HEMOSIL FACTOR II DEFICIENT PLASMA

K050661 · Instrumentation Laboratory CO · GJT · May 2, 2005 · Hematology

Device Facts

Record IDK050661
Device NameHEMOSIL FACTOR II DEFICIENT PLASMA
ApplicantInstrumentation Laboratory CO
Product CodeGJT · Hematology
Decision DateMay 2, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7290
Device ClassClass 2

Indications for Use

HemosIL Factor II Deficient Plasma is human plasma immunodepleted of factor II and intended for the in vitro diagnostic quantitative determination of factor II activity in citrated plasma, based on the prothrombin time (PT) assay, on IL Coagulation and ELECTRA Systems.

Device Story

HemosIL Factor II Deficient Plasma is a lyophilized human plasma reagent immunodepleted of Factor II (residual activity ≤1%). Used on IL Coagulation and ELECTRA systems, the reagent is mixed with patient citrated plasma and Owren’s Buffer to perform a modified prothrombin time (PT) clotting assay. Clotting time is inversely proportional to Factor II concentration; results are compared against a standard curve derived from normal pooled or reference plasma. The device assists clinicians in diagnosing Factor II deficiencies, which may cause bleeding tendencies or result from conditions like liver disease or DIC. The assay is performed in clinical laboratory settings by trained technicians.

Clinical Evidence

Bench testing only. Precision/reproducibility evaluated across multiple analyzers (ACL and ELECTRA series) per NCCLS EP5-T2, with %CV within acceptance criteria (Normal <10%, Special Test Level 2 <17%). Method comparison studies (n=60-61) showed high correlation (r=0.9855-0.9954) with predicate devices. Additional studies confirmed equivalent performance across different PT reagents.

Technological Characteristics

Lyophilized human plasma immunoadsorbed on a column of antibody specific for human Factor II; contains buffer and stabilizers. Principle: modified prothrombin time (PT) clotting assay. Storage: 2-8°C. Analyzers: IL Coagulation and ELECTRA systems.

Indications for Use

Indicated for quantitative determination of Factor II activity in citrated plasma for patients with suspected Factor II deficiency (congenital or acquired).

Predicate Devices

Submission Summary (Full Text)

{1} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K050661 B. Purpose for Submission: To seek clearance of a new assay C. Measurand: Factor II D. Type of Test: Clotting E. Applicant: Instrumentation Laboratory Co. F. Proprietary and Established Names: HemosIL Factor II Deficient Plasma G. Regulatory Information: 1. Regulation section: 21 CFR 864.7290 2. Classification: Class II 3. Product code: GJT 4. Panel: 81 Hematology {2} H. Intended Use: 1. Intended use(s): HemosIL Factor II Deficient Plasma is human plasma immunodepleted of Factor II for the quantitative determination of Factor II activity in citrated plasma, based on the prothrombin time (PT) assay, on IL Coagulation and ELECTRA™ Systems. 2. Indication(s) for use: 3. Special conditions for use statement(s): 4. Special instrument requirements: I. Device Description: The HemosIL Factor II Deficient Plasma kit consists of 5-1 mL vials of lyophilized human plasma that has been artificially depleted of Factor II and containing buffer and stabilizers. Residual Factor II activity is less than or equal to 1%, all other coagulation factors have normal levels. J. Substantial Equivalence Information: 1. Predicate device name(s): HemosIL Factor II Deficient Plasma on ACL Family of Analyzers Hemoliance Factor II Deficient Plasma on ELECTRA Series Analyzers 2. Predicate 510(k) number(s): K002400 K900133 3. Comparison with predicate: {3} | Similarities | | | | | --- | --- | --- | --- | | Item | Device | Predicate 1 | Predicate 2 | | Sample type | Citrated plasma | same | same | | Storage | 2-8°C | same | same | | Composition | Lyophilized human plasma immunoadsorbed on a column of antibody specific for human factor II | Same | same | | Intended use | Quantitative determination of Factor II activity based on the PT assay | Same | same | | Differences | | | | | --- | --- | --- | --- | | Item | Device | Predicate 1 | Predicate 2 | | Instrumentation | IL ACL and ELECTRA | IL Coagulation Systems | ELECTRA | | Manufacturer | IL, Orangeburg, NY | Trinity Biotech, Wicklow, Ireland | Trinity Biotech Wicklow, Ireland | # K. Standard/Guidance Document Referenced (if applicable): NCCLS EP5-T2, User Evaluation of Precision Performance of Clinical Chemistry Devices # L. Test Principle: Factor II (prothrombin), is a vitamin K-dependent plasma protein that is synthesized in the liver and circulates in plasma as an inactive molecule. During coagulation FII is converted to the active form, factor IIa (thrombin), by Factor Xa. Congenital deficiency of factor II is a very rate inherited disorder that causes in general a mild to moderate bleeding tendency. Factor II deficiencies may also be acquired secondarily due to other diseases such as hyperfibrinolysis, DIC and liver disease. {4} Factor II activity in a patient’s plasma is determined by performing a modified prothrombin time test. Patient plasma is diluted with Owren’s Buffer and mixed with a substrate plasma deficient in factor II. The mixture is assayed by the PT clotting procedure. The clotting time is dependent upon the concentration of factor II in the patient plasma. The clotting time is compared to a standard curve prepared prior to assay from dilutions of normal pooled plasma or commercial reference plasma mixed with the same factor II deficient plasma. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: a. Precision/Reproducibility: Precision was performed on an ACL Advance (K002400), the ACL Futura Coagulation Analyzer (K951891), the ACL 9000 (K000053), ACL 300 (K881367), ACL 3000 (K912087), ACL TOP (K033414), ELECTRA 1400C (K944227), ELECTRA 1600 (K931206), and the ELECTRA 1800C (K962664) using HemosIL RecombiPlasTin (K012768) PT reagent and 2 levels of controls: HemosIL Normal Control (K021023) and the HemosIL Special Test Control Level 2 (K040359). On the ACL 9000, ACL Futura, ACL TOP, and ELECTRA 1600C, each control was run in replicates of four twice a day for 10 days (n=80). On the ACL 300, ACL 3000, ACL Advance, ELECTRA 1400C, and ELECTRA 1800C, each control was run in replicates of four twice a day for five days (n=40). Within run, between run and total %CV was calculated per NCCLS EP5-T2. %CV were within the acceptance criteria (Normal &lt;10%, Special Test Level 2 &lt;17%). b. Linearity/assay reportable range: c. Traceability, Stability, Expected values (controls, calibrators, or methods): d. Detection limit: e. Analytical specificity: f. Assay cut-off: {5} 5 2. Comparison studies: a. Method comparison with predicate device: An in-house study was conducted using 60 patient citrated plasma samples with the HemosIL Factor II Deficient Plasma versus the predicate devices. Each sample was run in duplicate using HemosIL RecombiPlasTin reagent. | New HemosIL Factor II Deficient Plasma vs. Predicate HemosIL Factor II Deficient Plasma on ACL Family | | | | | --- | --- | --- | --- | | IL System | Slope | Intercept | r | | ACL 300 | 1.0357 | 1.0196 | 0.9921 | | ACL 6000 | 1.0464 | -2.1396 | 0.9954 | | ACL 9000 | 1.0458 | -1.6123 | 0.9953 | | ACL TOP | 1.0582 | -4.6831 | 0.9855 | | New HemosIL Factor II Deficient Plasma vs. Predicate Hemoliance Factor II Deficient Plasma on ELECTRA | | | | | E1600C | 1.063 | -0.2821 | 0.9917 | A field study was conducted using 61 patient citrated plasma samples with the HemosIL Factor II Deficient Plasma versus the predicate HemosIL Factor II Deficient Plasma on an ACL Futura. HemosIL RecombiPlasTin reagent was used in the testing. (y = 1.0602x -3.6113, r = 0.9946) b. Matrix comparison: 3. Clinical studies: a. Clinical Sensitivity: b. Clinical specificity: {6} c. Other clinical supportive data (when a. and b. are not applicable): The following studies were presented to demonstrate the equivalent performance of HemosIL Factor II Deficient Plasma with various PT reagents on representative IL instruments: 1. 49 patient samples spanning in range from 7 – 171% were test on an ACL Advance using HemosIL RecombiPlasTin (reference) vs. HemosIL PT-Fibrinogen HS (test) (K923921) reagent. Results are statistically similar (y = 0.8916 x -18.676, r= 0.9653). 2. The same testing pool was also run on an ACL Advance using HemosIL RecombiPlasTin (reference) vs. HemosIL PT-Fibrinogen HS Plus (test) (K933252) reagent. Results are statistically similar (y = 1.1194 x -13.560, r= 0.9846). 3. The same testing pool was also run on an ACL Advance using HemosIL RecombiPlasTin (reference) vs. HemosIL PT-Fibrinogen Recombinant (K981479) reagent. Results are statistically similar (y = 1.0551 x - 1.2247, r= 0.9796). 4. 40 patient samples spanning in range from 15 – 215% were test on an ELECTRA using HemosIL RecombiPlasTin (reference) vs. Hemoliance Brain Thromboplastin reagent. Results are statistically similar (y = 0.9650 x -2.1467, r= 0.9633). Results demonstrated that the performance of HemosIL Factor II Deficient Plasma is statistically equivalent between PT reagents. 4. Clinical cut-off: 5. Expected values/Reference range: N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...