ADVANCE HA COATED SPIKED TIBIAL BASE AND ADVANCE HA COATED MODULAR KEEL
Applicant
Wrightmedicaltechnologyinc
Product Code
MBH · Orthopedic
Decision Date
Apr 18, 2005
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 888.3565
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ADVANCE® HA Coated Components are indicated for use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2) inflammatory degenerative joint disease including rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and 5) treatment of fractures that are unmanageable using other techniques. The ADVANCE® HA Coated Components are for use without bone cement and are single use devices.
Device Story
ADVANCE® HA Coated Spiked Tibial Base and Modular Keel are orthopedic implants for knee arthroplasty. Components are manufactured from titanium alloy with titanium sintered bead porous coating and hydroxyapatite (HA) coating. Tibial base features 4 distal spikes; modular keel features 2 anterior and 2 posterior fins. Devices are intended for cementless fixation. Used by orthopedic surgeons in clinical settings to replace damaged joint surfaces, reduce pain, and restore knee function. Output is the physical implantation of the device to provide structural support and joint stability.
Clinical Evidence
No clinical data provided; substantial equivalence supported by bench testing and materials data.
Technological Characteristics
Materials: Titanium alloy, titanium sintered bead porous coating, hydroxyapatite (HA) coating. Design: Spiked tibial base (4 spikes), modular keel (2 anterior/2 posterior fins). Fixation: Cementless. Sterilization: Not specified. Energy source: None (mechanical implant).
Indications for Use
Indicated for skeletally mature patients undergoing knee arthroplasty for noninflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity correction, revision procedures, or fracture treatment unmanageable by other techniques.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text 'K043083 (pg 1 of)'. The text appears to be handwritten in black ink on a white background. The text likely represents a document identifier or page number.
Image /page/0/Picture/1 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a stylized graphic consisting of three parallel lines that are angled downwards and slightly overlap each other, creating a sense of depth. The letters are large and prominent, and the graphic is positioned to complement the text.
APR 1 8 2005
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe in accordance with the 1 000 and in conformance with 21 CRF 807, this information serves as a Moural Dovies Ave of 1998 and is a for the ADVANCE® HA Coated Spiked Tibial Base and the ADVANCE® HA Coated Modular Keel.
| Submitted By: | Wright Medical Technology, Inc. |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | November 5, 2004 |
| Contact Person: | Ehab M. Esmail<br>Sr. Manager, Regulatory Affairs |
| Proprietary Name: | ADVANCE® HA Coated Spiked Tibial Base and<br>the ADVANCE® HA Coated Modular Keel |
| Common Name: | HA COATED COMPONENTS |
| Classification Name and Reference: | 21 CFR 888.3565 Prosthesis, Knee,<br>Patello/femorotibial, Semi-Constrained,<br>Uncemented, Porous Coated,<br>Polymer/Metal/Polymer - Class II |
| Device Product Code and Panel Code: | Orthopedics/87/ MBH |
#### DEVICE INFORMATION
#### A. INTENDED USE
The ADVANCE® HA Coated Components are indicated for use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions:
- 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 2) inflammatory degenerative joint disease including rheumatoid arthritis;
- 3) correction of functional deformity;
- 4) revision procedures where other treatments or devices have failed; and
- 5) treatment of fractures that are unmanageable using other techniques.
#### headquarters Wright Medical Technology, Inc.
Arlington, TN 38002 901.867.9971 phone 5677 Airline Road
ww.wmt.com
international subsidiaries 011.32.2.378.3905 Belgium 011.39.0250.678.227 Italy
05.826.1600 Canada 011.81.3.3538.0474 Japan 011.33.1.45.13.24.40 France
011.44.1483.721.404 UK
011.49.4161.745130 Germany
{1}------------------------------------------------
K043083 (pg 2 of 2)
The ADVANCE® HA Coated Components are for use without bone cement and are single use devices.
## B. DEVICE DESCRIPTION
The ADVANCE® HA Coated Components are an HA Coated Spiked Tibial Base and an The ADVANCES In Feel. The design features of the components are described below.
ADVANCE® HA Coated Spiked Tibial Base
- 4 spikes on the distal surface .
- Manufactured from titanium alloy .
- Titanium sintered bead porous coated .
- HA coated ●
### ADVANCE® HA Coated Modular Keel
- 2 small anterior fins .
- 2 posterior fins .
- Manufactured from titanium alloy .
- HA coated .
The design features of ADVANCE® HA Coated Components are substantially equivalent to the design features of other devices previously cleared for market.
# C. SUBSTANTIAL EQUIVALENCE INFORMATION
The intended use, material, and design features of ADVANCE® HA Coated Components are substantially equivalent to the predicate devices previously cleared for market. The safety and effectiveness of ADVANCE® HA Coated Components are adequately supported by the substantial equivalence information, materials data, and testing results provided within the Premarket Notification.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/12 description: The image shows the logo for the Department of Health and Human Services, USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or bird with three stylized lines forming its body and wings.
APR 1 8 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Ehab M. Esmail Sr. Manager, Regulatory Affairs Wright Medical Technology, Inc. 5677 Airline Road Arlington, Tennessee 38002
Re: K043083
K043083
Trade/Device Name: ADVANCE® HA Coated Spiked Tibial Base and the ADVANCE® HA Coated Modular Keel Regulation Number: 21 CFR 888.3565 Regulation Number: 21 CFK 866.5503
Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prothesis Regulatory Class: II Product Code: MBH Dated: March 25, 2005 Received: March 28, 2005
Dear Mr. Esmail:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Security (10). "I substantially equivalent (for the indications
referenced above and have determined the device is substantial in interstate for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosule) to regary manced prought in the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior in May 28, 1970, the chance with the provisions of the Federal Food. Drug, devices that have been reciassified in accordance while approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act . The and Cosmetic Act (Act) that to hot require approvide on trols provisions of the Act. The You may, merelore, market the devree, becjocher to the one in annual registration, listing of
general controls provisions of the Act include requirements for any province and general controls provisions of the Feet neralies requires against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (sec above) movement major regulations affecting your device can
may be subject to such additional controls. Existing major regulations EDA may may be subject to such additional controller Entrological of 898. In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to Register be found in the Code of Pederal Registering your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualies of a subscription with other requirements of the Act
that FDA has made a determination that your device complies . You must that FDA nas made a delerimiation mac your as roos be species. You must of any list in or any Federal statutes and regulations daministered of registration and listing (21 l
eomply with all the Act's requirements, including, but not limited to set comply with all the Act S requirements, mercessed in the practice requirements as set CFR Part 807); labeling (21 CFR Party, good and 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Path, 21 CFR 1000 1050 forth in the quality systems (QD) regalantin (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000
{3}------------------------------------------------
Page 2 – Mr. Ehab M. Esmail
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin marketing your antial equivalence of your device to a legally
premarket notification. The FDA finding of substantial equivalias and thus, premarket notification. The FDA Inding of substanced by July 19, 2017 11:55
marketed predicate device results in a classification for your device and thus, permits your devic to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 81), please an If you desire specific advice for your active of the can and the regulation entitled, the regulation entitled, the international contact the Office of Compliance at (2 to 21 to a market notification the Division of Small "Misbranding by relerence to premarkst nothleadon (1) be Act from the Division of Small
other general information on your responsibilities under the Act from the (800) 638, 2 other general information on your responsion.com in toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-free warming on html Manufacturers, International and Consumer Fissionalite aroved childring stry/support/index.html.
Sincerely yours,
Stupt Plurde
S.E. Ph.D.
Airiam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K043083
#### Indications for Use
510(k) Number (if known):
Device Name: ADVANCE® HA Coated Spiked Tibial Base and ADVANCE® HA Coated Modular Keel
Indications For Use:
The ADVANCE® HA Coated Components are indicated for use in knee arthroplasty for The ADVANCE® HA Coated Components are mercion in skeletally mature patients with the following conditions:
- 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
-
- 3) correction of functional deformity;
- 3) correction of functional deforming,
4) revision procedures where other treatments or devices have failed; and
- 4) Tevision procedures where of the read of the rechniques.
The ADVANCE® HA Coated Components are for use without bone cement and are single use devices.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
typt Purbe
(Division Sign-Off)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Division of General, Restorative, and Neurological Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.