QMS VANCOMYCIN

K050419 · Seradyn, Inc. · LEH · Apr 1, 2005 · Clinical Toxicology

Device Facts

Record IDK050419
Device NameQMS VANCOMYCIN
ApplicantSeradyn, Inc.
Product CodeLEH · Clinical Toxicology
Decision DateApr 1, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3950
Device ClassClass 2

Indications for Use

The QMS® Vancomycin assay is intended for the quantitative determination of vancomycin in human serum or plasma on the Hitachi 717 analyzer. The results obtained are used in the diagnosis and treatment of vancomycin overdose and in monitoring levels of vancomycin to ensure appropriate therapy.

Device Story

QMS® Vancomycin is a homogeneous particle-enhanced turbidimetric immunoassay for quantitative measurement of vancomycin in human serum or plasma. It utilizes a two-reagent system: anti-vancomycin antibody (R1) and vancomycin-coated microparticles (R2). Principle of operation involves competition between drug in patient sample and drug on microparticles for antibody binding sites; in absence of drug, microparticles agglutinate. Presence of vancomycin inhibits agglutination, slowing absorbance change rate measured photometrically. The assay is performed on the Hitachi 717 analyzer in a clinical laboratory setting by trained personnel. Output is a concentration value (0-100 µg/mL) used by clinicians to adjust vancomycin dosing, ensuring therapeutic efficacy while avoiding toxicity. Benefits include rapid, automated monitoring of antibiotic levels to guide clinical decision-making.

Clinical Evidence

Bench testing only. Precision evaluated per NCCLS EP5-A (N=80 per level, total CV 5.12-8.84%). Accuracy/recovery (mean 99.61%) and linearity (mean 100.17%) confirmed. Analytical sensitivity (LDD) 0.46 µg/mL. Specificity testing showed <5% cross-reactivity with metabolite CDP-1 and <0.3% with common co-administered drugs. Interference testing (endogenous substances, HAMA) showed <10% error. Method comparison against predicate (N=146) yielded slope 1.031, y-intercept 1.115, R2=0.970.

Technological Characteristics

Homogeneous particle-enhanced turbidimetric immunoassay. Reagents: mouse monoclonal anti-vancomycin antibody (R1) and vancomycin-coated microparticles (R2). Energy source: photometric measurement on Hitachi 717 analyzer. Storage: 2-8 °C. Standards: NCCLS EP5-A, EP6-P, EP7-P, EP9-A.

Indications for Use

Indicated for the quantitative determination of Vancomycin in human serum or plasma to assist in the diagnosis and treatment of Vancomycin overdose and to monitor therapeutic levels in patients receiving Vancomycin therapy.

Regulatory Classification

Identification

A vancomycin test system is a device intended to measure vancomycin, an antibiotic drug, in serum. Measurements obtained by this device are used in the diagnosis and treatment of vancomycin overdose and in monitoring the level of vancomycin to ensure appropriate therapy.

Predicate Devices

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k050419 B. Purpose for Submission: New 510(k) C. Analyte: Vancomycin D. Type of Test: Quantitative Immunoassay E. Applicant: Seradyn, Inc. F. Proprietary and Established Names: Seradyn QMS® Vancomycin G. Regulatory Information: 1. Regulation section: 21 CFR 862.3950 Vancomycin Test System 2. Classification: Class II 3. Product Code: LEH 4. Panel: Toxicology (91) H. Intended Use: 1. Intended use(s): Refer to Indications for use. 2. Indication(s) for use: The QMS® Vancomycin assay is intended for the quantitative determination of Vancomycin in human serum or plasma on the Hitachi 717 analyzer. The results obtained are used in the diagnosis and treatment of Vancomycin overdose and in monitoring levels of Vancomycin to ensure appropriate therapy. 3. Special condition for use statement(s): N/A 4. Special instrument Requirements: The device is for use on the Roche Hitachi 717 Analyzer. {2} Page 2 of 11 I. Device Description: The QMS® Vancomycin assay consists of separately packaged reagents (R1 and R2) and calibrators. R1 is the antibody reagent that contains Vancomycin mouse monoclonal antibody (mouse ascites) in a buffer as stabilizer and sodium azide as a preservative. R2 is Vancomycin micro-reagent that contains Vancomycin coated microparticles in buffer containing a surfactant and sodium azide as a preservative. Both reagents are supplied in liquid form. The calibrators that are included with the assay were previously cleared in k872644 as the Innofluor Vancomycin Calibrator Set. The calibrators consist of six levels of one ml in liquid frozen form. The human serum used for the already cleared calibrators were tested by FDA approved methods and confirmed to be non-reactive for Hepatitis B surface Ag (HBsAg), HIV Type 1 and 2 Antibodies and Hepatitis C antibodies. J. Substantial Equivalence Information: 1. Predicate device name(s): TDX Vancomycin 2. Predicate K number(s): k813218 3. Comparison with predicate: Both devices are for measurement of the same analyte(s) in the same matrix, utilize the same test methodology and cutoff concentration. Both are for use on automated analyzers. | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | The QMS® Vancomycin assay is intended for the quantitative determination of Vancomycin in human serum or plasma on the Hitachi 717 analyzer. The results obtained are used in the diagnosis and treatment of Vancomycin overdose and in monitoring levels of Vancomycin to ensure appropriate therapy. | TDx Vancomycin is a reagent system for the quantitative measurement of serum or plasma Vancomycin levels in the management of patients receiving Vancomycin therapy. | | Analyte | Vancomycin | Vancomycin | | Matrix | Serum or Plasma | Serum or Plasma | | Storage | 2-8 °C | 2-8 °C | {3} Page 3 of 11 | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Methodology | Particle Enhanced Turbidimetric Immunoassay | Fluorescence Polarization (FPIA) | | Assay Range | 2.5 μg/L -100 μg/mL | 2.0 μg/L -100 μg/mL | ## K. Standard/Guidance Document Referenced (if applicable): The sponsor referenced the following guidance document(s) in their submission: NCCLS EP5-A: Evaluation of Precision Performance of Clinical Chemistry devices. NCCLS EP6-P: Evaluation of the Linearity of Quantitative Analytical Methods. NCCLS EP7-P: Interference Testing in Clinical Chemistry. NCCLS EP9-A: Method Comparison and Bias Estimation Using Patient Samples ## L. Test Principle: The QMS Vancomycin assay is a homogeneous particle-enhanced turbidimetric immunoassay. The assay is based on competition between drug in the sample and drug coated onto a microparticle for antibody binding sites of the Vancomycin antibody reagent. The vancomycin-coated microparticle reagent (R2) is rapidly agglutinated in the presence of the anti-vancomycin antibody reagent (R1) and in the absence of any competing drug in the sample. The rate of absorbance change is measured photometrically and is inversely proportional to the rate of agglutination of the particles. When a sample containing vancomycin is added, the agglutination reaction is partially inhibited, slowing down the rate of absorbance change. A concentration-dependent classic agglutination inhibition curve can be obtained with maximum rate of agglutination at the lowest vancomycin concentration and the lowest agglutination rate at the highest vancomycin concentration. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: a. Precision/Reproducibility: Precision was conducted and assessed according to NCCLS EP5-A. Two separate lots of the QMS Vancomycin reagents were run in duplicates for 20 non-consecutive days on the Hitachi 717 analyzers with QMS Vancomycin Calibrators and a commercially available control. Additionally, within run precision was conducted by rerunning the samples again after 2 hours. The acceptance criterion was a total CV of less than 10%. The results for both lots are shown in the two tables below. {4} | Precision for Lot 1 | | | Within Run | | Between Run | | Between Day | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | N | Mean | SD | CV% | SD | CV% | SD | CV% | SD | CV% | | Low Control | 80 | 7.6 μg/mL | 0.27 | 3.59 | 0.43 | 5.70 | 0.43 | 5.72 | 0.70 | 8.84 | | Mid Control | 80 | 20.8 μg/mL | 0.51 | 2.44 | 0.69 | 3.30 | 0.97 | 4.66 | 1.29 | 6.21 | | High Control | 80 | 33.7 μg/mL | 0.80 | 2.37 | 1.19 | 3.54 | 0.95 | 2.83 | 1.72 | 5.12 | | Precision for Lot 2 | | | Within Run | | Between Run | | Between Day | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | N | Mean | SD | CV% | SD | CV% | SD | CV% | SD | CV% | | Low Control | 80 | 6.3 μg/mL | 0.39 | 6.21 | 0.21 | 3.36 | 0.40 | 6.40 | 0.60 | 9.53 | | Mid Control | 80 | 17.8 μg/mL | 0.65 | 3.63 | 0.86 | 4.81 | 0.00 | 0.00 | 1.08 | 6.03 | | High Control | 80 | 31.5 μg/mL | 1.15 | 3.66 | 1.36 | 4.34 | 0.00 | 0.00 | 1.79 | 5.67 | Commercially available control levels | Controls | Vancomycin (μg/mL) | | --- | --- | | Level 1 | 6.9 (5.5-8.30) | | Level 2 | 22 (16-25) | | Level 3 | 35 (28-42) | Accuracy by recovery was determined by spiking human serum that was negative for vancomycin with USP traceable vancomycin that covered the assay range. The samples were analyzed in triplicate with two lots of the QMS Vancomycin assay. Linear regression for lot 1 was $y = 0.955x + 0.529$ ( $R^2 = 0.9958$ ) and for lot 2 was $y = 0.9632x + 1.153$ ( $R^2 = 0.9965$ ). # b. Linearity/assay reportable range: Linearity was conducted and assessed according to NCCLS EP6-A. The QMS Vancomycin Assay is designed to quantitate patient samples between $2.5\mu \mathrm{g / mL}$ to $100~\mu \mathrm{g / mL}$ . Calibrator A $(0\mu \mathrm{g / mL})$ was used as the diluent and Calibrator F $(100\mu \mathrm{g / mL})$ was diluted to 75.0, 37.5, 17.5, 7.5 and $2.5\mu \mathrm{g / mL}$ midpoint samples. The samples were run in triplicate with two lots of the assay. Linear regression plots of USP Vancomycin against recovered vancomycin for Lot 1 was $y = 1.000x - 0.6113$ ( $R^2 = 0.9997$ ) and for Lot 2 was $y = 1.0115x - 0.3769$ ( $R^2 = 0.9998$ ). {5} Page 5 of 11 c. Traceability (controls, calibrators, or method): The QMS Vancomycin assay does not include vancomycin controls or calibrators. Calibrators were cleared under k872644. The controls used in with this assay are commercially available. The sponsor is the current owner and manufacturer of the Innofluor line and continues to manufacture the calibrators using the same validated procedures established at the time of clearance for k872644. d. Detection limit: The sensitivity, or lowest detectable dose (LDD), is defined by the sponsor as the lowest concentration of analyte detectable from zero with 95% confidence when performing the assay. Sensitivity was established by determining the average concentration of analyte present in 20 replicates of the negative Calibrator A with two lots of the QMS Vancomycin assay. The LDD was calculated using the following formula: $$ \mathrm{LDD} = \frac{2 \times (\mathrm{SD\ mAbs\ of\ Zero\ Cal})}{\left(\mathrm{mAbs\ of\ Zero\ Cal} - \mathrm{mAbs\ of\ } 1^{\mathrm{st}}\ \mathrm{non-zero\ cal}\right)} \times \left( \mathrm{Conc\ of\ } 1^{\mathrm{st}}\ \mathrm{non-zero\ cal} \right) $$ | | Lot 1 | Lot 2 | | --- | --- | --- | | N | 20 | 20 | | Avg | 1.01 μg/mL | 0.22 μg/mL | | SD | 0.27 μg/mL | 0.19 μg/mL | | LDD | 0.53 μg/mL | 0.38 μg/mL | The average LDD was 0.46 μg/mL, which supports a claim of 0.55 μg/mL. e. Analytical specificity: Percent cross-reactivity was assessed according to the NCCLS EP7-A. The QMS Vancomycin assay utilizes a mouse derived (ascites) vancomycin monoclonal antibody directed against vancomycin. Specificity was conducted by determining cross-reactivity to the metabolites of vancomycin. In vivo, vancomycin degrades into its metabolite, CDP-1 (crystalline degradation product-1). CDP-1 has two forms, CDP-1 major and CDP-1 minor. The sponsor states difficulty in isolation of the major and minor atropisomers, isomers which arise by restricted rotation around a single bond, in sufficient {6} Page 6 of 11 amounts for a cross reactivity study, so a combined cross reactivity study of CDP-1 major: minor isomers was performed to obtain the total cross-reactivity of the metabolites. The metabolites at 100 µg/mL were spiked into serum with vancomycin at 25 µg/mL. The control sample was vancomycin in human serum with no metabolites. The control and spiked sample were assayed (n=5) using two lots of QMS Vancomycin assay in triplicate. Cross-reactivity was determined using the following equation: $$ = \frac{\text{(Conc. of vancomycin in spiked sample - conc. of vancomycin in control)}}{\text{Conc. of Cross-reactant}} \times 100 $$ If the numerator was less than a SD of the control, percent cross-reactivity was reported as not detectable. See the chart below. | Lot 1 | N | Mean | SD | CV | Numerator | % Cross Reactivity | | --- | --- | --- | --- | --- | --- | --- | | Control | 5 | 25.30 | 0.90 | 3.56 | - | - | | 100 µg/mL CDP-1 | 5 | 29.94 | 0.21 | 0.70 | 4.64 | 4.65 | | Lot 2 | N | Mean | SD | CV | Numerator | % Cross Reactivity | | Control | 5 | 24.98 | 1.32 | 5.28 | - | - | | 100 µg/mL CDP-1 | 5 | 29.95 | 0.86 | 2.87 | 4.61 | 4.61 | The result show that CDP-1 showed cross reactivity with the QMS Vancomycin assay that was less than 5%. The cross-reactivity for Teicoplanin (a structurally similar compound) was conducted. Four concentrations of Teicoplanin was spiked into human serum containing 25 µg/mL vancomycin and tested with two lots of the QMS Vancomycin assay. The result of the cross-reactivity data is shown in the chart below and showed no cross-reactivity with the QMS Vancomycin assay. | Teicoplanin Concentration (µg/mL) | % Cross-Reactivity | | --- | --- | | 100 | 0.68 | | 50 | 0.29 | | 25 | 0.20 | | 10 | 0.03 | Cross Reactivity of drugs that are commonly co-administered with vancomycin were tested in normal human serum. Cross-reactants were analyzed at 500 µg/mL in a vancomycin spiked serum pool at 25 µg/mL. Cross-reactant stock concentrations were either at 100x or 10x of the final concentration tested. Test samples were prepared {7} for each solvent system by combining 99 volumes of the stock analyte solution with 1 volume of $100\mathrm{x}$ or $10\mathrm{x}$ cross-reactant stock. Control aliquots were prepared for each solvent system by combining 99 volumes of the stock analyte solution with 1 volume of normal human whole blood. The test and control samples were run in triplicate and the cross-reactivities were calculated using the following formula. Percent Cross-Reactivity $= \left( {\left( {\mathrm{{Da}} - \mathrm{{Dt}}}\right) /\mathrm{C}}\right) \times {100}$ $\mathrm{Dt} =$ average observed concentration of the control solution $\mathrm{Da} =$ average of the observed concentration of the cross-reactant test solution $\mathrm{C} =$ concentration at which the cross-reactant is tested. The results of the cross-reactivity study for one lot is shown in the chart below. | Cross-reactant Drug | Avg. | SD | CV | Da-Dt | % Cross-reactivity | | --- | --- | --- | --- | --- | --- | | Acetaminophen | 24.90 | 0.58 | 2.33 | 0.24 | 0.05 | | Amikacin | 24.55 | 0.29 | 1.18 | -0.12 | -0.02 | | Amphotericin B | 27.53 | 0.08 | 0.29 | -0.37 | -0.07 | | Ampicillin | 24.83 | 0.20 | 0.82 | 0.40 | 0.08 | | Bendroflumethiazide | 24.40 | 0.11 | 0.47 | 0.23 | 0.05 | | Caffeine | 25.06 | 0.31 | 1.22 | 0.22 | 0.04 | | Carbenicillin | 25.45 | 0.20 | 0.79 | 0.32 | 0.06 | | Cefamandole Nafate | 23.83 | 0.16 | 0.67 | -0.83 | -0.17 | | Cefazolin | 24.47 | 0.18 | 0.74 | -0.19 | -0.04 | | Cephalexin | 24.56 | 0.42 | 1.69 | -0.54 | -0.11 | | Cephalosporin | 24.83 | 0.28 | 1.12 | -0.27 | -0.05 | | Cephalothin | 24.40 | 0.27 | 1.10 | -0.70 | -0.14 | | Chloramphenicol | 24.80 | 0.37 | 1.49 | 0.14 | 0.03 | | Chlorothiazide | 25.45 | 0.72 | 2.83 | 0.79 | 0.16 | | Clindamycin | 27.64 | 0.21 | 0.76 | 0.43 | 0.09 | | Erythromycin | 24.10 | 0.73 | 3.03 | -0.56 | -0.11 | | Ethacrynic Acid | 24.63 | 0.30 | 1.22 | -0.04 | -0.01 | | Ethambutol | 24.99 | 0.45 | 1.79 | 0.32 | 0.06 | | Fluorocytosine, 5- | 24.38 | 0.92 | 3.76 | -0.72 | -0.14 | | Furosemide | 24.89 | 0.25 | 0.99 | 0.23 | 0.05 | | Fusidic Acid | 24.90 | 0.05 | 0.20 | 0.23 | 0.05 | | Gentamicin | 27.22 | 0.88 | 3.24 | 0.04 | 0.01 | | Hydrochlorothiazide | 24.95 | 0.20 | 0.80 | 0.29 | 0.06 | | Ibuprofen | 24.31 | 0.26 | 1.08 | -0.35 | -0.07 | | Isoniazid | 24.53 | 0.34 | 1.38 | -0.24 | -0.05 | | Kanamycin-A | 25.08 | 0.87 | 3.48 | 0.23 | 0.05 | | Kanamycin-B | 25.20 | 0.50 | 2.00 | 0.35 | 0.07 | | Lincomycin | 25.14 | 0.55 | 2.19 | 0.48 | .010 | | Methotrexate | 27.13 | 0.07 | 0.27 | -0.55 | -0.11 | | Methyl Prednisolone | 26.99 | 0.14 | 0.52 | -0.22 | -0.04 | | Nalidixic Acid | 24.25 | 0.25 | 1.05 | -0.85 | -0.17 | | Naproxen | 24.53 | 0.27 | 1.10 | -0.14 | -0.03 | | Neomycin Sulfate | 25.46 | 0.10 | 0.40 | 0.62 | 0.12 | | Oxycodone | 24.90 | 0.10 | 0.40 | 0.62 | 0.12 | | Oxyphenidate | 24.90 | 0.10 | 0.40 | 0.62 | 0.12 | | Oxyphenidate + 2- | 24.90 | 0.10 | 0.40 | 0.62 | 0.12 | | Oxyphenidate + 2- | 24.90 | 0.10 | 0.40 | 0.62 | 0.12 | {8} | Cross-reactant Drug | Avg. | SD | CV | Da-Dt | % Cross-reactivity | | --- | --- | --- | --- | --- | --- | | Niacin | 24.84 | 0.60 | 2.43 | 0.17 | 0.03 | | Nitrofurantoin | 24.81 | 0.28 | 1.15 | 0.15 | 0.03 | | Oxytetracycline | 24.29 | 0.19 | 0.78 | -0.37 | -0.07 | | Penicillin G | 24.60 | 0.13 | 0.54 | -0.07 | -0.01 | | Penicillin V | 27.25 | 0.73 | 2.69 | 0.06 | 0.01 | | Phenacetin | 24.50 | 0.44 | 1.79 | -0.21 | -0.04 | | Prednisolone | 24.75 | 0.29 | 1.15 | 0.09 | 0.02 | | Prednisone | 24.56 | 0.85 | 3.46 | 0.15 | 0.03 | | Rifampicin | 25.04 | 0.47 | 1.88 | 0.37 | 0.07 | | Salicylic Acid | 25.16 | 0.74 | 2.96 | 0.49 | 0.10 | | Sisomicin | 27.52 | 0.37 | 1.33 | 0.34 | 0.07 | | Spectinomycin | 24.42 | 0.33 | 1.34 | -0.24 | -0.05 | | Sulfadiazine | 24.50 | 0.64 | 2.60 | -0.16 | -0.03 | | Sulfamethoxazole | 24.68 | 0.23 | 0.93 | 0.02 | 0.00 | | Sulfisoxazole | 24.56 | 0.74 | 3.01 | -0.10 | -0.02 | | Tetracycline | 24.85 | 0.22 | 0.89 | 0.19 | 0.04 | | Tobramycin | 24.05 | 0.28 | 1.18 | -0.79 | -0.16 | | Trimethoprim | 24.45 | 0.41 | 1.69 | -0.21 | -0.04 | *ND Data not obtained for sample; two replicates only Interference testing was conducted in triplicates with two lots of the QMS Vancomycin assay according to the NCCLS protocol EP7-A. For the following interfering substances at the concentrations listed, the QMS Vancomycin assay showed less than a $10\%$ error in detecting vancomycin. | Interfering Substances | Interference Concentration | N | Vancomycin (μg/mL) | Percent Recovery | | --- | --- | --- | --- | --- | | Albumin | 10 g/dL | 3 | 24.92 | 93.64 | | Bilirubin | 70 mg/dL | 3 | 27.00 | 100.07 | | Cholesterol | 500 mg/dL | 3 | 25.97 | 97.58 | | IgG | 6 g/dL | 3 | 25.90 | 97.34 | | Hemoglobin | 1150 mg/dL | 3 | 21.18 | 91.64 | | HAMA Type-1 | Normal human level | 3 | 28.27 | 105.31 | | Hama-Type-2 | Normal human level | 3 | 28.55 | 103.91 | | Heparin | 500 USP units/mL | 3 | 26.46 | 99.44 | | Triglyceride | 1000 mg/dL | 3 | 9.41 | 92.62 | | RF | 1100 IU/mL | 3 | 6.70 | 93.23 | f. Assay cut-off: N/A {9} Page 9 of 11 2. Comparison studies: a. Method comparison with predicate device: Method comparison was conducted according to the NCCLS EP9-A to determine equivalence between the QMS Vancomycin and Abbott TDx Vancomycin assays. Clinical serum samples (146 samples for lot 1 and 148 samples for lot 2) were tested with the predicate and the device that ranged from 0.04 to 100 µg/mL vancomycin. The relationship between vancomycin concentrations derived using each device was evaluated using Passing-Bablok Linear regression. Mean values for the TDx reference methods were plotted against those for the QMS on Hitachi 717. The results for both lots are shown in the table below. | | Lot 1 | Lot 2 | | --- | --- | --- | | N | 146 | 148 | | Slope | 1.03 (95% CI of 0.997 to 1.062) | 1.05 (95% CI of 1.017 to 1.092) | | Y Intercept | 1.11 (95% CI of 0.58 to 1.55) | 1.17 (95% CI of 0.56 to 1.73) | | R² | 0.970 | 0.964 | b. Matrix comparison: The sponsor demonstrated equivalence between serum and plasma by conducting the following studies: performance characteristics of the assay for serum and plasma studies, interferences when using different types of collection tubes and sample stability at 2 to 8 °C and -20 °C storage conditions. Blood were drawn from six healthy volunteers (no vancomycin therapy) into plasma (4 types) and serum (3 types) tubes. Plasma (K2 and K3 EDTA, Lithium and Sodium Heparin) and serum (SST, plastic tube with clot activator and a plastic tube with no additive) samples were divided into two equal parts. Vancomycin was spiked at low (7.5 µg/mL) and high (30 µg/mL) concentrations into the plasma and serum pools and analyzed on a Hitachi 717 analyzer. The results are shown in the chart below and so now significant difference between the recovery of vancomycin in serum or plasma. The collection tube evaluated showed no adverse effects on the vancomycin recovery and supports the claim for both serum and plasma samples with the QMS Vancomycin assay. {10} Page 10 of 11 | | Low Vancomycin (7.5 μg/ml) | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | K2 EDTA Tube | K3 EDTA Tube | Li Heparin Tube | Na Heparin Tube | Serum Separator Tube | Serum Tube with Clot Activator | Serum Tube (No Additive) Control | | Rep 1 | 7.08 | 5.67 | 6.36 | 5.96 | 6.08 | 5.63 | 6.18 | | Rep 2 | 6.69 | 5.13 | 6.12 | 5.60 | 5.69 | 5.59 | 6.14 | | Rep 3 | 6.80 | 5.37 | 6.12 | 5.50 | 5.92 | 5.61 | 6.00 | | Average | 6.86 | 5.39 | 6.20 | 5.69 | 5.90 | 5.61 | 6.11 | | Stdev | 0.20 | 0.27 | 0.14 | 0.24 | 0.20 | 0.02 | 0.09 | | % CV | 2.93 | 5.02 | 2.23 | 4.25 | 3.32 | 0.36 | 1.55 | | % Recovery | 112.28 | 88.22 | 101.53 | 93.13 | 96.56 | 91.82 | 100.00 | | | High Vancomycin (30 μg/ml) | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | K2 EDTA Tube | K3 EDTA Tube | Li Heparin Tube | Na Heparin Tube | Serum Separator Tube | Serum Tube with Clot Activator | Serum Tube (No Additive) Control | | Rep 1 | 23.10 | 21.69 | 22.11 | 22.24 | 22.63 | 23.23 | 22.27 | | Rep 2 | 23.22 | 22.14 | 22.42 | 21.96 | 22.49 | 24.10 | 22.32 | | Rep 3 | 22.66 | 21.44 | 22.56 | 22.41 | 23.09 | 23.85 | 21.75 | | Average | 22.99 | 21.76 | 22.36 | 22.20 | 22.74 | 23.73 | 22.11 | | Stdev | 0.29 | 0.35 | 0.23 | 0.23 | 0.31 | 0.45 | 0.321 | | % CV | 1.28 | 1.63 | 1.03 | 1.02 | 1.38 | 1.89 | 1.43 | | % Recovery | 103.98 | 98.42 | 101.13 | 100.41 | 102.82 | 107.30 | 100.00 | The samples from all tube types were divided into two portions and stored at 2-8 °C and -20 °C and retested at week one and week two to test anticoagulant interference. The results showed that there were no interferences with regard to recovery of vancomycin in serum or plasma for the 7 tube types tested. The results also showed that the samples are stable at both temperatures for at least 15 days. 3. Clinical studies: a. Clinical sensitivity: Not Applicable b. Clinical specificity: Not applicable. Clinical studies are not typically submitted for this device type. c. Other clinical supportive data (when a and b are not applicable): 4. Clinical cut-off: Not applicable. {11} Page 11 of 11 5. Expected values/Reference range: Therapeutic peak serum levels of 20- 40 µg/mL and trough levels of 5- 10 µg/mL have been reported for most strains of staphylococci and streptococci. However, therapeutic levels must be individually established based on patient differences and bacterial susceptibility. The risk of toxicity is appreciably increased by high concentration or prolonged therapy in patients with renal insufficiency. Toxic effects, such as ototoxicity and nephrotoxicity, have resulted when serum concentrations reach 80-100 µg/mL and are rarely seen when serum levels are maintained below 30 µg/mL. Sponsor references literature for these expected values/reference ranges. Wilhelm MP, Vancomycin. Mayo Clinic Proc 1991; 66:1165-1170. N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...