ORION DIAGNOSTICA UNIQ PINP RIA

K043125 · Orion Diagnostica, OY · CIN · May 4, 2005 · Clinical Chemistry

Device Facts

Record IDK043125
Device NameORION DIAGNOSTICA UNIQ PINP RIA
ApplicantOrion Diagnostica, OY
Product CodeCIN · Clinical Chemistry
Decision DateMay 4, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1050
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K043125 · May 4, 2005ORION DIAGNOSTICA UNIQ PINP RIAOrion Diagnostica, OYRetrospective analysis of serum samples from a Phase III Osteoporosis treatment study; Retrospective analysis of serum samples from a long-term hormone replacement therapy studyRetrospective analysis of patient serum samples from two clinical studies was used to compare the device's performance (PINP) against a predicate device (BAP) in monitoring bone mineral density (BMD) changes in postmenopausal women.Retrospective analysis; Osteoporosis; Hormone replacement therapy; Bone mineral density; Biochemical markers

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Phase III Osteoporosis Treatment Study (Retrospective Subset); Retrospective analysis of a subset of samples from a prospective, double-blind, placebo-controlled study; Follow-up/Duration: 12 monthsPostmenopausal women with diagnosed osteoporosis (at least 5 years since menopause); Sample Size: 124 patients (496 serum samples); Number of Sites: Multi-sitePlacebo control groupPercent change in PINP and BAP compared to BMD at 3, 6, and 12 months
Hormone Replacement Therapy Study (Retrospective Subset); Retrospective analysis of samples from a long-term hormone replacement therapy study; Follow-up/Duration: 12 monthsWhite postmenopausal women (aged 45-65, at least 3 years since last spontaneous bleeding, BMI < 30); Sample Size: 44 patients (132 serum samples)Not applicable for this studyPercent change in PINP and BAP compared to BMD at 6 and 12 months

Indications for Use

Orion Diagnostica UniQ™ PINP RIA is a quantitiative radioimmunoassay designed for the measurement of intact aminoterminal propeptide of type I procollagen, an indicator of osteoblastic activity, in human serum. The test is intended to be used as an aid in the management of postmenopausal osteoporosis. For in vitro diagnostic use.

Device Story

UniQ PINP RIA is an in vitro diagnostic radioimmunoassay (RIA) used to measure intact aminoterminal propeptide of type I procollagen in human serum. It serves as a biomarker for osteoblastic activity. The device is intended for use by laboratory professionals to assist clinicians in managing postmenopausal osteoporosis. The assay quantifies the propeptide concentration, providing data that clinicians use to assess bone turnover and monitor therapeutic response in patients. The device operates within a clinical laboratory setting.

Clinical Evidence

Retrospective analysis of two clinical studies. Study 1 (n=124) compared PINP and BAP response to BMD in postmenopausal women treated with alendronate vs. placebo over 12 months. Study 2 (n=44) evaluated PINP and BAP response to hormone replacement therapy over 12 months. Both studies demonstrated that PINP and BAP show similar decreases in response to treatment, correlating with increases in BMD. Analytical performance included precision (intra-assay CV 6.5-19.8%, inter-assay CV 6.0-10.1%), linearity (recovery 96-112%), and detection limit (2 µg/L).

Technological Characteristics

Radioimmunoassay (RIA) for the quantitative measurement of intact aminoterminal propeptide of type I procollagen in human serum. Classified under 21 CFR 862.1050 (Alkaline phosphatase or isoenzymes test system).

Indications for Use

Indicated for the quantitative measurement of intact aminoterminal propeptide of type I procollagen (PINP) in human serum to aid in the management of postmenopausal osteoporosis.

Regulatory Classification

Identification

An alkaline phosphatase or isoenzymes test system is a device intended to measure alkaline phosphatase or its isoenzymes (a group of enzymes with similar biological activity) in serum or plasma. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

Submission Summary (Full Text)

{1} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k043125 B. Purpose for Submission: New device C. Measurand: Aminoterminal propeptide of Type 1 procollagen D. Type of Test: Radioimmunoassay E. Applicant: Orion Diagnostica F. Proprietary and Established Names: Orion Diagnostica UniQ™ PINP RIA G. Regulatory Information: 1. Regulation section: 21 CFR 862.1050 2. Classification: Class II 3. Product code: CIN 4. Panel: 75 H. Intended Use: 1. Intended use(s): See Indications for use section below. 2. Indication(s) for use: Orion Diagnostica UniQ PINP RIA is a quantitative radioimmunoassay designed for the measurement of intact aminoterminal propeptide of type 1 procollagen, an indicator of osteoblastic activity, in human serum. The test is intended to be used as an aid in the management of postmenopausal osteoporosis. For in vitro diagnostic use. {2} 3. Special conditions for use statement(s): For Prescription Use Only 4. Special instrument requirements: N/A I. Device Description: The UniQ PINP RIA kit is a dual-antibody competitive radioimmunoassay that uses 125¹ labeled PINP, PINP RIA rabbit antiserum, seven calibrators, two lyophilized controls and a procollagen separation reagent. The calibrators and controls are provided in the UniQ PINP RIA kit and are not available separately. The controls used in the PINP calibrators and controls were tested and found negative for hepatitis B surface antigen, for anti-HIV 1 and 2 antibody and anti-HCV antibodies. J. Substantial Equivalence Information: 1. Predicate device name(s): Tandem-R Ostase Immunoradiometric Assay 2. Predicate 510(k) number(s): k961573 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Indications for Use | Orion Diagnostica UniQ PINP RIA is a quantitative radioimmunoassay designed for the measurement of intact aminoterminal propeptide of type 1 procollagen, an indicator of osteoblastic activity, in human serum. The test is intended to be used as an aid in the management of postmenopausal osteoporosis. For in vitro diagnostic use. | The Tandem-R Ostase immunoradiometric assay is an invitro device indicated for the quantitative measurement of skeletal alkaline phosphatase (sALP), an indicator of osteoblastic activity, in human serum. This device is intended to be used as a aid in the management of postmenopausal osteoporosis and Paget's disease. | | Matrix | Serum | Serum | {3} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Analyte | Intact Aminoterminal Propetide of type 1 procollagen | Skeletal Alkaline Phosphatase | # K. Standard/Guidance Document Referenced (if applicable): ECCLS Document Vol 3, No. 3: Guidelines for a User Laboratory, Evaluate and Select a Kit for its own use. NCCLS EP5-A: Evaluation of Precision Performance of Clinical Chemistry devices. NCCLS EP10-A: Preliminary Evaluation of Quantitative Clinical Laboratory Methods. NCCLS EP7-P: Interference Testing in Clinical Chemistry. NCCLS EP6-P: Evaluation of the Linearity of Quantitative Analytical Methods. # L. Test Principle: The UniQ PINP RIA (PINP) is based on the competitive radioimmunoassay technique. A known amount of labeled PINP and an unknown amount of unlabelled PINP in the sample compete for a limited number of high affinity binding sites of the polyclonal rabbit anti-PINP antibody. A second antibody, directed against rabbit IgG and coated to kaolin particles, is used to separate the antibody-bound PINP form free PINP. The radioactivity of the bound tracer antigen is measured using a gamma counter. The amount of labeled PINP in the sample tube is inversely proportional to the amount of PINP in the sampler. The concentrations in unknown samples are obtained from a calibration curve, which is based on the concurrent testing of the UNIQ PINP RIA calibrators that range between 1 and $250~\mu \mathrm{g / L}$ PINP. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: Precision was assessed and calculated according to NCCLS EP5-A. An intra-assay study was conducted by measuring 20 replicates of two sample pools and one control. An inter-assay study was conducted by measuring two sample pools and one control in duplicates with 10 separate assays. The results are summarized in the chart below. | Intra-assay precision of 20 replicates | | | Inter-assay precision of 10 duplicate determinations | | | | --- | --- | --- | --- | --- | --- | | Sample | Mean (μg/L) | CV (%) | Sample | Mean (μg/L) | CV (%) | | 1 | 5 | 19.8 | 6 | 6 | 10.1 | | 2 | 12 | 9.8 | 7 | 12 | 9.8 | | 3 | 32 | 6.5 | 8 | 32 | 8.3 | | 4 | 51 | 6.5 | 9 | 52 | 7.0 | | 5 | 173 | 10.2 | 10 | 167 | 6.0 | {4} A recovery study was conducted by adding known amounts of purified PINP antigen to three sample pools and measured with the UniQ PINP RIA assay. The samples were measured in duplicate and the recovery % was calculated. The recovery % equals the increase in concentration divided by the spike concentration multiplied by 100. Sample 1 (initial concentration of 16.5 µg/L) | Spike (μg/L) | Measured PINP (μg/L) | Increase (μg/L) | Recovery % | | --- | --- | --- | --- | | --- | 16.5 | | | | 38.8 | 50.0 | 33.4 | 86 | | 63.9 | 84.0 | 67.5 | 106 | | 95.1 | 106.4 | 89.9 | 95 | | Mean % recovery | | | 95 | Sample 2 (Initial Concentration of 24.5 µg/L) | Spike (μg/L) | Measured PINP (μg/L) | Increase (μg/L) | Recovery % | | --- | --- | --- | --- | | --- | 24.5 | | | | 38.8 | 62.1 | 37.6 | 97 | | 63.9 | 82.9 | 58.4 | 91 | | 95.1 | 103.2 | 78.7 | 83 | | Mean % recovery | | | 90 | Sample 3 (Initial Concentration of 35.0 µg/L) | Spike (μg/L) | Measured PINP (μg/L) | Increase (μg/L) | Recovery % | | --- | --- | --- | --- | | --- | 35.0 | | | | 38.8 | 67.1 | 32.2 | 83 | | 63.9 | 88.4 | 53.4 | 84 | | 95.1 | 125.5 | 90.5 | 95 | | Mean % recovery | | | 87 | A dilution study was conducted to compare the two possible sample diluents, the zero calibrator or 0.9% NaCl on two controls and one sample pool. The diluted and undiluted samples were measured with the UniQ PINP RIA assay in duplicates and % recovery was calculated from the diluted samples against the undiluted samples. The results supported that samples can be diluted with either the zero calibrator or 0.9% NaCl. b. Linearity/assay reportable range: Linearity studies were conducted according to NCCLS EP6-P. Linearity studies were conducted on five samples (mini-pools) and two controls. The concentrations of the diluted samples were calculated based on the dilution factors and compared to the result obtained directly from the undiluted sample. The samples were diluted with the 0 µg/L calibrator. {5} The percent recovery is calculated for each dilution using the equation: $$ \% \text{Recovery} = \frac{\text{Calculated for undiluted}}{\text{Measured concentration from undiluted sample}} \times 100 $$ The calculated for undiluted was obtained using the following equation: $$ \text{Calculated for undiluted} = \frac{\text{Measured concentration from diluted sample}}{\text{Dilution (in \% volume of the original sample)}} \times 100 $$ The results are listed in the table below. | Sample | Mean Concentration (μg/L) | Mean % Recovery | SD | %CV | | --- | --- | --- | --- | --- | | 1 | 74.7 | 96 | 5.5 | 7.4 | | 2 | 137.2 | 110 | 4.3 | 3.1 | | 3 | 156.5 | 102 | 8.3 | 5.3 | | 4 | 163.9 | 111 | 9.5 | 5.8 | | 5 | 172.9 | 106 | 6.5 | 3.8 | | 6 | 187.9 | 112 | 11.0 | 5.8 | | 7 | 240.9 | 101 | 20.9 | 8.7 | c. Traceability, Stability, Expected values (controls, calibrators, or methods): PINP antigen (that is used in controls and calibrators) has been purified from human ascites fluids and characterized by electrophoresis for purity. The value assignment of the first purified antigen stock was done by quantitative analysis of amino acids, after acid hydrolysis of the polypeptide. New preparations of the antigen preparations are conducted by testing the dilutions points are within the calibrator curve of PINP-RIA. The values were calculated as the mean of at least two dilution points. The antigen preparation was found to be negative for hepatitis B surface antigen, for anti-HIV 1 and 2 antibodies and anti-HCV antibodies. The in-house master calibrators (working calibrators) were prepared gravimetrically using the first calibrator antigen batch to obtain the 6 calibrator concentrations and are stored at $-70^{\circ}\mathrm{C}$. PINP working calibrators are in-house master calibrators used to calibrator each lot of manufacturer's products calibrator set, which is produced by diluting a purified PINP-preparation with a high PINP concentration in PINP calibrator buffer (calibrator 0). Product calibrators are stored at $-20^{\circ}\mathrm{C}$. Each new PINP product calibrator batch is tested against PINP working calibrators and the previous batch of PINP product calibrators in at least 10 acceptable tests. With each set of the calibrators, the calibrators curve is created and other calibrators are measured as samples. The acceptance criteria are that the deviation of the concentrations between the new product calibrators and {6} working calibrators is equal or smaller than 8% and that deviation of the cpm-values between the new product calibrators is equal or smaller than 3%. The validity of the results of each reagent lot is tested using controls (hi and low) and in-house patient sample pool and the measured in four replicates. The controls are lyophilized and stored at 2-8 °C and preserved with sodium azide. The in-house sample pool is stored at -20 °C. The target values for new batches of PINP controls are determined by performing at least twenty tests on both the new and present batch of controls in duplicates. The target values are determined by calculating the means of the results. The stability protocol for the calibrators and controls includes four test points. First test is performed at 0 time point, two tests between the 0 point and the expiry date (e.g., 8 and 16 months) and the last test is performed one month after the expiry date (25 months). Additionally, the reconstituted controls are tested after 7 weeks of storage at +2-+8 °C at each testing point. Bioburden of the calibrators and controls as well as the residual moisture of the lyophilized controls was determined. The UniQ PINP RIA controls are stable for 24 months in lyophilized state and for 6 weeks after reconstitution (+2 to +8 °C). The UniQ PINP RIA calibrators are stable for 24 months. d. Detection limit: The detection limit of the UniQ PINP RIA assay was assessed according to the ECCLS Document Vol 3. The detection limit was determined by measuring 20 replicates of the 0 calibrator and the calibration curve four times. The mean of the 4 results was used to establish the detection limit. The minimum detectable concentration of PINP in the assay was approximately 2 µg/l, defined as twice the standard deviation of the 0-binding value. The results are listed in the table below. | Calibrator | Test 1 | Test 2 | Test 3 | Test 4 | | --- | --- | --- | --- | --- | | Mean (AU) | 17879 | 18384 | 17657 | 17647 | | SD | 199 | 204 | 303 | 384 | | CV% | 1.1 | 2.1 | 1.9 | 2.8 | | N= | 20 | 20 | 20 | 20 | | Mean CPM-2xSD | 17480 | 17977 | 17052 | 16879 | | Sensitivity | 1.2 | 2.1 | 1.9 | 2.8 | | Mean of 4 test | 2.0 µg/L | | | | e. Analytical specificity: The effects of hemoglobin, bilirubin and triglycerides were studied. The tested components were added to three sample pools with suitable solvents and samples were assayed in duplicates with the UniQ PINP RIA. The results for all three samples are in the charts below and it was determined that hemoglobin, bilirubin and triglycerides don't interfere with the tested concentrations. {7} | Hemoglobin Effect | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | PINP (μg/L) n=2 | | | % PINP | | | | Added | Neat | +125 mg/dL | +250 mg/dL | +500 mg/dL | +125 mg/dL | +250 mg/dL | +500 mg/dL | | Sample 1 | 40.4 | 36.8 | 36.5 | 37.1 | 91 | 90 | 92 | | Sample 2 | 42.9 | 40.0 | 46.1 | 42.2 | 93 | 107 | 98 | | Sample 3 | 50.6 | 48.5 | 45.4 | 44.7 | 96 | 90 | 88 | | Average % Recovery | | | | | 93 | 96 | 93 | | Bilirubin Effect | | | | | | | | | | | PINP (μg/L) n=2 | | | % PINP | | | | Added | Neat | +5.8 mg/dL | +11.7 mg/dL | +23.4 mg/dL | +5.8 mg/dL | +11.7 mg/dL | +23.4 mg/dL | | Sample 1 | 38.3 | 36.6 | 37.7 | 38.8 | 95 | 98 | 101 | | Sample 2 | 38.8 | 37.7 | 37.4 | 37.8 | 97 | 96 | 97 | | Sample 3 | 45.6 | 46.5 | 45.2 | 46.3 | 102 | 99 | 102 | | Average % Recovery | | | | | 102 | 99 | 102 | | Triglyceride Effect | | | | | | | | | | | PINP (μg/L) n=2 | | | % PINP | | | | Added | Neat | +1000 mg/dL | +2000 mg/dL | +3000 mg/dL | +1000 mg/dL | +2000 mg/dL | +3000 mg/dL | | Sample 1 | 34.1 | 33.1 | 35.2 | 31.2 | 97 | 103 | 92 | | Sample 2 | 36.6 | 37.4 | 36.7 | 31.7 | 102 | 100 | 87 | | Sample 3 | 39.6 | 39.3 | 42.9 | 40.7 | 99 | 108 | 103 | | Average % Recovery | | | | | 99 | 108 | 103 | A cross-reactivity study was conducted of the PINP antiserum towards PIIINP (type III collagen N-terminal propeptide). PIIINP was added to the normal PINP RIA assay reaction to investigate whether if any cross reactivity would occur and would be seen as a reduction in CPM values. The inhibition curve of the PIIINP antigen was compared to PINP inhibition curve. The results showed no cross-reactivity to PIIINP. A second cross-reactivity study was conducted on the PINP antiserum towards N-terminal Col1-fragment of type I procollagen (Col1), which is a breakdown product of PINP. Col1 was added to the normal PINP RIA assay reaction to investigate any occurring cross reactivity that could be seen as a reduction in CPM values. The inhibition curve of Col1 was compared to the inhibition curve of PINP. The results showed that the cross-reactivity towards Col1 is approximately $1.2\%$ . f. Assay cut-off: The measurement range of the assay is $5 - 250~\mu \mathrm{g / L}$ 2. Comparison studies: a. Method comparison with predicate device: {8} Comparison was conducted with the Beckman Coulter Tandem-R Ostase Immunoradiometric assay (BAP). Since the two assays analyze two different analytes, the predicate and the device are able to be compared through the percent change of bone mineral density (BMD). In order to enable the comparability of the BAP and PINP assays, percent change data are calculated at each time point (3,6 and 12 months) from each individual baseline value. The sponsor states that for comparison purposes, a decrease equal to or greater than 25% is used as a cut-off between biochemical responders and non-responders. Retrospective analysis of samples from two studies were conducted by the sponsor. Study #1 was a Phase III Osteoporosis treatment study that used biphosphonate alendronate to treat postmenopausal women with diagnosed osteoporosis. Study #2 analyzed a long term study of hormone replacement therapy in the prevention of postmenopausal bone loss in healthy women. Both studies compare menopausal status, age, lumbar spine bone mineral density (BMD), PINP and BAP measurements. Normality of the data is studied by histogram analysis and by using Shapiro-WILK test. Differences between the means are compared with paired samples t-test (between different time points) or with independent samples t-test (between study groups). For receiver operator characteristics (ROC) analysis, cut-off in BMD for biochemical responder was set to greater than or less than 2% increase in the result. Similarly, for PINP and BAP the cut-off for biochemical responder was set to greater than or less than 25% decrease. Study #1 retrospectively analyzed a subset of samples from a placebo controlled Phase III Osteoporosis Treatment study. This study was a multi-site, prospective, double blind placebo controlled study in postmenopausal women of the US population with defined osteoporosis, at least 5 years since menopause. A subset of sera has been used to compare BAP to BMD at several sites. Based on a subset of the same serum samples, the sponsor study now compares the response of PINP and BAP to BMD at selected time points within and across the treatment group (TRM) and placebo control group (CTR). Percent change from 0 value at 3, 6, and 12 month time points were calculated and used for analyses. The sponsor stated that the sample size required confirming equivalence of BAP and PINP changes at confidence level of 0.05 and power of 0.80 is a minimum of 36 samples at each time point. 62 patients (248 serum samples) from TRM group and 62 patients (248 serum samples) from CTR group meet this requirement and gave a total of 124 patients and 496 serum samples. Results of this study are shown the chart below for BAP, PINP and BMD in the form of decreases or increases from the baseline values at time 3, 6 and 12 months. The result showed a similar decrease for BAP and PINP while the BMD values show an increase. The results are listed in the chart and figure below. {9} | BAP | 3 Months | 6 Months | 12 Months | | --- | --- | --- | --- | | TRM Group | ↓ 31.8 +/- 29.7% | ↓ 45.9 +/-16.2% | ↓ 43.4 +/- 19.9% | | CTR Group | ↓8.0 +/- 15.9% | ↓ 10.5 +/- 17.0% | ↓ 5.5 +/-20.8% | | PINP | 3 Months | 6 Months | 12 Months | | TRM Group | ↓ 55.4 +/- 34.0% | ↓ 71.6 +/- 15.6% | ↓ 70.8 +/- 15.5% | | CTR Group | ↓ 8.6 +/- 25.9% | ↓ 14.0 +/- 25.5% | ↓ 9.8 +/- 27.4% | | BMD | 3 Months | 6 Months | 12 Months | | TRM Group | ↑ 3.5 +/- 3.2% | ↑ 5.1 +/- 4.4% | ↑ 6.0 +/- 3.3% | | CTR Group | ↑ 0.3 +/- 3.1% | ↑ 1.2 +/- 3.3% | ↑ 0.1 +/- 3.2% | ![img-0.jpeg](img-0.jpeg) ![img-1.jpeg](img-1.jpeg) Response as mean percent change (± SE bars) in (A) BMD and (B) PINP concentration in patients receiving bisphosphonate (●) or placebo (○). Study #2 retrospectively analyzed samples from a postmenopausal osteoporosis hormone replacement therapy. The study analyzed the response of bone to HRT and was followed throughout the study by measuring BMD in the white postmenopausal women from the Finnish population with at least 3 years since the last spontaneous bleeding, aged 45 to 65 years (average age 55.9 years) and with a body mass index less than 30. Retrospectively, PINP and BAP as an aid in monitoring HRT were compared by the sponsor. Based on a subset of the same serum samples, the sponsor study now compares the response of PINP and BAP to BMD at selected time points within and across the treatment group (TRM). Percent change from the baseline (value 0) at 6 and 12 month time points were calculated and used for analyses. The sponsor stated that the sample size required confirming equivalence of BAP and PINP changes at confidence level of 0.05 and power of 0.80 is a minimum of 36 samples at each time point. BAP and PINP were analyzed blind by trained laboratory personnel at Orion Diagnostica by using the Tandem-R Ostase IRMA and the UniQ PINP RIA kits. Only patient samples that obtained the {10} lowest dose of the estradiol valerate (the treatment used in the study) were used, leaving a total of 44 patients and 132 serum samples. Since this was a dose finding study, placebo controls were not used. Results of this study are shown the chart below for BAP, PINP and BMD in the form of decreases or increases from the baseline values at times 6 and 12 months. The result showed a similar decrease for BAP and PINP while the BMD values show an increase. The results are listed in the chart and figure below. | BAP | 6 Months | 12 Months | | --- | --- | --- | | TRM Group | ↓ 11.5 +/- 29.4% | ↓ 19.1 +/- 29.0% | | PINP | 6 Months | 12 Months | | TRM Group | ↓ 12.3 +/- 39.6% | ↓ 19.8 +/- 46.5% | | BMD | 6 Months | 12 Months | | TRM Group | ↑ 1.4 +/- 2.3% | ↑ 3.1 +/- 3.8% | ![img-2.jpeg](img-2.jpeg) Response as mean percent change (± SE bars) in BMD (○) and PINP concentration (●) in postmenopausal women receiving low dose combined hormone replacement therapy. b. Matrix comparison: N/A. This assay is only used with serum. 3. Clinical studies: N/A 4. Clinical cut-off: N/A {11} 5. Expected values/Reference range: In order to determine the reference range, serum samples from healthy adults we collected and measured with the UniQ PINP RIA. The mean PINP concentrations for Finnish males, females and postmenopausal females' reference range, .95 CI for upper and lower limits are shown in the following table. | Reference Limits | Male | Female | Female (postmenopausal for 3 years)** | | --- | --- | --- | --- | | Number of Subjects | 323 | 210 | 383 | | Age Distribution | 22-65 | 25-44 | 52-62 | | Mean | 45 μg/L | 42 μg/L | 48 μg/L | | Reference Interval* | 22-87 μg/L | 19-83 μg/L | 16-96 μg/L | | .95Confidence Interval | | | | | Lower reference Limit | 20 - 25 μg/L | 14 - 23 μg/L | 15 - 19 μg/L | | Upper reference Limit | 77 - 94 μg/L | 71 - 96 μg/L | 87 - 103 μg/L | * Reference interval = 0.025 and 0.975 fractiles ** Concentrations are elevated compared to younger women (P=0.0005). N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...