COOK PEDIATRIC FLEXOR URETERAL ACCESS SHEALTH

K043418 · Cook Urological, Inc. · FED · Mar 31, 2005 · Gastroenterology, Urology

Device Facts

Record IDK043418
Device NameCOOK PEDIATRIC FLEXOR URETERAL ACCESS SHEALTH
ApplicantCook Urological, Inc.
Product CodeFED · Gastroenterology, Urology
Decision DateMar 31, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesPediatric

Indications for Use

Used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. The Cook® Pediatric Flexor Ureteral Access Sheath is intended for Pediatric Use in patients two years of age and older.

Device Story

Single-use, sterile ureteral access sheath; available in single or dual lumen configurations. Facilitates passage of endoscopes/instruments into urinary tract during endoscopic urological procedures. Features hydrophilic coating to reduce friction during insertion/removal. Used in clinical settings by urologists. Provides stable conduit for instrumentation; minimizes trauma during repeated instrument exchange. Benefits pediatric patients by enabling minimally invasive access to urinary tract.

Clinical Evidence

Bench testing only. Includes coefficient of friction testing, irrigation flow study, buckling test, and kink resistance testing. Biocompatibility testing performed. Clinical feedback and published articles cited to support safety and effectiveness in pediatric population.

Technological Characteristics

Single or dual lumen ureteral access sheath. Hydrophilic coating. Available in 9.5 Fr or 12 Fr inside diameters. Secondary lumen 3 Fr. Materials are standard medical-grade polymers. Sterilized device. No energy source. Standalone mechanical device.

Indications for Use

Indicated for pediatric patients aged two years and older undergoing endoscopic urological procedures requiring a conduit for endoscope or instrument passage into the urinary tract.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K0434/8 # COOK® MAR 3 1 2005 Cook Urological 1100 W. Morgan Street Spencer, IN 47460 USA Phone: 812-829-4891 Fax: 812-829-1801 www.cookgroup.com #### 510(k) SUMMARY l. #### Submitted By: Cindy Rumple Cook Urological 1100 West Morgan Street Spencer, Indiana 47460 (812) 829-4891 December 8, 2004 #### Device Trade Name: Cook® Pediatric Flexor Ureteral Access Sheath Proposed Classification Name: Sheath, for Endoscope, Ureteral Dilator Class II 78 FED 876.1500 #### Predicate Devices: The Cook® Pediatric Flexor Ureteral Access Sheath is similar with respect to indications for use and technology to exist predicate devices in commercial distribution. Specifically, the Cook® Pediatric Flexor Ureteral Access Sheath is similar to the AQ Hydrophilic Dilator (K961904) manufactured by Cook Urological, Incorporated, the Bard® Aqua Guide Ureteral Conduit (K030438) manufactured by C.R. Bard, Incorporated, the Bard® Aqua Guide Ureteral Access Sheath (K033778) manufactured by C.R. Bard, Incorporated, the Ureteral Access Sheath Set (K993650) manufactured by Applied Medical Resources Corporation, the Ureteral Access Sheath Set (K990775) manufactured by Applied Medical Resources Corporation, and the Ureteral Access Sheath Set II (K030956) manufactured by Boston Scientific Corporation (Microvasive). ### Device Description: The Cook® Pediatric Flexor Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. The Cook® Pediatric Flexor Ureteral Access Sheath is intended for Pediatric Use in patients two years of age and older. The Cook® Pediatric Flexor Ureteral Access Sheath is a single use sterile device that is offered as a single or dual lumen device. The Cook® Pediatric Flexor Ureteral Access Sheath will be offered in a 9.5 French inside diameter length or 12 French inside diameter 13 centimeter length single lumen version. The Cook® Pediatric Flexor Ureteral Access Sheath will also be offered in a 9.5 French inside diameter 13 centimeter length or 12 French inside diameter 13 {2}------------------------------------------------ centimeter length dual lumen version. The main lumen of the 9.5 French dual lumen version is 9.5 French inside diameter and the main lumen of the 12 French dual lumen version is 12 French inside diameter. The secondary lumen on both sizes is 3 French inside diameter. Both versions inside diamster. The occurrently a Hydrophilically coated which, when activated, allows easier insertion and removal of the sheath. The construction materials of the Cook® Pediatric Flexor Ureteral Access Sheath are all well known in the medical field. Biocompatibility testing has shown the materials to meet the test requirements. Bench top testing, coefficient of friction testing, irrigation flow study, buckling require test, kink resistance testing, risk analysis, complaints verses sales data, clinical feedback, and published articles support our claim that the device is as safe and effective in the pediatric population as in adult use. #### Substantial Equivalence: The device will be manufactured according to specified process controls and a Quality Assurance Program. The device will undergo packaging and sterilization procedures similar to devices currently marketed and distributed by Cook Urological, Incorporated. Being similar with respect to indications for use, materials, and physical construction to predicate devices, this device meets the requirements for section 510 (K) substantial equivalence. {3}------------------------------------------------ Image /page/2/Picture/2 description: The image shows a circular logo with a stylized eagle in the center. The eagle is facing left and has three lines extending from its head, resembling feathers or wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged around the top half of the circle, following the curve of the logo. The text is in uppercase and appears to be in a sans-serif font. MAR 3 1 2005 Food and Orug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Cindy Rumple Regulatory Affairs Technical Writer Cook® Urological 1100 W. Morgan Street SPENCER IN 47460 Re: K043418 Ro45418 Trade/Device Name: Cook® Pediatric Flexor Ureteral Access Sheath Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: February 23, 2005 Received: March 1, 2005 Dear Ms. Rumple: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to and exceeded to the enactment date of the Medical Device Amendments, or to devices that have been May 20, 1970, and channeling of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do ince require to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your Applo rary, it the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the I edolal statested including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ocgin marketing of substantial equivalence of your device to a legally prematics notification. "The PDA mainly of batisation for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please a If you desire specific advice for your device of our laboring numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation entitled, " risestanting on your responsibilities under the Act from the 807.97). You may obtain other general micronal and Consumer Assistance at its toll-free number (800) Division of Sthail Manufacturers, Internet address http://www.fda.gov/cdrh/dsmadsmamain.html Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### PREMARKET NOTIFICATION ## INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K043418 Device Name: Indications for Use: Cook® Pediatric Flexor Ureteral Access Sheath Used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. The Cook® Pediatric Flexor Ureteral Access Sheath is intended for Pediatric Use in patients two years of age and older. Prescription Use _ × (Part 21 CFR 801 Subpart D) OR Over the Counter Use_ (Part 21 CFR 807 Subpart C) (Please Do Not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Nancy C Brogdon (Division Sign-Of Division of Reproductive, Abe and Radiological Devic 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...