESS IMPLANT NEEDLE
ApplicantC.R. Bard, Inc.
Product CodeFBK · Gastroenterology, Urology
Decision DateApr 6, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Tegress™ Implant Needles are accessories for currently marketed endoscopes to allow delivery of injectable materials into tissues during an endoscopic procedure.

Device Story

The Tegress™ Implant Needle is a transurethral flexible needle accessory designed for use with existing endoscopes. It features a stainless steel non-coring needle tip, a stabilizing cannula, and a full-length needle guard. The device includes a semi-rigid catheter/tubing assembly and a working channel seal. It is used by clinicians during endoscopic procedures to facilitate the delivery of injectable materials into patient tissues. The device is compatible with DMSO.

Clinical Evidence

Bench testing only.

Technological Characteristics

Transurethral flexible needle with stabilizing cannula; stainless steel non-coring needle tip; semi-rigid catheter/tubing; working channel seal. Compatible with DMSO. Mechanical device; no software or energy source.

Indications for Use

Indicated for the delivery of injectable materials into tissues during endoscopic procedures. Intended for prescription use.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K050688 ## APR 6 2005 ## 510 (k) Summary of Safety and EFFECTIVENESS INFORMATION This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. - A. Submitter's Information: | Submitter's Name: | C. R. Bard, Inc., Urological Division | |-------------------------|---------------------------------------------------| | Address: | 8195 Industrial Blvd.<br>Covington, Georgia 30014 | | Contact Person: | Terri Morris | | Contact Person's Phone: | (770) 784-6774 | | Contact Person's Fax: | (770) 784-6419 | | Date of Preparation: | March 10, 2005 | - B. Device Name: | Trade Name: | Tegress™ Implant Needle | |----------------------|-------------------------| | Common / Usual Name: | Endoscopic Needle | | Classification Name: | Endoscope Accessory | - C. Predicate Device Name: | Trade Name: | Genyx Flexible Injection Needle<br>#K001146 | |-------------|---------------------------------------------| |-------------|---------------------------------------------| - D. Device Description: The Tegress™ Implant Needle is an accessory for currently marketed endoscopes and is used The Tegress - Implant Noodle is an account for tissue during an endoscopic procedure. The Tegress™ Implant Needle is a Transurethral Flexible Needle with Stabilizing Cannula that The Tegress ™ Implant Needle is a Transcritive catheter portion with longitundinal consists of a stainless steel non-coring needle at the tip. A full-length needle guard protects the stripe, and a stamess steel norFormig nedule at the species of a semi-rigid catheter/tubing and The Transurethral Needle and Stabilizing Cannula with Working Working Channel Seal. Working Channel Geal. - The "Transarenial and are compatible with DMSO. - E. Intended Use: The Tegress™ Implant Needles are accessories for currently marketed endoscopes to allow The Tegress - Implant Nobanoota to tissues during an endoscopic procedure. - F. Technological Characteristics Summary: The subject Tegress™ Implant Needle has the same intended use, design and fundamental scientific technology as the predicate device. - G. Performance Data Summary: The Tegress™ Implant Needle referenced in this submission is held to the same design, The Tegress™ Informance specifications as those needles currently manufactured by manufacture, and performance specification and validation activities for the modifications to the Tegress™ Implant Needle were conducted. {2}------------------------------------------------ Image /page/1/Picture/2 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged in a circular fashion around the eagle image. APR 6 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Terri Morris Regulatory Affairs Specialist C. R. Bard, Inc. Bard Urological Division 8195 Industrial Blvd. COVINGTON GA 30014-2655 Re: K050688 Trade/Device Name: Tegress™ Implant Needle Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FBK Dated: March 16, 2005 Received: March 17, 2005 Dear Ms. Morris: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) This letter witi anow you'ls ought mating of substantial equivalence of your device to a legally prematics notification - The Persons in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, picase note the regulation entired, formation on your responsibilities under the Act from the 001:37). Tou may ooually of the ational and Consumer Assistance at its toll-free number (800) DNSION of 301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsmamain.html Sincerely yours, Nancy C. brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Bard Urological Division, C.R. Bard, Inc. Tegress™ Implant Needle Premarket Notification [510(k)] KOSCLE 88 Indications for Use Statement 1.3 510(k) Number (if known): __ K050688 Device Name: ____Tegress™ Implant Needle Indications for Use: The Tegress™ Implant Needles are accessories for currently marketed endoscopes to allow The Tegress - Implant Neodios art accessed. delivery of injectable materials into tissues during an endoscopic procedure. > (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) concurrence of cdrh, office of device evaluation (ode) Over-The-Counter Use __ OR Prescription Use (Per 21 CFR 801.109) (Optional Format 1/2/96) Ylaucic bugden (Division Sign-Off) Division of Reproductive, Abdor and Radiological De 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...