K043148 · Alliance Tech Medical, Inc. · CAH · Dec 13, 2004 · Anesthesiology
Device Facts
Record ID
K043148
Device Name
ALL FLOW PULMONARY FUNCTION FILTER
Applicant
Alliance Tech Medical, Inc.
Product Code
CAH · Anesthesiology
Decision Date
Dec 13, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5260
Device Class
Class 2
Indications for Use
The ATM All Flow Pulmonary Function Filter is designed as a disposable and singleuse bi-directional filter for use in reducing possible bacterial and/or viral cross contamination of spirometers and pulmonary function testing instruments, associated valves and hoses, from aerosols and particulates, which may be present in a patient's exhaled gas. The device is indicated for diagnostic purposes.
Device Story
Disposable, single-use, bi-directional filter; protects spirometers and pulmonary function testing instruments from patient-exhaled bacterial/viral pathogens. Device consists of two molded plastic housings enclosing a disk of filter media; patient-side housing is opaque white; machine-side housing is color-coded by size. Available in six sizes to fit various sensor diameters. Used in clinical settings during diagnostic pulmonary function testing. Healthcare providers attach the filter between the patient and the testing instrument. By filtering aerosols and particulates while maintaining low airflow resistance, the device prevents pathogen transmission to equipment and subsequent patients, ensuring instrument hygiene and compliance with American Thoracic Society standards.
Clinical Evidence
Bench testing only. Bacterial Filtration Efficiency (BFE) > 99.99%; Viral Filtration Efficiency (VFE) > 99.99%. Airflow resistance measured at 0.23 cm H2O/L/sec. Filter dead space approximately 35 mL. Device meets American Thoracic Society Standardization of Spirometry recommendations.
Technological Characteristics
Bi-directional barrier filter; molded plastic housings; filter media disk. Six sizes available (25.7mm to 45.0mm ID). Passive filtration; no energy source. Non-sterile. Meets American Thoracic Society standards for airflow resistance and dead space.
Indications for Use
Indicated for diagnostic applications in patients undergoing pulmonary function testing or spirometry to reduce cross-contamination of equipment from exhaled aerosols and particulates.
Regulatory Classification
Identification
A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.
{1}------------------------------------------------
K043/48
Alliance Tech Medical, Inc. All Flow Pulmonary Function Filter 510(k) Submission
### 4.0 510(k) SUMMARY
DEC 1 3 2004
In accordance with 21 CFR section 807.92, Alliance Tech Medical, Inc. is submitting the following 510(k) summary:
#### 4.1 Submitter Information:
Alliance Tech Medical, Inc. Company Representative: John Silva, Exec. V.P. 5305 Mission Circle Granbury, TX 76049 USA FDA Registration No .: 3004476631 Owner / Operator No.: 9060999
### 4.2 Preparer of Submission and Contact for Information:
## Solutions MDI, Inc.
Gus Bock, Managing Director 40310 Calle de Suenos Murrieta, CA 92562 Telephone: (909) 698-1505 / Fax: (909) 677-3261
Keith Lowrey, Partner & Consultant (contact for correspondence and information) 611 South Schoolhouse Creek Rd. Grants Pass, OR 97526 Telephone: (541) 476-1628 / Cell: (541-601-3695
### 4.3 Name of Device:
| Proprietary Name: | All Flow Pulmonary Function Filter. |
|----------------------|------------------------------------------------------------|
| Common Name: | Disposable bacterial/viral filter. |
| Classification Name: | Breathing circuit bacterial filter, [21 CFR 868.5260(a)]. |
| Regulation Number: | 21 CFR 868.5260(a) for Breathing circuit bacterial filters |
| Product Code: | 73 CAH |
| Class: | Class II (performance standards) |
{2}------------------------------------------------
Alliance Tech Medical, Inc. All Flow Pulmonary Function Filter 510(k) Submission
### 4.4 Substantial Equivalence:
This submission establishes the substantial equivalence of the Alliance Tech Medical, Inc. All Flow Pulmonary Function Filter to five predicate devices:
- The Main Flow Bacterial/Viral PF Filter, K961914, SE letter: 08/14/96. (1)
- (2) The SpiroSafe Disposable PF Filter, K973314, SE letter: 11/21/97. (3) The KOKO Disposable PF Filter II, K934475, SE letter: 10/21/93. The MicroGard Disposable PF Filter, (4) K934272, SE letter: 11/24/93. The MultiSPIRO Disposable PF Filter, K951410, SE letter: 04/24/95. (5)
### 4.5 Description of the Device:
The All Flow Pulmonary Function Filter, manufactured by Alliance Tech Medical, Inc., resin that is a see barrier type, bi-directional filter fabricated from a plastic fittering modiment the customer packaged and non-sterile. Fabricated with a filtering medium that is highly effective in reducing the numbers of both bacterial and viral contaminates that may be present in a patient's exhaled gas, the device's design also minimizes the resistance to air flow. The product is intended to protect both the patient and pulmonary function instruments from potential transmission of pathogens by droplets and aerosols.
The device consists of two molded plastic housings enclosing a disk of filter media. To avoid confusion, the patient housing side is made from a white opaque of miller the and the machine housings are fabricated from the same plastic but with different colors. The colored end fits the machine and the white opaque end is for the patient side; therefore making clear definition as to which end is used for the patient and the machine.
Example of Filter Part No. 5551000 with patient and machine housing components separated:
Image /page/1/Picture/9 description: The image shows a dark, low-lit scene with two primary objects. On the left, there's a cylindrical object with an angled top, possibly made of cardboard or a similar material. To the right and slightly above, there's a round, reflective object, which could be a mirror or a metallic disc, catching the dim light and creating a bright spot in the image.
Opaque white colored patient housing
White colored machine housing
0012
{3}------------------------------------------------
Because spirometers can have different sensor diameters, the All Flow Pulmonary Function Filter is available in six different sizes and colors that correlate with their respective part numbers. The following chart lists the size and color of each model:
| ATM PF Filter Part No. | Color combination<br>(patient housing / machine housing) | I.D / O.D (mm) |
|------------------------|----------------------------------------------------------|----------------|
| 5551000 | Opaque white / solid white | 26.3 / 30.0 |
| 5552000 | Opaque white / solid red | 40.0 / 40.0 |
| 5553000 | Opaque white / solid orange | 45.0 / 48.5 |
| 5554000 | Opaque white / solid blue | 30 /34 |
| 5555000 | Opaque white / solid green | 30 / 33.5 |
| 5556000 | Opaque white / solid yellow | 25.7 / 28.3 |
# Filter Chart
### 4.6 Intended Use of the Device:
The ATM All Flow Pulmonary Function Filter is designed as a disposable and singleuse bi-directional filter for use in reducing possible bacterial and/or viral cross contamination of spirometers and pulmonary function testing instruments, associated valves and hoses, from aerosols and particulates, which may be present in a patient's exhaled gas. The device is indicated for diagnostic purposes.
### 4.7 Technological Characteristics in Comparison to the Predicates:
The All Flow Pulmonary Function Filter is substantial equivalent to the five predicate devices with respect to the following design characteristics and functions:
- The devices are intended for use in reducing the possible cross contamination 1. of spirometers and pulmonary function testing instruments from bacterial and/or viral pathogens by droplets, particulates, and/or aerosols.
- 2. The devices function as a barrier type, bi-directional filter, which is within a sealed double port assembly, with one port contacting the patient's mouth and the other port to be attached to the spirometers or pulmonary function-testing instrument.
- 3. The device housing components are fabricated from the same or similar plastic resin materials used in the predicate devices. The filter media is the same as or similar to that used in the predicate devices. These materials have long and extensive use in medical device applications.
- 4. The devices have been demonstrated to function effectively in reducing high challenge numbers of bacterial and viral contaminates.
- ട്. The devices provide air filtration while minimizing airflow resistance through the filter assemblies in the pulmonary test equipment.
- 6. The devices meet the recommendations of the American Thoracic Society's Standardization of Spirometry.
{4}------------------------------------------------
### 4.8 Conclusions drawn from the Non-Clinical Tests:
Data provided in this submission indicate that the basic functional characteristics of the All Flow Pulmonary Function Filter are substantially equivalent to those of the predicate devices.
- Bacterial Filtration Efficiency (BFE) testing demonstrated the All Flow 1. Pulmonary Function Filter, at an increased challenge, to be %BFE = > 99.99%.
Viral Filtration Efficiency (VFE) testing demonstrated the All Flow Pulmonary Function Filter, at an increased challenge, to be %VFE = > 99.99%.
- Airflow resistance testing demonstrated the device to an average of 2.625 cm 2. H₂O½SLPM at 720 L. (Ref. 760 mm HGA 70 °F), which is approximately 0.23 cm H2O/L/sec.
- Filter dead space was demonstrated to be approximately 35 mL. ന്
- The device meets the recommendations of the American Thoracic Society's র . Standardization of Spirometry.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human form or a bird in flight, composed of three curved lines.
Public Health Service
DEC 1 3 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Alliance Tech Medical, Incorporated C/O Mr. Keith Lowrey Solutions MDI, Incorporated 611 South Schoolhouse Creek Road Grants Pass, Oregon 97526
Re: K043148
Trade/Device Name: All Flow Pulmonary Function Filter Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: II Product Code: CAH Dated: November 11, 2004 Received: November 16, 2004
Dear Mr. Lowrey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
## Page 2 - Mr. Lowrey
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Chris
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Alliance Tech Medical, Inc. All Flow Pulmonary Function Filter 510(k) Submission
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): ___________________________________________________________________________________________________________________________________________________
Device Name: ·
All Flow Pulmonary Function Filter
-. '
Indications for Use:
The Alliance Tech Medical, Inc. All Flow Pulmonary Function Filter is designed as a disposable and single-use, bi-directional filter for use in reducing possible bacterial and/or viral cross contamination of spirometers and pulmonary function testing instruments, associated valves and hoses, from aerosols and particulates, which may be present in a patient's exhaled gas. The device is indicated for diagnostic applications.
(PLEASE NO NOT WRITE: BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CRDH, Office of Device Evaluation (ODE)
Cree Syekom
(Division Sign-Off (Division of Anesthesiology, General Hospital, Infection Control, Dental Device
510(k) Number
Prescription Use (per 21 CFR 801.109)
OR Over-the-Counter Use (Optional Format 1-2-96)
00299
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.