VITROS CHEMISTRY PRODUCT MALB REAGENT, CALIBRATOR KIT 24, MALB PERFORMANCE VERIFIERS I AND II
K042643 · Ortho-Clinical Diagnostics, Inc. · DCF · Nov 18, 2004 · Immunology
Device Facts
Record ID
K042643
Device Name
VITROS CHEMISTRY PRODUCT MALB REAGENT, CALIBRATOR KIT 24, MALB PERFORMANCE VERIFIERS I AND II
Applicant
Ortho-Clinical Diagnostics, Inc.
Product Code
DCF · Immunology
Decision Date
Nov 18, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5040
Device Class
Class 2
Indications for Use
VITROS Chemistry Products mALB Reagent: For in vitro diagnostic use only. VITROS Chemistry Products mALB Reagent is used to quantitatively measure albumin concentration in human urine (mALB). VITROS Chemistry Products Calibrator Kit 24: For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 24 is used to calibrate VITROS 5, 1 FS Chemistry Systems for the quantitative measurement of albumin in urine (mALB). VITROS Chemistry Products mALB Performance Verifiers I and II: For in vitro diagnostic use only. VITROS Chemistry Products mALB Performance Verifiers are assayed controls used to monitor the performance of mALB Reagent on VITROS 5, 1 FS Chemistry Systems.
Device Story
VITROS Chemistry Products mALB Reagent, Calibrator Kit 24, and Performance Verifiers I and II are used on the VITROS 5,1 FS Chemistry System; a fully automated clinical chemistry analyzer. The system processes human urine samples using immunoturbidimetry to determine albumin concentration. The analyzer combines liquid reagents with samples to perform the assay. Results are provided to clinicians to assist in diagnosing diabetic nephropathy, hypertension, and cardiovascular disease. The system is intended for professional use in clinical laboratory settings.
Clinical Evidence
Bench testing only. Precision evaluated per NCCLS EP5-A (within-lab CV 0.73-3.28%). Linearity confirmed across 0.6-19.0 mg/dL per NCCLS EP6-A. Specificity tested per NCCLS EP7-A against 20+ endogenous/exogenous substances; minor bias noted for furosemide and sulfamethoxazole. Method comparison (n=61) against predicate showed y = 0.93x + 0.03, r=0.977. Reference range established per NCCLS C28 using 129 healthy subjects.
Technological Characteristics
Liquid, dual-chambered reagent kit. Principle: immunoturbidimetric assay. Energy: spectrophotometric (340 nm). Instrumentation: VITROS 5, 1 FS Chemistry System. Traceability: IRMM/IFCC CRM 470. Reagents contain goat anti-sera to human albumin, polymers, and surfactants. Software: embedded in VITROS 5, 1 FS system for automated analysis and calibration curve storage.
Indications for Use
Indicated for quantitative measurement of human urine albumin (mALB) to aid in the diagnosis of diabetic nephropathy, hypertension, and cardiovascular disease. For prescription use on VITROS 5, 1 FS Chemistry Systems.
Regulatory Classification
Identification
An albumin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the albumin (a plasma protein) in serum and other body fluids. Measurement of albumin aids in the diagnosis of kidney and intestinal diseases.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
Predicate Devices
N Antiserum to Human Albumin assayed on the Dade Behring BN ProSpec System (K972929)
VITROS Chemistry Products Performance Verifiers I and II (K041720)
Submission Summary (Full Text)
{0}
Page 1 of 12
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE
A. 510(k) Number: k042643
B. Purpose For Submission:
Premarket Notification (510(k)) for Ortho-Clinical Diagnostics, Inc., of intention to manufacture and market the VITROS® Chemistry Product mALB Reagent, VITROS Chemistry Products Calibrator Kit 24, and the VITROS Chemistry Products mALB Performance Verifiers I and II.
C. Analyte: Microalbumin
D. Type of Test:
The VITROS Chemistry Products mALB Reagent is used for the quantitative in vitro diagnostic measurement of human urine albumin in conjunction with the VITROS Chemistry Products Calibrator Kit 24 on the VITROS 5, 1 FS Chemistry Systems. The basic test principle utilizes a turbidimetric inhibition immunoassay technology.
E. Applicant: Ortho-Clinical Diagnostics, Inc.
F. Proprietary and Established Names:
VITROS Chemistry Products mALB Reagent
VITROS Chemistry Products Calibrator Kit 24
VITROS Chemistry Products mALB Performance Verifiers I and II
G. Regulatory Information:
1. Regulation section:
21 CFR §866.5040, Albumin immunological test system
21 CFR §862.1660, Single (Specified) Analyte Controls (Assayed and Unassayed)
21 CFR §862.1150, Calibrator, Secondary
2. Classification:
Class II (assay) - Subject to limitations of exemptions, 21 CFR 866.9 (c)(5)
Class I (controls)
Class II (calibrator)
{1}
Page 2 of 12
3. Product Code:
DCF (assay)
JXX (controls)
JIT (calibrator)
4. Panel:
82 (Immunology)
75 (Chemistry)
75 (Chemistry)
H. Intended use(s):
1. Intended use(s)
VITROS Chemistry Products mALB Reagent: For in vitro diagnostic use only. VITROS Chemistry Products mALB Reagent is used to quantitatively measure albumin concentration in human urine (mALB).
VITROS Chemistry Products Calibrator Kit 24: For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 24 is used to calibrate VITROS 5, 1 FS Chemistry Systems for the quantitative measurement of albumin in urine (mALB).
VITROS Chemistry Products mALB Performance Verifiers I and II: For in vitro diagnostic use only. VITROS Chemistry Products mALB Performance Verifiers are assayed controls used to monitor the performance of mALB Reagent on VITROS 5, 1 FS Chemistry Systems.
2. Indication(s) for use:
VITROS Chemistry Products mALB Reagent: For in vitro diagnostic use only. VITROS Chemistry Products mALB Reagent is used to quantitatively measure albumin concentration in human urine (mALB). Measurement of urinary albumin aids in the diagnosis of diabetic nephropathy, hypertension and cardiovascular disease.
VITROS Chemistry Products Calibrator Kit 24: For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 24 is used to calibrate VITROS 5, 1 FS Chemistry Systems for the quantitative measurement of albumin in urine (mALB).
VITROS Chemistry Products mALB Performance Verifiers I and II: For in vitro diagnostic use only. VITROS Chemistry Products mALB Performance Verifiers are assayed controls used to monitor the performance of mALB Reagent on VITROS 5, 1 FS Chemistry Systems.
3. Special condition for use statement(s): For Prescription Use.
{2}
Page 3 of 12
4. Special instrument Requirements: VITROS 5, 1 FS Chemistry Systems
I. Device Description:
The VITROS 5, 1 FS Chemistry System is a fully automated clinical chemistry analyzer intended for use in the in vitro determination of various analytes in human specimens (serum, plasma, urine, and cerebrospinal fluid). The VITROS 5, 1 FS Chemistry System is designed for use with VITROS Chemistry Products MicroTip and Thin Film assays.
The system is comprised of four main elements:
1. The VITROS 5, 1 FS Chemistry System-instrumentation, which provides automated use of chemistry reagents. The VITROS 5, 1 FS Chemistry System was cleared for market by a separate 510(k) premarket notification (k031924).
2. The VITROS Chemistry Products MicroTip range of liquid reagent products (in this case VITROS Chemistry Products mALB Reagent, VITROS Chemistry Products Calibrator Kit 24, and VITROS Chemistry Products mALB Performance Verifiers I and II), which are combined by the VITROS 5, 1 FS Chemistry System to perform the VITROS mALB assay.
3. The VITROS Chemistry Products Thin Film range of dry products, which are dry, multilayered, analytical elements, coated on polyester supports. The thin film products each have there own 510(k) clearance numbers and were cleared for market use on the VITROS 5, 1 FS Chemistry System though submission of information required by the ODE Guidance Document: "Data For Commercialization Of Original Equipment Manufacturer, Secondary and Generic Reagents for Automated Analyzers". The required information was provided in the VITROS 5, 1 FS Chemistry System premarket notification (k031924)
4. Common reagents used by multiple assays on the VITROS System (in this case, VITROS Chemistry Products FS Diluent Pack 2). The VITROS System and reagents are designed specifically for use with the VITROS Chemistry Products range of products.
J. Substantial Equivalence Information:
1. Predicate device name(s): Dade Behring N Antiserum to Human Albumin, Prealbumin and Retinol-binding Protein
2. Predicate K number(s): k972929
3. Comparison with Predicate:
{3}
Page 4 of 12
The VITROS Chemistry Products mALB Reagent and VITROS Chemistry Products Calibrator Kit 24 are substantially equivalent to N Antiserum to Human Albumin assay on the Dade Behring BN ProSpec System (predicate device) which was 510(k) cleared in k972929 for IVD use.
The relationship between the VITROS mALB assay and the predicate device, determined by the Passing & Bablock linear regression is:
VITROS mALB assay = 0.93x + 0.03
with a correlation coefficient of 0.977,
where X is the result for the N Antiserum to
Human Albumin assay on the Dade Behring
BN ProSpec System.
In addition to the above mentioned correlation study, studies were performed to determine precision, analytical sensitivity, specificity and expected values of the VITROS mALB assay. The table below lists the characteristics of the VITROS mALB assay (new device) and the Dade N Antiserum to Human Albumin assay (predicate device).
| Device Characteristics | VITROS mALB assay (k402643) New Device | Dade Albumin assay (k972929) Predicate Device |
| --- | --- | --- |
| Intended Use | For in vitro diagnostic use only. VITROS Chemistry Products mALB Reagent is used to quantitatively measure albumin concentration in human urine (mALB) | In vitro diagnostic reagents for the quantitative determination of albumin in human urine using the BN Systems. |
| Method | Immunoturbidimetry | Rate Nephelometry |
| Reportable Range | 0.6 to 19 mg/dL | 0.0 to 34.0 mg/dL |
| Instrumentation | VITROS 5, 1 FS Chemistry Systems | Dade Behring BN ProSpec Systems |
| Sample Type | Urine | Urine |
| Reactive Ingredient | Goat anti-sera to human albumin | Rabbit antiserum to human albumin |
The VITROS Chemistry Products mALB Performance Verifiers I and II are substantially equivalent to the VITROS Chemistry Products Performance Verifiers (predicate device) which were 510(k) cleared in k041720 for IVD use. The table below lists the similarities and differences of the device characteristics between the VITROS mALB Performance Verifiers and the predicate device, VITROS Performance Verifiers I and II.
{4}
Page 5 of 12
| Device Characteristics | Performance Verifiers I and II (k042643) New Device | Performance Verifiers I and II (k041720) Predicate Device |
| --- | --- | --- |
| Intended Use | For in vitro diagnostic use only. VITROS Chemistry Products mALB Performance Verifiers are assayed controls used to monitor the performance of mALB Reagents on VITROS 5, 1 FS Chemistry Systems. | For in vitro diagnostic use only. VITROS Performance Verifier is an assayed control used to monitor performance on VITROS Chemistry Systems. |
| Fluid Matrix | A base matrix of processed human serum to which inorganic salt, buffers, protein, surfactant and preservatives have been added. | A base matrix of freeze-dried human serum to which enzymes, electrolytes Stabilizers, preservatives and other organic analytes have been added. |
| Analyte Levels | Low and High | Low and High |
## K. Standard/Guidance Document Referenced (if applicable):
Values assigned to the VITROS Chemistry Products Calibrator Kit 24 for urine albumin are traceable to IRMM/IFCC (Institute for Reference Methods and Materials/ International Federation of Clinical Chemistry and Laboratory Medicine) CRM 470 (RPPHS – Reference Preparation for Proteins in Human Serum) Reference Material.
FDA Guidance Document: “Data For Commercialization Of Original Equipment Manufacturer, Secondary and Generic Reagents for Automated Analyzers”
NCCLS Protocol EP5-A (Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline)
NCCLS EP6-A (Evaluation of the Linearity of Quantitative Analytical Methods-Approved Guideline)
NCCLS C28 (How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline-Second Edition)
NCCLS protocol EP7-A (Interference Testing in Clinical Chemistry; Approved Guideline (2002))
## L. Test Principle:
The quantitative measurement of urine albumin is performed using the VITROS Chemistry Products mALB Reagent in conjunction with the VITROS Chemistry Products Calibrator Kit 24 on VITROS 5, 1 FS Chemistry Systems. The VITROS Chemistry Products mALB Reagent is a dual chambered package containing ready-to-use liquid
{5}
Page 6 of 12
reagents. Samples, calibrators and controls are mixed with Reagent 1 containing a polymer and surfactant. Addition of antisera specific for human albumin (Reagent 2) produces an immunochemical reaction yielding antibody/antigen complexes. The light scattering properties of the antibody/antigen complexes increase solution turbidity proportional to albumin concentration in the sample. The turbidity is measured spectrophotometrically at 340 nm. Once a calibration has been performed for each reagent lot, the urinary albumin concentration in each unknown sample can be determined using the stored calibration curve and the measured absorbance obtained in the assay of the sample.
## Reaction Sequence
Step 1: Dilution of mALB Sample
Sample
R1 → Surfactant/ Polymer
Diluted Sample
Step 2: Formation of mALB/ Antibody Complexes
Diluted Sample
R2 → Antigen/ Antibody Complexes
Polyclonal antibodies
## M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision was evaluated with quality control materials on the VITROS 5, 1 FS Chemistry System following NCCLS Protocol EP5-A (Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline).
Two runs per day were performed on 22 different days, using three VITROS 5, 1 FS Chemistry Systems with three different reagent lots. Each run consisted of two control fluids to evaluate system precision and one human urine pool to simulate customer use. All fluids were run in duplicate. Runs within day were separated by at least two hours. The data were screened for outliers according to the statistical outlier test in NCCLS EP5-A. Two replicates were removed as a result of this analysis. The calibrator samples tested each week were used to generate a calibration for each week of the experiment, which was used to calibrate and predict all results from that week. This analysis introduces the variability inherent in the act of calibration.
The results of the test samples, for each matrix evaluated, were analyzed by ANOVA as detailed in NCCLS EP5-A. A nested ANOVA was performed to estimate within-run, run-to-run (within-day), day-to-day, week-to-week (cal-to-cal), and total within lab precision. Another nested ANOVA was performed to estimate within-day, day-to-day, week-to-week (cal-to-cal) and total within lab precision. The ANOVA results represent an approximation of the total, within-laboratory variability that would be observed using a single lot of reagents.
{6}
Testing was performed on three lots of reagent. Results from one lot were used to determine the product claim, since performance achieved with this lot defines the most conservative claim. See the table below:
VITROS Chemistry Products mALB Reagent Precision
| | Conventional Units (mg/dL) | | | SI Units (mg/dL) | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| System | Mean Conc. | Within Day SD* | Within Lab SD* | Mean Conc. | Within Day SD* | Within Lab CV** | Number of Observation | Number of Day | |
| VITROS 5, 1 FS Chemistry System | 4.4 | 0.05 | 0.07 | 43.6 | 0.50 | 0.73 | 1.67 | 88 | 22 |
| | 7.9 | 0.11 | 0.07 | 78.7 | 1.09 | 2.58 | 3.28 | 88 | 22 |
| | 2.6 | 0.03 | 0.007 | 25.8 | 0.34 | 0.73 | 2.83 | 88 | 22 |
* Within Day precision was determined using two runs per day with two replicates per run.
** Within Lab precision was determined using a single lot of reagent with one analyzer and four calibrations.
# b. Linearity/assay reportable range:
The linearity evaluation was performed based on NCCLS EP6-A (Evaluation of the Linearity of Quantitative Analytical Methods-Approved Guideline). A "high pool" was obtained by spiking a normal urine pool with human serum to an estimated albumin concentration of $190\mathrm{mg / L}$ . The low pool (normal urine pool) had an albumin concentration of $2.6\mathrm{mg / L}$ . Admixtures of the high pool and low pool were mixed to create 22 levels spanning the assay reportable range.
Four determinations of each of 22 pools were made together with 2 determinations each of VITROS Chemistry Products mALB Performance Verifiers I and II with each reagent lot on a VITROS 5, 1 FS Chemistry System. The VITROS Chemistry System operated within normal conditions.
The data were screened for outliers according to the statistical outlier test in NCCLS EP5-A. No data were rejected as statistical outliers. For each lot, the system responses were plotted against the percent high pool concentration to assess the degree to which the plotted curve conforms to a straight line.
The data presented define the linear range of the VITROS mALB Reagent assay when run on the VITROS 5, 1 FS Chemistry System. Based on the analysis of data from 21 fluids, the linear range of the assay was defined. Analysis by linear regression indicated that the assay is linear across the range of $0.6 - 19.0\mathrm{mg / dL}$ (6.0- $190.0\mathrm{mg / L}$ ) for all three lots tested.
# c. Traceability (controls, calibrators, or method):
Values assigned to the VITROS Chemistry Products Calibrator Kit 24 for urine albumin are traceable to IRMM/IFCC (Institute for Reference Methods and Materials/ International Federation of Clinical Chemistry and Laboratory Medicine) CRM 470 (RPPHS – Reference Preparation for Proteins in Human Serum) Reference Material.
{7}
# d. Detection limit:
The lower end of the reportable range is set at the value of the lowest calibrator which is $0.6\mathrm{mg / dL}$ or $6.0\mathrm{mg / L}$ .
# e. Analytical specificity:
To determine the effect of levels of exogenous and endogenous substances on the performance of the VITROS Chemistry Products mALB Reagent, NCCLS protocol EP7-A was used. A human urine pool spiked with human serum to obtain a target mALB of $25\mathrm{mg / L}$ and a simulated urine base spiked with human serum to a target mALB of $90\mathrm{mg / L}$ were used to test the following substances:
| Test Substance | Test Conc. Conv. Units | Test Conc SI Units |
| --- | --- | --- |
| Hemolysate | 500 mg/dL | 5 g/L |
| | 750 mg/dL | 7.5 g/L |
| | 1000 mg/dL | 10 g/L |
| Acetaminophen | 140 mg/dL | 9.3 mmol/L |
| Ammonia as Ammonium Sulfate | 570 mg/dL | 334.7 mmol/L |
| Amoxicillin | 200 mg/dL | 5.5 mmol/L |
| Ascorbic acid (L) | 500 mg/dL | 28.4 mmol/L |
| Ditaurobilirubin | 10 mg/dL | 0.17 mmol/L |
| | 20 mg/dL | 0.34 mmol/L |
| | 26 mg/dL | 0.44 mmol/L |
| Bovine Serum Albumin | 10 mg/dL | 0.1 g/L |
| Calcium as calcium chloride | 30 mg/dL | 7.5 mmol/L |
| Ceftriaxone, Na salt | 550 mg/dL | 9.91 mmol/L |
| Creatinine | 300 mg/dL | 26.5 mmol/L |
| Furosemide | 60 mg/dL | 1.8 mmol/L |
| Glucose | 4000 mg/dL | 222.0mmol/L |
| Hemolysate | 500 mg/dL | 5 g/L |
| | 750 mg/dL | 7.5 g/L |
| | 1000 mg/dL | 10 g/L |
| Human IgG | 200 mg/dL | 2 g/L |
| Ibuprofen | 50 mg/dL | 2.4 mmol/L |
| Magnesium As magnesium chloride | 60 mg/dL | 24.7 mmol/L |
| Propanol HCL | 55 mg/dL | 2.1 mmol/L |
| Salicylic Acid As sodium salicylate | 100 mg/dL | 7.3 mmol/L |
| Sulfamethoxazole | 120 mg/dL | 4.7 mmol/L |
| Trimethoprim | 2 mg/dL | 69 μmol/L |
| Urea | 3000 mg/dL | 499.5 mmol/L |
{8}
Page 9 of 12
| Uric Acid as sodium salt | 120 mg/dL | 7.14 mmol/L |
| --- | --- | --- |
| Thymol | 5 mL of 10% thymol per 1500 mL urine | |
| Boric Acid/Sodium Formate | 2g Boric Acid + 1g Sodium Formate per 100 mL urine | |
| Conc. HCL | 10 mL HCL per 1 L urine | |
| Glacial Acetic Acid | 10 mL glacial acetic acid per 1 L urine | |
Testing for endogenous and exogenous interferents followed the protocol outlined in EP7-A for paired-difference method.
Four to six determinations of each test substance and three determinations of the control fluids were carried out with two reagent lots on one VITROS 5, 1 FS Chemistry System.
For each substance tested, the "control" pool and the test substance pool replicate VITROS mALB assay results were calculated. The mean VITROS mALB assay results SD and CV (%) were calculated for each "control" pool and test substance pool. The bias was calculated as shown below:
Bias = (Mean conc. of interfering substance pool) – (Mean conc. of "control" pool).
## Known Interferences
The substances listed in the table, when tested at the concentrations indicated, cause the bias shown.
## Known Interfering Substances for mALB
| Interferent | Interferent | mALB Concentration | | Bias** | |
| --- | --- | --- | --- | --- | --- |
| | Concentration | Conv. SI (mg/dL) | (mg/L) | Conv. SI (mg/dL) | (mg/L) |
| Furosemide | 60 mg/dL 1.8 mmol/L | 2.4 | 23.8 | -0.5 | -4.7 |
| Sulfamethoxazole | 120 mg/dL 4.7 mmol/L | 2.3 | 23.3 | 0.5 | 5.0 |
** The bias is an estimate of the maximum difference observed.
A literature search was conducted to determine whether interferences from therapeutic drugs and their metabolites have been observed while
{9}
using immunoturbidimetric assay methods to quantify urinary albumin in human samples. No therapeutic drugs or metabolites were identified as causing analytical interference in urinary albumin assays.
The substances listed in the table below, at the concentrations shown, were tested according to NCCLS Protocol EP7-A. Bilirubin and hemoglobin were tested at urinary albumin concentrations of approximately 2.5 mg/dL (25 mg/L) and 10.0 mg/dL (100 mg/L) and found not to interfere, bias < 0.38 mg/dL (3.8 mg/L) and bias < 1.5 mg/dL (15 mg/L) respectively. The remaining compounds listed in the table were tested at a urinary albumin concentration of approximately 2.5 mg/dL (25 mg/L) and found not to interfere, bias < 0.38 mg/dL (3.8 mg/L).
Substances that do not interfere
| Compound | Concentration | |
| --- | --- | --- |
| Acetaminophen | 140 mg/dL | 9.3 mmol/L |
| Ammonia | 570 mg/dL | 334.7 mmol/L |
| Amoxicillin | 200 mg/dL | 5.5 mmol/L |
| Ascorbic acid (L) | 500 mg/dL | 28.4 mmol/L |
| Bilirubin | 26 mg/dL | 0.44 mmol/L |
| Bovine Serum Albumin (BSA) | 10 mg/dL | 0.1 g/L |
| Calcium | 30 mg/dL | 7.5 mmol/L |
| Ceftriaxone | 550 mg/dL | 9.9 mmol/L |
| Creatinine | 300 mg/dL | 26.5 mmol/L |
| Glucose | 4000 mg/dL | 222.0 mmol/L |
| Hemoglobin | 500 mg/dL | 5 g/L |
| Human IgG | 200 mg/dL | 2 g/L |
| Ibuprofen | 50 mg/dL | 2.4 mmol/L |
| Magnesium | 60 mg/dL | 24.7 mmol/L |
| Propanol | 55 mg/dL | 2.1 mmol/L |
| Salicylic acid | 100 mg/dL | 7.3 mmol/L |
| Trimethoprim | 2 mg/dL | 69μmol/L |
| Urea | 3000 mg/dL | 499.5 mmol/L |
| Uric acid | 120 mg/dL | 7.1 mmol/L |
Urine preservatives were tested with the VITROS Chemistry Products mALB Reagent at nominal concentrations of urinary albumin of approximately 2.5 mg/dL (25 mg/L) and found not to interfere, bias < 0.38 mg/dL (3.8 mg/L), at the concentrations shown.
f. Assay cut-off: Not applicable
{10}
Page 11 of 12
2. Comparison studies:
a. Method comparison with predicate device:
A total of 179 human urine samples were assayed using the VIIROS Chemistry Products mALB assay and the Dade Behring albumin assay (predicate device). Samples were analyzed by both methods on the same day. All samples were analyzed in triplicate on the VITROS 5, 1 FS Chemistry System analyzers and in duplicate on the ProSpec (predicate device). Testing was performed on two VITROS analyzers. Two different reagent lots were tested on both VITROS Systems.
For the quantitative comparison of methods, only those data within the reportable range of both the VITROS 5, 1 FS System (0.6 to 19.0 mg/dL) and the Dade Behring ProSpec (0.0 to 34.0 mg/L) were analyzed. One hundred and fifteen samples were removed from the regression analysis because they were out of the range on one of the assays tested.
The data were screened for outliers according to NCCLS Protocol EP5-A. A single replicate from one sample was excluded from the regression as a result of this analysis. The mean of the triplicates from the VITROS Chemistry Products mALB assay and the mean of the duplicates from the Dade Behring N Antiserum to Human Albumin assay were used in the analysis.
The relationship between the two methods, determined by Passing & Bablock linear regression, using 61 samples, the relationship between the two methods was as follows:
$$
\text{VITROS mALB assay} = 0.93 \times (\text{Dade Behring}) + 0.03 \, (\mathrm{mg/dL})
$$
$$
\text{Correlation coefficient}(r) = 0.977
$$
b. Matrix Comparison: Not applicable
3. Clinical studies:
a. Clinical sensitivity: Clinical studies are not typically submitted for this device type.
b. Clinical specificity: Clinical studies are not typically submitted for this device type.
c. Other clinical supportive data (when a and b are not applicable): Not applicable
4. Clinical cut-off: Not applicable
{11}
Page 12 of 12
# 5. Expected values/Reference range:
A total of 129 random urine specimens from apparently healthy adult subjects were assayed on two VITROS 5, 1 FS Chemistry System Analyzers. Two different reagent lots were used.
The evaluation of the results was performed according to NCCLS C28 (How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline-Second Edition.)
Histograms were created for each analyzer and reagent lot combination as well as the grand mean. The data were analyzed for outliers using box plots. Ten outliers were removed and the remaining data stored in ascending order of magnitude. Rank order numbers were assigned such that the lowest value has a rank of 1, and the highest number has a rank of n. The rank number of the $97.5^{\text{th}}$ percentile was computed as 0.975 $(n + 1)$ . The upper reference limit is equal to the value corresponding to the rank of the 0.975 fractile. The $90\%$ confidence interval around the upper reference limit was determined. The same data analysis was applied to compute the lower reference limit $(2.5^{\text{th}}$ percentile) with sample values predicted using an offline software analysis tool.
Results - The $97.5^{\text{th}}$ percentile value of the distribution and corresponding $90\%$ confidence interval was calculated for each analyzer and reagent lot as well as the grand mean across both analyzers and reagent lots. The combined results are summarized in the table below:
| Conventional Units (mg/dL) | SI Units (mg/L) |
| --- | --- |
| < 1.7 | < 16.7 (15.0 – 18.7) |
The results of this study are in the product labeling. In addition, a reference will be made to the following guidelines recommended by the American Diabetes Association.
| Category | 24-h collection (mg/24h) | Timed collection (μg/min.) | Spot collection (μg/mg creatinine) |
| --- | --- | --- | --- |
| Normal | <30 | <20 | <30 |
| Microalbuminuria | 30-300 | 20-200 | 30-300 |
| Clinical albuminuria | >300 | >200 | >300 |
5. American Diabetes Association. "Standards of Medical Care for Patients with Diabetes Mellitus" Diabetes Care, 23: Suppl. 1, S32-42, (2000).
# N. Conclusion:
The submitted material in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.