K043073 · Wrightmedicaltechnologyinc · KWA · May 4, 2005 · Orthopedic
Device Facts
Record ID
K043073
Device Name
PROCOTYL-E ACETABULAR SYSTEM
Applicant
Wrightmedicaltechnologyinc
Product Code
KWA · Orthopedic
Decision Date
May 4, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3330
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The PROCOTYL-E Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2. inflammatory degenerative joint disease such as rheumatoid arthritis; 3. correction of functional deformity; 4. revision procedures where other treatments or devices have failed; and, 5. treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. The PROCOTYL-E Acetabular System consists of single use components that are intended to accommodate for bone loss. This system is to be used in conjunction with associated WMT polyethylene and metal Acetabular liners articulating with associated WMT metal and ceramic femoral heads as part of an uncemented total hip arthroplasty.
Device Story
PROCOTYL-E Acetabular System; metal acetabular shells, plates, and hooks; designed for uncemented total hip arthroplasty. Features oblong dual hemispherical outer geometry with hemispherical offset to accommodate bone loss; includes eccentric design for screw fixation. Used in conjunction with WMT polyethylene/metal liners and WMT metal/ceramic femoral heads. Implanted by orthopedic surgeons in clinical settings to restore hip function and reduce pain in patients with degenerative joint disease or fractures. Provides stable anchorage via titanium alloy plates and hooks.
Clinical Evidence
No clinical data provided; substantial equivalence supported by materials data and bench testing results.
Technological Characteristics
Components: metal acetabular shells, plates, and hooks. Material: titanium alloy. Design: oblong dual hemispherical outer geometry with hemispherical offset. Fixation: uncemented, porous-coated. Intended for use with WMT polyethylene/metal liners and WMT metal/ceramic femoral heads.
Indications for Use
Indicated for skeletally mature patients requiring total hip arthroplasty for pain relief or improved function due to non-inflammatory/inflammatory degenerative joint disease (e.g., osteoarthritis, rheumatoid arthritis, avascular necrosis, ankylosis, protrusio acetabuli, hip dysplasia), functional deformity, revision of failed devices, or proximal femur fractures (nonunion, femoral neck, trochanteric) unmanageable by other techniques.
Regulatory Classification
Identification
A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).
Submission Summary (Full Text)
{0}------------------------------------------------
K043073
Image /page/0/Picture/1 description: The image shows the word "WRIGHT." in large, bold, black letters. Below the word are three stylized, textured, arrow-like shapes pointing to the left. The arrows are arranged in a descending order, with the top arrow being the largest and the bottom arrow being the smallest. The letters "IV" are visible below the arrows.
MAY - 4 2005
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of the PROCOTYLE-E.
Submitted By:
Date:
Contact Person:
Proprietary Name:
Common Name:
Classification Name and Reference:
Wright Medical Technology, Inc.
November 5, 2004
Ehab M. Esmail Senior Manager, Regulatory Affairs
#### PROCOTYLE-E
Acetabular Shell
21 CFR 888.3358 Hip joint metal/polymer/metal semi- constrained porous - coated uncemented prosthesis -- Class II
21 CFR 888.3353 Prosthesis, Hip, Semi-Constrained, metal/ ceramic/ polymer, cemented or non-porous uncemented- Class II
21 CFR 888.3330 Hip joint metal/ metal semiconstrained, with uncemented acetabular component prosthesis -- Class III
Device Product Code and Panel Code:
Orthopedics/87/LPH, KWA, LZO
### DEVICE INFORMATION
### A. INTENDED USE
The PROCOTYL-E Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- inflammatory degenerative joint disease such as rheumatoid arthritis; 2.
- correction of functional deformity; 3.
- 4. revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur 5. with head involvement that are unmanageable using other techniques.
Arlington, TN 38002 011.49.4161.745130 Germany
{1}------------------------------------------------
K043073 2/2
The PROCOTYL-E Acetabular System consists of single use components that are intended to accommodate for bone loss. This system is to be used in conjunction with associated WMT polyethylenc and metal Acetabular liners articulating with associated WMT metal and ceramic femoral heads as part of an uncemented total hip arthroplasty.
### B. DEVICE DESCRIPTION
The PROCOTYL-E Acetabular System consists of metal acetabular shells, plates and hooks.
The added design features of the PROCOTYL-E Acetabular Shell are summarized below:
- The PROCOTYL-E Acetabular shell feature an oblong dual hemispherical outer . geometry with a hemispherical offset to accommodate for bone loss.
- Eccentricity and offset are increased to accommodate for bone loss and screw fixation .
- The PROCOTYL-E shell allows use of a wide range of plates, hooks, all of which are . manufactured from titanium alloy, available in various types and sizes, thus guaranteeing a stable and secure anchorage of the shell to enhanced fixation and stability.
### C. SUBSTANTIAL EQUIVALENCE INFORMATION
The intended use, material, type of interface, and design features of the PROCOTYL-E Acetabular Shell are substantially equivalent to the competitive devices previously cleared for market. The safety and effectiveness of the PROCOTYL-E Acetabular Shell are adequately supported by the substantial equivalence information, materials data, and testing results provided within the Premarket Notification.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 4 2005
Mr. Ehab M. Esmail Senior Manager, Regulatory Affairs Wright Medical Technology, Inc. 5677 Airline Road Arlington, Tennessee 38002
Re: K043073
Trade/Device Name: PROCOTYL-E Acetabular System Regulation Numbers: 21 CFR 888.3330, 888.3353, 888.3358 Regulation Names: Hip joint metal/metal semi-constrained, with uncemented acetabular component prosthesis, Hip joint metal/ceramic/polymer semiconstrained cemented or nonporous uncemented prosthesis, Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: III Product Codes: KWA, LZO, LPH Dated: March 25, 2005 Received: March 28, 2005
Dear Mr. Esmail:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2 – Mr. Ehab M. Esmail
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I Dr is issuaires or our device complies with other requirements of the Act that FDA has made a dolorimiation administered by other Federal agencies. You must of ally redital statutes and regarations and limited to: registration and listing (21 Comply with an the rive orequentions)01); good manufacturing practice requirements as set CFK Fart 807), labeling (21 OF R Part 820); and if applicable, the electronic form in the quality bytellio (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin manesing your device of your device to a legally premits that notication. The President for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific aarroliance at (240) 276-0120. Also, please note the regulation entitled, Contact the Office of Company of Company of CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sincerery yours,
Stpt Rlurde
riam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a graphic of three stylized, overlapping shapes that resemble wings or chevrons. The "TM" symbol is located at the bottom right of the graphic.
### PROCOTYL-E Acetabular System INDICATIONS STATEMENT
1043073
510(k) Number (if known):
#### Device Name: PROCOTYL-E Acetabular System
Indications For Use:
The PROCOTYL-E Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular 1. necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- inflammatory degenerative joint disease such as rheumatoid arthritis; 2.
- correction of functional deformity; 3.
- revision procedures where other treatments or devices have failed; and, 4-
- treatment of fractures that are unmanageable using other techniques. న్.
The PROCOTYL-E Acetabular System consists of single use components that are intended to accommodate for bone loss. This system is to be used in conjunction with associated WMT polyethylene and metal Acctabular liners articulating with associated WMT metal and ceramic femoral heads as part of an uncemented total hip arthroplasty
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrent (Division Sign-Off)vice Evaluation (ODE) | | | |
|-----------------------------------------------------|--|--|--|
| Division of General, Restorative, | | | |
| and Neurological Devices | | | |
| headquarters<br>Wright Medical Technology, Inc. 5677 Airline Road | 510(k) Number | K043073 |
|-------------------------------------------------------------------|--------------------|-------------|
| Arlington, TN 38002 | 901.867.9971 phone | www.wmt.com |
| international subsidiaries | | | |
|----------------------------|--------------------------|-----------------------------|----------------------------|
| 011.32.2.378.3905 Belgium | 905.826.1600 Canada | 011.33.1.45.13.24.40 France | 011.49.4161.745130 Germany |
| 011.39.0250.678.227 Italy | 011.81.3.3538.0474 Japan | 011.44.1483.721.404 UK | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.