K050235 · Iris International, Inc. · GKL · Mar 23, 2005 · Hematology
Device Facts
Record ID
K050235
Device Name
IQ 200 URINE ANALYZER BODY FLUIDS MODULE
Applicant
Iris International, Inc.
Product Code
GKL · Hematology
Decision Date
Mar 23, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5200
Device Class
Class 2
Indications for Use
The iQ® 200 Urine Analyzer Body Fluids Module is an in-vitro diagnostic device used by a competent human observer to examine and count red blood cells and nucleated cells in cerebrospinal fluid and serous fluids. This module is a capability added to the iQ200 Urine Analyzer, a cleared urinalysis instrument (K022774).
Device Story
The iQ® 200 Urine Analyzer Body Fluids Module is an add-on to the iQ200 urinalysis instrument. It processes two aliquots of body fluid: one diluted in saline, one treated with lysing reagent to isolate nucleated cells. The device uses flow microscopy to capture and save particle images as samples pass a microscope objective. Images are automatically categorized by size on a video display. A human observer reviews these assignments, with the ability to manually reclassify particles; the system then recomputes and reports final cell concentrations. Used in clinical laboratory settings by trained personnel, the output provides quantitative cell counts to assist physicians in differential diagnosis and health assessment. The device benefits patients by automating the labor-intensive manual chamber counting process while maintaining clinical accuracy.
Clinical Evidence
Performance compared against manual chamber counting. Study included 304 samples for RBC and 299 for nucleated cells (NC). Regression analysis showed R2 of 0.992 (slope 0.906) for RBC and 0.967 (slope 1.015) for NC. Linearity confirmed from 0 to 10,000 particles/microliter per NCCLS EP6-A. Paired t-tests indicated no statistically significant difference between replicate counts.
Technological Characteristics
Automated flow microscopy system. Uses saline dilution and lysing reagent for cell identification. Features microscope objective for high-speed image capture, electronic particle concentration, and video display for observer review. Connectivity via integration with iQ200 urinalysis platform. Linear range 0-10,000 particles/microliter.
Indications for Use
Indicated for use by competent human observers to examine and count red blood cells and nucleated cells in cerebrospinal and serous fluids to assist physicians in diagnosing meningitis, intracranial hemorrhage, leukemias, malignancies, infections, and other central nervous system or serous fluid disorders.
Regulatory Classification
Identification
An automated cell counter is a fully-automated or semi-automated device used to count red blood cells, white blood cells, or blood platelets using a sample of the patient's peripheral blood (blood circulating in one of the body's extremities, such as the arm). These devices may also measure hemoglobin or hematocrit and may also calculate or measure one or more of the red cell indices (the erythrocyte mean corpuscular volume, the mean corpuscular hemoglobin, or the mean corpuscular hemoglobin concentration). These devices may use either an electronic particle counting method or an optical counting method.
Predicate Devices
Flow Microscopy of Pleural Fluid, Peritoneal Fluid and Peritoneal Dialysate in The Yellow IRIS Urinalysis Workstation (K914256)
Flow Microscopy of Cerebrospinal Fluid and Seminal Fluid in The Yellow IRIS Urinalysis Workstation (K934539)
Flow Microscopy of Synovial Fluid and Pericardial Fluid in The Yellow IRIS Urinalysis Workstation (K954006)
Sysmex XE-2100 Series Hematology Analyzer Body Fluid Application (K040073)
Submission Summary (Full Text)
{0}------------------------------------------------
## Iris Diagnostics, A Division of IRIS International, Inc.
9172 Eton Avenue, Chatsworth, CA 91311 818-709-1244 FAX 818-700-9661 800-776-4747
#### 510(k) SUMMARY
### 510(k) Submission: K050235
## iQ® 200 Urine Analyzer Body Fluids Module
#### General Information:
| Date of Submission: | January 31, 2005 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name of Device: | iQ® 200 Urine Analyzer Body Fluids Module |
| Common Name of Device: | Instrument for performing RBC and nucleated cell counts in Cerebrospinal and Serous Body Fluids. |
| Classification Name: | Automated cell counter, 21CFR 864.5200, Class II device |
| Submitter's Name: | Harvey L. Kasdan, Ph.D. Chief Scientist Iris Diagnostics, A Division of IRIS International, Inc. 9172 Eton Avenue, Chatsworth, CA 91311 |
| Indications for Use (Brief): | Cerebrospinal fluid analysis is ordered by physicians to diagnose meningitis, intracranial hemorrhage, leukemias, malignancies and central nervous system disorders. Serous fluid analysis is ordered by physicians to diagnose infections, hemorrhages, malignancies and other disorders. Cell count determination is a part of these analyses. The iQ® 200 Urine Analyzer Body Fluids Module is used by a competent human observer to examine and count red blood cells and nucleated cells in cerebrospinal fluid and serous fluids. |
| Intended Use: | The iQ® 200 Urine Analyzer Body Fluids Module is an in-vitro diagnostic device used by a competent human observer to examine and count red blood cells and nucleated cells in cerebrospinal fluid and serous fluids. This module is a capability added to the iQ200 Urine Analyzer, a cleared urinalysis instrument (K022774). |
{1}------------------------------------------------
| Substantial Equivalence<br>to Predicate Devices: | The iQ®200 Urine Analyzer Body Fluids Module is substantially<br>equivalent to Flow Microscopy of Pleural Fluid, Peritoneal Fluid<br>and Peritoneal Dialysate in The Yellow IRIS Urinalysis<br>Workstation (K914256), Flow Microscopy of Cerebrospinal Fluid<br>and Seminal Fluid in The Yellow IRIS Urinalysis Workstation<br>(K934539), Flow Microscopy of Synovial Fluid and Pericardial<br>Fluid in The Yellow IRIS Urinalysis Workstation (K954006), and<br>the Sysmex XE-2100 Series Hematology Analyzer Body Fluid<br>Application (K040073), in its principles and technology, its<br>automated functions and the participative involvement of a<br>competent human observer, as well as to manual body fluid<br>microscopic cell-count examination by a competent observer using<br>a light microscope. | | | | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|-------|-------|---------------|
| Summary of<br>Technological<br>Characteristics: | Two aliquots from each body fluid specimen sample are<br>prepared. One aliquot is diluted in normal saline to provide a<br>concentration in the linear range of the instrument. The second<br>aliquot is treated with a lysing reagent to allow unambiguous<br>identification of nucleated cells by eliminating RBC confusion.<br>Particle images are captured and saved electronically as the sample<br>flows past a microscope objective at a high speed, electronically<br>concentrating particles. Particle images are then ordered by size<br>into assigned categories on a video display. A competent human<br>observer may change machine assignments, after which particle<br>concentrations are recomputed and reported. | | | | |
| Performance Studies: | Microscopic cell counting performance of the iQ® 200 Analyzer<br>Body Fluids Module was compared with manual chamber<br>counting. | | | | |
| Conclusions Drawn<br>From Clinical Tests: | Clinical trial performance data demonstrated that the iQ® 200<br>Analyzer Body Fluids Module is substantially equivalent to the<br>manual chamber counting method used for body fluid cell count<br>determination. Regression analysis of all iQ®200 Analyzer Body<br>Fluids Module CSF and Serous fluid sample cell counts on<br>corresponding manual chamber counts yielded the following<br>results with outliers removed: | | | | |
| | Cell | Number of Samples | R2 | Slope | Slope 95% CI |
| | RBC | 304 | 0.992 | 0.906 | 0.896 - 0.915 |
| | Nucleated Cells (NC) | 299 | 0.967 | 1.015 | 0.993 - 1.037 |
(RBC R2 = 0.973 and NC R2 = 0.940 with outliers.)
Non-zero intercepts were not statistically significant. Similar results were obtained for CSF and Serous Fluid alone. Paired ttests showed that the mean difference between replicate cell counts was not statistically different from zero. NCCLS EP6-A protocol demonstrated linear response from 0 to 10,000 particles/microliter.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR 2 3 2005
Harvey L. Kasdan, Ph.D. Chief Scientist Iris Diagnostics Division of IRIS International, Inc. 9172 Eton Avenue Chatsworth, California 91311
Re:
k050235 Trade/Device Name: iQ® 200 Urine Analyzer Body Fluids Module Regulation Number: 21 CFR § 864.5200 Regulation Name: Automated Cell Counter Regulatory Class: II Product Code: GKL Dated: January 31, 2005 Received: February 1, 2005
Dear Dr. Kasdan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{3}------------------------------------------------
Page 2 -
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
lobetz Beckerh
Robert L. Becker, Jr., MD, PK Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K050235
Device Name: iQ® 200 Urine Analyzer Body Fluids Module
#### Indications for Use:
The iQ 200 Urine Analyzer Body Fluids Module is an in-vitro diagnostic device used by a competent human observer to examine and count red blood cells and nucleated cells in cerebrospinal fluid and serous fluids.
Cerebrospinal fluid analysis is ordered by physicians to diagnose meningitis, intracranial hemorrhage, leukemias, malignancies and central nervous system disorders. Serous fluid analysis is ordered by physicians to diagnose infections, hemorrhages, malignancies and other disorders. Cell count determination is a part of these analyses.
The information produced by the iQ 200 Urine Analyzer Body Fluids Module concerning the cell concentrations in CSF and serous body fluids is ordered at the discretion of the physician, and is part of a larger body of laboratory and other test results available to assist the physician in health assessments or differential diagnoses. Cell count findings are always subject to judgment and interpretation by physicians relative to the patient's overall clinical presentation and history.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) |
|--|-------------------------------------------------------------------|
|--|-------------------------------------------------------------------|
Stephanie Bautista
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of
510(k) K050233
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.