MODIFICATION TO PAJUNK STIMULONG PLUS CATHETER SETS
Applicant
Pajunk GmbH
Product Code
CAZ · Anesthesiology
Decision Date
Dec 21, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5140
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pajunk Stimulong Plus Catheter Sets are intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery is accomplished using the conduction catheter. To assist the physician pinpoint the area of application an electrical stimulus can be applied to the conduction needle and after placement of the conduction catheter to its tip via the catheter adapter.
Device Story
Single-use, sterile, non-pyrogenic conduction anesthesia set; includes conduction needle with tubing, catheter, and catheter adapter. Physician uses needle to access peripheral nerves/plexus; electrical stimulus applied to needle tip to assist in pinpointing application site. Once placed, catheter allows continuous anesthesia delivery for up to 72 hours. Stimulus can also be applied to catheter tip via adapter after placement. Used in clinical settings by physicians to provide regional anesthesia; benefits patient through targeted, continuous pain management.
Clinical Evidence
Bench testing only. Conformance to recognized consensus standards (DIN, ISO, EN, BS) verified for materials, sterilization, packaging, and mechanical properties. Biocompatibility testing performed for the conductive tip, wire, and alternate plastic material.
Technological Characteristics
Materials: Stainless steel (DIN 17440, DIN EN/ISO 9626), polypropylene. Components: Conduction cannula, catheter with conductive ring tip, catheter adapter. Energy: Electrical stimulation for needle/catheter placement. Connectivity: None. Sterilization: ETO (DIN EN 550). Standards: ISO 594 (Luer fittings), ISO 10993-1 (Biocompatibility), EN ISO 14971 (Risk Management).
Indications for Use
Indicated for patients requiring continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Used by physicians to assist in pinpointing application areas via electrical stimulation.
Regulatory Classification
Identification
An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.
{1}------------------------------------------------
DFC 21 2004
K043130
# 510(k) Summary of Safety and Effectiveness
### Submitter Information:
PAJUNK GmbH Karl Hall Straße 1 D78187 Geisingen Germany
## USA Contact:
Mr. Burk A. Brandt CE Consultancy, Inc. 5010 NW Crescent Valley Dr. Corvallis, OR 97330 - USA
Phone: (541) 752-3953 Fax: (541) 752-3972
## Device Name:
Trade Name: Stimulong Plus Catheter Sets Plexalong Sets Unipolar Needles Common Name: Anesthesia Conduction Kit Anesthesia Conduction Needle (W/Wo Introducer) Classification Name: Anesthesia Conduction Kit (Reference, 21CFR, 868.5140, April 1, 2003), Anesthesia Conduction Needle (Reference, 21CFR, 868.5140, April 1, 2003)
#### Predicate Devices:
The Stimulong Plus Catheter sets and Plexalong sets consist of a Pajunk Unipolar needle (conduction cannula with nerve stimulus connector and tubing), and an open ended conduction catheter and a catheter adapter. The catheter in the Stimulong Plus Catheter is the same as the catheter in the Plexalong sets except for the conductive ring tip. The Stimulong Plus Catheter sets have been cleared for market by the FDA under 510(k) number K033018. The Plexalong sets have been cleared under K013041 and the Unipolar needles under 510(k) number K000722 (facet and Sprotte tip conduction cannula and tubing). Unipolar needles with a Tuohy tip were cleared for market under 510(k) number K023218.
The packaging materials used to package the Stimulong Plus Catheter, Pajunk Plexolong sets, and cannula has not changed. The contract sterilizer, other than a company name change (was Griffith Micro Science, now Sterigenics) and sterilizing process is the same.
{2}------------------------------------------------
## Device Description:
The Paiunk Stimulong Plus Catheter Sets and Plexalong sets are single use, sterile, non-pyrogenic and latex free conduction anesthesia sets intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. The Stimulong Plus Catheter sets and Plexalong sets consist of a single use sterile, non-pyrogenic conduction needle with tubing, a catheter and catheter adapter. The catheter in the Stimulong Plus Catheter is the same as the catheter in the Plexalong sets except for a conductive ring tip. Continuous delivery is accomplished using the catheter. To assist the physician pinpoint the area of application an electrical stimulus can be applied to the tip of the conduction needle and after placement of the conduction catheter (Stimulong Plus catheter) to its tip via the catheter adapter.
#### Intended Use:
The Pajunk Stimulong Plus Catheter Sets are intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery is accomplished using the conduction catheter. To assist the physician pinpoint the area of application an electrical stimulus can be applied to the conduction needle and after placement of the conduction catheter to its tip via the catheter adapter.
# Technology Characteristics:
Except for the insulating material used to coat the needles, the Pajunk conduction cannula, including the physical dimensions, connector, tubing, metal and plastics, have been cleared under 510(k) numbers K000722. K013041 and K023218. The materials used to manufacture the Pajunk catheter and catheter adapter except for the conductive catheter tip and wire are identical to the materials used to manufacture the catheter and catheter adapter of the predicate devices described earlier in this 510(k) Summary of Safety and Effectiveness. Biocompatibility testing of the conductive tip, wire and alternate plastic material is located in Section 7 of this submission. Use of a conduction catheter was cleared by FDA under 510(k) No. K030937. The Stimulong Plus Catheter Sets are supplied in polypropylene containers that are sealed to assure sterility.
{3}------------------------------------------------
# Summary of Performance Testing
The Pajunk Stimulong Plus Catheter Sets were designed to conform to the applicable sections of the following recognized consensus standards. The testing included verifying conformance to these standards.
| Standard | Issue Date | Title |
|--------------------|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DIN 13090/ISO 594 | 08.1984 | Luer fittings w/wo locking feature |
| DIN 13097 Part 1 | 01.1980 | Medical injection cannula |
| DIN 13097 Part 3 | 11.1979 | Medical cannula |
| DIN 17442/ISO 9626 | 10.1977 | Steel for medical instruments |
| DIN EN 550 | 07.1993 | Sterilization of med. Prod.; Validation & routine controls<br>for sterilization with ETO |
| DIN EN 556 | 01.1995 | Sterilization of medical products, requirements for<br>medical products that are labeled "sterile" |
| DIN EN 724 | 12.1994 | Guidance on the application of EN29001 and EN46001<br>for non-active medical products |
| PrEN 868-1 | 10.1996 | Packaging materials for the sterilization of packaged<br>goods. Part 1: general requirements for the validation<br>of the packaging of sterilized end-packaged products |
| DIN EN 868-2 | 03.1993 | Packaging materials for the sterilization of packaged<br>goods. Part 2: sterilization packaging, requirements<br>and tests. |
| DIN EN 980 | 08.1996 | Graphic symbols for marking medical products |
| DIN EN 1441 | 08.1994 | Risk analysis for medical products |
| EN ISO 14971 | 2000 | Risk Management |
| DIN EN 1707 | 01.1997 | 6% Luer cone connections for injection cannula and<br>particular medical equipment |
| DIN EN/ISO 9626 | 06.1995 | Cannula tube of non-rusting steel (SS) for the<br>manufacture of medical products |
| DIN EN ISO 10993-1 | 08.2003 | Biological evaluation of medical products - instructions<br>for selection of tests |
| DIN EN 46001 | 12.1993 | Particular requirements for medical products |
| DIN 17440 | 09.1996 | Stainless Steels |
| BS 4843 | | Single entry IV cannula |
# Conclusion:
The PAJUNK Stimulong Plus Catheter Sets are as safe and effective as the predicate devices when used according to the instructions in the directions for use supplied with the devices.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge of the circle. In the center of the seal is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 1 2004
Pajunk GmbH c/o Mr. Burk A. Brandt President CE Consultancy, Incorporated 5010 NW Crescent Valley Drive Corvallis, Oregon 97330
Re: K043130
Trade/Device Name: Pajunk Stimulong Plus Catheter Sets Regulation Number: 868.5140 Regulation Name: Anesthesia Conduction Kit Regulatory Class: II Product Code: CAZ, BSP Dated: December 3, 2004 Received: December 6, 2004
Dear Mr. Brandt:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becater of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device interstate connifetee prior to ria, 20, 177, 1978, 1978, 1998, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 199 Alliendinens, or to de rises and in tic Act (Act) that do not require approval of a premarket the rederal I vou, Drug, and Ocommay, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls provisions of an annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III If your device is classified (soch additional controls. Existing major regulations affecting (1 MA), it may of subject to back of Federal Regulations, Title 21, Parts 800 to 898. In your device can be found in the weat nouncements concerning your device in the Federal Register.
{5}------------------------------------------------
# Page 2 - Mr. Brandt
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements modifined into may Federal statutes and regulations administered by other Federal agencies. or the Act of all, I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and ifsuing (21 CF read in the quality systems (QS) regulation (21 CFR Part 820); and if requirements as between as product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a premained with a medicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), rr you contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the maj South Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address lttp://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):K043130
Device Name: Pajunk Stimulong Plus Catheter Sets
Indications For Use:
The Pajunk Stimulong Plus Catheter Sets are intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery is accomplished using the conduction catheter. To assist the physician pinpoint the area of application an electrical stimulus can be applied to the conduction needle and after placement of the conduction cathter to its tip via the catheter adapter.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sta Senille
Division Sign-Off
Anes ു. General Hospital. Infection Control. Jer tal Devices
510(k) Number: _ Ko 9313
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.