MGP 15 DICOM THEATER

K042942 · Barco NV Barcoview · LLZ · Nov 22, 2004 · Radiology

Device Facts

Record IDK042942
Device NameMGP 15 DICOM THEATER
ApplicantBarco NV Barcoview
Product CodeLLZ · Radiology
Decision DateNov 22, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2

Indications for Use

The MGP 15 Dicom Theater is intended to be used as a tool in displaying and viewing digital images for review and analysis by trained medical practitioners.

Device Story

MGP 15 Dicom Theater is a digital image display system; functions as a medical projector for high-resolution visualization of digital images. Used by trained medical practitioners in clinical settings to review and analyze medical imagery. Device facilitates clinical decision-making by providing a platform for viewing diagnostic data. Benefits include enhanced visualization capabilities for medical professionals.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Digital image display system/medical projector designed for high-resolution visualization. Technical specifications and materials not detailed beyond classification as an image processing system.

Indications for Use

Indicated for use by trained medical practitioners for the display, review, and analysis of digital medical images.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K042942 NOV 2 2 2004 ## 510(K) SUMMARY | Manufacturer: | Barco NV Barcoview<br>Theodoor Sevenslaan 106<br>8500 Kortrijk<br>Belgium | |--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitted By: | Ferguson Medical<br>Consultant to Barco NV | | Contact Information: | Phone: +32(0) 56 23 32 11<br>FAX: +32(0) 56 23 3 74 | | Classification Name: | System, image processing | | Common/Usual Name: | Dicom compliant projection system, medical<br>projector and others | | Proprietary Name: | MGP 15 Dicom Theater | | Classification Number: | 21 CFR 892.2050/Procode 90LLZ | | Substantial Equivalence: | Dicom Theater (K033153) and others | | Device Description: | The MGP 15 Dicom Theater device is a digital<br>image display system | | Intended Use: | The MGP 15 Dicom Theater device is intended<br>to be used in displaying and viewing digital<br>images for review by trained medical<br>practitioners. | | Technological Characteristics: | The MGP 15 Dicom Theater consists of<br>components to provide high resolution<br>visualization of digital images. | {2}------------------------------------------------ Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## NOV 2 2 2004 Barco NV Barcoview % Mr. Frank Ferguson Official Correspondent FERGUSON MEDICAL 12200 Academy Road NE #931 ALBUQUERQUE NM 87111 Re: K042942 Trade/Device Name: MGP 15 Dicom Theater Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: October 10, 2004 Received: October 25, 2004 .Dear Mr. Ferguson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 510(k) Number (If known): Device Name: MGP 15 Dicom Theater Indications For Use: The MGP 15 Dicom Theater is intended to be used as a tool in displaying and viewing digital images for review and analysis by trained medical practitioners. ## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use_ XX (Per 21 CFR 801.109) OR Over-The- Counter Use ________________________________________________________________________________________________________________________________________________________ Nancy C. brordon (Division Sign-Off)() Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K042942
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...