← Product Code [PAG](/productcode/PAG) · K042949

# PELVILACE TRANSOBTURATOR BIOURETHRAL SUPPORT SYSTEM (PELVILACE TO SYSTEM) (K042949)

_C.R. Bard, Inc. · PAG · Nov 9, 2004 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K042949

## Device Facts

- **Applicant:** C.R. Bard, Inc.
- **Product Code:** [PAG](/productcode/PAG.md)
- **Decision Date:** Nov 9, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The PelviLace™ Transobturator BioUrethral Support System is intended for use as a soft tissue patch to reinforce soft tissue membranes. It is specifically indicated for the surgical repair of tissue weakness resulting from damaged or ruptured soft tissue and for the treatment of stress urinary incontinence in women resulting from urethral hypermobility and/or intrinsic sphincter deficiency.

## Device Story

PelviLace™ TO BioUrethral Support System consists of Pelvicol implant and introducers; used for surgical treatment of stress urinary incontinence in women. Implant provides natural backboard for urethra during abdominal pressure increases; reinforces soft tissue membranes. Used in clinical setting by physicians; manual placement of implant via introducers.

## Technological Characteristics

Consists of Pelvicol implant and introducers. Functions as a soft tissue patch for reinforcement. No software or electronic components.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- InnerLace™ BioUrethral Support System
- Uretex™ TO Transobturator Urethral Support System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K042949 1 of 1

C. R. Bard, Inc. 8195 Industrial Blvd. Covington, GA 30014

NOV - 9 2004

Image /page/0/Picture/3 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are all capitalized and connected to each other. The word is black and the background is white. 

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATIO

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

- A. SUBMITTER INFORMATION:

| Submitter's Name: | C. R. Bard, Inc., Urological Division |
|-------------------|---------------------------------------|
| Address:          | 8195 Industrial Blvd.                 |
|                   | Covington, GA 30014                   |

John C. Knorpp Contact Person: 770-784-6451 Contact Person's Telephone Number: 770-784-6419 Contact Person's Fax: October 25, 2004 Date of Preparation:

## в. DEVICE NAME:

| Trade Name:           | PelviLace™ TO BioUrethral Support System |
|-----------------------|------------------------------------------|
| Common/Usual Name:    | Suburethral Sling                        |
| Classification Names: | PAG Mesh, Surgical, Polymeric            |

### C. PREDICATE DEVICE NAME:

Trade Names: InnerLace™ BioUrethral Support System and Uretex™ TO Transobturator Urethral Support System

### D. Device Description:

The Bard® PelviLace™ TO BioUrethral Support System consists of the PelviLace Pelvicol implant and introducers. The introducers are used to place the implant which provides a natural backboard for the urethra during abdominal pressure increases.

## E. INTENDED USE:

The PelviLace™ TO BioUrethral Support System is used for the treatment of stress urinary incontinence.

- F. TECHNOLOGICAL CHARACTERISTICS SUMMARY:
The subject PelviLace™ TO BioUrethral Support System has the same intended use, design and fundamental scientific technology as the predicate devices.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. John C. Knorpp Senior Regulatory Affairs Specialist C.R. Bard, Inc., Urological Division 8195 Industrial Boulevard COVINGTON GA 30014

SEP 2 8 2012

K042949 Trade/Device Name: PelviLace™ Transobturator BioUrethral Support System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: PAG Dated: October 25, 2004 Received: November 1, 2004

Dear Mr. Knorpp:

Re:

This letter corrects our substantially equivalent letter of November 9, 2004.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{2}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Sincerely yours,

Benjamin R. Fuchs

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Bard Urological Division, C.R. Bard, Inc. PelviLace™ TO System Premarket Notification [510(k)]

Indications for Use Statement 1.3

510(k) Number (if known): ___K042949

Device Name: ___PelviLace™ Transobturator BioUrethral Support System

Indications for Use:

The PelviLace™ Transobturator BioUrethral Support System is intended for use as a soft The PelvilLace™ Transolurator Dioorcanal Oupport of the surgical repair of tissue patch to remiorce soft tissue membranes. It is specifically in weap soculting from damaged or ruptured sont ussue incribition of stress urinary incontinence in women resulting from publumbling oling fity and/or intrinsic sphincter deficiency.

X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

Miriam C. Provost

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

**510(k) Number** K042949

(Recommended Format 11/13/2003)

---

**Source:** [https://fda.innolitics.com/device/K042949](https://fda.innolitics.com/device/K042949)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
