RELIADENT DENTAL IMPLANT SYSTEM

K043428 · Bioinfera, Inc. · NHA · Jul 12, 2005 · Dental

Device Facts

Record IDK043428
Device NameRELIADENT DENTAL IMPLANT SYSTEM
ApplicantBioinfera, Inc.
Product CodeNHA · Dental
Decision DateJul 12, 2005
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.

Device Story

Reliadent Dental Implant System comprises endosseous root-form implants, healing caps, restoration abutments, and surgical instruments. Implants are conical or cylindrical, 3.6mm-6.0mm diameter, 8mm-17mm length; feature internal/external hexagonal stability design; some variants HA-coated to enhance osseointegration. Implants composed of medical grade 4 titanium; abutments composed of medical grade 2 titanium. System used by dental professionals in clinical settings for surgical placement into jawbone to support artificial teeth. Surgical kit includes drills and hand tools compatible with various implant devices. Device provides structural foundation for prosthetic restoration, enabling patient chewing function.

Clinical Evidence

No clinical data provided; substantial equivalence based on design, material, and intended use comparisons to the predicate device.

Technological Characteristics

Materials: Medical grade 4 titanium (implants), medical grade 2 titanium (abutments). Design: Conical/cylindrical endosseous root-form; internal/external hexagonal connection. Surface: HA-coated variants available. Dimensions: 3.6-6.0mm diameter, 8-17mm length. System includes surgical instruments (drills, hand tools).

Indications for Use

Indicated for patients requiring dental implants in the upper or lower jaw to support prosthetic devices and restore chewing function.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ JUL 12 2005 K043428 Page 1 A 510(k) Summary Pertaining to the Safety and Effectiveness of the Reliadent Dental Implant System #### Submitter Information: Chan Q. Wang President Bioinfera. Inc. 21205 Halworth, Rd Beachwood, Oh 44122 Phone: (216) 226-2680 E-mail:jxh27 a`vahoo.com Date Summary Prepared: December 7, 2004 ## Device Name: Proprietary name - Reliadent Dental Implant System Common/Usual name - Implant, Endosseous, Root Form Trade name - Reliadent Dental Implant System This device has been classified Class II Special Controls. Classification number: DZE. Regulation Number: 21 CFR 872.3640. Classification Advisory Committee: Dental # Predicate Device: MIS Dental Implant System 510(k) - K040807 Cleared - June 6, 2004 ## Description of the Device: The Reliadent Dental Implant System includes surgical implants, healing cap, restoration abutments, and surgical instruments. #### Implants: Includes one or two step dental implant devices which are designed as conical and cylindrical shapes. They consist of various diameters with a range of 3.6mm-6.0mm and lengths with a range of 8mm-17mm. The internal and external hexagonal shapes are important to maintain the implant stability. One group of the products is HA coated. This surface has been modified to enhance the osseointegration. The dental implants are composed of medical grade 4 pure titanium. {2}------------------------------------------------ # Cover screw and healing cap: The cover screw and healing caps are supplied together with the implants. They cover the internal body of implant to allow gum free space for connection between the implants and restoration abutments. # Abutment: Abutments and accessories are the support part to prosthetic restoration. Screw retained abutments are included in the Reliadent Dental Implant System. The abutments are composed of medical grade 2 titanium. # Surgical Instrument Kit: The surgical instruments include the implants installation set, abutment connections, and drill set. There are hand tools, drill bits of different sizes, and handles in these sets. The instrument kit is designed to be used with a wide range of commercially available implant devices. # Indications for Use: Indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. ## Substantial Equivalence: The Reliadent Dental Implant System has the same intended use as the MIS Dental Implant System from MIS Technologies, Inc. Elmwood Park, NJ 07407, cleared under 510(k) Number: K040807. The MIS Dental Implant System has equivalent performance characteristics in its intended use, material and design. The MIS Dental Implant System contain implants, cover screw and healing caps, abutments and the applicable surgical instruments. The Reliadent Dental Implant System is substantially the same as the currently marketed MIS dental implant with a modified surface. This surface promotes osseointegration. All other technological characteristics are similar and both devices show equivalent performance capabilities. # Conclusion: The evaluation of the Reliadent Dental Implant System does not raise any additional concerns regarding safety and effectivity and may therefore be considered substantially equivalent to the predicate devices. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) in the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines representing the serpent entwined around a staff. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the caduceus. The seal is presented in black and white. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 2 2005 Mr. Chan Q. Wang President Bioinfera, Incorporated 21205 Halworth, Road Beachwood, Ohio 44122 Re: K043428 Trade/Device Name: Reliadent Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Implant Regulatory Class: II Product Code: NHA Dated: July 1, 2005 Received: July 5, 2005 Dear Mr. Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Wang Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and iloting (21 cert forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), nlease contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Cintron vintbon Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use #### 510(k) Number (if known): K043428 Device Name: Reliadent Dental Implant System Indications for Use: Indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. Prescription Use __ X AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Kain Muelly for ASR (Division Sign-Off (Division Sign-On) Division of Anesthesiology, General Hospital Infection Control, Dental D 510(k) Number. K 043428 Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...