← Product Code [EHD](/productcode/EHD) · K043557

# MODEL PC-1000 PANORAMIC/CEPHALOMETRIC X-RAY SYSTEMS (K043557)

_Panoramic Corp. · EHD · Jan 13, 2005 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K043557

## Device Facts

- **Applicant:** Panoramic Corp.
- **Product Code:** [EHD](/productcode/EHD.md)
- **Decision Date:** Jan 13, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.1800
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

For diagnostic radiographic use in dental, oral surgery, and orthodontic practices.

## Device Story

PC-1000 Panoramic/Cephalometric X-ray system provides panoramic dental images; covers 70-80% more anatomy than standard full-mouth series. Device features inscribed lines and angled mirror for Frankfort/midsagittal plane alignment; numbered temple supports for skull density measurement to optimize kVp settings. System is free-standing; plugs into standard 115V outlet. Operated by dental/orthodontic personnel in clinical settings. Output is a diagnostic radiographic image used by clinicians to assess dental/oral anatomy and support treatment planning. Benefits include reduced guesswork in radiography, improved anatomical visualization, and simplified installation/operation compared to previous models.

## Clinical Evidence

Bench testing only. No clinical data provided. Compliance with 21 CFR Part 1020 and international standards (IEC 60601-1, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-1-2) confirmed safety and effectiveness.

## Technological Characteristics

Extraoral X-ray system; 90 kVp max output; 0.5 mm x 0.5 mm focal spot; 6 mA tube current; 12-second exposure time. Energy source: 115-120V AC. Control system: 3 microprocessors (upgraded from analog hardware). User interface: Touch controls. Standards: 21 CFR Part 1020, IEC 60601-1 (1988-12), IEC 60601-2-7, IEC 60601-2-28, IEC 60601-1-2. UL listed.

## Regulatory Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

## Predicate Devices

- Model PC-1000 Panoramic / Cephalometric X-ray System ([K882436](/device/K882436.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K043557
Page 1 of 2

JAN 1 3 2005

EXHIBIT 2 510(k) Summary PANORAMIC CORP. 4321 Goshen Rd. Fort Wayne, IN 46818 800.654.2027 260.489.2291 Fax 260.489.5683 www.pancorp.com December 14, 2004 Contact: Doug Pack, Director of Operations.

- 1. Identification of the Device: Proprietary-Trade Name: Model PC-1000 Panoramic / Cephalometric X-ray Systems Classification Name: Unit, x-ray, extraoral with timer Product Code: EHD Common/Usual Name: Panoramic Dental X-Ray System
- 2. Equivalent legally marketed devices: Model PC-1000 Panoramic / Cephalometric X-ray System K882436
- 3. Indications for Use (intended use) For diagnostic radiographic use in dental, oral surgery, and orthodontic practices.
- 4. Description of the Device: The PC-1000 panoramic X-ray machine provides a panoramic picture, showing 70-80% more anatomy than a standard series of full-mouth X-rays. Inscribed lines and an angled mirror aid in accurately establishing the Frankfort plane and the midsagittal plane. In addition, the numbered temple supports allow accurate measurement of skull density for proper kVp settings, eliminating the guess work typical in panoramic radiography. Panoramic's quick and easy installation allows more time to be spent on demonstration. As your new Panoramic X-ray is installed, our service representative will test and calibrate the equipment, as well as teach your personnel how to use it. Panoramic provides you with a professional operator's manual in both print and on video tape so you can teach new personnel the operation of the PC-1000 as well as refresh the memory of your current staff. Each free-standing PC-1000 is shipped and delivered almost completely assembled, avoiding the noise and hours of disruption that is common with unassembled models. Also, the PC-1000 plugs into a standard 115-volt outlet eliminating the need for expensive rewiring or modifications to one's office.
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench, user, and standards testing indicates that the new device is as safe and effective as the predicate devices.

{1}------------------------------------------------

K043551
Paylerz

|  | J. Substantial Equivalence Chart, Model PC-1000 Panoramic / Cephalometric X-ray Systems<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |  |
|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|

| Characteristic             | Model PC-1000 Panoramic /<br>Cephalometric X-<br>ray Systems<br>K882436                 | Modified Model<br>PC-1000 Panoramic<br>/ Cephalometric X-<br>ray Systems |
|----------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Intended Use:              | Diagnostic Dental<br>X-Rays (Panoramic<br>and Cephalometric)                            | SAME                                                                     |
| Energy Source:             | 120 V Ac 20 A                                                                           | SAME                                                                     |
| User Interface             | Dedicated Controls                                                                      | Touch Controls                                                           |
| Maximum output             | 90 kVp                                                                                  | SAME                                                                     |
| Focal spot                 | 0.5 mm x 0.5 mm                                                                         | SAME                                                                     |
| Tube Current               | 6 ma                                                                                    | SAME                                                                     |
| Exposure time              | 12 seconds                                                                              | SAME                                                                     |
| Method of Control          | Analog Hardware                                                                         | (3) Microprocessors                                                      |
| Performance<br>Standard    | Complies with 21<br>CFR Part 1020.                                                      | SAME                                                                     |
| International<br>Standards | IEC 60601-1 (1988-<br>12) & amend.<br>IEC 60601-2-7 IEC<br>60601-2-28,<br>IEC 60601-1-2 | SAME                                                                     |
| Safety                     | UL Listing                                                                              | SAME                                                                     |

## 7. Conclusion

After analyzing bench, user, and standards testing data, it is the conclusion of Panoramic Corporation that the Model PC-1000 Panoramic / Cephalometric X-ray Systems are as safe and effective as the predicate device, have few technological differences, and have no new indications for use, thus rendering them substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

JAN 1 3 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Panoramic Corp. % Mr. Daniel Kamm, P.E. Regulatory Engineer Kamm & Associates PO Box 7007 DEERFIELD IL 60015

Re: K043557 Trade/Device Name: Model PC-1000 Panoramic/ Cephalometric X-ray Systems Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: 90 EHD

Dated: December 24, 2004 Received: December 27, 2004

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use strect in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing provisions of the Al and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a deternination that your device complies with other requirements of the Act any Federal statutes and regulations administered by other Federal agencies. You must comply with the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807) vilan and (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems(QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Nancy C. brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: Model PC-1000 Panoramic / Cephalometric X-ray Systems . (Modification to K882436)

Indications For Use: For diagnostic radiographic use in dental, oral surgery, and orthodontic practices.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Nancy C. Brydon

(Division Sign-Off) (Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

---

**Source:** [https://fda.innolitics.com/device/K043557](https://fda.innolitics.com/device/K043557)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).