← Product Code [NBT](/productcode/NBT) · K042579

# BINDAZYME HUMAN ASCA IGG AND IGA EIA DIAGNOSTIC TEST KIT (K042579)

_The Binding Site · NBT · Jul 28, 2005 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K042579

## Device Facts

- **Applicant:** The Binding Site
- **Product Code:** [NBT](/productcode/NBT.md)
- **Decision Date:** Jul 28, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5785
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

This assay is intended for the in vitro semi-quantitative measurement of ASCA IgA anti-Saccharomyces cerevisiae antibody in human serum. The presence of ASCA may aid in the diagnosis of patients with Crohn's disease. The test results should be used in conjunction with clinical findings and other laboratory tests. The ASCA IgA should not be used alone as a screening test for ASCA. ASCA IgA test should be used to complement, but not to replace or to substitute the ASCA IgG antibody test since some Crohn's disease subgroup patients may not have IgA antibodies.

## Device Story

Bindazyme Human ASCA IgG and IgA assays are semi-quantitative ELISA tests for detecting anti-Saccharomyces cerevisiae antibodies in human serum. The device uses microplate strips coated with mannan isolated from S. cerevisiae. Patient serum samples, calibrators, and controls are added to wells; autoantibodies bind to mannan. After washing, peroxidase-labeled rabbit anti-human IgG or IgA conjugate is added, followed by TMB substrate. The resulting color intensity, measured at 450 nm via microplate reader, is proportional to antibody concentration. Used in clinical laboratories by trained personnel to aid in Crohn's disease diagnosis. Results are interpreted alongside clinical findings and other laboratory tests. The device provides a diagnostic aid for clinicians to identify patients with Crohn's disease, particularly those who may not express IgA antibodies.

## Clinical Evidence

Clinical performance evaluated using 105 Crohn's disease (CD) patients, 96 ulcerative colitis (UC) patients, and 120 normal donors. For CD patients, 44% were ASCA IgG positive and 37% were ASCA IgA positive. Method comparison against predicate devices showed 88.7% overall agreement for IgG and 95.0% for IgA. Specificity was assessed in celiac and ANA-positive samples.

## Technological Characteristics

Enzyme immunoassay (EIA) for semi-quantitative detection of antibodies in human serum.

## Regulatory Identification

The Anti-Saccharomyces cerevisiae (S. cerevisiae ) antibody (ASCA) test system is an in vitro diagnostic device that consists of the reagents used to measure, by immunochemical techniques, antibodies to S. cerevisiae (baker's or brewer's yeast) in human serum or plasma. Detection of S. cerevisiae antibodies may aid in the diagnosis of Crohn's disease.

## Special Controls

*Classification.* Class II (special controls). The special control is FDA's “Guidance for Industry and FDA Reviewers: Class II Special Control Guidance Document for Anti-*Saccharomyces cerevisiae* (*S. cerevisiae* ) Antibody (ASCA) Premarket Notifications.”

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k042579

B. Purpose for Submission:
New Devices

C. Measurand:
Anti-Saccharomyces Cerevisiae Antibody (ASCA), IgG and IgA

D. Type of Test:
Semi-quantitative ELISA

E. Applicant:
The Binding Site, Ltd.

F. Proprietary and Established Names:
Bindazyme™ Human ASCA IgG
Bindazyme™ Human ASCA IgA

G. Regulatory Information:

1. Regulation section:
21 CFR 866.5785 Anti-Saccharomyces Cerevisiae (ASCA) test system

2. Classification:
II

3. Product code:
NBT, Antibodies, Saccharomyces Cerevisiae (S. Cerevisiae)

4. Panel:
Immunology 82

H. Intended Use:

1. Intended use(s):
The Bindazyme Human ASCA IgG and the Bindazyme Human ASCA IgA assays are intended for the in vitro semi-quantitative measurement of ASCA IgG (Product Code MK053) or ASCA IgA (Product Code MK054), anti-Saccharomyces cerevisiae antibody, in human serum. The presence of ASCA may aid in the diagnosis of patients with Crohn's disease. The test results should be used in conjunction with clinical findings and other laboratory tests. The ASCA IgA should not be used alone as a screening test for ASCA. ASCA IgA should be used to complement, but not to replace or to substitute for the ASCA

{1}

IgG antibody test since some Crohn's disease subgroup patients may not have IgA antibodies. For in vitro diagnostic use only.

2. Indication(s) for use:
The Bindazyme Human ASCA IgG assay is intended for the in vitro semi-quantitative measurement of ASCA IgG, anti-Saccharomyces cerevisiae antibody, in human serum. The presence of ASCA may aid in the diagnosis of patients with Crohn's disease. The test results should be used in conjunction with clinical findings and other laboratory tests. For in vitro diagnostic use only.
The Bindazyme Human ASCA IgA assay is intended for the in vitro semi-quantitative measurement of ASCA IgA, anti-Saccharomyces cerevisiae antibody, in human serum. The presence of ASCA may aid in the diagnosis of patients with Crohn's disease. The test results should be used in conjunction with clinical findings and other laboratory tests. The ASCA IgA should not be used alone as a screening test for ASCA. ASCA IgA should be used to complement, but not to replace or to substitute for the ASCA IgG antibody test since some Crohn's disease subgroup patients may not have IgA antibodies. For in vitro diagnostic use only.

3. Special conditions for use statement(s):
The devices are for prescription use only.

4. Special instrument requirements:
Microplate reader capable of measuring OD at 450 nm.
Microplate washing device (250-350μL repeating, or multichannel pipette, or automated system).

I. Device Description:
Each device contains the following: microplate strips with breakaway microwells coated with mannan; five levels of calibrators (100, 33.3, 11.1, 3.7, 1.23 U/mL); positive, negative and cut-off controls (diluted human serum); wash buffer concentrate; Type III sample diluent; rabbit anti-human immunoglobulin (IgG or IgA) horseradish peroxidase conjugate; 3, 3', 5, 5' Tetramethylbenzidine (TMB) substrate; 3M phosphoric acid stop solution, and Q.C. Certificate for ASCA IgG and ASCA IgA.

J. Substantial Equivalence Information:
1. Predicate device name(s):
INOVA QUANTA Lite™ ASCA IgG
INOVA QUANTA Lite™ ASCA IgA

2. Predicate 510(k) number(s):
k000732
k000733

{2}

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | New Device | Predicate Device  |
|  Intended use | To aid in the diagnosis of Crohn's disease. | Same  |
|  Technology | ELISA | Same  |
|  Assay Format | Semi-quantitative | Same  |
|  High Positive and Negative controls | Ready to use | Same  |
|  Substrate | TMB | Same  |
|  Incubation times | 30-30-30 minutes | Same  |
|  Platform | 96 well microtiter plates | Same  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Antigen | Mannan isolated from S.cerevisiae | Partially purified S.cerevisiae  |
|  Enzyme-Conjugate | Horseradish Peroxidase rabbit anti-human IgG or IgA | Horseradish Peroxidase goat anti-human IgG or IgA  |
|  Calibrators | 5 levels: 100, 33.3, 11.1, 3.7, 1.23 U/mL | None  |
|  Low positive control | None | Ready to use  |
|  Cut-off control | Ready to use | None  |
|  Sample type and dilution | Serum at 1:100 | Serum at 1:101  |
|  Sample volume required | 10 μL | 5 μL  |
|  Sample diluent | Tris Buffered saline with Tween 20 | Phosphate buffer with Tween 20. pH 7.2  |
|  Wash buffer | 20x concentrate | 40x concentrate  |
|  Stop solution | 0.344M sulfuric acid | 3M phosphoric acid  |
|  OD measurement | Within 30 minutes | Within 1 hour  |
|  OD reading | 450 nm | 450nm/620nm  |
|  ASCA IgG Results Interpretation | Negative: ≤ 22.0 U/mL Positive: > 22 U/mL | Negative: 0.0-20.0 Units Equivocal:20.1-24.9 Positive: ≥ 25.0 Units  |
|  ASCA IgA Results Interpretation | Negative: ≤ 7.0 U/mL Positive: > 7.0 U/mL | Negative: 0.0-20.0 Units Equivocal: 20.1-24.9 Positive: ≥ 25.0 Units  |

K. Standard/Guidance Document Referenced (if applicable):
None referenced.

{3}

4

L. Test Principle:

The Microwells are pre-coated with mannan isolated from *Saccharomyces cerevisiae*. Calibrators, controls and diluted patient samples are added to the wells and autoantibodies recognizing mannan bind during the first incubation. After washing the wells to remove all unbound proteins, peroxidase labeled rabbit anti-human IgG or IgA antibody (γ or α chain specific) conjugate is added. The conjugate binds to the captured human antibody and the excess unbound conjugate is removed by a further wash step. The bound conjugate is visualized with 3,3',5,5'-tetramethylbenzidine (TMB) substrate which gives a blue reaction product, the intensity of which is proportional to the concentration of autoantibody in the sample. Phosphoric acid is added to each well to stop the reaction. This produces a yellow end point color, which is read at 450nm.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

The intra-assay precision was measured using six samples, with ASCA IgG concentration level ranges from 12.1 to 67.1 U/mL, and ASCA IgA concentration level ranges from 2.9 to 67.8 U/mL. The ASCA sample ranges tested were within the two devices calibration range of 1.23 through 100 U/mL. These six samples were tested in duplicate twenty times. The % C.V. for each sample was below 10% and they are listed as follows:

INTRA-ASSAY PRECISION

|  ASCA IgG  |   |   |
| --- | --- | --- |
|  n=20 | Concentration (U/mL) | % C.V.  |
|  Sample 1 | 12.1 | 3.7  |
|  Sample 2 | 14.2 | 4.2  |
|  Sample 3 | 33.0 | 4.1  |
|  Sample 4 | 46.8 | 2.3  |
|  Sample 5 | 67.1 | 6.4  |
|  Sample 6 | 81.3 | 6.7  |
|  ASCA IgA  |   |   |
|  n=20 | Concentration (U/mL) | % C.V.  |
|  Sample 1 | 2.9 | 7.3  |
|  Sample 2 | 3.1 | 4.7  |
|  Sample 3 | 10.5 | 4.1  |
|  Sample 4 | 15.0 | 2.0  |
|  Sample 5 | 30.9 | 4.7  |
|  Sample 6 | 67.8 | 2.9  |

The inter-assay precision was measured using eight clinical samples tested in duplicate six times for three days. The % C.V. for each sample was ≤12% and they are listed as follows:

{4}

INTER-ASSAY PRECISION

|  ASCA IgG  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  n=6 | Conc (U/mL) | % C.V. | n=6 | Conc (U/mL) | % C.V.  |
|  Sample 1 | 10.2 | 8.9 | Sample 5 | 38.4 | 7.8  |
|  Sample 2 | 12.7 | 10.4 | Sample 6 | 48.7 | 6.6  |
|  Sample 3 | 28.7 | 7.5 | Sample 7 | 67.9 | 7.9  |
|  Sample 4 | 37.0 | 8.8 | Sample 8 | 80.0 | 4.7  |
|  ASCA IgA  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  n=6 | Conc (U/mL) | % C.V. | n=6 | Conc (U/mL) | % C.V.  |
|  Sample 1 | 2.9 | 5.9 | Sample 5 | 10.2 | 4.7  |
|  Sample 2 | 3.0 | 7.7 | Sample 6 | 14.9 | 7.2  |
|  Sample 3 | 6.5 | 9.2 | Sample 7 | 63.2 | 7.3  |
|  Sample 4 | 6.6 | 12.0 | Sample 8 | 89.4 | 7.3  |

b. Linearity/assay reportable range: The measuring range of the assay is 1.23 – 100 U/mL.

i. Linearity study: Three known Crohn's positive samples with different levels of ASCA IgG or ASCA IgA were serially diluted after the initial dilution of 1:100. The values were plotted and the correlation coefficients were calculated. The ASCA IgG R² values on the three samples were 0.9999, 0.9985, and 0.9994, and the ASCA IgA R² values were 0.9991, 0.9990, and 0.9995. All samples correlate to their expected values based on the dilution and had comparable linearity.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

i. Traceability: There is no recognized standard or reference material for ASCA.

d. Detection limit: The minimum level of detection of the assay was measured the diluent 20 times. The sensitivity was based on the mean plus two standard deviations of the mean OD and was then converted to U/mL. The detection limits for ASCA IgG and ASCA IgA are 0.61 U/mL and 0.93 U/mL respectively.

e. Analytical specificity:

Interference: The substances that were tested in the interference study were Bilirubin F (Free), Bilirubin C (Conjugate), Haemoglobin, Chyle, and Rheumatoid Factor. The potentially interfering substance or corresponding blank material (water) was added at a 1/10 dilution to both a negative sample and a positive Crohn's disease sample. The samples were then assayed to determine whether the interfering substance had any effect on the assay value.

{5}

None of the above interfering substances interfered with the assay.

An additional three samples were performed on the ASCA IgA interference study, to demonstrate that results close to the cut-off of 7.0 U/mL would not be affected by the above interfering substances. Three samples ranging from 4.1 to 5.6 U/mL were tested. None of the interfering substances caused the result to change to positive. All values are within ± 20% of the sample with water added.

f. Assay cut-off: The cut-off value of ASCA IgG (≤ 22.0 U/mL) and ASCA IgA (≤7.0 U/mL) levels were established in serum from 200 normal blood donors. For the ASCA IgG, ten of the 200 samples (5%) were above the cut-off at 22 U/mL (approx. mean + 2SD). Using an alternative ASCA IgG kit, three of these positive samples with values of 99.2, 643.1 and 30.9 U/mL were also tested positive while the remaining 7 samples had values between 23.0 - 34.8 U/mL. This study showed that 91.5% were below the ASCA IgG cut-off. For the ASCA IgA, none of the 200 samples tested were above the cut-off at 7.0 U/mL (approx. mean + 5SD).

2. Comparison studies:

a. Method comparison with predicate device:

Comparison was determined against the predicate ASCA IgG and IgA EIA kits using 120 normal and 105 samples from patients with clinically confirmed Crohn's disease. Samples in the equivocal range in alternative kits (4 samples) were excluded from the analysis. Results are summarized below.

|   | QUANTA Lite™ ASCA IgG  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  + | Equivocal* | -  |
|  BINDAZYME | + | 47 | 0 | 5  |
|  ASCA IgG | - | 20 | 4 | 149  |
|  Total |   | 67 | 4 | 154  |
|  Positive Percent Agreement |   | 70.2%  |   |   |
|  Negative Percent Agreement |   | 96.8%  |   |   |
|  Overall Agreement |   | 88.7%  |   |   |
|   | QUANTA Lite™ ASCA IgA  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  + | Equivocal* | -  |
|  BINDAZYME | + | 32 | 3 | 8  |
|  ASCA IgA | - | 3 | 1 | 178  |
|  Total |   | 35 | 4 | 186  |
|  Positive Percent Agreement |   | 91.4%  |   |   |
|  Negative Percent Agreement |   | 95.7%  |   |   |
|  Overall Agreement |   | 95.0%  |   |   |

{6}

b. Matrix comparison:

Both assays use serum as the matrix.

# 3. Clinical studies:

a. Clinical sensitivity:

Study on other GI Inflammatory Disease Groups: One hundred and five samples from clinically confirmed Crohn's disease (CD) and 96 clinically confirmed ulcerative colitis (UC) patients were tested. Of the 31 samples positive for both ASCA IgG and IgA, 27  $(87\%)$  were from the Crohn's Disease group. The incidence of positive results in the normal sample group supports the reported incidence in the normal range study of the Bindazyme ASCA EIA kits (refer to the Assay Cut-off section). The table below shows the results:

|  BINDAZYME™ ASCA EIA Result  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Group | N | IgG Pos (n) | IgA Pos (n) | IgG or IgA positive (n) | IgG and IgA Pos (n)  |
|  Crohn's Disease | 105 | 44% (46) | 37% (39) | 55% (58) | 26% (27)  |
|  UC | 96 | 14% (13) | 6% (6) | 17% (16) | 3% (3)  |
|  Normals | 120 | 5% (6) | 3% (4) | 8% (9) | 1% (1)  |

b. Clinical specificity:

ASCA Study on Celiac and ANA positive samples: Study results on the incidence of ASCA in samples from patients positive for antibody markers of celiac disease and samples positive for antinuclear antibodies are listed in the table below:

|  Sample | N | ASCA IgG pos | ASCA IgA pos  |
| --- | --- | --- | --- |
|  Gliadin IgG pos | 11 | 2 | 3  |
|  Gliadin IgA pos | 14 | 6 | 2  |
|  Transglutaminase IgA pos | 9 | 3 | 2  |
|  Transglutaminase IgA pos | 4 | 0 | 0  |
|  Antinuclear antibody pos | 10 | 0 | 0  |
|  GBM positive | 10 | 2 | 1  |

c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.

# 4. Clinical cut-off:

Same as assay cut-off.

{7}

5. Expected values/Reference range:
Expected values in the normal population using the Bindazyme ASCA assays are ≤22.0 U/mL for ASCA IgG and ≤7.0 U/mL for ASCA IgA.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

8

---

**Source:** [https://fda.innolitics.com/device/K042579](https://fda.innolitics.com/device/K042579)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
