← Product Code [LCX](/productcode/LCX) · K043443

# ONE STEP HCG URINE PREGNANCY TEST (K043443)

_Guangzhou Wondfo Biotech Co., Ltd. · LCX · Apr 8, 2005 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K043443

## Device Facts

- **Applicant:** Guangzhou Wondfo Biotech Co., Ltd.
- **Product Code:** [LCX](/productcode/LCX.md)
- **Decision Date:** Apr 8, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1155
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Guangzhou Wondfo Biotech Co., Ltd. One Step HCG Urine Pregnancy Test is intended for non-professional, over-the-counter use and for professional and laboratory use for the qualitative identification of the elevated levels of human Chorionic Gonadotrophin (hCG) in urine to aid in the determination of pregnancy.

## Device Story

Solid-phase, sandwich immunochromatographic assay; detects hCG in urine. User soaks absorbent pad with urine; sample migrates via capillary action through membrane coated with mouse monoclonal anti-α-hCG antibody and pad containing mouse monoclonal anti-β-hCG antibody colloidal gold conjugate. Presence of hCG forms colored line in test region; control line indicates valid test. Used in home or clinical settings; operated by patients or healthcare professionals. Visual readout at 5 minutes; positive result indicates elevated hCG levels consistent with pregnancy. Benefits include rapid, qualitative pregnancy confirmation.

## Clinical Evidence

No clinical data provided; substantial equivalence established via bench testing and comparison to legally marketed predicate devices.

## Technological Characteristics

Solid-phase, sandwich immunochromatographic assay. Materials: membrane coated with mouse monoclonal anti-α-hCG antibody; pad with mouse monoclonal anti-β-hCG antibody colloidal gold conjugate; goat anti-mouse IgG control. Formats: strip, cassette, midstream. Storage: 4–30 °C. Traceable to WHO 3rd International Standard. No electronic components or software.

## Regulatory Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

## Predicate Devices

- WHPM One-Step Pregnancy Test (k041273)
- Acon Quick-Check II Home pregnancy Test (k032510)
- Acon Quick-Check II Home pregnancy Test Strip (k033041)
- Accutest hCG-Combo (k032987)
- E.P.T. Pregnancy Test (k033658)
- Unimark Home Pregnancy Test Device (k032992)
- Good Morning Test (k031798)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
k043443

B. Purpose for Submission:
Clearance of new device

C. Measurand:
Human Chorionic Gonadotrophin (hCG)

D. Type of Test:
Qualitative chromatographic immunoassay

E. Applicant:
Guangzhou Wondfo Biotech Co., Ltd.

F. Proprietary and Established Names:
One Step HCG Urine Pregnancy Test

G. Regulatory Information:
1. Regulation section:
21 CFR §862.1155, Human chorionic gonadotropin (HCG) test system
2. Classification:
Class II
3. Product code:
LCX, kit, test, pregnancy, hCG, over the counter
JHI, radioimmunoassay, human chorionic gonadotropin
4. Panel:
Clinical Chemistry (75)

H. Intended Use:
1. Intended use(s):
The Guangzhou Wondfo Biotech Co., Ltd. One Step HCG Urine Pregnancy Test

{1}

is intended for non-professional, over-the-counter use and for professional and laboratory use for the qualitative identification of the elevated levels of human Chorionic Gonadotrophin (hCG) in urine to aid in the determination of pregnancy.

2. Indication(s) for use:
See Intended Use above.

3. Special conditions for use statement(s):
For over-the-counter and professional use

4. Special instrument requirements:
None

I. Device Description:
The One Step HCG Urine Pregnancy Test will be sold in three formats: cassette, test strip, and midstream. The test strip and midstream kits consist of one test device and a package insert. The cassette kit consists of one test device and a disposable plastic dropper, and a package insert. Each test device contains mouse monoclonal anti-α-hCG antibody coated membrane and a pad containing mouse monoclonal anti-β-hCG antibody colloidal gold conjugate. The control antibodies are goat anti-mouse IgG.

J. Substantial Equivalence Information:

1. Predicate device name(s):
WHPM One-Step Pregnancy Test
Acon Quick-Check II Home pregnancy Test
Acon Quick-Check II Home pregnancy Test Strip
Accutest hCG-Combo
E.P.T. Pregnancy Test
Unimark Home Pregnancy Test Device
Good Morning Test

2. Predicate 510(k) number(s):
k041273
k032510
k033041
k032987
k033658
k032992
k031798

3. Comparison with predicate:

{2}

The device and the predicate devices share the same intended use and test methodology.

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Formats Available | Strip, Cassette, Midstream | Strip, Cassette, Midstream  |
|  Specimen type | Urine | Urine  |
|  Antibodies | Goat, mouse | Goat, mouse  |
|  Cutoff | 25 mIU/mL | 25 mIU/mL  |
|  Storage Temperature | 4 – 30 °C | 4 – 30 °C  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Read time | 3 – 5 minutes | 3 – 10 minutes  |

## K. Standard/Guidance Document Referenced (if applicable):

None

## L. Test Principle:

The device is a solid phase, sandwiched immunochromatographic assay. Users are instructed to soak the absorbent pad with urine. The urine will migrate via capillary action toward the result and control window. If hCG is present in the urine it reacts with an anti-HCG-colloidal gold particle conjugate to form a colored line in the test region of the strip. A colored line in the control region of the device indicates adequate sample volume and capillary action. Absence of a colored line in the control region is an indication of an invalid result. Users are instructed to read the device at 5 minutes.

## M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:

Precision of the device was evaluating by testing negative urine samples spiked with hCG (to 500 IU/ml, 200 IU/ml, 100 IU/ml, 5 IU/ml, 1000 mIU/ml, 500 mIU/ml, 100 mIU/ml, 50 mIU/mL, 25 mIU/ml, 20 mIU/ml, 15 mIU/ml, and 10 mIU/ml hCG), one positive urine sample (containing 29 mIU/mL hCG), and one negative urine sample 10 times with each of three lots of product (total n/sample = 30). The results are summarized below:

{3}

|  hCG | Tested | Lot I | Lot II | Lot III  |
| --- | --- | --- | --- | --- |
|  500 IU/mL | 1 | - | - | -  |
|   |  2 | - | - | -  |
|   |  3 | - | - | -  |
|   |  4 | - | - | -  |
|   |  5 | - | - | -  |
|   |  6 | - | - | -  |
|   |  7 | - | - | -  |
|   |  8 | - | - | -  |
|   |  9 | - | - | -  |
|   |  10 | - | - | -  |
|  |   |   |   |   |
|  200 IU/mL | 1 | + | + | +  |
|   |  2 | + | + | +  |
|   |  3 | + | + | +  |
|   |  4 | + | - | +  |
|   |  5 | + | + | +  |
|   |  6 | + | + | +  |
|   |  7 | + | + | +  |
|   |  8 | - | + | -  |
|   |  9 | + | + | +  |
|   |  10 | + | + | +  |
|  |   |   |   |   |
|  100 IU/mL | 1 | + | + | +  |
|   |  2 | + | + | +  |
|   |  3 | + | + | +  |
|   |  4 | + | + | +  |
|   |  5 | + | + | +  |
|   |  6 | + | + | +  |
|   |  7 | + | + | +  |
|   |  8 | + | + | +  |
|   |  9 | + | + | +  |
|   |  10 | + | + | +  |
|  |   |   |   |   |
|  5 IU/mL | 1 | + | + | +  |
|   |  2 | + | + | +  |
|   |  3 | + | + | +  |
|   |  4 | + | + | +  |
|   |  5 | + | + | +  |
|   |  6 | + | + | +  |
|   |  7 | + | + | +  |
|   |  8 | + | + | +  |
|   |  9 | + | + | +  |
|   |  10 | + | + | +  |
|  |   |   |   |   |
|  25 MIU/ml | 1 | + | + | +  |
|   |  2 | + | + | +  |
|   |  3 | + | + | +  |
|   |  4 | + | + | +  |
|   |  5 | + | + | +  |
|   |  6 | + | + | +  |
|   |  7 | + | + | +  |
|   |  8 | + | + | +  |
|   |  9 | + | + | +  |
|   |  10 | + | + | +  |
|  |   |   |   |   |
|  20 MIU/ml | 1 | - | - | -  |
|   |  2 | + | - | -  |
|   |  3 | - | - | -  |
|   |  4 | - | - | -  |
|   |  5 | - | - | -  |
|   |  6 | - | - | -  |
|   |  7 | - | - | -  |
|   |  8 | - | - | -  |
|   |  9 | - | - | -  |
|   |  10 | - | - | -  |

{4}

|  hCG | Tested | Lot I | Lot II | Lot III  |
| --- | --- | --- | --- | --- |
|  1000 mIU/mL | 1 | + | + | +  |
|   |  2 | + | + | +  |
|   |  3 | + | + | +  |
|   |  4 | + | + | +  |
|   |  5 | + | + | +  |
|   |  6 | + | + | +  |
|   |  7 | + | + | +  |
|   |  8 | + | + | +  |
|   |  9 | + | + | +  |
|   |  10 | + | + | +  |
|  |   |   |   |   |
|  500 mIU/mL | 1 | + | + | +  |
|   |  2 | + | + | +  |
|   |  3 | + | + | +  |
|   |  4 | + | + | +  |
|   |  5 | + | + | +  |
|   |  6 | + | + | +  |
|   |  7 | + | + | +  |
|   |  8 | + | + | +  |
|   |  9 | + | + | +  |
|   |  10 | + | + | +  |
|  |   |   |   |   |
|  100 mIU/mL | 1 | + | + | +  |
|   |  2 | + | + | +  |
|   |  3 | + | + | +  |
|   |  4 | + | + | +  |
|   |  5 | + | + | +  |
|   |  6 | + | + | +  |
|   |  7 | + | + | +  |
|   |  8 | + | + | +  |
|   |  9 | + | + | +  |
|   |  10 | + | + | +  |
|  hCG | Tested | Lot I | Lot II | Lot III  |
| --- | --- | --- | --- | --- |
|  15 mIU/ml | 1 | - | - | -  |
|   |  2 | - | - | -  |
|   |  3 | - | - | -  |
|   |  4 | - | - | -  |
|   |  5 | - | - | -  |
|   |  6 | - | - | -  |
|   |  7 | - | - | -  |
|   |  8 | - | - | -  |
|   |  9 | - | - | -  |
|   |  10 | - | - | -  |
|  |   |   |   |   |
|  10 mIU/ml | 1 | - | - | -  |
|   |  2 | - | - | -  |
|   |  3 | - | - | -  |
|   |  4 | - | - | -  |
|   |  5 | - | - | -  |
|   |  6 | - | - | -  |
|   |  7 | - | - | -  |
|   |  8 | - | - | -  |
|   |  9 | - | - | -  |
|   |  10 | - | - | -  |
|  |   |   |   |   |
|  Negative (0 mIU/mL) | 1 | - | - | -  |
|   |  2 | - | - | -  |
|   |  3 | - | - | -  |
|   |  4 | - | - | -  |
|   |  5 | - | - | -  |
|   |  6 | - | - | -  |
|   |  7 | - | - | -  |
|   |  8 | - | - | -  |
|   |  9 | - | - | -  |
|   |  10 | - | - | -  |

# b. Linearity/assay reportable range:

High dose hook effect was evaluated by spiking increasing hCG concentrations into negative urine samples and evaluating the test result lines. The sponsor concludes that the test can be used for samples up to  $100,000\mathrm{mIU / mL}$ . Results are summarized below (units = mIU/mL).

{5}

|  hCG concentration | Test Result  |
| --- | --- |
|  12.5 | +/-  |
|  25 | +  |
|  100 | ++  |
|  hCG concentration | Test Result  |
|  500 | +++  |
|  1000 | +++  |
|  5000 | ++  |
|  25,000 | +  |
|  100,000 | +  |
|  200,000 | +/-  |
|  500,000 | -  |

The sponsor repeated the high dose hook effect evaluation during the precision studies and the following results were obtained (see expanded preision data above):

|  hCG concentration | Test result  |
| --- | --- |
|  10 mIU/ml | -  |
|  15 mIU/ml | -  |
|  20 mIU/ml | -/+  |
|  25 mIU/ml | +  |
|  100 mIU/ml | ++  |
|  500 mIU/ml | +++  |
|  1000 mIU/ml | +++  |
|  5 IU/ml | ++  |
|  25 IU/ml | +  |
|  100 IU/ml | +  |
|  200 IU/ml | -/+  |
|  500 IU/ml | -  |

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

This device is traceable to the WHO  $3^{\mathrm{rd}}$  International standard.

Protocols and acceptance criteria were described for stability testing including validation of shipping stability.

d. Detection limit:

The sensitivity of the device was tested by spiking 60 negative urine samples with varying concentrations of hCG from a standard sample. Results are summarized below (units = mIU/mL).

{6}

|  Concentration | Positive | Negative | Total n  |
| --- | --- | --- | --- |
|  0 | 0 | 5 | 5  |
|  12.5 | 0 | 5 | 5  |
|  18.75 | 0 | 5 | 5  |
|  25 | 29 | 1 | 30  |
|  50 | 10 | 0 | 10  |
|  Concentration | Positive | Negative | Total n  |
|  100 | 5 | 0 | 5  |
|  Total | 44 | 16 | 60  |

All results (100%) below 18.75 mIU/mL tested negative. Fourteen (14) out of 15 samples (98.8%) at 25 mIU/mL tested positive. All samples (100%) above 50 mIU/mL tested positive. The sponsor claims a cutoff for positive of 25 mIU/mL.

## e. Analytical specificity:

To evaluate potential cross-reactivity with similar endogenous compounds, 90 negative urine samples were divided into 3 aliquots each and were supplemented with increasing concentrations of luteinizing hormone (hLH), follicle stimulating hormone (hFSH), or thyroid stimulating hormone (hTSH). Results were read in 5 minutes. Results are summarized below (total n per concentration = 30, units = mIU/mL).

|  (mIU/mL) | Lot I |   | Lot II |   | Lot III  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  + | - | + | - | + | -  |
|  hLH | 100 | 0 | 30 | 0 | 30 | 0 | 30  |
|   |  300 | 0 | 30 | 0 | 30 | 0 | 30  |
|   |  500 | 1 | 29 | 1 | 29 | 0 | 30  |
|  hFSH | 100 | 0 | 30 | 0 | 30 | 0 | 30  |
|   |  300 | 0 | 30 | 0 | 30 | 0 | 30  |
|   |  500 | 0 | 30 | 1 | 29 | 0 | 30  |
|  hTSH | 750 | 0 | 30 | 0 | 30 | 0 | 30  |
|   |  1000 | 0 | 30 | 0 | 30 | 0 | 30  |
|   |  1250 | 0 | 30 | 0 | 30 | 1 | 29  |

The sponsor claims that there is no interference at 300 mIU/mL hLH, 300 mIU/mL hFSH, or 1000 mIU/mL hTSH.

To evaluate the potential for interference by certain exogenous compounds, samples with varying hCG concentrations were spiked with potential interferants and tested. No interferences were observed at the concentrations tested. Results are summarized below (units = mIU/mL).

{7}

|  Substance | 0 mIU/mL hCG | 25 mIU/mL hCG | 50 mIU/mL hCG  |
| --- | --- | --- | --- |
|  Acetominophen 20 mg/dL | - | + | +  |
|  Acetylsalicylic acid 20 mg/dL | - | + | +  |
|  Ascorbic acid 20 mg/dL | - | + | +  |
|  Atropine 20 mg/dL | - | + | +  |
|  Caffeine 20 mg/dL | - | + | +  |
|  Centesic acid 20 mg/dL | - | + | +  |
|  Glucose 2 g/dL | - | + | +  |
|  Hemoglobin 20 mg/dL | - | + | +  |
|  Substance | 0 mIU/mL hCG | 25 mIU/mL hCG | 50 mIU/mL hCG  |
|  Tetracycline 20 mg/dL | - | + | +  |
|  Ampicillin 20 mg/dL | - | + | +  |
|  Albumin 20 mg/dL | - | + | +  |

The sponsor evaluated the device and found that samples with pH 4 - 9 or with densities between  $1.000 - 1.050$  were not adversely affected and produced the expected results.

# f. Assay cut-off:

The cutoff for a positive test is  $25\mathrm{mIU / mL}$ . See Detection Limit section above.

# 2. Comparison studies:

# a. Method comparison with predicate device:

The device and the predicate were used to evaluate 619 urine samples. A portion of the samples (245) were obtained from pregnant women (aged 20 - 35). The remainders (374) were obtained randomly from apparently healthy people. These samples were tested using all three strip formats of the device. Results are summarized below:

|  STRIP | Device |   | Predicate  |   |
| --- | --- | --- | --- | --- |
|   |  Positive | Negative | Positive | Negative  |
|  Pregnant | 243 | 2 | 244 | 1  |
|  Other | 1 | 373 | 0 | 374  |
|  CASSETTE | Device |   | Predicate  |   |
| --- | --- | --- | --- | --- |
|   |  Positive | Negative | Positive | Negative  |
|  Pregnant | 243 | 2 | 243 | 2  |
|  Other | 1 | 373 | 1 | 373  |

{8}

|  MIDSTREAM | Device |   | Predicate  |   |
| --- | --- | --- | --- | --- |
|   |  Positive | Negative | Positive | Negative  |
|  Pregnant | 243 | 2 | 243 | 2  |
|  Other | 1 | 373 | 0 | 374  |

b. Matrix comparison:

Not applicable.

3. Clinical studies:

a. Clinical Sensitivity:

Urine samples from 100 patients who were clinically confirmed to be pregnant and 166 randomly chosen, non-pregnant female patients were tested with the strip form of the device. All 100 samples from pregnant women tested positive with the device, and all 166 non-pregnant women tested negative.

To test the useability of the strip format of the device, 87 women completed self-tests and recorded the results. The women also collected urine samples for testing at a hospital lab by a healthcare professional. The same women were clinically followed and confirmed "pregnant" or "not pregnant." Results are summarized below:

|   | Device |   | Total  |
| --- | --- | --- | --- |
|   |  Positive | Negative  |   |
|  Positive | 50 | 1 | 51  |
|  Negative | 0 | 36 | 36  |
|  Total | 50 | 37 | 87  |

To test the useability of the cassette format of the device, 54 women completed self-tests and recorded the results. The women also collected urine samples for testing at a hospital lab by a healthcare professional. The same women were clinically followed and confirmed "pregnant" or "not pregnant." Results are summarized below:

|   | Device |   | Total  |
| --- | --- | --- | --- |
|   |  Positive | Negative  |   |
|  Positive | 31 | 0 | 31  |
|  Negative | 0 | 23 | 23  |
|  Total | 31 | 23 | 54  |

To test the useability of the midstream format of the device, 48 women completed self-tests and recorded the results. The women also collected urine samples for testing at a hospital lab by a healthcare professional. The same women were clinically followed and confirmed "pregnant" or "not pregnant." Results are summarized below:

{9}

|   | Device |   | Total  |
| --- | --- | --- | --- |
|   |  Positive | Negative  |   |
|  Positive | 28 | 0 | 28  |
|  Negative | 0 | 20 | 20  |
|  Total | 28 | 20 | 48  |

b. Clinical specificity:

See above

c. Other clinical supportive data (when a. and b. are not applicable):

Patients (n = 189) who participated in the lay-user studies conducted testing using only the package insert as a guide. In addition, each patient was given a questionnaire to assess the readability of the labeling. The results of the questionnaire showed that the consumers found the test easy to use and that they did not have trouble understanding the labeling and interpreting the results.

4. Clinical cut-off: Not applicable

5. Expected values/Reference range:

All healthy non-pregnant females should test negative for hCG. Although hCG levels vary in normal newly pregnant women, the sponsor claims that the test should detect hCG levels one day after the woman's missed menstrual cycle.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K043443](https://fda.innolitics.com/device/K043443)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
