TAMPER EVIDENT ULTRASAFE PASSIVE DELIVERY SYSTEM

K042712 · Safety Syringes, Inc. · MEG · Oct 15, 2004 · General Hospital

Device Facts

Record IDK042712
Device NameTAMPER EVIDENT ULTRASAFE PASSIVE DELIVERY SYSTEM
ApplicantSafety Syringes, Inc.
Product CodeMEG · General Hospital
Decision DateOct 15, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesPediatric

Indications for Use

The UltraSafe Passive™, X-Series, Needle Guard is intended for use as a safety mechanism designed to reduce the occurrence of accidental needlesticks to healthcare professionals during disposal of a used syringe and needle assembly. The UltraSafe Passive, X-Series, Needle Guard is snapped onto the barrel of an ISO Standard pre-filled glass syringe, in a retracted position. Upon completing the injection, the UltraSafe Passive, X-Series, Needle Guard passively activates and covers the exposed needle. When the Passive Needle Guard is activated, it provides protection from an accidental needlestick. The healthcare professional disposes the used syringe and needle assembly in a sharps container. The intended patient population is unrestricted and includes children and adults. The intended environment of use is where healthcare professionals are required to administer medication, including vaccines, by means of a syringe.

Device Story

UltraSafe Passive X-Series Needle Guard is a passive safety accessory for ISO standard pre-filled glass syringes (0.5 mL to 5.0 mL). Device snaps onto syringe barrel in retracted position; automatically activates upon injection completion to cover exposed needle. Used by healthcare professionals in clinical settings to prevent accidental needlesticks during disposal. Passive activation mechanism eliminates manual engagement required by predicate B-Series. Healthcare professional performs injection; guard deploys automatically; user disposes of assembly in sharps container. Benefits include reduced risk of occupational needle-stick injuries.

Clinical Evidence

Controlled simulated clinical use test conducted with nurses comparing X-Series (subject) to B-Series (predicate). Study evaluated safety, performance, usability, and clarity of instructions. Results indicated X-Series was rated equal to or better than the predicate on all parameters.

Technological Characteristics

Accessory for ISO 11040-4 compliant pre-filled glass syringes. Passive mechanical activation mechanism. Dimensions accommodate 0.5 mL to 5.0 mL fill volumes. Single-use, non-sterile (accessory).

Indications for Use

Indicated for use as a single-use accessory with ISO Standard pre-filled glass syringes to protect healthcare professionals from accidental needlesticks. Used for parenteral administration in a wide range of patients, including children and adults.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K042712 ## 510(k) Summary OCT 15 2004 ### Special 510(k) summary of safety and effectiveness #### Company information Safety Syringes, Inc. 1925 Palomar Oaks Way, Suite 204 Carlsbad, CA 92008 USA #### Device Identification Trade Name - UltraSafe Passive™, X-Series, Needle Guard - Models X50, X100, X100L, X150, X225, X300, X500 Classification Name - Syringe, Piston (accessory) Classification - Class II Product Code - 80 FMF #### Predicate Device The UltraSafe® Injection System Needle Guard - Models B50, B100L. B150, B225, and B300 #### Device Description The UltraSafe Passive™, X-Series, Needle Guard is an accessory for ISO Standard pre-filled glass syringes. ISO Standard pre-filled glass syringes are supplied in multiple sizes. They fit with ISO Standard pre-filled glass syringes ranging from 0.5 mL to 5.0 mL of fill volume. The guards are designed to meet the requirements specified in ISO 11040-4 Prefilled syringes - Part 4 Glass barrels for injectables (March 1995). #### Intended Use The UltraSafe Passive™, X-Series, Needle Guard is intended for use as a safety mechanism designed to reduce the occurrence of accidental needlesticks to healthcare professionals during disposal of a used syringe and needle assembly. The UltraSafe Passive, X-Series, Needle Guard is snapped onto the barrel of an ISO Standard pre-filled glass syringe, in a retracted position. Upon completing the injection, the UltraSafe Passive, X-Series, Needle Guard passively activates and covers the exposed needle. When the Passive Needle Guard is activated, it provides protection from an accidental needlestick. The healthcare professional disposes the used syringe and needle assembly in a sharps container. The intended patient population is unrestricted and includes children and adults. {1}------------------------------------------------ K042712 292 The intended environment of use is where healthcare professionals are required to administer medication, including vaccines, by means of a syringe. #### Indications for Use UltraSafe Passive™, X-Series, Needle Guards are single use devices that are indicated for use as an accessory with ISO Standard pre-filled glass syringes to protect healthcare professionals from accidental needle sticks. This device is used on a wide range of patients, including children and adults, and parenteral methods of administration. #### Comparison of the UltraSafe® Injection System Needle Guard - B-Series to the UltraSafe Passive™, X-Series, Needle Guard The UltraSafe Passive™ X-Series, Needle Guards are modifications of the previously cleared (K002070) UltraSafe Injection System Needle Guard, B-Series. The B-Series are manually activated needle guards. The X-Series are passively (automatically) activated needle guards. The basic design, function, intended use, and indications for use of the B-Series and X-Series are similar. Safety Syringes, Inc. conducted iterative user evaluations of the X-Series needle guard. A controlled simulated clinical use test was conducted with the UltraSafe Passive, X-Series, Needle Guard to validate the product in the hands of typical users. The UltraSafe Injection System Needle Guard, B-Series, was used as the control. In the controlled simulated clinical use test, each of the specified acceptance criteria were either met or exceeded. The nurses involved in this study rated the X-Series to be equal to, or better than, the B-Series on all of the parameters related to safety, performance, and usability. In addition to evaluating safety, performance, and usability parameters, there were questions of the study participants regarding the clarity and adequacy of the written and illustrated Directions for Use, and the participants' response to the need for in-service training. On each of the parameters, the X-Series rated as equal to, or better, than the B-Series. - K9722878 Thomas L. Hale 10/07/04 03 10 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three horizontal lines curving upwards, resembling a person with outstretched arms. OCT 1 5 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Thomas L. Hall Director, Regulatory Affairs and Quality Assurance Safety Syringes, Incorporated 1939 Palomar Oaks Way, Suite A, Carlsbad, California 92009 Re: K042712 Trade/Device Name: Tamper Evident UltraSafe® Passive™ Delivery System Syringe, Piston (Accessory) Regulation Number: 880.5860 Regulation Name: Antistick Syringe, Piston Syringe Regulatory Class: II Product Code: MEG, FMF Dated: September 27, 2004 Received: September 30, 2004 Dear Mr. Hall: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Hall Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements nean that 1 Dr mas made statutes and regulations administered by other Federal agencies. or the Act of ally I outstar the Act's requirements, including, but not limited to: registration r ou must comply with and 807); labeling (21 CFR Part 801); good manufacturing practice alle listing (21 er result in the quality systems (QS) regulation (21 CFR Part 820); and if requirements as be former product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) Prinstet notification. The FDA finding of substantial equivalence of your device to a promative noticate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), rr you decire spiee of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the maj count Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susan Runner Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K0427/2 # Indications for Use Statement 510(k) Number: _ Tamper Evident Device Name: UltraSafe® Passive™ Delivery System Syringe, Piston (Accessory) Indications for Use: Single Use devices that are indicated for use as an accessory with pre-filled \SO le from that are indicated for use of he diberary professionals from Single Ose devices that are indication for ass as the althcare professionals from accidental needlesticks. These devices can be used on a wide range of patients, including children and adults, and parenteral methods of administration. Anew Navesse for Adw 10/14/04 (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number: K042712 ## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation Prescription Use × (Per 21 CFR 801.109 OR Over the Counter______________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...