← Product Code [KWP](/productcode/KWP) · K043355

# SINTEA BIOTECH POSTERIOR LUMBAR SYSTEM WITH POLYAXIAL AND RECOVERY SCREWS, MODEL PLS.01.T5.X (K043355)

_Sintea Biotech, Inc. · KWP · Dec 17, 2004 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K043355

## Device Facts

- **Applicant:** Sintea Biotech, Inc.
- **Product Code:** [KWP](/productcode/KWP.md)
- **Decision Date:** Dec 17, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Sintea Posterior Lumbar System is a posterior, nonpedicle screw system of the noncervical spine indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, tumor, pseudarthrosis, and failed previous fusion. The Sintea Biotech Posterior Lumbar System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The Sintea Biotech Posterior Lumbar System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-SI vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

## Device Story

Sintea Biotech Posterior Lumbar System is a spinal fixation orthosis; provides immobilization and stabilization of thoracic, lumbar, and sacral spine segments. System components include multi-axial screws, recovery screws, and screws with rings. Multi-axial screws allow up to 25 degrees of angulation to accommodate patient anatomy; recovery screws utilize a locking ring instead of a locking cap. Used by surgeons in clinical settings to treat spinal instabilities or deformities. Device acts as an adjunct to fusion; implants are attached to the spine to facilitate stabilization. Benefits include structural support for bone healing in patients with degenerative conditions or trauma. System is a line extension of existing predicate hardware; shares materials, design, and mechanical properties.

## Clinical Evidence

No clinical data. Biomechanical bench testing was performed to demonstrate that the modified screw constructs are safe and effective for spinal fixation.

## Technological Characteristics

Spinal fixation system; components include multi-axial screws, recovery screws, and locking rings. Materials are identical to predicate system. Multi-axial design allows 25-degree angulation. Compliance with voluntary standards including ASTM and ISO; manufacturing under ISO 9001 quality system.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- Sintea Biotech Posterior Lumbar System ([K020085](/device/K020085.md))
- Xia Spinal System Polyaxial Screw ([K031893](/device/K031893.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# DEC 17 2004

#### 510(k) SUMMARY

## Sintea Biotech Posterior Lumbar System Multi-axial Screw and Recovery Screw with Ring Prepared November 23, 2004 Revised December 10, 2004

Sintea Biotech Posterior Lumbar System Multi-axial Screw and Recovery Trade Name: Screw with Ring Common Name: Spinal Interlaminal Fixation Orthosis Classification Name(s): Spinal Interlaminal Fixation Orthosis and Pedicle Screw Spinal System, Class II. Classification(s): § 888.3050 - Spinal Interlaminal Fixation Orthosis; 888.3070 -Pedicle Screw Spinal System, Class II. Device Class: Class II for all requested indications Classification Panel: Orthopedic Device Panel Product Code(s): KWP, MNH, MNI

Applicant Name & Address: Sintea Biotech, Inc. 407 Lincoln Road Suite 10L Miami Beach, FL 33139 (305) 673-6226 FAX (305) 673-3312

Company Contact: Mr. Guido Zorzoli Sintea Biotech Inc. 407 Lincoln Road Miami Beach, FL 33139 (305) 673-6226 FAX (305) 673-3312

Predicate Device:

As a special 510(k) submission, the predicate device to which we are claiming equivalence is our own product. Sintea Biotech's Posterior Lumbar System with standard screw (K020085, decision date 12/10/2002). This special 510(k) submission represents a modification to the predicate, in that the standard (monoaxial) screw has been modified to be a multi-axial screw and a recovery screw with ring.

Additionally, we are claiming substantial equivalence to the Xia Spinal System's Polyaxial Screw, manufactured by Howmedica Osteonics, K031893, decision date 7/17/2003.

Page 1 of 3

{1}------------------------------------------------

### Description of Device Modification:

This submission is intended to address a line extension to the Sintea Biotech Posterior Lumbar System. The extension consists of the addition of a multi-axial screw, a recovery screw, and a screw for ring.

## Intended Use and Indications for Use:

The Sintea Posterior Lumbar System is a posterior, nonpedicle screw system of the noncervical spine indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, tumor, pseudarthrosis, and failed previous fusion.

The Sintea Biotech Posterior Lumbar System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

The Sintea Biotech Posterior Lumbar System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-SI vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

## Device Description:

PLS recovery screws and screws for ring are available with the same material, shape, fillet, diameters and length of PLS standard screws. They have the same characteristics as PLS standard screws with the exception of a locking ring adopted in place of a locking cap.

The PLS Multi-axial screws are made of the same material as the PLS standard (monoaxial) screws. The somatic thread is the same as the thread of the PLS standard screws. The lengths and diameters of the multi-axial screws are the lengths and diameters of the PLS standard screws. The design of this screw allows for up to 25 degrees of angulation in any direction to accommodate different anatomic conditions.

Please refer to the body of this submission for an extensive description of the multi-axial screw, recovery screw, screw for ring, and ring.

#### Statement of Technological Comparison:

The subject components share the same materials of construction, intended use and basic design characteristics as the predicate devices. Mechanical testing demonstrated comparable mechanical properties to the predicate devices.

Summary Basis for Equivalence and Comparison Table:

{2}------------------------------------------------

Biomechanical studies conducted on the Sintea Biotech Posterior Lumbar System multiaxial screw, recovery screw, screw for ring and ring implant constructs demonstrate that the device system is safe, effective, and suitable for use as a spinal fixation device system. Based on the available information concerning the referenced comparison devices (the Sintea Biotech Posterior Lumbar System, the Xia Spinal System), these devices are similar in that:

- The devices have the same intended use and indications for use; .
- The devices are made of the same implant alloy; and .
- The devices have similar form, function, components, instruments, . dimensions, geometry and features.

### Performance Standards:

Food and Drug Administration mandated Performance Standards for Spinal Interlaminal Fixation Orthosis, Spondylolisthesis Spinal Fixation Device System Class II, and Pedicle Screw Spinal System Class II devices are not in effect. Sintea Biotech, Inc. intends to comply with all voluntary Performance Standards applicable to the Sintea Biotech Posterior Lumbar System. At the present time, various performance standards such as ASTM, ISO, QSR/GMP and in-house SOP standards are used. Sintea Biotech, Inc. also complies with the general controls authorized under Sections 501, 502, 510, 516, 518, 519 and 520 of the Food, Drug and Cosmetic Act. In addition, Sintea Biotech, Srl., which is the location of the manufacturing facility for this device, has earned the CE Mark (number 0546) using the ISO 9001 quality system model, and is in good standing with IQNet, their international certification body.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 17 2004

Mr. Guido Zorzoli Chief Executive Officer Sintea Biotech, Inc. 407 Lincoln Road Suite 10L Miami Beach, Florida 33139

Re: K043355

Trade/Device Name: Sintea Biotech Posterior Lumbar System with multiaxial screw and recovery screw/ring Regulation Number: 21 CFR 888.3050, 888.3070 Regulation Name: Spinal interlaminal fixation orthosis, Pedicle screw spinal system Regulatory Class: II Product Code: KWP, MNH, MNI Dated: November 22, 2004 Received: December 6, 2004

Dear Mr. Zorzoli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 – Mr. Guido Zorzoli

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and in yourse of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you a the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Sincerely yours,

Mark A. Milliken

Celia M. Witten, Ph.D., M.D. Director

Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): K043355

Device Name: Sintea Biotech Posterior Lumbar System with multiaxial screw and recovery screw/ring

Indications for Use:

The Sintea Posterior Lumbar System is a posterior, nonpedicle screw system of the The Olifical Footoner Lated for degenerative disc disease (defined as back pain of nonocrood opine malsate reration of the disc confirmed by history and radiographic discogence ongin finen degencreture, tumor, pseudoarthrosis, and failed previous fusion.

The Sintea Biotech Posterior Lumbar System is a pedicle screw system intended to The Online Dioton't Posterior of spinal segments in skeletally mature patients provide inimobilization and element of the following acute and chronic instabilities or as an adjariet to fablem ic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

The Sintea Biotech Posterior Lumbar System is a pedicle screw system indicated for the

treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-SI vertebra in skeletally

mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

*Mark N. Mikess*
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices

510(k) Number K043355

---

**Source:** [https://fda.innolitics.com/device/K043355](https://fda.innolitics.com/device/K043355)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).