← Product Code [LKD](/productcode/LKD) · K043003

# TCM40 MONITORING SYSTEM (K043003)

_Radiometer Medical Aps · LKD · Mar 1, 2005 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K043003

## Device Facts

- **Applicant:** Radiometer Medical Aps
- **Product Code:** [LKD](/productcode/LKD.md)
- **Decision Date:** Mar 1, 2005
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 868.2480
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Pediatric

## Indications for Use

The TCM4 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures. It is indicated for use on neonates, pediatrics, and adults not under gas anesthesia. The TCM40 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures as well as of oxygen saturation of arterial hemoglobin and pulse rate. It is indicated for use on neonates, pediatrics, and adults not under gas anesthesia.

## Device Story

TCM4/40 Monitoring Systems consist of a monitor unit, combined oxygen/carbon dioxide interface module, electrode assembly, and (for TCM40) an oxygen saturation/pulse module. Device inputs include transcutaneous gas signals and pulse oximetry data. The system processes these inputs to provide continuous real-time monitoring of pO2, pCO2, SpO2, and pulse rate. Used in clinical settings by healthcare providers to monitor patient physiological status. Output is displayed on a touch screen, allowing clinicians to assess patient respiratory and circulatory health. Benefits include non-invasive, continuous monitoring of gas exchange and oxygenation status, aiding in clinical decision-making for patients not under gas anesthesia.

## Clinical Evidence

No clinical data. Safety and performance were established through bench testing and compliance with consensus standards including IEC-60601-1, IEC-60601-1-2, IEC-60601-2-23, IEC-60601-3-1, EN-865, and IEC-60601-1-8.

## Technological Characteristics

Monitor unit includes cabinet, touch screen, electronic controls, power supply, and battery. Uses combined tcpO2/tcpCO2 electrodes and pulse/SpO2 sensors. Connectivity includes integrated monitoring unit. Safety standards: IEC-60601-1, IEC-60601-1-2, IEC-60601-2-23, IEC-60601-3-1, EN-865, IEC-60601-1-8.

## Regulatory Identification

A cutaneous carbon dioxide (PcCO2) monitor is a noninvasive heated sensor and a pH-sensitive glass electrode placed on a patient's skin, which is intended to monitor relative changes in a hemodynamically stable patient's cutaneous carbon dioxide tension as an adjunct to arterial carbon dioxide tension measurement.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Cutaneous Carbon Dioxide (PcCO2) and Oxygen (PcO2) Monitors; Guidance for Industry and FDA.” See § 868.1(e) for the availability of this guidance document.

## Predicate Devices

- TCM400 Monitoring System
- TCM3 Monitoring System
- OxiMAX N-550 Pulse Oximetry System
- TOSCA PCO2, SpO2 and Pulse Rate Monitoring System
- Perimed Transcutaneous PO2 and PCO2 Monitor (PF5040)
- MicroGas 7650 Transcutaneous Monitor
- Siemens Transcutaneous TpO2/CO2 Gas Module

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Premarket Notification; TCM4/40; K043003

RADIOMETER COPENHAGEN

February 16, 2005

MAR ] - 2005

# 510(k) Summary - TCM4/40 Monitoring Systems

| Submitter:      | Radiometer Medical ApS                  |
|-----------------|-----------------------------------------|
| Address:        | Åkandevej 21, DK-2700 Brønshøj, Denmark |
| Phone:          | +45 3827.3827                           |
| Fax:            | +45 3827.2736                           |
| Contact Person: | Pierre Pelletier                        |

| Trade Name:           | TCM4/40 Monitoring Systems                           |
|-----------------------|------------------------------------------------------|
| Common name:          | Transcutaneous pO2/pCO2/SpO2/pulse monitoring system |
| Classification Names: | 21CFR§868.2500: Monitor, Oxygen, Cutaneous           |
|                       | 21CFR§868.2480: Monitor, Carbon Dioxide, Cutaneous   |
|                       | 21CFR§870.2700: Oximeter                             |
| Predicate Devices:    | TCM400 Monitoring System                             |
|                       | TCM3 Monitoring System                               |
|                       | OxiMAX N-550 Pulse Oximetry System                   |
|                       | TOSCA PCO2, SpO2 and Pulse Rate Monitoring System    |
|                       | Perimed Transcutaneous PO2 and PCO2 Monitor (PF5040) |
|                       | MicroGas 7650 Transcutaneous Monitor                 |
|                       | Siemens Transcutaneous TpO2/CO2 Gas Module           |

#### Device Description

The TCM4/40 Monitoring Systems essentially consist in a monitor unit, a combined oxygen/carbon dioxide interface module along with the combined electrodes assembly and (for the TCM40 only) an oxygen saturation/pulse module with a selection of three sensors.

#### Intended use

The TCM4 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures. It is indicated for use on neonates, pediatrics, and adults not under gas anesthesia.

The TCM40 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures as well as of oxygen saturation of arterial hemoglobin and pulse rate. It is indicated for use on neonates, pediatrics, and adults not under gas anesthesia.

{1}------------------------------------------------

RADIOMETER COPENHAGEN

# February 16, 2005

## Technological characteristics

The monitoring unit (including software, cabinet, touch screen, electronic controls, power supply and battery) is an adaptation of Radiometer's TCM400 Monitoring System, with only one combined tcp02/tcpCO2 electrode and (for the TCM40 only) one pulse/SoO2 sensor (the TCM400 has up to 6 separate tcg0> electrodes). The combined tcp02/tcpCO2 electrode is identical to the electrode used in Radiometer's previous generation TCM3 Monitoring System. The pulse/SpO2 sensors are identical to the sensors used with Puritan Nellcor Bennett's OxiMAX N-550 Pulse Oximetry System.

## Safety and performance evaluation studies

In order to ensure TCM4/40 Monitoring Systems is safe and effective for its intended use, it has been designed and tested to the requirements of the following standards:

- . IEC-60601-1:1988 Medical electrical equipment - Part 1: General requirements for safety
- IEC-60601-1-2:2001 . Medical Electrical Equipment - Part 1-2: General Requirements for Safety Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- IEC-60601-2-23:1999 . Medical electrical equipment - Part 2-23: Particular reguirements for the safety, including essential performance, of transcutaneous partial pressure monitoring equipment
- IEC-60601-3-1:1996 . Medical Electrical Equipment Part 3-1: Essential performance requirements for transcutaneous oxygen and carbon dioxide partial pressure monitoring equipment
- EN-865:1997 . Pulse oximeters - Particular requirements

. IEC-60601-1-8:2003

Medical electrical equipment - Part 1-8: General requirements for safety Collateral standard: General requirements, tests and quidance for alarm systems in medical electrical equipment and medical electrical systems

Based upon the extensive use of consensus standards covering all aspects of safety and performance, no clinical evaluation was deemed necessary to assess the product performance levels.

## Conclusions from safety and performance evaluation studies

Compliance with the requirements in the above-named safety and performance standards demonstrates that the TCM4/40 Monitoring Systems are state-of-the-art transcutaneous oxygen/carbon dioxide monitors and pulse oximeters that have the necessary safety mechanisms and performs up to the requirements of the relevant international consensus standards.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract design of an eagle's head and neck, rendered in black lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle design. The text is in a sans-serif font and is also black.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 - 2005

Mr. Pierre Pelletier Regulatory Affairs Manager Radiometer Medical ApS Åkandevej 21, Brønshøj Denmark DK-2700

Re: K043003

Trade/Device Name: TCM4/40 Monitoring Systems Regulation Number: 21 CFR 868.2480 Regulation Name: Cutaneous Carbon Dioxide (PcCO2) Monitor Regulatory Class: II Product Code: LKD, LPP, KLK, DQA Dated: January 20, 2005 Received: January 24, 20045

Dear Mr. Pelletier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Ecderal Register.

{3}------------------------------------------------

#### Page 2 - Mr. Pelletier

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-frec number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Susan Runne
Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K043003

TCM4/40 Monitoring Systems Device Name:

#### Indications for Use:

The TCM4 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and The TCM4 Monitoring System is mended for continued for use on neonates, pediatrics, and adults not under gas anesthesia.

The TCM40 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures as well as of oxygen saturation of arterial hemoglobin and pulse Carbon uloxide partial pressures as well as of thyger of the mot under gas anesthesia.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

# Concurrence of CDRH, Office of Device Evaluation (ODE)

and the comments of the comments of the comments of the comments of the comments of the comments of

Cicia Fowler

sion Sign-O
son of Anesthesiology, General Hospital.
action Control. Dental Devices

in Number

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K043003](https://fda.innolitics.com/device/K043003)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
