VITROS CHEMISTRY PRODUCTS TOBRA REAGENT, CALIBRATOR KIT 14 AND TDM PERFORMANCE VERIFIERS I, II AND III
K043384 · Ortho-Clinical Diagnostics, Inc. · LDO · Feb 7, 2005 · Clinical Toxicology
Device Facts
Record ID
K043384
Device Name
VITROS CHEMISTRY PRODUCTS TOBRA REAGENT, CALIBRATOR KIT 14 AND TDM PERFORMANCE VERIFIERS I, II AND III
Applicant
Ortho-Clinical Diagnostics, Inc.
Product Code
LDO · Clinical Toxicology
Decision Date
Feb 7, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3900
Device Class
Class 2
Indications for Use
VITROS Chemistry Products TOBRA Reagent: For in vitro diagnostic use only, VITROS Chemistry Products TOBRA Reagent is used on the VITROS 5,1 FS Chemistry System to quantitatively measure tobramycin (TOBRA) concentration in human serum and plasma. Serum or plasma tobramycin measurements are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to ensure appropriate therapy. VITROS Chemistry Products Calibrator Kit 14: For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 14 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of tobramycin (TOBRA). VITROS Chemistry Products TDM Performance Verifier I, II and III: For in vitro diagnostic use only. VITROS TDM Performance Verifier is an assayed control used to monitor performance of ACET, CRBM, DGXN, PHBR, PHYT and TOBRA on VITROS Chemistry Systems.
Device Story
VITROS Chemistry Products Tobramycin Reagent is a quantitative homogeneous enzyme immunoassay for use on the VITROS 5,1 FS Chemistry System. The device measures tobramycin concentration in human serum or plasma samples. Principle of operation involves competition between tobramycin in the patient sample and labeled tobramycin (glucose-6-phosphate dehydrogenase) for binding to polyclonal sheep antibodies. The resulting enzyme activity is monitored spectrophotometrically at 340 nm. The system uses a stored calibration curve to determine unknown sample concentrations. The device is intended for clinical laboratory use by trained personnel to assist in therapeutic drug monitoring and overdose diagnosis. Output is a quantitative concentration value (ug/mL) used by clinicians to adjust dosage or assess toxicity. The system includes calibrators and assayed performance verifiers to ensure analytical accuracy and precision.
Clinical Evidence
Bench testing only. Equivalence demonstrated through comparative analysis with predicate device using patient samples. Correlation study yielded a slope of 1.01 and correlation coefficient of 0.997. Additional bench studies included precision, linearity, and specificity assessments.
Indicated for quantitative measurement of tobramycin in human serum and plasma to diagnose/treat overdose and monitor therapeutic levels. Used on VITROS 5,1 FS Chemistry System.
Regulatory Classification
Identification
A tobramycin test system is a device intended to measure tobramycin, an aminoglycoside antibiotic drug, in plasma and serum. Measurements obtained by this device are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to ensure appropriate therapy.
Predicate Devices
Syva® EMIT® 2000 Tobramycin Assay and calibrators (k003341)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k043384
B. Purpose for Submission:
New assay and calibrators. Changes to calibrator verifiers.
C. Measurand:
Tobramycin
D. Type of Test:
Quantitative homogeneous enzyme immunoassay
E. Applicant:
Ortho-Clinical Diagnostics Inc.
F. Proprietary and Established Names:
VITROS Chemistry Products Tobramycin Reagent, Calibrator Kit 14, TDM Performance Verifier I, II and III.
G. Regulatory Information:
1. Regulation section:
Tobramycin Test System (21CFR862.3900)
Assayed controls (21 CFR 862.3280)
Calibrators (21 CFR 862.3200)
2. Classification:
Class II
3. Product code:
91LDO, 91DLJ, 91DIF
{1}
4. Panel:
91, Toxicology
H. Intended Use:
1. Intended use(s): See indications for use below.
2. Indication(s) for use:
For *in vitro* diagnostic use only. VITROS Chemistry Products TOBRA Reagent is used on the VITROS FS 5,1 Chemistry System to quantitatively measure tobramycin (TOBRA) concentration in human serum and plasma. Serum or plasma tobramycin measurements are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to help ensure appropriate therapy.
For *in vitro* diagnostic use only. VITROS Chemistry Products Calibrator Kit 14 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of tobramycin (TOBRA).
For *in vitro* diagnostic use only. VITROS TDM Performance Verifier is an assayed control used to monitor performance of ACET, CRBM, DGXN, PHBR, PHYT, and TOBRA on VITROS Chemistry Systems.
3. Special conditions for use statement(s):
See specificity concerning drugs that may interfere with the assay. The device is for use with serum and plasma (lithium heparin and EDTA). It is not for use with other anticoagulants.
4. Special instrument requirements: For use on the VITROS 5,1 FS
I. Device Description:
Reagent 1 contains tobramycin labeled with glucose-6-phosphate dehydrogenase and other non-reactive components, including mouse monoclonal antibodies. Reagent 2 contains polyclonal sheep antibodies to tobramycin, NAD, glucose-6-phosphate and other non-reactive components.
Calibrator Kit 14 is an aqueous solution containing tobramycin, buffer, proteins, salts, surfactants and preservatives. Nominal values of tobramycin (ug/mL): 0, 0.6, 2, 4, 6, 10.
The performance verifiers are assayed controls prepared from bovine serum to which therapeutic drugs, salts and preservatives are added.
{2}
J. Substantial Equivalence Information:
1. Predicate device name(s):
Syva® EMIT® 2000 Tobramycin Assay and calibrators; VITROS Chemistry Products Performance Verifiers
2. Predicate 510(k) number(s):
k003341 (reagents), k042476 (performance verifiers)
3. Comparison with predicate:
The devices are similar in intended use and methodology. Both devices are homogeneous enzyme immunoassays. The predicate device is for use on Syva Analyzer Systems; the new device is for use on the VITROS 5,1 FS Chemistry System.
The VITROS Chemistry Products TDM Performance Verifiers are substantially equivalent to VITROS Chemistry Products TDM Performance Verifiers, currently in commercial distribution (K042476). Tobramycin has been added and labeling updated to add assigned values.
K. Standard/Guidance Document Referenced (if applicable):
NCCLS Guideline – EP5-A, Evaluation of Precision Performance of Clinical Chemistry Devices
NCCLS Guideline – EP6-A, Evaluation of the Linearity of Quantitative Analytical Methods
NCCLS Guideline – EP7-A, Interference Testing in Clinical Chemistry
NCCLS Guideline – EP9-A2, Method Comparison and Bias Estimation Using Patient Samples
L. Test Principle:
Patient sample is added to reagent 1, which contains tobramycin labeled with glucose-6-phosphate dehydrogenase, followed by reagent 2, which contains antibody reactive to tobramycin, glucose-6-phosphate and NAD. The assay is based on competition between tobramycin in the sample and labeled tobramycin. Tobramycin in the sample, related to enzyme activity, is monitored spectrophotometrically at 340 nm. Unknown sample concentrations are determined using the (stored) calibration curve.
3
{3}
M. Performance Characteristics (if/when applicable):
Analytical studies were performed on 3 lots.
1. Analytical performance:
a. Precision/Reproducibility:
Within-day and within laboratory precision was determined using the system's bovine serum-based QC materials. The evaluation followed NCCLS EP-5A, with 2 replicates per run, two runs per day for 22 days, n=88 observations. Within-day runs were separated by at least 2 hours. Calibration was performed once each week. The sample order was randomized. Results for within-day and within-lab are shown. Testing of multiple lots and instruments yielded similar results.
| Sample | Mean (ug/mL) | N | Within-day SD (ug/mL) | Within-lab (ug/mL) | Within-lab %CV |
| --- | --- | --- | --- | --- | --- |
| Control level 1 | 1.35 | 88 | 0.035 | 0.053 | 3.9 |
| Control level 2 | 3.44 | 88 | 0.049 | 0.082 | 2.4 |
| Control level 3 | 7.12 | 88 | 0.054 | 0.112 | 1.6 |
Within-run precision was also estimated for patient serum pools at concentrations near the low end of the reportable range. Standard deviations were calculated based on 5 replicates, for each of 3 reagent lots, at tobramycin levels ranging from 0.36-1.1 ug/mL (i.e. total of 15 observations at each level). Results near the limit of detection were all within the within-run standard deviation acceptance limits of 0.088 ug/mL.
b. Linearity/assay reportable range:
The evaluation followed NCCLS EP-6A. Serum pools with tobramycin concentrations at 12 levels spanning the reportable range were evaluated. Each level was tested in replicates of 5 and average values of observed/expected concentrations were determined. The assay reportable range is 0.6-10 ug/mL. In this range deviations from expected concentrations were less than the bias limits of +/-0.26 ug/mL near the low end and +/-1.3 ug/mL near the high end of the reportable range.
Recovery after sample dilution with the recommended diluent was also evaluated. Serum samples with tobramycin concentrations in the range of approximately 8 ug/mL were diluted 2x and 4x. Recoveries (from triplicate analyses) were all within acceptance criteria of 88-112%.
{4}
c. Traceability, Stability, Expected values (controls, calibrators, or methods): VITROS Chemistry Products Calibrator Kit 14 and performance verifiers I, II and III are for use with this device.
Value assignment for calibrators: Values assigned to calibrators are traceable to USP tobramycin reference standard. Stock solution concentrations are confirmed by GCMS or HPLC methods. Expanded uncertainties for product calibrators (the 95% confidence interval around the assigned value mean) are 0.09 for 0.6 ug/mL and 0.41 for 10 ug/mL. Correlation analyses (described in the 510(k)) to confirm that trueness is transferred, indicate slopes and intercepts approaching 1 and 0, respectively, and correlation coefficients > 0.99.
Calibrator long-term stability: Calibrators are stored at 9 degrees C and evaluated at intervals up to and past the expiration date. Testing compares results obtained using the test calibrators to those obtained using reference calibrators, on the Syva 30R Analyzer. The samples used in testing were sample pools at medical decision points, and bias is determined for the reference versus the test calibrators. Acceptance criteria for bias are: +/- 0.169 ug/mL for tobramycin concentrations < 2 ug/mL, and +/-0.0754[tobramycin] + 0.0184 ug/mL for concentrations > 2 ug/mL.
Calibrator opened-bottle stability: Opened bottle stability is evaluated at intervals up to and past recommended expiration date and restored at 2-8 degrees C. Bias is determined relative to previously unopened calibrators. Acceptance limits are: +/- 0.069 ug/mL for tobramycin concentrations < 2 ug/mL, and +/-0.0307[tobramycin] + 0.0075 ug/mL for concentrations > 2 ug/mL.
Value assignment for calibrator verifiers: Performance verifier lots are tested with each reagent to establish target values. Allowable ranges applied reflect estimates of expected lab to lab variability. The range of means is calculated from precision data from multiple systems (minimum 5) in multiple laboratories. A pooled SD is determined based on total SD's from each site and the range is calculated based on 3x pooled SD. Range of means for controls are provided on performance verifier assay sheets.
Verifier stability: Long-term stability for controls is evaluated using vials stored at -18 degrees C at intervals up to and past the expiration date. Opened stability is evaluated using vials opened and stored at 2-8 degrees C up to 7 days. Observed tobramycin concentrations (mean of replicates) are compared to those at "time 0", at intervals up to and past expiration. Examples of calculated acceptance criteria, in terms of bias, relative to time 0, are shown:
5
{5}
| | Long-term acceptance limit (ug/mL) (+/-) | Opened vial acceptance limit (ug/mL) (+/-) |
| --- | --- | --- |
| Level 1 2 ug/mL | 0.35 | 0.18 |
| Level 2 4 ug/mL | 0.69 | 0.32 |
| Level 3 7 ug/mL | 1.21 | 0.54 |
d. Detection limit:
The limit of detection, 0.6 ug/ml, was determined based on tobramycin-negative serum samples from 10 human donors, as well as from low level calibrator material. Three reagent lots, 2 calibrator kit lots and 2 instrument systems, were used in testing. The lower limit of detection was calculated as:
3.3X√Calibration Error Variance (SD)² + Pooled replicate Variance (SD)²
Results based on these calculations support the detection limit of 0.6 ug/ml.
The limit for deviation from linearity at samples near the low end of the reportable range is 0.26 ug/mL (see Linearity, above.) Within-run standard deviations were 0.088 ug/mL for samples with concentrations near the low end of the reportable range (see Precision, above.)
e. Analytical specificity:
Testing followed NCCLS EP-7A for the paired-difference method. Human serum pools with tobramycin concentrations approximately 4 and 9 ug/mL, were individually spiked with bilirubin (45 mg/dL), intralipid (1000 mg/dL) and hemoglobin (1000 mg/dL).
Serum pools containing 4 ug/mL tobramycin were spiked with other commonly co-administered drugs and tested for interference.
Concentrations of drugs tested were as follows: carbenicillin, 500 ug/mL; cephalothin, 500 ug/mL; chloramphenicol, 1000 ug/mL; clindamycin, 500 ug/mL; erythromycin, 1000 ug/mL; neomycin, 100 ug/mL; netilimicin, 100 ug/mL; penicillin G, 500 ug/mL; sisomicin, 100 ug/mL; streptomycin, 100 ug/mL; sulfamethoxazole, 600 ug/mL; tetracycline, 500 ug/mL; trimethoprim, 25 ug/mL and vanomycin, 200 ug/mL.
Bias was calculated as the mean tobramycin concentration observed for the test substance pool minus the mean tobramycin concentration of the control pool. Acceptance criteria are bias < 1.08 ug/mL for the 9 ug/ml pool and bias < 0.5 ug/mL for the 4 ug/mL pool. Based on the criteria above, the drugs listed above did not interfere under the conditions tested. However, high intralipid concentrations may cause sample turbidity and
{6}
an absorbance value that exceeds 3.0 AU, causing an analyzer condition code and suppressed results.
The drugs shown in the table below cross-reacted. Percent cross-reactivity is defined: (mean concentration with substance – mean concentration without substance) X (100) /(concentration of substance).
Results for samples from patients receiving these drugs may be falsely elevated.
| Drug | % cross-reactivity | Tobramycin concentration (ug/mL) |
| --- | --- | --- |
| Amikacin | 1% | 7 |
| Dibekacin | 55% | 4 |
| Gentamicin | 2% | 4 |
| Kanamycin | 23% | 7 |
f. Assay cut-off:
NA. This is a quantitative assay.
2. Comparison studies:
a. Method comparison with predicate device:
One hundred and eleven human serum samples were evaluated with the VITROS Chemistry Products Tobramycin Reagent and the Syva EMIT® Tobramycin Plus. Samples were selected to be patient serum samples of 2 mL or more that contained tobramycin across the reportable range of the assay. Each sample was measured in triplicate. The analysis was also performed using singlicates and results of this analysis are shown below. Similar results were obtained for 3 lots tested. Slope = 1.02, intercept = 0.0, r = 0.99, sy/x = =0.26.
b. Matrix comparison:
Serum and plasma (EDTA and lithium heparin) samples were evaluated by paired difference testing of samples ranging from approximately 1-9 ug/mL. The bias between the mean values of replicates (n=3) was defined as: bias = test sample-serum sample.
Biases observed ranged from approximately -3% to -2% for lithium heparin plasma tubes and -10% to 2% for full EDTA plasma tubes and were within acceptance limits. Results support use of the assay with serum and plasma (EDTA and lithium heparin) samples. The device is not for use with other anticoagulants.
7
{7}
3. Clinical studies:
a. Clinical Sensitivity:
N/A. (Not typically reviewed for this type of test)
b. Clinical specificity:
N/A. (Not typically reviewed for this type of test)
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
N/A. See expected values
5. Expected values/Reference range:
Guidelines for reference ranges from the literature are provided.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.