COOK IRELAND SONNET POLYPECTOMY SNARE

K050294 · Cook Ireland, Ltd. · FDI · Mar 31, 2005 · Gastroenterology, Urology

Device Facts

Record IDK050294
Device NameCOOK IRELAND SONNET POLYPECTOMY SNARE
ApplicantCook Ireland, Ltd.
Product CodeFDI · Gastroenterology, Urology
Decision DateMar 31, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sonnet TM Polypectorny Snare is a monopolar electrosurgical device used with an electrical unit for endoscopic polypectomy.

Device Story

Sonnet Polypectomy Snare is a monopolar electrosurgical instrument for endoscopic polypectomy. Device comprises 240cm catheter, proximal handle, and braided stainless steel wire loop attached to stainless steel drive-wire. Loop is extendable and retractable from catheter distal tip. Handle includes active cord connection for electrosurgical unit interface. Used by clinicians during endoscopic procedures to excise polyps. Device functions by capturing tissue within wire loop and applying electrosurgical current to perform resection. Benefits include effective tissue removal during endoscopy.

Clinical Evidence

Bench testing only; electrical testing performed.

Technological Characteristics

Monopolar electrosurgical snare; braided stainless steel wire loop; stainless steel drive-wire; 240cm single-lumen catheter; proximal handle with active cord connection. Manual actuation.

Indications for Use

Indicated for use in endoscopic polypectomy procedures requiring a monopolar electrosurgical device. Prescription use only.

Regulatory Classification

Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # COOK® ### MAR 3 1 2005 #### Cook Ireland Ltd. O'Halloran Road, National Technology Park, Limerick, IRELAND Phone: +353 61 334440 Fax: +353 61 334441 www.cookgroup.com K050294 Pg 1 of 2 #### 510(k) Summary #### Submitted By: Cook Ireland Ltd. O'Halloran Road, National Technology Park, Limerick, Ireland #### Contact Information: Emmett Devereux QA/RA Manager Cook Ireland Limited O'Halloran Road National Technology Park Limerick, Ireland Phone: +353-61-334440 Fax: +353-61-334441 Email: edevereux@cook.ie April Lavender Cook Incorporated 750 Daniels Way Bloomington, IN 47401 U.S.A Phone: 812-339-2235 Fax: 812-332-0281 Email: LAVENDER@COOK.INC.COM #### Date Prepared: February 3, 2005 510(k) #: #### Device: Trade Name: Common/Usual Name: Class: Product Code: Sonnet TM Polypectomy Snare Snare Class II FDI #### Intended Use: The Sonnet TM Polypectorny Snare is a monopolar electrosurgical device used with an electrical unit for endoscopic polypectomy. {2}------------------------------------------------ # COOL K050294 Pg 2 of 2 #### Device Description: The Sonnet™ Polypectomy Snare consists of a 240cm catheter attached to a handle at the proximal end and incorporates a braided stainless steel wire in a loop configuration that is attached to a stainless steel drive-wire and is extendable and retractable with respect to the catheter distal tip. The handle provides an active cord connection. The catheter has a single lumen. #### Test Data: Tests include electrical testing #### Substantial Equivalence: The subject device is similar with respect to intended use and/or design features to the predicate devices in terms of section 510(k) substantial equivalence. The subject device is safe and effective and is substantially equivalent to the predicate devices. | Manufacturer | Device | 510(k) | |------------------------------|----------------------------|--------| | Wilson-Cook | Wilson-Cook Snare | K85195 | | Horizons International Corp. | Horizons Polypectomy Snare | K01166 | {3}------------------------------------------------ Image /page/2/Picture/2 description: The image shows a logo with a stylized bird in flight. The bird is facing left, and its body is represented by three curved lines. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES". The logo is black and white. MAR 3 1 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MO 20850 Mr. Emmett Devereux QA/RA Manager Cook® Ireland Limited O'Halloran Road National Technology Park LIMERICK IRELAND Re: K050294 Trade/Device Name: Sonnet™ Polypectomy Snare Regulation Number: 21 CFR §876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: FDI Dated: February 3, 2005 Received: February 7, 2005 Dear Mr. Devereux: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to the cheroped of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that recaussition in acoroval of a premarket approval application (PMA). You may, therefore, market the do not require approvate or a controls provisions of the Act. The general controls provisions of the Act as not regurements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket 11 your ab room be subject to such additional controls. Existing major regulations affecting your Applorary, It ifaly of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I rease of a determination that your device complies with other requirements of the Act or any I Drivas mades and regulations administered by other Federal agencies. You must comply with all the r cacial starses and registration and limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter witi anow you to begin maneting your and equivalence of your device to a legally premarket notheation. The PDA mainly of saosannal vice and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice tor your do 1706 of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, prease note the regulation entired, consibilities under the Act from the 807.97). Tou may obtain other general missance at its toll-free number (800) DVVSlon of 301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsmamain.html Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): _K 05 c 3 94 Device Name: Sonnet™ Polypectomy Snare Indications for Use: The Sonnet ™ Polypectomy Snare is a monopolar electrosurgical device used with an electrical unit for endoscopic polypectomy. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use Only Counter (Per 21 CFR § 801.109 OR Over-the- Nancy C Brazdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K050294
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...