K043216 · Confirma, Inc. · LLZ · Jan 6, 2005 · Radiology
Device Facts
Record ID
K043216
Device Name
CADSTREAM VERSION 4.0
Applicant
Confirma, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 6, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
CADstream is a computer aided detection (CAD) system intended for use in analyzing magnetic resonance imaging (MRI) studies. CADstream automatically registers serial patient image acquisitions (in 2D or 3D) to minimize the impact of patient motion, segments and labels tissue types based on enhancement characteristics (parametric image maps, referred to as angiogenesis maps), and provides quantitative measurements of morphological features of the segmented tissues. CADstream performs other user-defined post-processing functions (image subtractions, multiplanar and oblique reformats, maximum intensity projections, image averaging, removal of cardiac artifact). CADstream also can be used to provide accurate and reproducible measurements of the segmented tissue volumes of interest (VOI)). These measurements include longest diameter, longest in-plane diameters, volume measurement-(reported in cc), ratio of breast volume to VOI, distance of VOI to anatomical landmarks, and 3D renderings of the VOI. The system includes an optional remote image viewer (CADalyst™), optimized for viewing breast MR studies processed by CADstream. CADalyst also displays and telecommunicates images from a number of other medical modalities, including ultrasound and digitized mammographic images, processed data from FDA-cleared third party image processing systems, and FDA-approved systems for computer-aided detection and advanced image processing. The ultrasound and digitized mammographic images displayed on the CADalyst remote image viewer must not be used for primary diagnostic interpretation. CADstream™ includes software to support the use of interventional breast coils and MR stereotactic localization devices to perform MR-guided breast interventional procedures (SureLocTM). Using information from MR images regarding the coordinates of a user-specified region of interest, and fiducial coordinates, the software provides an automatic calculation of the location and depth of the targeted region of interest, such as a lesion or suspected lesion. CADstream also includes the option to add annotations based on the American College of Radiology's Breast Imaging Reporting and Data System (BI-RADS®) Breast Imaging Atlas. When interpreted by a skilled physician, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made based solely on the results of CADstream analysis.
Device Story
CADstream analyzes dynamic breast MRI studies; inputs include serial 2D/3D MR image acquisitions. System uses multivariate pattern recognition to segment/classify tissue based on signal intensity enhancement (angiogenesis maps). Outputs include parametric maps, quantitative morphological measurements (VOI, diameter, volume), and 3D renderings. Operates on off-the-shelf PC hardware; includes server software for analysis and CADalyst viewer for clinician review. Used in clinical settings by radiologists/physicians to assist in screening and diagnosis. SureLoc feature calculates coordinates for MR-guided breast biopsies/interventions. Provides supplemental information for clinical decision-making; patient management decisions must not rely solely on device output.
Clinical Evidence
Bench testing only. No clinical data provided. System performance validated through risk analysis, design verification, and validation testing processes to ensure comparability to predicate devices.
Technological Characteristics
System consists of proprietary software on off-the-shelf PC hardware. Features include DICOM query/retrieve, multi-modality image processing, and automated registration. Connectivity via network to MR scanners and PACS. Software performs multivariate pattern recognition for tissue segmentation. Includes SureLoc for interventional coordinate calculation.
Indications for Use
Indicated for use by physicians to analyze MRI studies, specifically dynamic breast MRI, for tissue segmentation, classification, and quantitative measurement of morphological features. Supports MR-guided breast interventional procedures via coordinate calculation. Not for primary diagnostic interpretation of ultrasound or digitized mammographic images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
# 7
# 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87 (h).
# 1. Identification of Submitter:
| Submitter: | Confirma, Inc. |
|----------------|-------------------------------------------|
| Address: | 821 Kirkland Avenue<br>Kirkland, WA 98033 |
| Phone: | 425-576-1226 |
| Fax: | 425-576-9295 |
| Contact: | Patricia A. Milbank |
| Title: | Regulatory Consultant |
| Phone: | 425-894-9733 |
| Fax: | 425-865-9023 |
| Date Prepared: | November 19, 2004 |
# 2. Identification of Product:
| Trade Name: | CADstream™ Version 4.0 |
|----------------------|-----------------------------------------------|
| Common Name: | Picture Archiving and Communications System |
| Classification Name: | Image Processing System (LLZ) 21 CFR 892.2050 |
| Manufacturer: | Confirma, Inc. |
|---------------|---------------------|
| | 821 Kirkland Avenue |
| | Kirkland, WA 98033 |
### 3. Marketed Devices
The multi-modality review workstation provided in CADstream Version 4.0 is substantially equivalent to the devices listed below:
| Model: | Seno Advantage |
|-----------------|-----------------------------------|
| Manufacturer: | GE Medical Systems |
| 510 (k) Number: | K033400 |
| Model: | Sectra IDS5 Radiology Workstation |
| Manufacturer: | Sectra-Imtec Ab |
| 510 (k) Number: | K033712. K040376 |
{1}------------------------------------------------
# 4. Device Description:
The CADstream device relies on the assumption that pixels having similar MR signal intensities represent similar tissues. The CADstream software simultaneously analyzes the pixel signal intensities from multiple MR sequences and applies multivariate pattern recognition methods to perform tissue segmentation and classification. CADstream is designed to analyze dynamic breast MRI studies.
The CADstream system consists of proprietary software developed by Confirma installed on off-the-shelf personal hardware. The CADstream System consists of the following key components:
- A PC Server: a desk-side or rack-mount PC capable of running the . Server software, Client User Interface software, and Study Viewer software
- Server software: performs CADstream analysis and processing .
- . Client User Interface software: an administrative web page hosted on the Server
- Study Viewer software (CADalyst): an Image Viewer component, . optimized for viewing breast MR studies processed by CADstream
- . Archive System: a PC with a combination CD/DVD burner and printer
To initiate CADstream analysis and processing, images are acquired and pushed to the CADstream system from the MR scanner. CADstream is configured to automatically process the images and create additional series. In this configuration, CADstream forwards the processed and original images to the existing softcopy reading station.
CADstream Version 4.0 includes the following software features:
- CADstream™, which allows the user to select a study to view, define . settings for CADstream processes, process and export CADstream results, and perform administrative tasks.
- CADalyst™ (the study viewer), which allows the user to view . CADstream results, interact with the volumes, curves, and AngioMap overlays, and print and save a CADstream Portfolio. This component can be accessed from a standalone CADstream workstation, a remote computer connected to the same network as the CADstream server, such as a PACS, or a PC.
- . CADalyst includes multi-modality viewing capabilities, allowing the user to view ultrasound and digitized mammographic images. The ultrasound and digitized mammographic images displayed on the CADalyst remote image viewer must not be used for primary diagnostic interpretation.
{2}------------------------------------------------
- SureLoc™ is a guidance tool that assists the user in calculating target ● coordinates for biopsy, wire localization, or other interventional procedures.
- DICOM query/retrieve and print, which allows retrieval of DICOM . data from any DICOM-compliant device that is configured for CADstream query and retrieval.
#### 5. Indications for Use
CADstream is a computer aided detection (CAD) system intended for use in analyzing magnetic resonance imaging (MRI) studies. CADstream automatically registers serial patient image acquisitions (in 2D or 3D) to minimize the impact of patient motion, segments and labels tissue types based on enhancement characteristics (parametric image maps, referred to as angiogenesis maps), and provides quantitative measurements of morphological features of the segmented tissues. CADstream performs other user-defined post-processing functions (image subtractions, multiplanar and oblique reformats, maximum intensity projections, image averaging, removal of cardiac artifact).
CADstream also can be used to provide accurate and reproducible measurements of the segmented tissue volumes of interest (VOI)). These measurements include longest diameter, longest in-plane diameters, volume measurement-(reported in cc), ratio of breast volume to VOI, distance of VOI to anatomical landmarks, and 3D renderings of the VOI.
The system includes an optional remote image viewer (CADalyst™), optimized for viewing breast MR studies processed by CADstream. CADalyst also displays and telecommunicates images from a number of other medical modalities, including ultrasound and digitized mammographic images, processed data from FDA-cleared third party image processing systems, and FDA-approved systems for computer-aided detection and advanced image processing. The ultrasound and digitized mammographic images displayed on the CADalyst remote image viewer must not be used for primary diagnostic interpretation.
CADstream™ includes software to support the use of interventional breast coils and MR stereotactic localization devices to perform MR-guided breast interventional procedures (SureLocTM). Using information from MR images regarding the coordinates of a user-specified region of interest, and fiducial coordinates, the software provides an automatic calculation of the location and depth of the targeted region of interest, such as a lesion or suspected lesion.
{3}------------------------------------------------
CADstream also includes the option to add annotations based on the American College of Radiology's Breast Imaging Reporting and Data System (BI-RADS®) Breast Imaging Atlas.
When interpreted by a skilled physician, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made based solely on the results of CADstream analysis.
# 6. Comparison with Predicate Devices
The CADalyst multi-modality review workstation provided with CADstream Version 4.0 is substantially equivalent to the following image processing systems used by radiologists:
| Seno Advantage Workstation | |
|----------------------------|--------------------|
| Manufacturer: | GE Medical Systems |
| 510 (k) Number: | K033400 |
Sectra IDS5 Radiology Workstation Manufacturer: Sectra-Imtec Ab K033712, K040376
Each of these workstations allows easy selection, review, processing, archive, printing and media interchange of multi-modality images from a variety of diagnostic imaging systems.
### 7. Conclusions
CADstream Version 4.0, with the remote CADalyst multi-modality review workstation, provides additional features to further integrate radiology department workflow. The potential hazards have been studied and controlled as part of the product development process, including risk analysis, test and design considerations, and planned verification and validation testing processes. The CADstream system provides images comparable to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 6 2005
Ms. Patricia A. Milbank Regulatory Consultant Confirma, Inc. 821 Kirkland Avenue, Suite 100 KIRKLAND WA 98033
Re: K043216 Trade/Device Name: CADstream™ Version 4.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: November 19, 2004 Received: November 22, 2004
Dear Ms. Milbank:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis loker will and wys a substantial equivalence of your device of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html
Sincerely yours,
Nancy C. brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indication(s) for Use Statement 6
510(k) Number:
To be assigned by FDA
CADstream™ Version 4.0 Device Name:
Indications for Use:
CADstream is a computer aided detection (CAD) system intended for use in analyzing magnetic resonance imaging (MRI) studies. CADstream automatically registers serial patient image acquisitions (in 2D or 3D) to minimize the impact of patient motion, segments and labels tissue types based on enhancement characteristics (parametric image maps, referred to as angiogenesis maps), and provides quantitative measurements of morphological features of the segmented tissues. CADstream performs other user-defined post-processing functions (image subtractions, multiplanar and oblique reformats, maximum intensity projections, image averaging, removal of cardiac artifact).
CADstream also can be used to provide accurate and reproducible measurements of the segmented tissues volumes (volumes of interest (VOI)). These measurements include longest diameter, longest in-plane diameters, volume measurement (reported in cc), ratio of breast volume to VOI, distance of VOI to anatomical landmarks, and 3D renderings of the VOI.
The system includes an optional remote image viewer (CADalyst™), optimized for viewing breast MR studies processed by CADstream. CADalyst also displays and telecommunicates images from a number of other medical modalities, including ultrasound and digitized mammographic images, processed data from FDA-cleared third party image processing systems, and FDA-approved systems for computer-aided detection and advanced image processing. The ultrasound and digitized mammographic images displayed on the CADalyst remote image viewer must not be used for primary diagnostic interpretation.
CADstream™ includes software to support the use of interventional breast coils and MR stereotactic localization devices to perform MR-guided breast interventional procedures (SureLoc™). Using information from MR images regarding the coordinates of a user-specified region of interest, and fiducial coordinates, the software provides an automatic calculation of the location and depth of the targeted region of interest, such as a lesion or suspected lesion.
{7}------------------------------------------------
CADstream also includes the option to add annotations based on the American College of Radiology's Breast Imaging Reporting and Data System (BI-RADS®) Breast Imaging Atlas.
When interpreted by a skilled physician, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made based solely on the results of CADstream analysis.
Over-The-Counter Use Prescription Use _ AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーー
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Legmann
---
(Division Sign-Off)
(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.