← Product Code [KNT](/productcode/KNT) · K043258

# MODIFIED TFX MEDICAL SAFETY NEEDLE WITH INTRODUCER, PART NO. T4634-02 (K043258)

_Teleflex Medical · KNT · Dec 16, 2004 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K043258

## Device Facts

- **Applicant:** Teleflex Medical
- **Product Code:** [KNT](/productcode/KNT.md)
- **Decision Date:** Dec 16, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Modified TFX Medical Safety Needle with Introducer is intended to be used for guidewire introduction during gastrointestinal procedures such as PEG (Percutaneous Endoscopic Gastrostomy), PEJ (Percutaneous Endoscopic Jejunostomy) or other endoscopic gastrointestinal procedures requiring placement of a guidewire.

## Device Story

Device consists of safety needle with passive sharps protection and sheath introducer; used for guidewire introduction during gastrointestinal procedures (e.g., PEG, PEJ). Needle incorporates blunter technology for sharps protection; sheath facilitates guidewire placement (0.015"-0.052"). Available in peelable and non-peelable configurations. Operated by clinicians in clinical settings. Material change from high-density polyethylene to polypropylene for cannula sheath; increased outer diameter for rigidity. Device provides mechanical access for guidewire placement; sharps protection mechanism reduces risk of needle-stick injuries to healthcare providers.

## Clinical Evidence

Bench testing only. No clinical data presented.

## Technological Characteristics

Cannula sheath material: Polypropylene (formerly high-density polyethylene). Passive sharps protection mechanism. Accommodates guidewires 0.015"-0.052". Configurations: peelable and non-peelable. Conformance to ANSI/AAMI/ISO 10993-1 for biological evaluation.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- TFX Medical Introducer Needle ([K851140](/device/K851140.md))
- PUNCTUR-GUARD Blood Collection Needle ([K895024](/device/K895024.md))
- TFX Medical Safety Needle with Introducer ([K000665](/device/K000665.md))
- Modified TFX Medical Safety Needle with Introducer ([K021034](/device/K021034.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

TFX Medical Special 510(k) Submission: Modified TFX Medical Safety Needle with Introducer

DEC 1 6 2004

November 16, 2004

K043258
Page 1 of 3

## Attachment 4 510(k) Summary

## Substantial Equivalence

In accordance with the requirements of 21 CFR § 807, this summary is formatted with the Agency's final rule "... 510(k) Summaries and 510(k) Statements ..." and can be used to provide equivalence summary to anyone requesting it from the Agency.

Teleflex Medical Inc. Manufacturer 50 Plantation Drive Tall Pines Park Jaffrey, NH 03452

- Contact Person Susan Kagan Regulatory Affairs Specialist Phone: (508) 677-6675 (888) 273-6897 Fax: e-mail: skagan@teleflexmedical.com
November 16, 2004 Date Prepared

## Device Information

| Common Name:         | Guidewire Introduction Safety Needle with<br>Introducer |
|----------------------|---------------------------------------------------------|
| Proprietary Name:    | Modified TFX Medical Safety Needle with<br>Introducer   |
| Classification Name: | Tube, Gastrointestinal and Accessories                  |

## Indications for Use

The Modified TFX Medical Safety Needle with Introducer is intended to be used for guidewire introduction during gastrointestinal procedures such as PEG (Percutaneous Endoscopic Gastrostomy), PEJ (Percutaneous Endoscopic Jejunostomy) or other endoscopic gastrointestinal procedures requiring placement of a guidewire.

Confidential

Page 26

{1}------------------------------------------------

November 16, 2004

K043258
Page 2 of 3

## Device Description

This Intended Use of the Modified Safety Needle with Introducer device is identical to the Modified TFX Medical Safety Needle with Introducer cleared by the FDA in K021034. In order to provide a more robust design, the material of construction has been changed from High Density Polyethylene to Polypropylene. An increase in the outer diameter has also been implemented to add additional rigidity to the cannula sheath.

The Modified TFX Medical Safety Needle with Introducer will continue to allow placement of guidewires ranging from 0.015" - 0.052". The variance in sizes and lengths is due to the specific procedure, physician preference and patient body type.

This product consists of the following two components:

- Safety Needle (Needle with Passive Sharps Protection)- The Safety Needle, has the 1. same blunter technology as the Bio-Plexus, Punctur-Guard Blood Collection needle (K895024).
- Sheath Introducer- The functionality of the peelable, splitable introducer is identical 2. to the existing introducer sold by TFX Medical, which was initially cleared under K920908 and subsequently cleared under K021034 to be used for guidewire introduction during gastrointestinal procedures.

The sheath will also be made available in a non-peelable configuration.

## Substantial Equivalence

The device is similar in intended use, design, and performance characteristics to the currently cleared Modified TFX Medical Safety Needle with Introducer cleared under K021034.

The determination of substantial equivalence for this device was based on a detailed device description, performance testing and conformance with voluntary performance standards, e.g. ANSI/AAMI/ISO 10993-1 Biological Evaluation of Medical Devices.

#### 1. Identification of the legally marketed device to which the submitter claims equivalence:

The Modified TFX Medical Safety Needle with Introducer is substantially equivalent in design and materials to:

### Confidential

{2}------------------------------------------------

K643258
Page 3 of 3

- The TFX Medical Introducer Needle -- K851140 .
- The PUNCTUR-GUARD Blood Collection Needle of Bio-Plexus, Inc, for the . activation mechanism - K895024
- The TFX Medical Safety Needle with Introducer K000665 .
- The Modified TFX Medical Safety Needle with Introducer-K021034 .

#### Summary of Technological Characteristics: 2.

The device is equivalent technologically to the devices mentioned on pages 2. The change in sheath material from High Density Polyethylene to Polypropylene, plus the inclusion of a non-peelable sheath configuration is the reason for this submission.

The new material for the cannula, Polypropylene, is in the same family, Olefins, as the current material of the cannula, High Density Polyethylene. Therefore they behave very similarly under processing conditions such as molding, sterilization and aging.

The only difference between the peelable and non-peelabe introducers is that the skiving operation, which allows for the splitting of the sheath, is not performed during manufacturing. Otherwise all characteristics are identical.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal for the Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three wavy lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 6 2004

Ms. Susan Kagan Regulatory Affairs Specialist Teleflex Medical 50 Plantation Drive Tall Pines Park JAFFREY NH 03452

Re: K043258

Trade/Device Name: Modified TFX Medical Safety Needle with Introducer Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: 78 KNT Dated: November 16, 2004 Received: November 24, 2004

Dear Ms. Kagan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Nancy C. brogdon

Nancy C. Brogdon

Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation . Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

November 16, 2004

TFX Medical Special 510(k) Submission: Modified TFX Medical Safety Needle with Introducer

# Indications for Use

510(k) Number (if known): K_043258 Device Name: Modified TFX Medical Safety Needle with Introducer

Indications for Use :

## Indications for Use:

The Modified TFX Medical Safety Needle with Introducer is intended to be used for guidewire introduction during gastrointestinal procedures such as PEG (Percutaneous Endoscopic Gastrostomy), PEJ (Percutaneous Endoscopic Jejunostomy) or other endoscopic gastrointestinal procedures requiring placement of a guidewire.

Prescription Use X (Part 21 CFR 801 Subpart D)

. AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

In Magazz

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

Page 1 of 1 (Posted November 13, 2003)

Confidential

---

**Source:** [https://fda.innolitics.com/device/K043258](https://fda.innolitics.com/device/K043258)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
