The Triage® D-Dimer Test is a fluorescence immunoassay to be used with the Triage® Meter Plus for the quantitative determination of cross-linked fibrin degradation products containing D-dimer in EDTA whole blood and plasma specimens. The test is used as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism.
Device Story
The Triage D-Dimer Test is a fluorescence immunoassay designed for use with the Triage Meter Plus. It analyzes EDTA whole blood or plasma specimens to quantitatively measure cross-linked fibrin degradation products (D-dimer). The device is intended for professional laboratory use to assist clinicians in assessing patients suspected of thromboembolic events, such as pulmonary embolism, or disseminated intravascular coagulation. By providing quantitative D-dimer levels, the test aids in evaluating abnormal clot burden, supporting clinical decision-making regarding patient diagnosis and management.
Clinical Evidence
Bench testing only. Precision: 6.0-15.4% CV. Linearity: 100-5000 ng/mL (recovery 80-115%). Analytical specificity: interference limits established for hemoglobin, lipids, bilirubin, fibrinogen, and fragments D/E. Method comparison (N=180) vs. Dade Behring STRATUS CS DDMR yielded slope 0.999, intercept -85.89, cc 0.92. Matrix comparison (N=22) between whole blood and plasma showed R2=0.98. Reference range study (N=208) established 90th percentile at 400 ng/mL and 95th percentile at 600 ng/mL.
Technological Characteristics
Fluorescence immunoassay; quantitative measurement of D-dimer; utilizes Triage Meter Plus for signal detection; specimen types include EDTA whole blood and plasma; professional laboratory use.
Indications for Use
Indicated for patients suspected of having disseminated intravascular coagulation or thromboembolic events, including pulmonary embolism. Used for quantitative determination of D-dimer in EDTA whole blood and plasma specimens.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number: #K042890
B. Purpose for Submission: Device modification
C. Measurand: D-dimer
D. Type of Test: Quantitative
E. Applicant: Biosite, Inc.
F. Proprietary and Established Names: Triage® D-Dimer Test; Fibrin Split Products
## G. Regulatory Information:
1. Regulation section: 21 CFR 864.7320 – Fibrin(ogen) Degradation Products
2. Classification: Class II
3. Product code: DAP, GHH
4. Panel: Hematology (81)
## H. Intended Use:
1. Intended use(s): The Triage® D-Dimer Test is a fluorescence immunoassay to be used with the Triage® Meter Plus for the quantitative determination of cross-linked fibrin degradation products containing D-dimer in EDTA whole blood and plasma specimens. The test is used as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism.
2. Indication(s) for use: Same as Intended Use.
3. Special conditions for use statement(s): For professional and laboratory use.
4. Special instrument requirements: Biosite Triage® Meter Plus
## I. Device Description:
The Triage® D-Dimer Test is identical in principle, reagents and procedure to the Biosite Triage® Profiler S.O.B. Panel, which contains an assay for the D-dimer analyte. The D-Dimer Test consists of a murine monoclonal antibody (MAB) against D-dimer. The MAB is labeled with a fluorescent dye and immobilized on
{1}
a solid phase with stabilizers. The Kit contains (25), each, tests and transfer pipettes; and (1), each, reagent code chip and roll of printer paper.
# J. Substantial Equivalence Information:
1. Predicate device name(s): Biosite Triage® Profiler S.O.B. Test; Dade Behring STRATUS CS DDMR Test
2. Predicate 510(k) number(s): #K040437; #K022976
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Analyte | D-dimer | Same |
| Principle | Fluorescence immuno-assay | Same |
| Instrumentation | Triage® Meter Plus | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Analyte(s) | Single | Multiple |
# K. Standard/Guidance Document Referenced (if applicable): N/A
L. Test Principle: When a whole blood sample is added to the sample port of the Triage® D-Dimer Test, the blood cells are separated from the plasma via a filter in the device. The plasma reacts with the fluorescent antibody conjugates in the reaction chamber. Following incubation, the reaction mixture flows down the detection lane. Analyte and antibody conjugate are captured in a binding assay. Analyte concentration in the sample is directly proportional to the detected fluorescence. Results are read on the Triage® Meter Plus instrument (#K973547).
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
a. Precision/Reproducibility: Three levels of controls were tested x (10)/day, over a 10-day period. Average within-day precision ranged from $6.0 - 14.4\%$ CV. The average total precision ranged from $6.1 - 15.4\%$ CV.
{2}
b. Linearity/assay reportable range: Four donor plasmas were spiked with purified D-dimer, and diluted with unspiked plasma to cover the measuring range of 100 – 5000 ng/mL.
Actual vs expected results yielded a recovery between 80 – 115%, with an observed measuring range of 0.10 – 4.61 ug/mL. Regression statistics yielded slopes that ranged 0.95 – 1.03; and ‘R’ values of 0.99 – 1.00.
c. Traceability, Stability, Expected values (controls, calibrators, or methods): The calibrator is traced to an in-house purified protein preparation of D-dimer, based upon the mass (concentration) of analyte present in plasma.
d. Detection limit: A ‘zero’ calibrator was tested (20) times on (3) lots of reagent, using (2) different meters, over a 5-day period. The lower limit of detection (LLOD) was established as <100 ng/mL, with 95% confidence.
e. Analytical specificity: A multiplicity of common pharmaceuticals, proteins and biologics were tested for interference. The following limits for interference were established for hemoglobin (5 mg/mL); lipids (30 mg/mL); bilirubin (0.15 mg/mL); fibrinogen (1 mg/mL); Fragment D (20 ug/mL); Fragment E (20 ug/mL); and hematocrit (<30% or >55%).
f. Assay cut-off: N/A
2. Comparison studies:
a. Method comparison with predicate device: The Triage® D-Dimer Test was compared to the Dade Behring STRATUS CS DDMR Test on patient samples (N = 180). Samples included: Healthy (N = 111), PE (N = 17), MI (N = 32), Angina (N = 11), CHF (N = 4), and Chest pain (N = 5), with values ranging from < 100 ng/mL to > 5000 ng/mL.
The study yielded these regression statistics:
Slope = 0.999
Intercept = (-) 85.89
cc = 0.92
A bias plot of the difference between methods vs the mean of all methods, demonstrated several outliers (> ± 1000).
b. Matrix comparison: A matrix comparison study was performed using samples (N = 22), in triplicate, of whole blood and plasma (EDTA). Samples ranged in value from 157 – 2997 ng/mL. The regression statistics from the study were:
y = 0.91x + 54.32; R2 = 0.98. The ‘p’ value = >0.05.
{3}
3. Clinical studies:
a. Clinical Sensitivity: N/A
b. Clinical specificity: N/A
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off: N/A
5. Expected values/Reference range: A reference range study was performed on healthy donors (N = 208), ranging in value from 10 – 1850 ng/mL. Donor samples were from males (N = 131) and females (N = 77), who ranged 19-79 years of age. The 90th % = 400 ng/mL; and the 95th % = 600 ng/mL.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.