DIADEXUS PLAC TEST

K050523 · Diadexus, Inc. · NOE · Jun 15, 2005 · Immunology

Device Facts

Record IDK050523
Device NameDIADEXUS PLAC TEST
ApplicantDiadexus, Inc.
Product CodeNOE · Immunology
Decision DateJun 15, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5600
Device ClassClass 2

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Ischemic stroke risk prediction——Hazard ratio 1.81 (95% CI 1.22–2.69, p=0.0034) for highest tertile (>420 ng/mL) vs lowest tertile (<310 ng/mL) in multivariate model.——Case-cohort study of 956 patients (194 ischemic stroke cases, 762 controls) from a multi-center epidemiologic study.—

Indications for Use

The diaDexus PLAC™ test is an enzyme immunoassay for the quantitative determination of Lp-PLA2 (lipoprotein-associated phospholipase A2) in human plasma, to be used in conjunction with clinical evaluation and patient risk assessment as an aid in predicting risk for coronary heart disease, and ischemic stroke associated with atherosclerosis.

Device Story

The diaDexus PLAC™ test is a sandwich enzyme immunoassay (EIA) used to quantify Lp-PLA2 levels in human plasma. The kit utilizes monoclonal anti-Lp-PLA2 antibodies (2C10 and 4B4) coated on microwell strips and conjugated to horseradish peroxidase. The device is operated by laboratory personnel in a clinical setting. Input consists of human plasma samples; the assay produces an absorbance signal measured via spectrophotometry. A standard curve is generated using provided calibrators, and sample concentrations are interpolated using point-to-point curve fitting software. Results assist clinicians in assessing patient risk for coronary heart disease and ischemic stroke associated with atherosclerosis, providing prognostic information beyond traditional risk factors.

Clinical Evidence

Clinical evidence provided via a case-cohort study of 956 patients (47-69 years old) from a multi-center epidemiologic study. Participants were followed for ~9 years for development of CHD or ischemic stroke. Analysis included 194 ischemic stroke cases and 762 controls. Cox regression models (univariate and multivariate adjusted for demographics, CHD, and other prognostic factors) demonstrated that Lp-PLA2 levels >420 ng/mL (highest tertile) are a predictor of risk for ischemic stroke associated with atherosclerosis compared to levels <310 ng/mL (lowest tertile). Hazard ratios ranged from 1.62 to 2.09 depending on the model.

Technological Characteristics

Sandwich-type enzyme immunoassay; mouse monoclonal antibody-coated microtiter stripwells; horseradish peroxidase-labeled antibody; colorimetric substrate; spectrophotometric detection at 450 nm; requires microtiter plate reader (bandwidth ≤10 nm, optical density range ≥3).

Indications for Use

Indicated for use in conjunction with clinical evaluation and patient risk assessment to aid in predicting risk for coronary heart disease and ischemic stroke associated with atherosclerosis in adult patients.

Regulatory Classification

Identification

A low-density lipoprotein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the low-density lipoprotein in serum and other body fluids. Measurement of low-density lipoprotein in serum may aid in the diagnosis of disorders of lipid (fat) metabolism and help to identify young persons at risk from cardiovascular diseases.

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k050523 B. Purpose for Submission: Clearance of new intended use for previously cleared device C. Measurand: Lipoprotein-Associated Phospholipase A₂ (Lp-PLA₂) D. Type of Test: Quantitative Enzyme Immunoassay E. Applicant: diaDexus, Inc. F. Proprietary and Established Names: diaDexus PLAC™ Test G. Regulatory Information: 1. Regulation section: 21 CFR §866.5600, Low-density lipoprotein immunological test system 2. Classification: Class II 3. Product code: NOE, test, system, immunoassay, lipoprotein-associated phospholipase A₂ 4. Panel: Immunology (82) {1} H. Intended Use: 1. Intended use(s): See indications for use below. 2. Indication(s) for use: The diaDexus PLAC™ Test is an enzyme immunoassay for the quantitative determination of Lp-PLA2 (Lipoprotein-Associated Phospholipase A₂) in human plasma, to be used in conjunction with clinical evaluation and patient risk assessment as an aid in predicting risk for coronary heart disease and ischemic stroke associated with atherosclerosis. 3. Special conditions for use statement(s): For professional use only To be used in conjunction with clinical evaluation and patient risk assessment. 4. Special instrument requirements: Microtiter plate reader with a bandwidth of 10 nm or less and an optical density range of 3 or greater at 450nm I. Device Description: The diaDexus PLAC™ Test contains mouse monoclonal antibody coated microtiter stripwells (96), calibration materials (6), buffers, sample diluent, conjugate, substrate and stopping solution. J. Substantial Equivalence Information: 1. Predicate device name(s): diaDexus PLAC™ Test 2. Predicate 510(k) number(s): k030477, k040101 3. Comparison with predicate: The device is exactly the same as the predicate device except that a new intended use has been included (prediction of the risk for ischemic stroke associated with atherosclerosis). {2} K. Standard/Guidance Document Referenced (if applicable): None were referenced. L. Test Principle: The diaDexus PLAC™ Test is a sandwich-type enzyme immunoassay that uses two anti-Lp-PLA₂ monoclonal antibodies. One antibody is immobilized on the bottom of microtiter wells. Sample and buffer are incubated in the wells and the unbound fractions are washed off. The second antibody (labeled with horseradish peroxidase) is added and incubated to bind with the Lp-PLA₂ bound to the surface of the well. The excess is washed away, and a colorimetric substrate is added and allowed to react with the peroxidase to form a blue color. A stop solution converts the resulting color from blue to yellow which is measure spectrophotometrically at 450 nm. The measured absorbance is directly proportional to the amount of Lp-PLA₂ in the sample. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Analytical performance was established under k030477. b. Linearity/assay reportable range: Analytical performance was established under k030477. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Analytical performance was established under k030477. d. Detection limit: Analytical performance was established under k030477. e. Analytical specificity: Analytical performance was established under k030477. f. Assay cut-off: See clinical studies in 3 c. below. {3} 2. Comparison studies: a. Method comparison with predicate device: Not applicable. b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. See clinical information below. b. Clinical specificity: Not applicable. See clinical information below. c. Other clinical supportive data (when a. and b. are not applicable): A study was conducted by the manufacturer using 956 samples from patients who were part of a large multi-center epidemiologic study, sponsored by the National Heart, Lung and Blood Institute. Patients, 47 to 69 years of age, were free from coronary heart disease or stroke at the start of the study and followed for the development of coronary heart disease (CHD) or stroke for ~9 years. This was a case-cohort study where samples from all the ischemic stroke cases associated with atherosclerosis (194) were tested together with 762 appropriately matched participants without CHD or stroke (controls). Cox regression models were used to evaluate the association of Lp-PLA2 and ischemic stroke associated with atherosclerosis, using a univariate analysis (Model 1), a univariate analysis adjusted for demographics (Model 2), and a multivariate model adjusted for demographics and other prognostic factors (Model 3). The same cutpoints of Lp-PLA2 as were used in the original CHD risk indication were investigated in this study (420 and 310 ng/mL, the 67th and 33rd percentiles, respectively) to divide study participants into tertiles. The hazard ratios of the Cox regression analyses demonstrated that Lp-PLA2 may be used as a predictor of risk for ischemic stroke associated with atherosclerosis for the highest tertile when compared to the lowest tertile of Lp-PLA2. See tables below for additional information. 4 {4} Results of Cox Regression Models – Ischemic Stroke Associated with Atherosclerosis Analysis Population | Lp-PLA_{2} (ng/mL) | Lp-PLA_{2} Risk Ratio (95 % CI, p-value)* | | | | --- | --- | --- | --- | | | < 310 | 310 - 420 | < 420 | | # ischemic stroke cases associated with atherosclerosis / total subjects | 47 / 283 (16.6 %) | 44 / 305 (14.4 %) | 103 / 368 (28 %) | | Model 1 | 1.0 | 0.85 (0.57 – 1.29, p=0.45) | 1.79 (1.27 – 2.52, p=0.0010) | | Model 2 | 1.0 | 0.89 (0.59 – 1.35, p=0.58) | 2.09 (1.46 – 3.01, p=0.0001) | | Model 3 | 1.0 | 0.89 (0.58 – 1.36, p=0.59) | 1.81 (1.22 – 2.69, p=0.0034) | *the lowest tertile with Lp-PLA<sub>2</sub> values &lt;310 ng/mL is used as a reference group Model 1: univariate analysis Model 2: adjusted for age, race, and gender Model 3: adjusted for age, race, gender, diabetes, LDL, HDL, blood pressure, smoking, BMI, and CRP Further analyses were performed to determine if Lp-PLA<sub>2</sub> was predictive of ischemic stroke associated with atherosclerosis across the complete range of systolic blood pressure (SBP) in the population, and to determine whether blood pressure and Lp-PLA<sub>2</sub> were additive in assessing risk for ischemic stroke. Systolic blood pressure tertile cutpoints were assigned by the 33rd and 67th percentiles of the population (113 and 130 mm Hg, respectively). The study population was divided into the low, mid, and high range (1st, 2nd, and 3rd tertile) of SBP and the low, mid, and high range of Lp-PLA<sub>2</sub>. The relative risk of each group was compared to the risk of events associated with the group in the 1st tertile of the two variables. Risk Ratios of Ischemic Stroke Associated with Atherosclerosis: Additive Effects of Lp-PLA<sub>2</sub> and Systolic Blood Pressure | Lp-PLA_{2} (ng/mL) | | | | | | --- | --- | --- | --- | --- | | | | <310 | 310-420 | >420 | | SBP (mm Hg) | # ischemic stroke cases associated with atherosclerosis/total subjects in category | 47/283 (16.6%) | 44/305 (14.4%) | 103/368 (28.0%) | | <113 | 29/270 (10.7%) | 1.00 | 1.00 (0.34-2.99, p=0.99) | 2.45 (0.96-6.25, p=0.06) | | 113-130 | 60/337 (17.8%) | 2.49 (0.98-6.31, p=0.05) | 1.42 (0.53-3.84, p=0.49) | 3.72 (1.56-8.91, p=0.003) | | >130 | 105/349 (30.1%) | 3.67 (1.50-8.99, p=0.004) | 3.99 (1.64-9.69, p=0.002) | 6.40 (2.75-14.86, p<0.0001) | {5} CHD was a strong predictor of ischemic stroke associated with atherosclerosis in all three models (see Tables below). However, the third tertile of $\mathrm{Lp - PLA_2}$ remained a predictor even when CHD was incorporated in all three models with risk ratios decreasing by less than 0.2 units in all cases. Results of Cox Regression Models – Ischemic Stroke Associated with Atherosclerosis Analysis Population (additional adjustment for CHD included in all 3 models) | Lp-PLA2 (ng/mL) | Lp-PLA2 Risk Ratio (95% CI, p-value)* | | | | --- | --- | --- | --- | | | <310 | 310-420 | >420 | | # ischemic stroke cases associated with atherosclerosis / total subjects | 47/283 (16.6%) | 44/305 (14.4%) | 103/368 (28.0%) | | Model 1 | 1.0 | 0.82 (0.55-1.24, p=0.36) | 1.62 (1.14-2.29, p=0.0069) | | Model 2 | 1.0 | 0.89 (0.59-1.34, p=0.57) | 2.03 (1.41-2.93, p=0.0001) | | Model 3 | 1.0 | 0.86 (0.56-1.31, p=0.48) | 1.75 (1.18-2.60, p=0.0057) | * The lowest tertile with Lp-PLA2 values $&lt; {310}\mathrm{\;{ng}}/\mathrm{{mL}}$ is used as the reference group Model 1: adjusted for CHD Model 2: adjusted for age, race, gender, CHD Model 3: adjusted for age, race, gender, diabetes, LDL, HDL, blood pressure, smoking, BMI, CRP, CHD Results of Cox Regression Models –Ischemic Stroke Associated with Atherosclerosis Analysis Population Predictive power of CHD diagnosis | | CHD Risk Ratio (95% CI, p-value)* | | | --- | --- | --- | | CHD Status | No CHD | CHD | | # ischemic stroke cases associated with atherosclerosis / total subjects | 146/849 (17.2%) | 48/107 (44.9%) | | Model 1 | 1.0 | 2.94 (2.11-4.09, p<0.0001) | | Model 2 | 1.0 | 2.90 (2.07-4.06, p<0.0001) | | Model 3 | 1.0 | 2.26 (1.58-3.23, p<0.0001) | * "No CHD" is used as the reference group Model 1: adjusted for Lp-PLA2 Model 2: adjusted for age, race, gender, Lp-PLA2 Model 3: adjusted for age, race, gender, diabetes, LDL, HDL, blood pressure, smoking, BMI, CRP, Lp-PLA2 {6} 4. Clinical cut-off: The hazard ratios of the Cox regression analyses demonstrated that Lp-PLA2 may be used as a predictor of risk for ischemic stroke associated with atherosclerosis for the highest tertile (&gt;420 ng/mL) when compared to the lowest tertile (&lt;310 ng/mL) of Lp-PLA2. See tables above for additional information. 5. Expected values/Reference range: See k030477. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...