ARC SURGICAL RESORBABLE COMPRESSION SCREW SYSTEM

K043176 · Arc Surgical, LLC · HWC · Jun 17, 2005 · Orthopedic

Device Facts

Record IDK043176
Device NameARC SURGICAL RESORBABLE COMPRESSION SCREW SYSTEM
ApplicantArc Surgical, LLC
Product CodeHWC · Orthopedic
Decision DateJun 17, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ARC Surgical Resorbable Compression Screw is a bioabsorbable PLLA screw intended to provide fixation and/or reduction of bone fragments and bones. Specifically, as listed below, the screw is intended to fix osteochondral fragments, osteotomies and fractures in the foot, and limited fractures in the wrist, elbow, and ankle. - Osteochondral fragments - Femoral condyle - Talus - Navicular - Hallux Valgus correction - Tarsal fractures - Distal Radius fractures - Radial Head fractures.

Device Story

Bioabsorbable bone fixation system; composed of poly-L-lactic acid (PLLA) screws; headless design with varying thread pitch for compression; tapered square cannulation; square-geometry driver. Used by surgeons for internal fixation of bone fragments/fractures in foot, wrist, elbow, ankle. Provided sterile; injection molded from PLLA pellets. Device provides mechanical stability to bone fragments during healing; material resorbs over time.

Clinical Evidence

No clinical data. Assessment of performance data and clinical/non-clinical tests stated as not applicable.

Technological Characteristics

Material: Poly-L-lactic acid (PLLA) (Boehringer-Ingleheim Resomer 210S). Polymerization: Ring-opening polymerization using stannous (tin) octoate catalyst. Dimensions: 4.3mm-4.8mm diameter, 16mm-24mm length. Design: Headless, varying thread pitch, tapered square cannulation. Sterilization: Provided sterile.

Indications for Use

Indicated for fixation and/or reduction of bone fragments and bones, including osteochondral fragments (femoral condyle, talus, navicular), hallux valgus correction, tarsal fractures, distal radius fractures, and radial head fractures in the foot, wrist, elbow, and ankle.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K043176 pg. 1 of 2 ## JUN 1 7 2005 ## ARc SURGICAL LLCr 5885 N.W. Cornelius Pass Road Suite 100 Hillsboro, Oregon 97124 Tel (503) 645-9300 ## 510(k) Summary This summary regarding 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. Submitter Information: Arc Surgical LLC 5885 N.W. Cornelius Pass Road, Suite 100 Hillsboro, OR 97124 USA Phone: (503) 645-9300 FAX: (503)645-9304 Contact: Ed Boehmer, Regulatory Supervisor | Classification Name: | Smooth or threaded metallic bone fixation fastener | |----------------------------------------|---------------------------------------------------------------------------------------------------------| | Common Name: | Screw, Fixation, Bone | | Proprietary Name: | Arc Surgical Resorbable Compression Screw System | | Proposed Regulatory Class: | Class II, 21 CFR 888.3040 | | Device Product Code: | HWC | | Legally Marketed Equivalent Device(s): | Arthrex Bio-Compression Screw K032098<br>Bionx Smart Screw K003077<br>Biomet ReUnite Bone Screw K992301 | Device Description: The ARC Surgical Resorbable Compression Screw System is composed of screws injection molded from poly-L-lactic acid (PLLA) with diameter (4.3mm to 4.8mm) and length (16mm to 24mm). Headless screw compression is obtained from the varying pitch of the threadless screw has a tapered square cannulation and the driver has a square-shaped geometry to drive the screw. The screws are provided sterile. Intended Use: The ARC Surgical Resorbable Compression Screw is a bioabsorbable PLLA screw intended to provide fixation and/or reduction of bone fragments and bones. Specifically, as listed below, the screw is intended to fix osteochondral fragments, osteotomies and fractures in the foot, and limited fractures in the wrist, elbow, and ankle. - . Osteochondral fragments - Femoral condyle o - Talus 0 - Navicular ্ - Hallux Valgus correction ● - Tarsal fractures . - Distal Radius fractures ● - Radial Head fractures . These are similar to intended use of predicate devices and do not raise new issues of safety and effectiveness. {1}------------------------------------------------ K043176 page 2 of 2 Technological Characteristics: The screws are being injection molded from PLLA pellets (Boehringer-Ingleheim Resomer 210S). The PLLA pellets are polymerized from L-lactide monomers through the process of ring-opening polymerization. Stannous (tin) octoate is utilized as the catalyst. The predicates devices listed use either PLLA or LactoSorb (82% PLLA/18%PGA). ## An assessment of performance data is not applicable. A discussion of clinical and non-clinical tests is not applicable. Based upon the similarities of the ARC Surgical Resorbable Compression Screw System and the predicate devices studied, the safety and effectiveness of the ARC Surgical Resorbable Compression Screw System is substantially equivalent to the predicate devices referenced. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo. Public Health Service JUN 17 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Ed Boehmer ARC Surgical LLC 5885 N.W. Cornelius Pass Road, Suite 100 Hillsboro, Oregon 97229 Re: K043176 Trade/Device Name: Arc Surgical Resorbable Compression Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: May 9, 2005 Received: May 10, 2005 Dear Mr. Boehmer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 – Mr. Ed Boehmer This letter will allow you to begin marketing your device as described in your Section 510(k) I this letter will anow you to oegin manteeing your antial equivalence of your device to a legally premits et notheation. The PDA miaing of casification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac not 2011-01-20 2017 - 120 - 120 - 12-2 - 12 "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Misorallumg by Iciercice to premation nothilities under the Act may be obtained from the Other general information on Joan responsional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Signature Miriam Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 1 Page of K043176 510(k) Number (if known): _ Arc Surgical Resorbable Compression Screw System Device Name: _ Indications For Use: System is intended to fix The Arc Surgical Resorbable Compression Screw The Arc Surgical Resorband Collipression "Ocrow" - System" - System" - 12 osteochondral fragments, osteotonies and fractures in the foot, and limited fractures in the wrist, elbow, and ankle, as listed: - Osteochondral fragments . - o Femoral condyle - Talus o - o Navicular - Hallux Valgus correction . - Tarsal fractures . - Distal Radius fractures . - Radial Head fractures . Prescription Use _X_ (Part 21 CFR 801 Subpart D) AND/OR No Over-The-Counter Use _ (Part 21 CFR 807 Subpart C) (Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Signature ាំ​ន​ចោ និង​ ១-Of Cision of General Gestorative d Neurological Levices 0431
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%