← Product Code [GKZ](/productcode/GKZ) · K050057

# COULTER LH 750 HEMATOLOGY ANALYZER (K050057)

_Beckman Coulter, Inc. · GKZ · Feb 10, 2005 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K050057

## Device Facts

- **Applicant:** Beckman Coulter, Inc.
- **Product Code:** [GKZ](/productcode/GKZ.md)
- **Decision Date:** Feb 10, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The LH 750 Series Body Fluid Application is a procedure for obtaining in vitro quantitative determinations of leukocytes (WBC) and erythrocytes (RBC) in cerebrospinal fluid, serous fluids, and synovial fluid, using a Coulter LH 700 Series Hematology Analyzer.

## Device Story

Coulter LH 750 is a quantitative hematology analyzer; performs CBC, leukocyte differential, reticulocyte analysis, and nucleated RBC counts. For body fluid analysis, samples are manually aspirated and diluted in separate WBC and RBC baths. Device utilizes Coulter Principle (electrical impedance) to detect/measure resistance changes as cells pass through apertures; pulses are channelized, corrected for coincidence, and calibrated to yield WBC and RBC counts. System uses isotonic diluent, lytic reagents, and instrument cleaner. Operated in manual mode by laboratory personnel. Output provides quantitative cell counts for clinical diagnostic use. Modification in this submission updates sample stability claims for synovial and serous fluids.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, risk analysis, and Declaration of Conformity with Design Controls confirming product specifications were met.

## Technological Characteristics

Uses Coulter Principle (electrical impedance) and Coulter VCS (volume, conductivity, light scatter) technology. Reagents include isotonic diluent, lytic reagents, and instrument cleaner. System performs automated dilution and mixing. Connectivity via barcode scanner or manual input. Software-controlled analysis.

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- Coulter LH 750 Body Fluids Application ([K030606](/device/K030606.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE

A. 510(k) Number:
K050057

B. Purpose for Submission:
Special 510(k) for a modification of sample stability claims

C. Manufacturer and Instrument Name:
Beckman Coulter Coulter® LH 750 Hematology Analyzer

D. Type of Test or Tests Performed:
The Coulter® LH 750 Hematology Analyzer is capable of providing a complete blood cell analysis (CBC) including a differential leukocyte cell count. This application is for the enumeration of leukocytes (WBC) and erythrocytes (RBC) in body fluids.

E. System Descriptions:

1. Device Description: The Coulter® LH 750 is a quantitative hematology analyzer with complete blood count (CBC), leukocyte differential, reticulocyte analysis, and nucleated RBC capability. It uses a reagent system consisting of isotonic diluent, lytic reagents to lyse the red cells without significantly affecting the white cells and an instrument cleaner. Additionally, the system includes reagents used for reticulocyte staining and analysis.

2. Principles of Operation: The Coulter® LH 750 uses the Coulter Principle for enumerating and sizing blood cells, automatic diluting and mixing for sample processing and a single beam photometer for hemoglobinometry. It utilizes Coulter VCS (volume, conductivity, and light scatter) technology for WBC differential analysis and classification and reticulocyte analysis.

For body fluid analysis, specimen is aspirated into the LH 700 via the manual mode and is diluted in separate WBC and RBC baths. The Coulter Method of counting cells is used to detect and measure changes in electrical resistance when a cell passes through a small aperture. The resistance generates an electrical pulse. The accumulation of electrical pulses are channelized, processed for coincidence correction, and multiplied by a calibration factor, yielding the WBC and RBC counts.

{1}

2

3. Modes of Operation:
Manual mode.

4. Specimen Identification:
Bar code scanner or manual input

5. Specimen Sampling and Handling:
Samples are manually mixed and handled

6. Calibration:
User calibration with whole blood and/or commercial calibration materials.

7. Quality Control:
Commercial quality control material is available for daily monitoring of the instrument.

8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X or No ☐
The software was reviewed and cleared in a previous submission for this instrument (K032342)

F. Regulatory Information:

1. Regulation section:
21 CFR 864.5220

2. Classification:
II

3. Product code:
GKZ

4. Panel:
81

{2}

G. Intended Use:

1. Indication(s) for Use:

The LH 750 Series Body Fluid Application is a procedure for obtaining in vitro quantitative determinations of leukocytes (WBC) and erythrocytes (RBC) in cerebrospinal fluid, serous fluids, and synovial fluid, using a Coulter LH 700 Series Hematology Analyzer.

2. Special Conditions for Use Statement(s):

H. Substantial Equivalence Information:

1. Predicate Device Name(s) and 510(k) numbers:

Coulter LH 750 Body Fluids Application - K030606

2. Comparison with Predicate Device:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | The LH 750 Series Body Fluid Application is a procedure for obtaining in vitro quantitative determinations of leukocytes (WBC) and erythrocytes (RBC) in cerebrospinal fluid, serous fluids, and synovial fluid, using a Coulter LH 700 Series Hematology Analyzer. | same  |
|  Reagent systems | Coulter® LH 700 Series or Isoton 4 diluent, Coulter Lyse S® III or Lyse S® 4, Coulter Clenz®, Coulter Latron™ Primer and Latron Control, Coulter LH 700 Series Pak | same  |
|  Operating principle | Coulter principle and Coulter VCS | same  |

{3}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Sample stability | 8 hrs for synovial fluid analysis and 8 hrs for serous fluids | 1 hr for synovial fluid and 24 hrs for serous fluids  |

I. Special Control/Guidance Document Referenced (if applicable):

J. Performance Characteristics:

1. Analytical Performance:
a. Accuracy:
b. Precision/Reproducibility:
c. Linearity:
d. Carryover:
e. Interfering Substances:

2. Other Supportive Instrument Performance Data Not Covered Above:

As required for a Special 510(k), the Sponsor has provided a risk analysis as well as a Declaration of Conformity with Design Controls indicating that development activities were conducted under appropriate design controls procedures, and the overall product specifications were met.

{4}

K. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

L. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

5

---

**Source:** [https://fda.innolitics.com/device/K050057](https://fda.innolitics.com/device/K050057)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
