← Product Code [JWH](/productcode/JWH) · K043440

# SMITH & NEPHEW, INC. REVISION KNEE SYSTEM (K043440)

_Smith & Nephew, Inc. · JWH · Feb 18, 2005 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K043440

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Feb 18, 2005
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Revision Knee System Components are indicated for rheumatoid arthritis; post-traumatic arthritis; osteoarthritis; degenerative arthritis; and failed osteotomies, unicompartmental replacement; or total knee arthroplasties. The components are designed for use in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent. The Revision Knee Components are for single use only and are intended for implantation with bone cement.

## Device Story

Revision Knee System consists of femoral components, tibial components, angled/offset couplers, stem extensions, and wedges. Designed for primary and revision knee arthroplasty in patients with ligamentous instability (absent/incompetent posterior cruciate and collateral ligaments). Implanted by orthopedic surgeons in clinical/hospital settings using bone cement. System provides structural replacement for damaged knee joints; restores joint function and stability. Single-use device.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on design and material similarity to legally marketed predicate devices.

## Technological Characteristics

System includes femoral and tibial components, couplers, stems, and wedges. Materials are consistent with previously cleared Genesis II and Revision Knee System components (metal/polymer). Designed for cemented implantation. No electronic or software components.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Smith & Nephew Revision Knee System ([K041106](/device/K041106.md))
- Smith & Nephew Genesis II Total Knee System ([K951987](/device/K951987.md))
- Smith & Nephew Genesis II Total Knee System ([K953274](/device/K953274.md))
- Smith & Nephew Genesis II Total Knee System ([K962137](/device/K962137.md))

## Reference Devices

- Zimmer Legacy Constrained Condylar Knee (L-CCK)
- Biomet Oncology Salvage System (OSS)
- Sulzer Orthopedics MOST System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K043440

# 510(K) Summary SMITH & NEPHEW REVISION KNEE SYSTEM

| SUBMITTER'S NAME:                 | Smith & Nephew, Inc., Orthopaedic Division                                              |
|-----------------------------------|-----------------------------------------------------------------------------------------|
| SUBMITTER'S ADDRESS:              | 1450 Brooks Road, Memphis, TN 38116                                                     |
| SUBMITTER'S TELEPHONE NUMBER:     | 901-399-6707                                                                            |
| CONTACT PERSON:                   | Gino J. Rouss                                                                           |
| DATE SUMMARY PREPARED:            | December 13, 2004                                                                       |
| TRADE OR PROPRIETARY DEVICE NAME: | Revision Knee System                                                                    |
| COMMON OR USUAL NAME:             | Femoral Components, Tibial Components, Couplers, Wedges, and Stem Extensions            |
| CLASSIFICATION NAME:              | Knee Joint Patellofemorotibial Metal/Polymer/Metal Semi-Constrained Cemented Prosthesis |
| DEVICE CLASS:                     | Class II                                                                                |
| PANEL CODE:                       | Orthopedics/87                                                                          |

# DEVICE INFORMATION:

#### A. INTENDED USE:

The Revision Knee System Components are indicated for rheumatoid arthritis; post-traumatic arthritis; osteoarthritis; degenerative arthritis; and failed osteotomies, unicompartmental replacement; or total knee arthroplasties. The components are designed for use in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent. The Revision Knee Components are for single use only and are intended for implantation with bone cement.

These indications are the same as currently used for the Genesis II Constrained Knee System and Revision Knee System cleared via K962137 and K041106.

### B. DEVICE DESCRIPTION:

The Smith & Nephew Revision Knee System contains femoral knee components, tibial components, angled and offset couplers, stem attachments and wedges. The designs of these devices are based upon existing components of the Genesis II Total Knee System and Revision System, previously cleared for market under K951987, K953274, K962137, and K041106.

## C. SUBSTANTIAL EQUIVALENCE INFORMATION:

The Smith & Nephew Revision Knee System Components are similar to the following commercially available devices regarding design features, overall indications, and materials:

- 토 Smith & Nephew Revision Knee System (K041106)
- Smith & Nephew Genesis II Total Knee System (K951987, K953274, K962137) 발
- l Zimmer Legacy Constrained Condylar Knee (L-CCK)
- Biomet Oncology Salvage System (OSS) 트
- . Sulzer Orthopedics MOST" System

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 8 2005

Mr. Gino J. Rouss Regulatory Affairs Specialist Smith & Nephew, Inc. 1450 E. Brooks Road Memphis, Tennessee 38116

Re: K043440

Trade/Device Name: Revision Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semiconstrained cemented prosthesis Regulatory Class: II Product Code: JWH

Dated: February 13, 2005 Received: February 14, 2005

Dear Mr. Rouss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, de need and have Act (Act) that do not require approval of a premarket approval application (PMA). and Oosmete ! Ios ( 10) wat the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 – Mr. Gino J. Rouss

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,
Mark N Milken

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): K043440

Device Name: Revision Knee System

Indications for Use:

The Revision Knee System Components are indicated for:

- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity levels are compatible with an adequate long-term result.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. The constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent.

The Revision Knee Components are for single use only and are intended for implantation with bone cement.

| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|---|
| AND/OR                                          |   |
| Over-The-Counter Use<br>(21 CFR 807 Subpart C)  |   |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

**(Division Sign-Off)**
Division of General, Restorative,
and Neurological Devices

Page 1 of 1

510(k) Number K043440

---

**Source:** [https://fda.innolitics.com/device/K043440](https://fda.innolitics.com/device/K043440)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
