HL 188 & HL 189 BODY FAT METER

K043414 · Health & Life Co., Ltd. · MNW · Feb 9, 2005 · Cardiovascular

Device Facts

Record IDK043414
Device NameHL 188 & HL 189 BODY FAT METER
ApplicantHealth & Life Co., Ltd.
Product CodeMNW · Cardiovascular
Decision DateFeb 9, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2770
Device ClassClass 2
AttributesPediatric

Intended Use

The HEALTH & LIFE HL 188 & HL 189 BODY FAT METER is intended for estimation of the body fat of percentage in the home.

Device Story

Hand-held, non-sterile, reusable bioimpedance meter; estimates body fat percentage. Operates via bioelectrical impedance analysis; user inputs personal data; device measures impedance; calculates body fat percentage. Used in home setting by individuals for self-monitoring. Output displayed to user; assists in tracking body composition. Benefits include convenient, non-invasive home monitoring of body fat.

Clinical Evidence

Clinical comparison study performed against DEXA (Dual energy X-ray absorptiometry) technology. Subjects grouped by gender and age. Study demonstrated statistically and clinically acceptable repeatability across all age, weight, and height groups.

Technological Characteristics

Hand-held bioimpedance meter. Conforms to IEC 60601-1 (safety) and IEC 60601-1-2 (EMC). Non-invasive measurement principle.

Indications for Use

Indicated for estimation of body fat percentage in healthy individuals aged 10 to 80 years in a home environment.

Regulatory Classification

Identification

An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.

Special Controls

*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K043414 PAGE 1 OF 2 ## FEB - 9 2005 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92 | 1.0 Submitter's Name: | Health & Life Co., Ltd. | |--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | 9 F, No. 168, Jian Yi Road, Chung Ho City, Taipei County, Taiwan<br>ZIP code : 235 | | Phone: | 886-2-82271300 | | Fax: | 886-2-82271301 | | Contact: | Mr. Paul Yang / President | | 2.0 Device Name: | HEALTH & LIFE HL 188 & HL 189 BODY FAT METER | | 3.0 Predicate Device: | • LONG WELL LW-6Dxx Body Fat Analyzer (K030203) marketed<br>by LONG WELL ELECTRONICS CORP. &<br>• Omron HBF-306 Body Fat Analyzer (K011652) marketed by<br>Omron Healthcare INC. | | 4.0 Device Description: | HEALTH & LIFE HL 188 & HL 189 BODY FAT METER is a<br>hand-held, non-sterile, reusable Body Fat Analyzer intended for<br>estimation of the body fat of percentage in the home. | | 5.0 Intended Use: | The HEALTH & LIFE HL 188 & HL 189 BODY FAT METER is<br>intended for estimation of the body fat of percentage in the home. | | 6.0 Performance Summary: | In terms of operating specification, Safety & EMC requirements,<br>the device conforms to applicable standards included IEC 60601-1<br>and IEC 60601-1-2 requirements. Moreover, A comparison study<br>with device that use DEXA(Dual energy X-ray absorptiometry)<br>technology was performed to validate the performance of the<br>HEALTH & LIFE HL 188 & HL 189 BODY FAT METER.<br><br>Subjects were grouped as male/ female, ages. The comparison<br>study demonstrated that the clinical repeatability of HEALTH &<br>LIFE HL 188 & HL 189 BODY FAT METER is statistically and<br>clinically acceptable in all age/weight/height groups. | . {1}------------------------------------------------ ## 7. Conclusions: The HEALTH & LIFE HL 188 & HL 189 BODY FAT METER have the same intended use and similar technological characteristics as LONG WELL LW-6Dxx Body Fat Analyzer (K030203) marketed by LONG WELL ELECTRONICS CORP. and Omron HBF-306 Body Fat Analyzer( K011652 )marketed by Omron Healthcare INC.. Moreover, bench testing contained in this submission and clinical testing supplied demonstrate that any differences in their technological characteristics do not raise and new questions of safety or effectiveness. Thus, the HEALTH & HL 188 & HL 189 BODY FAT METER is substantially equivalent to the predicate devices. LIFE {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three horizontal lines representing its wings. The eagle is positioned to the right of a circular emblem containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged around the circumference of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 HEALTH & LIFE CO., LTD. c/o Ms. Jennifer Reich Harvest Consulting Corporation 3892 South America West Trail FLAGSTAFF AZ 86001 Re: K043414 Trade/Device Name: HEALTH & LIFE HL 188 & HL 189 BODY FAT METER Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: 74 MNW Dated: December 3, 2004 Received: December 13, 2004 FEB - 9 2005 Dear Ms. Reich: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally prematics notificate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you atter office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsmadsmamain.html Sincerely yours, Nancy C. brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): _______K043414 ## Device Name: HEALTH & LIFE HL 188 & HL 189 BODY FAT METER Health & Life Co., Ltd. Indications For Use: The device is a noninvasive bioimpedance meter used to estimate body fat percentage in the home setting. The device is for normal, healthy people only, and the applicable age range is 10 to 80 years old. The device is to be used in a home environment having normal temperature and humidity conditions. Prescription Use (Part 21 CFR 801 Subpart D) 510(k) Number AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nancy Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 40/12/14 Abdominal.t Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...