Prospective, longitudinal clinical trial (originally for predicate device K031588)
The sponsor uses clinical trial data from the predicate device to demonstrate that the CTC count threshold of 5 or more per 7.5 mL of blood is predictive of shorter progression-free and overall survival in metastatic breast cancer patients, supporting the clinical utility of the new device.
Multi-center prospective, longitudinal clinical trial; Prospective, longitudinal clinical trial; Follow-up/Duration: Mean 4.5 ± 2.4 weeks for first follow-up
Metastatic breast cancer patients with measurable disease starting a new line of therapy; Sample Size: 177; Number of Sites: Multi-center
Not applicable for this study
Progression-free survival (PFS) and overall survival (OS)
Indications for Use
The CellSearch™ Circulating Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood. The presence of CTC in the peripheral blood, as detected by the CellSearch™ Circulating Tumor Cell Kit, is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast cancer. A CTC count of 5 or more per 7.5mL of blood is predictive of shorter progression free survival and overall survival.
Device Story
The CellSearch™ Circulating Tumor Cell Kit uses immunomagnetic selection to isolate epithelial cells from 7.5 mL of whole blood. Anti-EpCAM monoclonal antibodies conjugated to ferrofluid particles bind to target cells; a magnetic field concentrates these cells onto a cartridge surface. Fluorescent reagents (CK-PE, CD45-APC, DAPI) are added to label cytokeratin, leukocytes, and nuclei, respectively. The cartridge is processed on the CellTracks® Analyzer II, which acquires fluorescent images. Software identifies potential tumor cells (EpCAM+, CK+, DAPI+, CD45-) and presents them in a gallery for operator classification. The operator confirms CTCs by checking composite images; the software tallies the count. Used in clinical laboratories by trained technicians. The CTC count (threshold 5 cells/7.5 mL) assists clinicians in monitoring metastatic breast cancer patients, providing prognostic information regarding progression-free and overall survival.
Clinical Evidence
Prospective longitudinal trial (N=177) of metastatic breast cancer patients. Primary endpoints: Progression-Free Survival (PFS) and Overall Survival (OS) based on baseline and 1st follow-up CTC counts. Patients with ≥5 CTC/7.5 mL had significantly shorter median PFS (11.7 vs 30.3 weeks, p=0.0001) and OS (43.3 vs >80 weeks, p<0.0001) compared to those with <5 CTC. Multivariate Cox regression confirmed CTC count as the strongest predictor of PFS and OS.
Technological Characteristics
Immunomagnetic selection using anti-EpCAM ferrofluid particles. Fluorescent labeling (CK-PE, CD45-APC, DAPI). System includes CellTracks® AutoPrep System and CellTracks® Analyzer II. Connectivity: Standalone system. Software: Automated image acquisition and analysis. Sterilization: Not specified.
Indications for Use
Indicated for patients treated for metastatic breast cancer. The presence of CTC in peripheral blood is associated with decreased progression-free and overall survival. A CTC count of 5 or more per 7.5 mL of blood is predictive of shorter progression-free and overall survival.
Regulatory Classification
Identification
An immunomagnetic circulating cancer cell selection and enumeration system is a device that consists of biological probes, fluorochromes, and other reagents; preservation and preparation devices; and a semiautomated analytical instrument to select and count circulating cancer cells in a prepared sample of whole blood. This device is intended for adjunctive use in monitoring or predicting cancer disease progression, response to therapy, and for the detection of recurrent disease.
Special Controls
The device is classified as Class II under regulation 21 CFR 866.6020 with special controls. The special control guidance document "Immunomagnetic Circulating Cancer Cell Selection and Enumeration System" is available at https://www.fda.gov/cdrh/oivd/guidance/1531.pdf
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Immunomagnetic Circulating Cancer Cell Selection and Enumeration System.” See § 866.1(e) for availability of this guidance document.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K050245
B. Purpose for Submission:
Reagent formulation changes and changes in procedural incubation times to deliver the same assay performance characteristics on the new automated version of the CellPrep semi-automated cell preparation instrument, now called CellSearch AutoPrep. Larger volumes of reagents are needed for automated use.
C. Measurand:
EpCam, Cytokeratins 8, 18 and/or 19, and CD45 to identify Circulating Tumor Cells (CTC) (Cells appearing to look like tumor cells with epithelial cell markers and no lymphocyte marker on their surfaces).
D. Type of Test:
A semi-automated qualitative immunomagnetic-capture immunofluorescent detection image analysis test.
E. Applicant:
Veridex, LLC, A Johnson and Johnson Company
F. Proprietary and Established Names:
CellSearch™ Circulating Tumor Cell Kit
G. Regulatory Information:
1. Regulation section:
21 CFR 866.6020-Immunomagnetic Circulating Cancer Cell Selection and Enumeration System
2. Classification:
Class II
3. Product code:
NQI System, Immunomagnetic, Circulating Cancer Cell, Enumeration
4. Panel:
Immunology 82
H. Intended Use:
1. Intended use(s):
The CellSearch™ Circulating Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood.
2. Indication(s) for use:
The presence of CTC in the peripheral blood, as detected by the CellSearch™ Circulating Tumor Cell Kit, is associated with decreased progression free survival
{1}
and decreased overall survival in patients treated for metastatic breast cancer. A CTC count of 5 or more per 7.5 mL of blood is predictive of shorter progression free survival and overall survival.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
The CellTracks® AutoPrep system and the CellTracks® Analyzer II. The CellTracks® Analyzer II is a semi-automated fluorescence microscope intended to enumerate fluorescently labeled cells that are immunomagnetically selected and distributed over a viewing surface
# I. Device Description:
The CellSearch™ Circulating Tumor Cell Kit contains a ferrofluid-based capture reagent and immunofluorescent reagents. The ferrofluid reagent consists of nanoparticles with a magnetic core surrounded by a polymeric layer coated with antibodies targeting the EpCAM antigen for capturing CTC. After immunomagnetic capture and enrichment, fluorescent reagents are added for identification and enumeration of CTC. The fluorescent reagents include the following: anti-CK-Phycoerythrin (PE) specific for the intracellular protein cytokeratin (characteristic of epithelial cells), DAPI which stains the cell nucleus, and anti-CD45-Allophycocyanin (APC) specific for leukocytes
The reagent/sample mixture is dispensed by the CellTracks® AutoPrep System into a cartridge that is inserted into a MagNest® cell presentation device. The strong magnetic field of the MagNest® device attracts the magnetically labeled epithelial cells to the surface of the cartridge. The CellTracks® Analyzer II or CellSpotter® Analyzer automatically scans the entire surface of the cartridge, acquires images and displays any event to the user where CK-PE and DAPI fluorescence are co-located. Images are presented to the user in a gallery format for final classification. An event is classified as a tumor cell when its morphological features are consistent with that of a tumor cell and it exhibits the phenotype EpCAM+, CK+, DAPI+ and CD45-
# J. Substantial Equivalence Information:
1. Predicate device name(s):
CellSearch™ Epithelial Cell Kit
2. Predicate 510(k) number(s):
K031588
3. Comparison with predicate of modifications made to reagents:
| Similarities | | |
| --- | --- | --- |
| Item | Predicate | Device |
| Summary of changes | CellSearch™ Epithelial Cell Kit | CellSearch™ Circulating Tumor Cell Kit |
| Cell Fixative There is no change to the Cell Fixative reagent between systems | 2.5 mL bottle Cell Fixative: Contains PBS, 25% proprietary ingredients, BSA and | 3.0 mL bottle Cell Fixative: Contains PBS, 25% proprietary ingredients, 0.1% BSA and 0.1% sodium azide. (red cap) |
{2}
| Similarities | | |
| --- | --- | --- |
| Item | Predicate | Device |
| | sodium azide. This bottle is color-coded with a **red** cap. | |
| Differences | | |
| Item | Predicate | Device |
| **Anti-EpCAM Ferrofluid**No changes were made to theanti-Ep-CAM antibody used on the ferrofluid. Neither were any changes made to the formulation of or manufacture of Anti-Ep-CAM ferrofluid; except for the change in final concentration required to accommodate the CellTracks® AutoPrep System required differences in volume of reagent added (150 μL versus 100 μL) and the slightly shorter incubation time | **1.1 mL vial anti-EpCAM-Ferrofluid**: Contains a suspension of 0.03% magnetic nanoparticles conjugated to a mouse monoclonal antibody that is specific for a cell surface marker present on epithelial cells in a buffer containing bovine serum albumin (BSA) and ProClin 300 preservative. | **3.0 mL bottle Anti-EpCAM Ferrofluid**: Contains a suspension of 0.022% magnetic nanoparticles conjugated to a mouse monoclonal antibody specific for the cell surface marker EpCAM present on epithelial cells in a buffer containing 0.03% bovine serum albumin (BSA) and 0.05% ProClin 300 preservative. (**brown** cap) |
| **Nucleic Acid DyeDAPI** is the same dye for both systems. A modification was made in the formulation of the Nucleic Acid Dye in order to accommodate the CellTracks® AutoPrep System/ CellTracks® Analyzer II. The modification was to formulate in Normal Saline for AutoPrep instead of distilled water. This was done because adding 20 μL of distilled water to 260 μL of PBS-BSA, as was done in CellPrep, would not change the ionic strength of the media very much. In AutoPrep however, adding 150 μL of water to 700 μL of PBS-BSA would have had a large effect on the ionic strength. Therefore, the DAPI was formulated in | **0.25 mL vial Nucleic Acid Dye**: Contains 0.005% 4', 6-diamidino-2-phenylindole, dihydrochloride (DAPI) and ProClin 300. This vial is color-coded with a **purple** inset in the cap. | **3.0 mL bottle Nucleic Acid Dye**: Contains 0.005% 4', 6-diamidino-2-phenylindole, dihydrochloride (DAPI) and ProClin 300. (**blue** cap) |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Predicate | Device |
| normal saline. Cap color change | | |
| anti-CK-PE and anti-CD45-APCThe original 3 Staining reagents have been combined into one Staining reagent. Same materials and small optimization change in monoclonal antibody concentration. The Staining reagent contains three fluorescent-probe-conjugated antibodies. The three conjugated antibodies are C11-PE, CK19-PE, and anti-CD45-APC. In AutoPrep, the exact same antibodies are used as are used for CellPrep. Each antibody is purified using the same procedures, conjugated to the same fluorochrome in the same 1:1 F/P ratio, using the same conjugation chemistry and procedures, and formulated in the same diluent as was used for the original CellSearch-CellPrep reagents. In CellSearch-CellPrep each antibody was formulated into the same dilution buffer, but was supplied as three separate reagents, each of which had to be added to the test sample in 20 μL aliquots. For AutoPrep, all three antibodies are instead formulated into one bottle so that addition of a single 150 μL of antibody staining reagent delivers all three antibodies. The concentrations of the antibodies as delivered to the test sample are slightly lower than was used in CellSearch-CellPrep. The final | 0.25 mL vial anti-CK-PE A: Contains 0.001% mouse monoclonal antibodies specific to cytokeratins 8 and 18 conjugated to phycoerythrin (PE) in phosphate buffered saline (PBS) containing BSA, surfactant and sodium azide. This vial is color-coded with a yellow inset in the cap.
0.25 mL vial anti-CK-PE B: Contains 0.001% mouse monoclonal antibodies specific to cytokeratin 19 conjugated to PE in PBS containing BSA, surfactant and sodium azide. This vial is color-coded with a blue inset in the cap.
0.25 mL vial anti-CD45-APC: Contains 0.005% mouse monoclonal antibodies specific to CD45 conjugated to allophycocyanin (APC) in PBS containing BSA, surfactant and sodium azide. This vial is color-coded with a red inset in the cap. | 3.0 mL bottle Staining Reagent: Contains 0.0008% mouse monoclonal antibodies specific to cytokeratins conjugated to phycoerythrin (PE); 0.0012% mouse anti-CD45 mouse monoclonal antibody conjugated to allophycocyanin (APC) in phosphate buffered saline (PBS) containing 0.5% BSA, and 0.1% sodium azide. (white cap) |
4
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Predicate | Device |
| combination of antibodies was formulated to provide a robust mixture of antibody conjugates that would provide performance comparable to that of the individually formulated CellSearch-CellPrep reagents. Cap color is now white | | |
| **Aggregation agent in Buffer** No material changes. The active ingredient of this reagent is Streptavidin. The Streptavidin reacts with desthiobiotin on the surface of the Ferrofluid particles to initiate the proprietary controlled aggregation technology. Streptavidin is present in great excess. For CellPrep, this aggregation reagent was formulated in dilution buffer and 6 mLs was added to bring the total volume of the blood sample to approximately 10 mLs. In AutoPrep, the aggregation reagent is formulated in PBS-BSA and the same amount (in μg) of Streptavidin is added (150 μL) to the 10 mL blood sample that already contains dilution buffer. | **60 mL bottle AB Dilution Buffer**: Contains phosphate buffered saline (PBS), BSA, 0.6% other animal protein, 0.006% proprietary aggregation agent and sodium azide. | • **3.0 mL bottle Capture Enhancement Reagent**: Contains PBS, 0.5% BSA, 0.02% proprietary reagent for controlled ferrofluid aggregation and 0.1% sodium azide. (clear cap) |
| **Permeabilization Reagent** The matrix used and the active ingredient of Permeabilization Reagent is the same for both CellPrep and AutoPrep. Due to the longer incubation time used on AutoPrep the concentration of Saponin is slightly lower than was used in CellPrep. In addition, for AutoPrep biotin was added to | • **2.5 mL bottle Permeabilization Reagent**: Contains phosphate buffered saline (PBS), 0.05% proprietary permeabilization reagent and sodium azide. This bottle is color-coded with a **green** cap. | • **3.0 mL bottle Permeabilization Reagent**: Contains phosphate buffered saline (PBS), 0.011% proprietary permeabilization reagent and 0.1% sodium azide (green cap) |
{5}
| Similarities | | |
| --- | --- | --- |
| Item | Predicate | Device |
| Permeabilization Reagent. Biotin is inert with respect to permeabilization and is also used in the CellSearch-CellPrep assay. The same component, biotin, is used in both CellPrep and AutoPrep to affect the deaggregation, and the action and mechanism remains unchanged between CellSearch-CellPrep and CellSearch-AutoPrep reagents. | | |
K. Standard/Guidance Document Referenced (if applicable):
The CellSearch™ Circulating Tumor Cell kit was developed in conformance to the following standards and guidances.
ISO 14971 Medical Devices- Application of Risk Management to Medical Devices
Guidance for Industry and FDA Staff Class II Special Controls Guidance Document: Circulating Cancer Cell Selection and Enumeration System (May 11,2004)
Guidance for Industry and FDA Staff: Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use (November 30, 2004)
The CellTracks Analyzer II performance data were developed in conformance to the following standards.
EP5-A NCCLS document: Evaluation of Precision Performance of Clinical Chemistry Devices
EP9-A NCCLS document: Method comparison and Bias Estimation Using Patient Samples
All requirements for these standards were met. EP9-A testing was performed using donor spiked samples rather than actual cancer patient samples.
L. Test Principle:
Epithelial cells are immunomagnetically labeled by targeting the Epithelial Cell Adhesion Molecule (EpCAM) antigen. Anti-EpCAM monoclonal antibodies conjugated to ferrofluid particles are colloidal and, when mixed with a sample containing the target epithelial cells, bind to the EpCAM antigen associated with the epithelial cells. After immunomagnetic selection of epithelial cells from 7.5 mL of blood, fluorescent reagents are added at this time to discriminate between the immunomagnetically selected cells. Anti-Cytokeratin – Phycoerythrin (CK-PE) stains the intracellular cytoskeleton cytokeratin proteins expressed in cells of epithelial origin, anti-CD45-Allophycocyan (CD45-APC) stains leukocytes and DAPI stains DNA present in the cell nucleus.
{6}
The processed reagent/sample mixture is dispensed by the CellTracks® AutoPrep System into a cartridge that is inserted into a MagNest® cell presentation device. The strong magnetic field of the MagNest® device causes the magnetically-labeled target cells to move to the surface of the cartridge. The cartridge is then placed on the CellTracks® Analyzer II for data acquisition and analysis. The CellTracks Analyzer II scans the entire surface of the cartridge with a series of fluorescence filters that are defined for a given assay and acquires images of PE, APC and DAPI fluorescence staining of the entire viewing surface.
After data acquisition is completed, the images are analyzed for any event where cytokeratin-PE and DAPI are within a specified space in the cartridge, i.e. indicating the possible presence of a cell with a nucleus that expresses cytokeratin. Images from each fluorescent color as well as a composite image of the cytokeratin staining (green) and the nuclear staining (purple) are presented to the user in a gallery for final cell classification. A cell is classified as a tumor cell when it its EpCAM+ (i.e., it is captured), CK+, DAPI+ and CD45-. A check mark placed by the operator next to the composite images classifies the event as a Circulating Tumor Cell (CTC) and the software tallies all the checked boxes to obtain the CTC count.
The sponsor's data demonstrate that metastatic breast cancer patients with 5 or more CTC/per 7.5 mL of blood have a significantly greater probability for shorter progression free and overall survival than patients who have fewer than 5 CTC per 7.5 mL of blood.
This test methodology is new. It is hoped that further clinical studies reported in the scientific literature will corroborate the clinical study performed by the sponsor for submission to the FDA for the predicate device.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
(1) System Reproducibility with CellSearch™ Circulating Tumor Cell Control. Three separate CellSearch™ Circulating Tumor Cell Control samples were prepared and processed each day for over 30 days, per the long run method of NCCLS guideline EP5-A². Each single-use sample bottle contains a low and a high concentration of cells from a fixed cell line that have been pre-stained with two different fluorochromes. Summary statistics for the high and low control cells is presented below.
Table 1. Summary of Precision Analyses
| | Low | High |
| --- | --- | --- |
| N | 99 | 99 |
| Mean cell count | 48 | 969 |
| Total Precision Standard Deviation (S_{T}) % CV | 18% | 5% |
(2) System Reproducibility with Patient Samples
A total of 163 duplicate samples were collected from 47 patients over the course of the clinical study. These samples were processed at multiple sites to determine the reproducibility of CTC measurements. The regression equation for the
{7}
comparison of these 163 duplicate samples was $\mathrm{Y} = 0.98\mathrm{x} + 0.67$ , $\mathrm{R}^2 = 0.99$ . Table 2 shows the summary of the data for replicates where the average of the two CTC results was $< 5$ compared to those where the average (avg.) was $\geq 5$ .
Table 2. Reproducibility of CTC Counts in Duplicate Samples (n=163) with an Average of $< 5$ or $\geq 5$ CTC per $7.5\mathrm{mL}$ of blood.
| | CTC <5 | CTC ≥5 |
| --- | --- | --- |
| Number of Duplicates | 123 | 40 |
| Mean CTC Count of Duplicates | 0.7 | 210 |
| Avg. Duplicate Standard Deviation | 0.5 | 12 |
| Avg. %CV of Duplicates | 60% | 20% |
# Accuracy/Recovery:
Blood samples from a single healthy donor were pooled and five of six 7.5 mL aliquots were spiked with 5, 20, 81, 325 and 1300 cultured breast cancer cells (SK-Br-3). The sixth tube was unspiked pooled blood and served as a zero point. These samples were processed on the CellTracks® AutoPrep System with the CellSearch™ Circulating Tumor Cell Kit and CTC counts were determined on the CellTracks® Analyzer II. The experiment was repeated for four additional donors. The observed cell counts were plotted against the results of the expected cell count. The results are summarized in Table 3.
Table 3. Percent Detection Estimates.
| Expected Tumor Cell Count | Mean Observed Tumor Cell Count | Range of Percent Recovery |
| --- | --- | --- |
| 1300 | 1215 | 91 to 95% |
| 325 | 308 | 82 to 101% |
| 81 | 85 | 80 to 136% |
| 20 | 22 | 95 to 140% |
| 5 | 7 | 120 to 200% |
To determine the overall, or least squares fit, for the comparison of the observed and expected cell counts across all the data, linear regression analysis was performed. The regression equation for these 30 samples was $\mathrm{Y} = 0.93\mathrm{x} + 3.87$ , $\mathrm{R}^2 = 0.999$ . The results of this study indicate that on average over the tested CTC range the recovery, as derived from regression analysis, is $93\%$ .
Given the linear response of the tumor cell counts, one would expect the slope of the observed versus expected plot to be 1.0. However, the slope was 0.93. This is because the CellTracks® AutoPrep System with CellSearch™ CTC Kit involves the capture and fluorescent labeling of cells followed by their
{8}
detection and enumeration by the CellTracks® Analyzer II. The loss of cells could therefore be attributed to one of the following possibilities; 1) the recovery of only 93% of the tumor cells spiked into 7.5 mL of blood by the CellTracks® AutoPrep System, 2) the detection of only 93% of the tumor cells present in the sample chamber by the CellTracks® Analyzer II or 3) a combination of both of these sources of error.
b. Linearity/assay reportable range:
Another way to examine the previous data is to analyze it as a dilution series to evaluate test linearity. We removed the confounding variable of percent recovery by using the observed value of the original sample divided by the dilution factors to determine the expected values for the dilution series for each patient sample. Regression of all of these numbers of observed tumor cells versus the numbers of expected tumor cells yielded a slope of 1.007, an intercept of 3.0, an r² = 0.99 and r = 0.995. Therefore, once the percent recovery (cell loss) was factored out of the CTC values of each of the original samples, this analysis of the data demonstrated that the detection of CTC was linear over the reportable range of 0 to 1238 tumor cells.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
No recognized reference material or method.
d. Detection limit:
One CTC per 7.5 mL can be detected by the CellTracks® Analyzer II resulting in a limit of detection of 1 CTC in a cartridge. Linear regression shows that on average, 93% of CTC present in a 7.5 mL blood sample are recovered using the CellTracks® AutoPrep System (see Recovery section). The loss of approximately 7% of the CTC in the sample is not sufficient to reduce the limit of detection of 1 CTC.
e. Analytical specificity:
The CellSearch™ Circulating Tumor Cell Kit contains a ferrofluid-based capture reagent and immunofluorescent reagents. The ferrofluid reagent consists of nanoparticles with a magnetic core surrounded by a polymeric layer coated with antibodies targeting the EpCAM antigen for capturing CTC. After immunomagnetic capture and enrichment, fluorescent reagents are added for identification and enumeration of CTC. The fluorescent reagents include the following: anti-CK-Phycoerythrin (PE) specific for the intracellular protein cytokeratin (characteristic of epithelial cells), DAPI which stains the cell nucleus, and anti-CD45-Allophycocyanin (APC) specific for leukocytes
Interfering Substances:
SK-BR-3 cells spiked into blood samples were exposed to potential interfering substances and compared to untreated controls. Toxic levels (5 times therapeutic index) of the following cancer drugs, over-the-counter drugs, and other exogenous substances were tested: cyclophosphamide, Mitomycin C®, Procrit®, biotin, 5-fluorouracil, methotrexate, tamoxifen citrate, paclitaxel, Arimidex®, acetaminophen, acetylsalicylic acid, caffeine, dextromethorphan, Aredia®, Human Anti-Mouse Antibody (HAMA) type 1, HAMA type 2, Herceptin®, and ibuprofen. No significant differences in SK-BR-3 cell
9
{9}
numbers were detected, indicating that these substances do not interfere with the CellSearch™ kit.
Samples spiked with toxic levels of doxorubicin resulted in aberrant staining of leukocytes as cytokeratin and CD45 dual positive cells, due to the doxorubicin being a fluorescent compound that is incorporated into nucleated cells. If seen, the staining pattern of all cells being CD45 positive and cytokeratin positive is obvious and easily identified by the operator as a known interference staining profile. If blood is drawn after the recommended 7-day washout period, following doxorubicin infusion, this interference is unlikely to be observed in clinical practice given controlled therapeutic levels and rapid drug clearance.
Potential interference from lipemia was studied by adding Intralipid to samples to a concentration of 2.6%, which corresponds to greater than 1000 mg/dl triglyceride. Samples were lysed to simulate total hemolysis. Bilirubin at 7.4 mg/dL, HAMA 1/HAMA 2 and hematocrit from 18-60% were studied. Lipemia, hemolysis, icterus and a broad range of hematocrit values do not interfere with the CellSearch™ test. HAMA 1 and HAMA 2 also do not interfere, indicating that individuals receiving mouse Ig by parenteral routes can be tested successfully with the CellSearch™ test.
f. Assay cut-off:
Results are reported as the number of CTC / 7.5 mL of blood. A CTC count of 5 or more per 7.5 mL of blood is predictive of shorter progression free survival and overall survival. This cut-off was established in the 510(k) of the previous version of this assay, the predicate device, K031588.
2. Comparison studies:
a. Method comparison with predicate device:
To demonstrate comparable performance between the new assay and the predicate system, a study was performed using duplicate samples split between both the predicate device, CellSearch™ Epithelial Cell System (K031588), and the new device, CellSearch™ Circulating Tumor Cell System (K050245). The study consisted of spiking normal donor whole blood samples with three different tissue culture lines (SKBr-3, PC3-9 and MCF-7) at three different levels (~5, ~50, and ~1000) for 5 days. The three cell lines (SK-Br-3, MCF-7, or PC3-9) were chosen to cover a broad range of EpCAM and Cytokeratin antigen density representing the capture and detection portions of the assay respectively. Three spike levels of each cell line were chosen to cover a range of potential clinical values. Of the three cell lines tested, the PC3-9 cell line has the lowest Cytokeratin antigen density. SK-Br-3 cells demonstrate an uneven bimodal population consisting primarily of moderate level Cytokeratin antigen density cells and a smaller population of higher expressing cells. MCF-7 cells demonstrate the highest level of consistent Cytokeratin expression. The Cytokeratin antigen is the target of the detection reagent for tumor cells in the CellSearch™ Circulating Tumor Cell kits. The design and execution of this study is consistent with the NCCLS guideline EP9-A. A total of 45 samples were analyzed on each of the two platforms
10
{10}
For MCF-7 cells the slope of the regression line = 1.03, an intercept of 1.5 and an r² = 0.994. For SKBr-3 cells, the slope of the regression line = 1.01 with an intercept of 2.9 and an r² = 0.984. For PC3-9 cells, the slope of the regression line = 1.19 with an intercept of 10.5 and an r² = 0.963. Analysis of data from all 3 tumor cell lines combined shows a slope of the regression line = 1.09 with an intercept of 1.5 and an r² = 0.966.
The slope of 1.19 for PC3-9 cells may be due to an improved dynamic range of the AutoPrep / CellTracks Analyzer II system resulting in a flattening out of the response curve at higher cell numbers. In other words, the recovery of CTC by the AutoPrep / CellTracks Analyzer II platform at high numbers of cells may be somewhat more sensitive than recovery by the CellPrep / CellSpotter platform, particularly with lower EpCAM antigen density cells as is the case with PC3-9 cells (Figure 1). This difference could also be attributable to increased reliability and/or stability of the AutoPrep as compared to the CellPrep for sample preparation. Regardless of this potential difference however, there appears to be no difference between the AutoPrep / CellTracks Analyzer II platform and the CellPrep / CellSpotter platform at the medical decision level of 5-50 CTC’s.
b. Matrix comparison:
Since whole blood is the only matrix for this test., no matrix comparison studies were performed.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
All of the above new analytical studies with the CellSearch™ Circulating Cell Kit/AutoPrep/CellTracks® Analyzer II system demonstrate that the detection of tumor cells by the CellSearch™ Circulating Cell Kit/ AutoPrep/CellTracks® Analyzer II system is substantially equivalent to the predicate system. Therefore, the following clinical data generated using the predicate system (K035188) is applicable to the new device (K050245).
Metastatic Breast Cancer Patients
A multi-center prospective, longitudinal clinical trial was conducted. Results were used to determine whether the number of CTC predict disease progression and survival. Patients with measurable disease and who were starting a new line of therapy were enrolled (N=177). Clinical data were analyzed on an intent-to-treat basis.
{11}
Table 4. Patient Demographics
| Age at Baseline (Median) | 58.0 ± 13.4 (58) |
| --- | --- |
| Race: White | 153 (84%) |
| Black | 14 (8%) |
| Hispanic | 7 (4%) |
| Unknown | 3 (2%) |
| ER/PR + | 121 (68%) |
| ER/PR - | 54 (31%) |
| Unknown | 2 (1%) |
| Her-2/neu - | 91 (52%) |
| Her-2/neu 1+ | 12 (7%) |
| Her-2/neu 2+ | 18 (10%) |
| Her-2/neu 3+ | 27 (15%) |
| Unknown | 29 (16%) |
| Line of Therapy | 1st83 (47%) |
| | 2nd25 (14%) |
| | ≥3rd67 (38%) |
| | Unk.*2 (1%) |
| Type of Therapy | Hormone 47 (26%) |
| | Chemo 87 (49%) |
| | Immu/C/H 28 (16%) |
| | H / C 10 (6%) |
| | No Tx**4 (2%) |
| | Unk.*1 (1%) |
*Unk. = Information not available **No Tx. = No treatment information obtained C or Chemo = Chemotherapy, H or Hormone = Hormone Therapy, I or Immuno = Immunotherapy
Baseline CTC count was determined prior to initiation of a new line of therapy. A first follow-up CTC count was determined after the initiation of therapy. For the baseline analyses, Progression Free Survival (PFS) was measured from the time of the baseline blood draw to the diagnosis of progression by CT scans and/or clinical signs and symptoms, and Overall Survival (OS) was measured from the time of baseline blood draw to the time of death. For the first follow-up analyses, PFS was measured from the time of $1^{\text{st}}$ follow-up blood draw (mean $4.5 \pm 2.4$ weeks following enrollment) to diagnosis of progression or death, and OS was measured from the time of $1^{\text{st}}$ follow-up blood draw to the time of death.
# Progression Free Survival (PFS) Analysis
# PFS Using Baseline CTC Results
All 177 patients had a baseline CTC test performed. For Kaplan-Meier analysis, patients were segmented into two groups based upon their CTC count at baseline:
- The Favorable group $(N = 90)$ , represented in green, consisted of patients with $< 5$ CTC.
- The Unfavorable group $(N = 87)$ , represented in red, consisted of patients with $\geq 5$ CTC.
Median PFS was 30.3 weeks (~7.0 months) for the Favorable group and 11.7 weeks (~2.7 months) for the Unfavorable group. The difference in PFS between the two groups is highly significant (Log-rank p=0.0001, Cox Hazards Ratio=1.9547, chi-square=15.33, p = 0.0001). These results are illustrated in Figure 1.
{12}
13
Figure 1. PFS of Patients with < 5 or > 5 CTC at Baseline (N=177)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.