PIEZOSURGERY DEVICE

K043408 · Piezosurgery S.R.L. · DZI · Jun 8, 2005 · Dental

Device Facts

Record IDK043408
Device NamePIEZOSURGERY DEVICE
ApplicantPiezosurgery S.R.L.
Product CodeDZI · Dental
Decision DateJun 8, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4120
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Piezosurgery device is a bone cutting instrument intended for use in oral surgery.

Device Story

Piezosurgery device utilizes piezoelectric ultrasonic technology to perform bone cutting; minimizes trauma to soft tissue. System includes sharp, smoothing, and blunt insert tips. Operated by clinicians in oral surgery settings. Device transforms electrical energy into mechanical ultrasonic vibrations at the insert tip; enables precise osteotomies. Benefits include reduced risk of tissue heating, osteonecrosis, and soft tissue trauma compared to traditional rotary instruments. Clinician uses tactile control to guide the handpiece; requires limited pressure to achieve desired cutting action.

Clinical Evidence

Performance and safety evaluations based on clinical and histologic data. Evaluated incision characteristics, cutting precision, surgical tactile control, risk of adjacent tissue damage, mineralized tissue heating, and integrity of osteotomized surfaces. Results demonstrated desired cutting action and reduced risk of soft tissue trauma.

Technological Characteristics

Piezoelectric ultrasonic technology; handpiece with interchangeable sharp, smoothing, and blunt insert tips. Class II bone cutting instrument. Operates via ultrasonic vibration for mechanical bone cutting.

Indications for Use

Indicated for use in oral surgery, including implantology, periodontal surgery, and surgical orthodontics.

Regulatory Classification

Identification

A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Piezosurgery 510(k) Summary of Safety and Effectiveness in accordance with 21 CFR 807.92 (c) | K number: | K043408 | |-----------|---------| |-----------|---------| 1. Submitter's information: | | Tomaso Vercellotti, M.D., D.D.S.<br>Piezosurgery s.r.l.<br>Molo Ponte Morosini 41/7<br>16126 Genoa, Italy<br>Phone: +39 010 254 1934<br>E-mail: perfler@piezosurgery.com | |--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------| 2. Contact person: | | Maria E. Donawa, M.D.<br>Donawa Consulting<br>P.zza Albania, 10<br>00153 Rome, Italy<br>Phone: +39 06 578 2665<br>E-mail: medonawa@donawa.com | |--|-----------------------------------------------------------------------------------------------------------------------------------------------| |--|-----------------------------------------------------------------------------------------------------------------------------------------------| 3. Date summary prepared: | | June 6, 2005 | |--|--------------| |--|--------------| 4. Device name and classification: | a. Trade/device name: | Piezosurgery® | |--------------------------|-----------------------------------------------------------------------| | b. Classification name: | Bone Cutting Instrument and Accessories (per 21 CFR section 872.4120) | | c. Classification panel: | Dental | | d. Regulatory class: | Class II | | e. Product code: | DZI | ### 5. Device description: 5. Device dessmithom: The Piezosurgery® device uses piezoelectric ultrasonic technology to generate mechanical The riezdsurgery done cutting, with minimal trauma to soft tissue. The device is supplied mith sharp, smoothing and blunt insert tips for use in oral surgery, including implantology, periodontal surgery and surgical orthodontics. #### 6. Intended use: o. The Piezosurgery device is a bone cutting instrument intended for use in oral surgery. #### 7. Predicate device: Micro-motor system MD20 Nouvag (K042434) #### 8. Substantial equivalence comparison: o. Technical oquivals.comparisons presented in this submission support a Technical, clinical and niotologic compansent prosurgery and other bone cutting devices muling of oubotantial of the US. In addition, this technology has been shown to allow precise osteotomies with minimal risk of tissue heating and osteonecrosis damage. {1}------------------------------------------------ # 9. Performance evaluations: 9. Performance evaluations. Performance and safety evaluations were based on clinical and histologic data on the Performance and salely evaluations were bacal on and surgical tactile control, risk of adjacent tissue incision characteristics, cutting precision and surgical tactile contro Incision characteristics, cutting probision and ourgious technized surfaces, and postdamage, mineralized tissue heating, integrity of the osted insinces, and postdamage, mineralized tissue neating, megnity of the octosty delineated incisions and the operative nealling. The evaluations domenation pve the desired cutting action, need for limited pressure on the handpiece to achieve the desired to ask tigen heed for limited pressure on the nanspioo to dellier sibility of trauma to soft tissue. #### 10. Conclusion: 10. Conclusion: The performance and safety data presented in this premarket notification support a finding and The performance and salety uata proomed in this promation of the predicate device specified in this submission. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings. The eagle is positioned within a circle, and the text "U.S. Department of Health and Human Services, Inc." is written around the circle's perimeter. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 8 - 2005 Piezosurgery Srl C/O Ms. Maria E. Donawa President Donawa Consulating Srl Piazza Albania, 10 00153 Rome, Italy Re: K043408 Trade/Device Name: Piezosurgery® Regulation Number: 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: II Product Code: DZI Dated: April 28, 2005 Received: May 2, 2005 Dear Ms. Donawa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Donawa Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a Iegally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jnette G. Michael m.D Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement | 510(k) Number: | K043408 | |----------------------|----------------------------------------------------------------------------------------| | Device Name: | Piezosurgery® | | Indications for Use: | The Piezosurgery device is a bone cutting instrument intended for use in oral surgery. | Prescription Use_ × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ## Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Runne (Division Sign-Off) Division of Anesthesiology, General Hespita. Infection Control, Dental Devices Kaus408 510(k) Number __ Page 1_of 1__
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...