K043517 · Nihon Kohden America, Inc. · DRT · Feb 3, 2005 · Cardiovascular
Device Facts
Record ID
K043517
Device Name
TRANSMITTER, ZS-940PA SERIES
Applicant
Nihon Kohden America, Inc.
Product Code
DRT · Cardiovascular
Decision Date
Feb 3, 2005
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device, ZS-940PA series, is intended to transmit physiological data, such as: electrocardiogram (ECG), respiration, blood oxygen saturation (SpO2), noninvasive blood pressure (NIBP), and pulse waveform from a patient to a Nihon Kohden monitor for continuous monitoring applications. The transmitter can change channels when connected to the QI-910PK channel writer. The front LCD displays SpO2%, NIBP, pulse rate, pulse waveform amplitude, electrode condition mark, battery condition and NIBP measuring mode and interval. The system communicates in the new WMTS band as required by FCC. The monitored physiological parameters communicated via radiofrequency can be displayed, stored and printed at a Nihon Kohden Central monitor, such as WEP-4000 or ORG-9700/CNS-9700. The device will be available for sale only upon the order of a physician or licensed health care provider and it will be available for use by medical personnel on all patient populations within a medical facility, including ICU, CCU, recovery room and general ward.
Device Story
ZS-940PA Transmitter is a wearable device for continuous physiological monitoring. It collects ECG, respiration, SpO2, NIBP, and pulse waveform data from patients. The device is placed in a non-latex blood pressure cuff pocket on the patient's arm. It transmits physiological signals via radio frequency (WMTS band) to a Nihon Kohden central monitor (e.g., WEP-4000, ORG-9700/CNS-9700) for display, storage, and printing. The device features a front LCD showing SpO2%, NIBP, pulse rate, pulse waveform amplitude, electrode status, and battery condition. It is operated by medical personnel in clinical settings like ICUs and general wards. The system facilitates real-time patient monitoring, allowing healthcare providers to observe vital signs and make informed clinical decisions. It benefits patients by enabling continuous, remote monitoring of critical physiological parameters without tethering them to a bedside monitor.
Clinical Evidence
Bench testing only. The device underwent electromagnetic, environmental, safety, and performance testing to verify intended operation. Design validation of software and hardware was performed to document safety and efficacy. No clinical data was required as the device does not directly contact patients and uses established accessories.
Technological Characteristics
Wireless transmitter operating in the WMTS band. Features an LCD for local data display. Powered by battery. Non-invasive design; placed in a blood pressure cuff pocket. Complies with IEC 60601-1 and 21 CFR Part 898. No sterilization required. Software-based monitoring and transmission functionality.
Indications for Use
Indicated for continuous monitoring of ECG, respiration, SpO2, NIBP, and pulse waveform in medical facilities (ICU, CCU, recovery, general ward). Use by medical personnel on all patient populations, including neonates, children, and adults.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Reference Devices
WEP-4000
ORG-9700/CNS-9700
QI-910PK channel writer
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) NOTIFICATION ZS-940PA Transmitter
14-3511
## FEB - 3 2005
## SECTION 2 - 510(K) SUMMARY
# Name and Address of Applicant
Phone: (949) 580-1555 Nihon Kohden America, Inc. Dial 9 and then Ext. 4401 Fax: (949) 580-1550 90 Icon St. Attn: Serrah Namini, Foothill Ranch, Ca 92610 Regulatory Affairs Assoc. Dir.
### Indications for Use:
The device, ZS-940PA series, is intended to transmit physiological data, such as: electrocardiogram (ECG), respiration, blood oxygen saturation (SpO2), noninvasive blood pressure (NIBP), and pulse waveform from a patient to a Nihon Kohden monitor for continuous monitoring applications. The transmitter can change channels when connected to the QI-910PK channel writer. The front LCD displays SpO2%, NIBP, pulse rate, pulse waveform amplitude, electrode condition mark, battery condition and NIBP measuring mode and interval. The system communicates in the new WMTS band as required by FCC. The monitored physiological parameters communicated via radiofrequency can be displayed, stored and printed at a Nihon Kohden Central monitor, such as WEP-4000 or ORG-9700/CNS-9700.
The device will be available for sale only upon the order of a physician or licensed health care provider and it will be available for use by medical personnel on all patient populations within a medical facility, including ICU, CCU, recovery room and general ward.
The device complies with IEC 60601-1 sub-clause 56.3(c) implemented by 21 CFR Part 898 Performance Standard for Electrode Lead Wires and Patient Cables. The device is also in compliance with applicable sections of IEC standard as listed in this application.
Sterilization is not required for the device.
The device does not directly contact patients; It is place in a pocket of a non-latex blood pressure cuff on a patient's arm. Accessories that contact patients, such as SpO2 probes, are the same accessories as used with other legally marketed products or are comprised of the same component materials as the accessories of predicate devices. Therefore, good laboratory practice studies required per 21 CFR Part 58 are not applicable for this submission.
The device was subject to electromagnetic, environmental, safety and performance testing procedures. These tests verified the intended operation of the device. Safety and efficacy was documented through Design validation of the software and hardware of the device.
Therefore, Nihon Kohden believes that the device is substantially equivalent to the Nihon Kohden predicate devices.
{1}------------------------------------------------
K043517
p2/2
## SECTION 3 - PROPOSED LABELING
#### A. Intended Use
The device is intended to monitor, display and transfer physiological data. Specifically, the device is intended to monitor heart rate, pulse rate, blood oxygen saturation (SpO22), respiratory rate, and non- invasive blood pressure (NIBP). This device may be used to transmit physiological signals via radio frequency. The device will be available for use by medical personnel on all patient populations within a medical facility.
#### Device/Package Labels B.
The proposed product labels for the device are located in Attachment # A .
#### C. Proposed Packaging
Packaging drawings/ specifications for the device is depicted in Attachment # A.
#### D. Instructions for Use
The proposed instructions for use is provided with each packaged device and is presented in Attachment # H, Operator's Manual .
#### Advertisement/Promotional Literature E.
To date, no advertisement or promotional literature has been created for the distribution in the United States.
#### F. Contraindications, Precautions & Warnings
Warnings and Cautions listed in the Operator's Manual are depicted in Attachment # B.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 3 2005
Nihon Kohden America, Inc. c/o Ms. Serrah Namini RA Associate Director 90 Icon St. Foothill Ranch, CA 92610
Re: K043517
Trade Name: ZS-940PA Transmitter Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: DRT Dated: December 16, 2004 Received: December 20, 2004
Dear Ms. Namini:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section >10(x) premainter as abstantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosure) to tegary manated president Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior to May 26, 1976, the enaordance with the provisions of the Federal Food, Drug.
In a lead have been reclassified in accordance with the provisions of the United devices that have been recuire approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a presisions of the Act . Je and Costheric Ac. (Act) that do not require approvince subject to the general controls provisions of the Act. The You may, merefore, mance the do rece, solo, e requirements for annual registration, listing of the general controls provisions of the flor labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can
may be subject to such additional controls. Existing major may be subject to such additional bonatorior Linkstong Collections, Title 21, Parts 800 to 898. In addition, FDA may be found in the Ood of Pouch oncerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Serrah Namini
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that IDA's issualled of a succession with other requirements of the Act
that FDA has made a determination that your device complies with other requirements that FDA has made a dolorimiation administered by other Federal agencies. You must or any Federal statutes and regalations and adding, but not limited to: registration and listing (21 comply with an the Act 3 requirements, more and manufacturing practice requirements as set CFR Part 807), labeling (21 CFR Part 807), good if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Declons our device as described in your Section 510(k) This letter will anow you to ogin manseans your and equivalence of your device to a legally premarket notheation: "The PDF mining of easification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acvice for your as (240) 276-0120 Also, please note the regulation entitled, Comact the Office of Compularse an (21CFR Part 807.97). You may obtain Misbranding by reference to premantonitiiiiiiiis under the Act from the Division of Small other general information on your responses Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
B. Zimmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## NIHON KOHDEN AMERICA, INC.
510(k) NOTIFICATION ZS-940PA Transmitter
G. Indications for Use Statement
K043517 510(k) Number (if known): __
Device Name: ZS-940PA Series
Indications for Use:
The device is intended to transmit physiological data, such as: electrocardiogram (ECG), respiration, blood oxygen saturation (SpO2), noninvasive blood pressure (NIBP), and pulse waveform from a patient to a Nihon Kohden monitor for continuous monitoring. The transmitter can change channels when connected to the QI-910PK channel writer. The front LCD displays SpO2%, NIBP, pulse rate, pulse waveform amplitude, electrode condition mark, battery condition and NIBP measuring mode and interval. The system communicates in the new WMTS band as required by FCC. The device will be available for use within a medical facility, including ICU, CCU, recovery room and general ward by medical personnel on adults and children for NIBP and all other functions for all patient populations, including neonates.
Prescription Use X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K043517 B/Ammer
i Cardiovascular Devices 510(k) Number
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.