← Product Code [LNH](/productcode/LNH) · K050199

# MAGNETOM SYMPHONY A TIM SYSTEM (K050199)

_Siemens Medical Solution · LNH · Feb 18, 2005 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K050199

## Device Facts

- **Applicant:** Siemens Medical Solution
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Feb 18, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The MAGNETOM Symphony a Tim System is intended for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis. The MAGNETOM Symphony a Tim System may also be used for imaging during interventional procedures performed with MR compatible devices such as in room display and MR safe biopsy needles.

## Device Story

MAGNETOM Symphony a Tim System; magnetic resonance diagnostic device (MRDD). Inputs: magnetic resonance signals from patient anatomy. Transformation: system hardware processes signals to generate transverse, sagittal, coronal, and oblique cross-sectional images and spectroscopic data. Output: visual images/spectra for physician interpretation. Usage: clinical environment; operated by trained physicians/technicians. Supports interventional procedures using MR-compatible accessories (e.g., biopsy needles, in-room displays). Clinical benefit: provides internal structural/functional visualization to assist in medical diagnosis.

## Clinical Evidence

Bench testing only. Device performance and safety parameters verified against FDA-recognized NEMA standards and international IEC standards for Magnetic Resonance Imaging Devices.

## Technological Characteristics

1.5T Magnetic Resonance Diagnostic Device. Hardware components consistent with predicate MAGNETOM Avanto and Symphony systems. Supports cross-sectional imaging and spectroscopy. Compatible with MR-safe interventional accessories. Conforms to NEMA and IEC safety/performance standards.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Siemens MAGNETOM 1.5 T Avanto ([K032428](/device/K032428.md))
- Siemens MAGNETOM 1.5 T Symphony ([K971684](/device/K971684.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 12 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in This summary of 910(R) barry and of Safe Medical Device Act 1990 and 21 CFR § 807.92.

I. General Information

| Establishment       | Siemens Medical Solutions. Inc.<br>51 Valley Stream Parkway<br>Malvern. PA 19355                                                                               |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number | 2240869                                                                                                                                                        |
| Manufacturer        | Siemens AG. Bereich Med<br>Henkestrasse 127<br>D-91052 Erlangen. Germany                                                                                       |
| Registration Number | 8010024                                                                                                                                                        |
| Contact Person      | Ms. Ana Ladino<br>Technical Specialist, Regulatory Submissions<br>51 Valley Stream Parkway<br>Malvern. PA 19355<br>Phone: (610)448-1785<br>Fax: (610) 448-1787 |
| Device Name         | Trade Name: MAGNETOM Symphony a Tim System                                                                                                                     |
|                     | Classification Name: Magnetic Resonance Diagnostic Device                                                                                                      |
|                     | CFR Code: 21 CFR § 892.1000                                                                                                                                    |
|                     | Classification: Class II                                                                                                                                       |

#### Performance Standards

None established under Section 514 the Food, Drug, and Cosmetic Act.

Classification:

CONFIDENTIAL

{1}------------------------------------------------

### III. Safety and Effectiveness Information Supporting Substantial Equivalence.

#### Intended Use

The MAGNETOM Symphony a Tim System is intended for use as a magnetic resonance The MANDI on MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.

The MAGNETOM Symphony a Tim System may also be used for imaging during interventional procedures performed with MR compatible devices such as in room display and MR safe biopsy needles.

#### Device Description

Device Deseripati designed scanner. It consists of the same types of hardware that are currently available with the MAGNETOM Avanto and Symphony systems.

#### Substantial Equivalence

The system is substantially equivalent to the following cleared medical devices:

| Predicate Device Name           | FDA Clearance<br>Number | FDA Clearance Date |
|---------------------------------|-------------------------|--------------------|
| Siemens MAGNETOM 1.5 T Avanto   | K032428                 | 10/16/03           |
| Siemens MAGNETOM 1.5 T Symphony | K971684                 | 08/05/97           |

#### General Safety and Effectiveness Concerns:

Operation of the MAGNETOM Symphony a Tim System is substantially equivalent to the commercially available MAGNETOM 1.5 T Avanto System and 1.5 T Symphony System.

The MAGNETOM Symphony a Tim System will conform to the FDA recognized NEMA Standards for the measurement of performance and safety parameters and the international IEC standard for safety issues with Magnetic Resonance Imaging Devices. This will assure that the performance of this device can be considered safe and effective with respect to the currently available MAGNETOM Avanto and Symphony systems.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.

FEB 1 8 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Ana Ladino Technical Specialist, Regulatory Submissions Siemens Medical Solutions, USA, Inc. 51 Valley Stream Parkway MALVERN PA 19355

Re: K050199 Trade/Device Name: MAGNETOM Symphony a Tim System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: January 26, 2005 Received: January 28, 2005

Dear Ms. Ladino:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have teviewed your becaren be (a) p.
above and have determined the device is substantially equivalent (for the indications for use stated in above and have decemininarketed predicate devices marketed in interstate commerce prior to the chelosure) to regary mannent date of the Medical Device Amendments, or to devices that have been May 20, 1 770, the Chactinent dato of the Federal Food, Drug, and Cosmetic Act (Act) that recrassmen in accordantes internet approval application (PMA). You may, therefore, market the do not require approval of a promance appear. The Act. The general controls provisions of the Act uceries, subject to the generation, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket n your de rice is classinia (coo acch additional controls. Existing major regulations affecting your Apploval), it may of bacycer to and Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA de roce but of total announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I lease be advised mat rest in that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the I cacan statues and regalative, but not limited to registration and listing (21 CFR Part 807); labeling Act 3 requirements, moractice requirements as set forth in the quality systems (QS) (21 CFR Part 8017, good x20); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin hanceing of substantial equivalence of your device to a legally premarked notheation: The PDA maing of basisification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your cerros of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, picase note the regulation entitled, formation on your responsibilities under the Act from the 607.77). Tou may obtain outer general and Consumer Assistance at its toll-free number (800) DVISion of 301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsmadsmam.html

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## 3 Indications for Use Statement

Koro199 510(k) Number (if known)

Device Name: MAGNETOM Symphony a Tim System

Indications for Use:

The "MAGNETOM Symphony a Tim System" is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.

The "MAGNETOM Symphony a Tim System" may also be used for imaging during interventional procedures performed with MR compatible devices such as, in room display and MR safe biopsy needles.

(please do not write below this line- continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation Prescription Use OR Over-The-Counter Use Division Sinn-Of Division of Reproductive. Ab and Radiological Device 510(k) Number

Siemens 510(k) Premarket Notification MAGNETOM Symphony a Tim System

January 26, 2005

Page 14

CONFIDENTIAL

---

**Source:** [https://fda.innolitics.com/device/K050199](https://fda.innolitics.com/device/K050199)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
