EZ SMART

K043340 · Vip Diagnostics, LLC · NBW · Jun 2, 2005 · Clinical Chemistry

Device Facts

Record IDK043340
Device NameEZ SMART
ApplicantVip Diagnostics, LLC
Product CodeNBW · Clinical Chemistry
Decision DateJun 2, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2

Indications for Use

The EZ Smart Blood Glucose Test Strips are used with the EZ Smart Meter to measure Glucose (sugar) in the whole blood. The EZ Smart Test strips are for testing outside the body (in vitro diagnostic use). The EZ Smart Blood Glucose Monitoring System is intended for use in the home and in the professional settings to monitor blood glucose levels for better glucose level control among diabetics. The EZ Smart Blood Glucose Monitoring System is indicated for use with capillary whole blood samples drawn from the fingertips and forearm.

Device Story

System uses electrochemical biosensor technology to measure glucose in fresh capillary whole blood. User applies blood drop to test strip; glucose oxidase enzyme triggers chemical reaction producing current; meter measures current to calculate glucose concentration. Results displayed in 10 seconds. Used in home or professional settings by patients or clinicians. Provides quantitative glucose values (20-600 mg/dL) to assist in diabetes management and glucose control. No physical changes from predicate; submission adds alternative site testing (forearm) capability.

Clinical Evidence

No clinical data provided; bench testing only. The submission relies on the established performance characteristics of the predicate device, with modifications limited to labeling.

Technological Characteristics

Glucose test system (21 CFR 862.1345). System consists of a meter and test strips for in vitro diagnostic use. No changes to materials, sensing principles, or energy sources were reported.

Indications for Use

Indicated for diabetics to monitor blood glucose levels using capillary whole blood from fingertips or forearm. Not for use during rapid glucose changes (e.g., post-meal, post-insulin, post-exercise) or when hypoglycemia is suspected.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k043340 B. Purpose for Submission: The manufacturer submitted a 510(k) amendment to the EZ Smart Blood Glucose Monitoring System. This amendment addresses only labeling and indication changes for alternative site testing of the forearm. No physical changes were made to the originally cleared EZ Smart Blood Glucose Monitoring System (k040848) C. Measurand: Glucose D. Type of Test: Quantitative electrochemical biosensor technology E. Applicant: VIP International Wholesalers Corp. F. Proprietary and Established Names: EZ Smart Blood Glucose Monitoring System G. Regulatory Information: 1. Regulation section: 21 CFR §862.1345, Glucose test system 2. Classification: Class II 3. Product code: NBW, CGA {1} 4. Panel: 75 H. Intended Use: 1. Intended use(s): See Indications for Use below. 2. Indication(s) for use: The EZ Smart Blood Glucose Test Strips are used with the EZ Smart Blood Glucose Meter to measure glucose (sugar) in whole blood. The EZ Smart Test Strips are for testing outside the body (in vitro diagnostic use). The EZ Smart Blood Glucose Monitoring System is intended for use in the home and in professional settings to monitor blood glucose levels for better glucose level control among diabetics. The EZ Smart Blood Glucose Monitoring System is indicated for use with capillary whole blood samples drawn from the fingertips and forearm. 3. Special conditions for use statement(s): EZ Smart Blood Glucose Test Strips and EZ Smart Control Solutions are to be used only with the EZ Smart Blood Glucose Meter to test glucose in fresh capillary whole blood only. This meter can be used for Alternate Site Testing of the forearm only. This meter is not to be used for Neonatal Testing 4. Special instrument requirements: EZ Smart Blood Glucose Meter. I. Device Description: The EZ Smart Blood Glucose Test Strips are used with the EZ Smart Blood Glucose Meter to quantitatively measure glucose in capillary whole blood. When the edge of the EZ Smart test strip is touched to a drop of blood, the test strip draws the blood into the sample chamber and the glucose reading is displayed on the meter after 10 seconds. The test measures glucose from 20 mg/dL (1.1mmol/L) to 600 mg/dL (33.3 mmol/L). The EZ Smart Test Strip is calibrated to display the equivalent of plasma glucose values to allow the comparison of results with laboratory methods. 2 {2} 3 J. Substantial Equivalence Information: 1. Predicate device name(s): VIP Diagnostics EZ Smart Blood Glucose Monitoring System 2. Predicate 510(k) number(s): k040848 3. Comparison with predicate: The VIP Diagnostics LLC, EZ Smart Blood Glucose Monitoring System in this submission is the same VIP Diagnostics EZ Smart Blood Glucose Monitoring System previously cleared under (k040848). This submission was submitted to provide clinical studies of alternative site testing of the forearm by lay users. Substantial Equivalence Comparison Similarities | Item | Predicate Device EZ Smart (k040848) | Proposed Device EZ Smart (k043340) | | --- | --- | --- | | Similarities | 1. Monitors glucose using whole blood. 2. Directly displays results without requiring calculation. 3. Test principle includes measuring a current produced by a chemical reaction. 4. Test principle: Uses glucose oxidase reaction. 5. Measuring range: 20 to 600 ng/dL. | 1. Monitors glucose using whole blood. 2. Directly displays results without requiring calculation. 3. Test principle includes measuring a current produced by a chemical reaction. 4. Test principle: Uses glucose oxidase reaction. 5. Measuring range: 20 to 600 mg/dL. | K. Standard/Guidance Document Referenced (if applicable): The EZ Smart Blood Glucose Monitoring System has been tested with the listed standards and found in compliance with the council EMC directive 89/336/EEC. {3} 4 | Test Standards | | | | --- | --- | --- | | EN 60601-1-2/1993 | Medical electrical equipment – Electromagnetic compatibility | | | | EN 55011/1998 + A1/1999 | Emissions, Limits and methods of measurement of radio disturbance characteristics of industrial, scientific and media (ISM) radio-frequency equipment. | | | IEC 801-3/1984 | Immunity, Electrostatic discharge | | | IEC 801-3/1984 ENV 50204/1995 | Immunity, Radio-frequency electromagnetic field. | CDRH documents: “Write it Right” and “Recommendations for Developing User Instruction Manuals for Medical Devices Used in Home Health Care” NCCLS Guideline GP-14P- Labeling for Home Use In Vitro Testing Products, NCCLS Guideline EP6-A, SMOG Readability Formula ## L. Test Principle: The test principle is based on electrochemical biosensor technology using glucose oxidase. The strip uses the enzyme glucose oxidase to produce a current that will stimulate a chemical reaction. This reaction is measured by the meter and displayed as the blood glucose result. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Within-run precision of EZ Smart test strips was measured with venous blood samples in the laboratory. The pooled precision data for three hundred seventy five test strip lots (n=375) is shown in the table below: Within-Run Precision | Average Glucose Concentration (mg/dL) | 49.7 | 85 | 121 | 221 | 362 | | --- | --- | --- | --- | --- | --- | | SD (mg/dL) | 2.1 | 1.6 | 4.0 | 6.8 | 9.0 | | CV% | 4.2 | 1.8 | 3.3 | 3.1 | 2.5 | #### b. Linearity/assay reportable range: Linearity (analytical range) studies were designed in accordance with NCCLS Guideline EP6-A. Venous blood was drawn from healthy {4} volunteers and collected into lithium heparin Vacutainer® tubes. The blood was then placed in a room temperature environment overnight, until glycolysis took place to reduce the glucose concentration to nearly zero. The blood was then pooled and allocated to lithium heparin tubes. A small amount of high concentration glucose (10,000 mg/dL) was added to each tube to obtain the desired blood glucose levels needed to perform the test. Recovery tests were performed by confirming the blood glucose concentration with the YSI 2300. Blood samples with glucose concentrations ranging from 20 to 600 mg/dL were then tested with the EZ Smart Blood Glucose Monitoring System. Three lots of EZ Smart Test Strips, chosen at random, were used during the test. The test results were evaluated to describe accuracy over the entire range of blood glucose values. A linear regression analysis was performed by the method of least squares (Y = 0.9717X + 3.66, R² = 0.9987). The sensitivity of the system was determined by the slope of the regression line whereas the linearity was determined by the correlation coefficient of the regression line. All measurements determined by the EZ Smart System are within a 15% bias of the reference (YSI 2300) results (glucose concentration > 100 mg/dL) and a 15 mg/dL bias of the reference (YSI 2300) results (glucose concentration < 100 mg/dL) see table below: | YSI (mg/dL) | 20 | 75 | 152 | 258 | 359 | 482 | 600 | | --- | --- | --- | --- | --- | --- | --- | --- | | EZ Smart Results Lot R030212 | 25 | 77 | 148 | 269 | 358 | 478 | 595 | | | 23 | 82 | 139 | 258 | 354 | 459 | 589 | | | 24 | 77 | 158 | 254 | 368 | 486 | 570 | | | 20 | 71 | 156 | 245 | 348 | 469 | 591 | | | 21 | 74 | 147 | 249 | 347 | 465 | 600 | | | 28 | 76 | 155 | 243 | 345 | 462 | 597 | | | 30 | 83 | 148 | 255 | 350 | 471 | 582 | | | 26 | 76 | 155 | 248 | 351 | 467 | 591 | | MEAN | 24.63 | 77.00 | 150.75 | 252.63 | 352.63 | 469.63 | 589.38 | | SD | 3.38 | 3.93 | 6.36 | 8.37 | 7.44 | 8.78 | 9.55 | | CV% | 13.72% | 5.10% | 4.22% | 3.31% | 2.11% | 1.87% | 1.62% | | Bias% | 23.13% | 2.67% | -0.82% | -2.08% | -1.78% | -2.57% | -1.77% | c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability has been referenced by the manufacturer to ISO 17511. d. Detection limit: The detection range is from 20 – 600 mg/dL (1.1 to 33.3 mmol/L). See linearity/assay detection limit above. {5} e. Analytical specificity: Interference testing was conducted to determine the effect of selected endogenous and exogenous substances. The following study was used to elucidate whether the EZ Smart Blood Glucose Monitoring System is capable of a precise reading in blood samples containing the interferences. Interference studies were conducted according to NCCLS EP7-P. The Interfering Effect of Drugs and Chemical Compounds on EZ Smart Test Strips. | Interfering Compounds | Physiological Levels Test Conc. | Test Concentration | No Interference Level of EZ Smart Test Strip | | --- | --- | --- | --- | | Acetone | | 6 mg/ml | 6 mg/ml | | Acetaminophen | 20 μg/ml | 40 μg/ml | 30μg/ml | | Ascorbic Acid | 12 μg/ml | 150 μg/ml | 75 μg/ml | | Alcohol | | 35 μg/ml | 35 mg/ml | | Barbital | | 1 mg/ml | 1 mg/ml | | Benzoic Acid | | 14.4 mg/ml | 14.4 mg/ml | | Bilirubin | | 200 μg/ml | 200 μg/ml | | Caffeine | | 1 mg/ml | 1 mg/ml | | Cholesterol | 3 mg/ml | 5 mg/ml | 5 mg/ml | | Cholic Acid | | 6 μg/ml | 6 μmol/L | | Creatinine | 15 μg/ml | 300 μg/ml | 300 μg/ml | | EDTA | | 4 mg/ml | 4 mg/ml | | Ephedrine | 0.1 μg/ml | 100 μg/ml | 100μg/ml | | Ethylene Glycol | | 40 μg/ml | 40 mg/ml | | Erythromycin | | 2 mg/ml | 2 mg/ml | | Glycerol | | 1 mmol/L | 1 mmol/L | | Ibuprofen | 42 μg/ml | 400 μg/ml | 400μg/ml | | L-DOPA | 3 μg/ml | 20 μg/ml | 10 μg/ml | | Lecithin | | 50 μg/ml | 50 mg/ml | | Potassium Chloride | | 10 μg/ml | 10 mmol/L | | Salicylate | 0.3 mg/ml | 1.25 mg/ml | 1.25 mg/ml | | Sodium Bicarbonate | | 40 mmol/L | 30 mmol/L | | Sodium Fluoride | | 5 mg/ml | 3.75 mg/ml | | Tetracycline | 4 μg/ml | 40 μg/ml | 40 μg/ml | | Tolazamide | 25 μg/ml | 50 μg/ml | 37.5 mg/ml | | Tolbutamide | 0.1 mg/ml | 1 mg/ml | 1 mg/ml | | Triglyceride | 1.9 mg/ml | 30 mg/ml | 22.5 mg/ml | {6} f. Assay cut-off: Not applicable for this type of device. ## 2. Comparison studies: a. Method comparison with predicate device: 51 patients were tested for blood glucose levels using the EZ Smart Blood Glucose Meter and Strips. Blood glucose levels were determined from tests taken on the index fingers of each patient and compared to results from tests taken on their lower forearms. A trained technician performed the comparison tests initially, followed by the patients on themselves. The technician tested an index finger on the right hand and the right lower forearm within a time limit of five minutes. The patient tested a finger on the left hand and the left lower forearm also within the five minute time limit. Comparative results were plotted on Clarke Error Grids and statistical analysis was performed to determine the correlation between the patient performed finger results to forearm results. The same comparison was calculated for the technician performed finger/forearm results. A comparison was also made between the technician performed test and the patient performed test. The results of the statistical analysis of the Alternate site (Forearm vs. Finger stick) using EP Evaluator (3.23J) are tabulated below: | X / Y | Corr. Coeff. (R) | Slope Regular | Slope Deming | Intercept Regular | Intercept Deming | | --- | --- | --- | --- | --- | --- | | Tech Finger/Tech Alternate | 0.9925 | 1.006 (0.9716 – 1.041) | 1.014 (0.9792 – 1.049) | -1.0 (-6.6 – 4.6) | -2.2 (-7.8 – 3.4) | | Tech Alternate/Patient Alternate | 0.9810 | 0.9311 (.08798 – 0.9824) | 0.9482 (0.8967 – 0.9997) | 9.3 (0.9 – 17.6) | -2.2 (-7.8 – 3.4) | | Tech Finger/Patient Finger | 0.9839 | 0.9146 (0.8683 – 0.9680) | 0.9284 (0.8820 – 0.9748) | 9.8 (2.3 – 17.3) | 7.7 (0.2 – 15.2) | | Patient Finger/Patient Alternate | 0.9862 | 1.021 (0.9731 – 1.069) | 1.036 (0.9878 – 1.084) | -0.8 (-8.4 – 6.7) | -3.0 (-10.6-4.5) | (95% Confidence Intervals are in parentheses). {7} The results of the Clarke Error Grid analysis are summarized below: | X / Y | Region | N | Percent | Cum Percent | | --- | --- | --- | --- | --- | | Tech Finger/ Patient Alternate | A | 51 | 100 | 100 | | Tech Alternate/ Patient Alternate | A | 49 | 96 | 96 | | | B | 2 | 4 | 100 | | Patient Finger/ Patient Alternate | A | 51 | 100 | 100 | | Tech Finger/ Patient Finger | A | 49 | 96 | 96 | | | B | 2 | 4 | 100 | | Tech Finger/ Tech Alternate | A | 51 | 100 | 100 | b. Matrix comparison: Not applicable for this type of device. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not applicable for this type of device 5. Expected values/Reference range: The normal fasting glucose range for a non-diabetic adult is 70 to 110 mg/dL. (3.9 to 6.1 mmol/L)¹. One to two hours after meals, normal glucose values should be less than 120 mg/dL (6.7 mmol/L)². 1. Burtis CA Ashwood ER, eds: Tietz Textbook of Clinical Chemistry. 2nd Edition. W.B. Saunders. Philadelphia. 1994. p. 2190. {8} 2. Krall LP and Beaser RS: Joslin Diabetes Manual. Lea and Febiger. Philadelphia 1989. p. 138. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 9
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...