K050226 · Smith & Nephew, Inc. · HWC · Feb 11, 2005 · Orthopedic
Device Facts
Record ID
K050226
Device Name
ASIAN INTRAMEDULLARY HIP SCREW NAILS
Applicant
Smith & Nephew, Inc.
Product Code
HWC · Orthopedic
Decision Date
Feb 11, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intramedullary Hip Screws are indicated for intracapsular fractures of the femoral neck; trochanteric or subtrochanteric fractures; osteotomies for patients with diseases or deformities of the hip; hip arthrodesis; supracondylar fractures and distal femoral fractures using a supracondylar plate; ipsilateral femoral shaft/neck fractures; intertrochanteric fractures; femoral neck fractures; subcapital fractures; comminuted neck and shaft fractures; femur reconstruction following tumor resection; leg length discrepancies secondary to femoral inequality; and prophylactic nailing of impending pathologic fractures. Asian Intramedullary Hip Screws and accessories are for single use only.
Device Story
Intramedullary hip screw system for proximal femur fracture management. Device consists of intramedullary nails (various diameters, lengths, neck-shaft angles; 4-degree medio-lateral bend for trochanteric insertion) and accessories (sliding/subtrochanteric lag screws, centering sleeves, set screws, compression screws, nail caps). Used by orthopedic surgeons in clinical/surgical settings to provide internal fixation for femoral fractures and reconstructions. Device provides mechanical stabilization of bone segments to facilitate healing. Single-use only.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Metallic bone fixation appliance. Includes intramedullary nails with medio-lateral bend and various lag screw options. Principle of operation is mechanical fixation identical to predicate devices.
Indications for Use
Indicated for patients with femoral fractures (intracapsular, trochanteric, subtrochanteric, supracondylar, distal, shaft, neck, subcapital, comminuted), hip deformities/diseases requiring osteotomy, hip arthrodesis, femur reconstruction post-tumor resection, leg length discrepancies, and prophylactic nailing for impending pathologic fractures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
| FEB 11 2005 | | 510(k) Summary of Safety and Effectiveness<br>Asian Intramedullary Hip Screw Nails | K 050226 |
|-------------------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|----------|
| Submitted by: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 Brooks Road<br>Memphis, TN 38116 | | |
| Date: | January 31, 2005 | | |
| Contact Person: | David Henley<br>Senior Regulatory Affairs Specialist | | |
| Proprietary Name: | Asian Intramedullary Hip Screw Nails | | |
| Common Name: | Intramedullary Hip Screw Nails and Accessories | | |
| Classification Name and Reference: | 21 CFR 888.3030 - Single/multiple component metallic<br>bone fixation appliances and accessories, Class II | | |
| Device Product Code and Panel Code: | KTT and HWC / Orthopaedics / 87 | | |
#### Device Description:
ray
Intramedullary Hip Screws and accessories provide an intramedullary approach to the management of fractures of the proximal femur. Devices include intramedullary nails with various diameters, lengths and neck shaft angles. The long nails are available in a right or left hand configuration. Each nail includes a 4 " medio-lateral bend to allow for greater trochanteric insertion. Each nail is used with a choice of specially designed sliding lag screws or subtrochanteric lag screws in various lengths. Additional accessories are available such as a centering sleeve, a set screw, a compression screw and nail caps.
#### Intended Use:
Smith & Nephew, Inc. Asian Intramedullary Hip Screws are indicated for intracapsular fractures of the femoral neck, trochanteric or subtrochanteric fractures; osteotomies for patients with directures on deformities of the hip; hip arthrodesis; supracondylar fractures and distal femoral fractures using a supracondylar plate; ipsilateral femoral shaft/neck fractures; intertrochanteric fractures; femoral neck fractures; subcapital fractures; comminuted neck and shaft fractures; femur reconstruction following; tumor resection; leg length discrepancies secondary to femoral inequality; and prophylactic nailing of impending pathologic fractures. Asian Intramedullary Hip Screws and accessories are for single use only.
#### Technological Characteristics:
The principle of operation of the subject devices is identical to that of the predicates. There are no changes in intended use, performance specifications or method of operation. A review of the test data for the subject devices indicates that they are equivalent to the predicate devices currently in clinical use and are capable of withstanding expected in vivo loading without failure.
#### Substantial Equivalence Information:
Substantial equivalence for the Asian Intramedullary Hip Screw is based on its similarities in indications for use, design features, operational principles, and material composition when compared to the predicate devices cleared under the following submissions: K040656, Smith & Nephew In the production K993289, Smith & Nephew Trauma Internal Fixation System; K040212, TriGen Internation System; and K952697, Ti Classic Compression Hip Screw.
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle. The logo is black and white.
FEB 1 1 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David Henley Senior Regulatory Affairs Specialist Smith & Nephew, Inc. Orthopedic Division 1450 Brooks Road Memphis, Tennessee 38116
Re: K050226
Trade/Device Name: Asian Intramedullary Hip Screw Regulation Number: 21 CFR 888.3030,21 CFR 888.3040 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories, Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC, KTT Dated: January 31, 2005 Received: February 1, 2005
Dear Mr. Henley:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass stated in to May 28, 1976, the enactment date of the Medical Device Amendments, or to conimeres prior to 110) 2011 11:15 accordance with the provisions of the Federal Food, Drug, de Moos mat have t (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Per (110) market the device, subject to the general controls provisions of the Act. The r ou mayy aross provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may be subject to back additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease oe actived that I Dr unation that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. David Henley
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
sincerely yours,
Mark A. Millman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Premarket Notification Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Asian Intramedullary Hip Screw
Indications for Use:
Intramedullary Hip Screws are indicated for intracapsular fractures of the femoral neck; trochanteric or subtrochanteric fractures; osteotomies for patients with diseases or deformities of the hip; hip arthrodesis; supracondylar fractures and distal femoral fractures using a supracondylar plate; ipsilateral femoral shaft/neck fractures; intertrochanteric fractures; femoral neck fractures; subcapital fractures; comminuted neck and shaft fractures; femur reconstruction following tumor resection; leg length discrepancies secondary to femoral inequality; and prophylactic nailing of impending pathologic fractures.
Asian Intramedullary Hip Screws and accessories are for single use only.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ and/or (Per 21 CFR 801, 109)
Over-the-Counter Use (Optional Format 1-2-96)
## (Please do not WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
C. Mark N. Mulhausen
al. Restorative, and Neurological Devices
510(k) Number K05022
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.