← Product Code [HAW](/productcode/HAW) · K050651

# MEDTRONIC ORTHOPAEDIC TRAUMA APPLICATION (K050651)

_Medtronic Navigation, Inc. · HAW · Jul 27, 2005 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K050651

## Device Facts

- **Applicant:** Medtronic Navigation, Inc.
- **Product Code:** [HAW](/productcode/HAW.md)
- **Decision Date:** Jul 27, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Orthopaedic Trauma Application is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The Orthopaedic Trauma Application is indicated for trauma procedures in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, a long bone, pelvis or vertebra can be identified relative to a CT or MR based model, digitized landmarks or fluoroscopy images of the anatomy. - Acetabular, femoral and tibia fractures - Guide Wire placement - Implant/Hardware Removal - Intertrochanteric Fractures - Intramedullary Nailing - Long-bone Fracture Fixation - Long-bone Fracture Reduction - Pelvic Fracture Fixation - Pelvic Fracture Reduction - Screw/Implant Placement - Tibial, Femoral and Acetabular Osteotomies

## Device Story

The Medtronic Orthopaedic Trauma Application is a software-based stereotaxic instrument used to assist surgeons in orthopaedic trauma procedures. It processes inputs from CT/MR models, digitized anatomical landmarks, or fluoroscopy images to provide real-time navigation. The system enables surgeons to precisely locate anatomical structures during open or percutaneous procedures, such as fracture fixation, reduction, and hardware placement. Used in the operating room, the device provides visual guidance to the surgeon, facilitating accurate instrument and implant positioning. By integrating imaging data with real-time tracking, the system aids in complex orthopaedic interventions, potentially improving surgical accuracy and patient outcomes.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and risk analysis.

## Technological Characteristics

Software-based stereotaxic navigation system. Integrates with existing surgical navigation hardware. Inputs include CT/MR imaging, digitized landmarks, and fluoroscopy. Operates as an aid for anatomical localization during orthopaedic surgery. Connectivity via standard surgical navigation platform interfaces.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Medtronic Orthopaedic Hip ([K021980](/device/K021980.md))
- FluoroNav (K99014)
- Orthopaedic Application ([K012448](/device/K012448.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# nary of Safety and Effectiveness Medtronic Orthopaedic Trauma Application

#### Manufacturer l.

Medtronic Navigation, Inc. 826 Coal Creek Circle Louisville, Colorado 80027 USA Telephone Number: (720) 890-3217 Fax Number: (720) 890-3517

#### Contact 11.

Tina Dreiling Associate Regulatory Affairs Specialist Medtronic Navigation, Inc.

#### Product Name / Classification lll.

- Stereotaxic instrument Common Name: Instrument, Stereotaxic Classification Name: Medtronic Orthopaedic Trauma Application Trade Name: Stereotaxic instrument - Class II as described in 21 CFR § 882.4560 Product Code: HAW
#### Date Summary Submitted IV.

March 11, 2005

#### Description of Device Modification V.

Description of Device Modification software combines existing FluoroNav (K99014) The Meditorile Orthopaedic Tradina Applications used to assist surgeons with the stabilization and repair of orthopaedic fractures.

#### Substantial Equivalence VI.

Substantial Equivalence
The Medtronic Orthopaedic Trauma Application is substantially equivalent to the substantially The Meditonic Orthopaedic Hip (K021980) applications and is also substantially the Huordvav (K5902 14) and Ofthopacition (K012448). As required by risk equivalent to the Brancab Vooter Helenchvities performed by designated individuals and the results demonstrated substantial equivalence.

#### Indications for Use VII.

Indications for Use
The Orthopaedic Trauma Application is intended as an aid for precisely locating anatomical The Orthopaeuic Trauma Application is in the Orthopaedic Trauma Application is indicated for trauma procedures in which the use of stereotactic surgery may be appropriate, and indicated to trauma proceduces in which the ass of ever the skull, a long bone, pelvis or vertebra where reletence to a nglo andonifour stransmission of alleritized landmarks or fluoroscopy images of the anatomy

- Acetabular, femoral and tibia 1 fractures
- Guide Wire placement .
- Implant/Hardware Removal 트
- Intertrochanteric Fractures I
- Intramedullary Nailing 동
- Long-bone Fracture Fixation I
- Long-bone Fracture . Reduction
- Pelvic Fracture Fixation I
- Pelvic Fracture Reduction 제
- Screw/Implant Placement .
- Tibial. Femoral and ■ Acetabular Osteotomies

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo consists of a stylized eagle with three curved lines representing its wings, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged in a circular pattern around the eagle. The eagle is black, and the text is also black.

JUL 2 7 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Tina Dreiling Associate Regulatory Affairs Specialist Medtronic Navigation, Inc. 826 Coal Creek Circle Louisville, Colorado 80027

Re: K050651

Trade/Device Name: Medtronic Orthopaedic Trauma Application Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: July 5, 2005 Received: July 6, 2005

Dear Ms. Dreiling:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Tina Dreiling

This letter will allow you to begin marketing your device as described in your Section 510(k) I his icter with anow you to ocgin maining of substantial equivalence of your device to a legally prematics notineation. The Pro mixessification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acriously - J at (240) 276-0115. Also, please note the regulation entitled, Colliact the Office of Compullion in (21 transmit) (21CFR Part 807.97). You may obtain of Small other gelleral information on your response Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Stupt Edwards

Mark N. Melkerson, MS Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications For Use

510(k) Number (if known): k050651

### Device Name: Medtronic Orthopaedic Trauma Application

Indications For Use: The Orthopaedic Trauma Application is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The Orthopaedic Trauma Application is indicated for trauma procedures in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, a long bone, pelvis or vertebra can be identified relative to a CT or MR based model, digitized landmarks or fluoroscopy images of the anatomy

- Acetabular, femoral and tibia fractures 이
- Guide Wire placement ●
- t Implant/Hardware Removal
- Intertrochanteric Fractures ●
- Intramedullary Nailing 해
- Long-bone Fracture Fixation .
- Long-bone Fracture Reduction .
- Pelvic Fracture Fixation I
- Pelvic Fracture Reduction 체
- Screw/Implant Placement 트
- Tibial, Femoral and Acetabular Osteotomies 1

Prescription Use
(Part 21 CFR 801 Subpart D) X

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(ODE)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)

| Concurrence of CDRH, Office of Device Evaluation |
|--------------------------------------------------|
|--------------------------------------------------|

| (Division Sign-Off) |
|---------------------|
|---------------------|

| Division of General, Restorative and Neurological Devices |
|-----------------------------------------------------------|
|-----------------------------------------------------------|

Page 1 of 1

| 510(k) Number | K050651 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda.innolitics.com/device/K050651](https://fda.innolitics.com/device/K050651)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
